Validation of electrical velocimetry in resuscitation of patients undergoing liver transplantation. Observational study

Authors: Ahmed M. Mukhtar, Mohamed Elayashy, Amr H. Sayed, Gihan M. Obaya, Akram A. Eladawy, Mai A. Ali, Hisham M. Dahab, Dina Z. Khalaf, Mostafa A. Mohamed, Amr H. Elfouly, Gad M. Behairy, Amr A. Abdelaal

Published in: Journal of Clinical Monitoring and Computing | Issue 2/2020

Login to get access

Abstract

Major hemodynamic changes are frequently noted during liver transplantation (LT). We evaluated the performance of electrical velocimetry (EV) as compared to that of TEE in SV optimization during liver transplantation. This was an observational study in 32 patients undergoing LT. We compared SV values measured simultaneously by EV (SVEV) and TEE (SVTEE) at baseline 30 min after induction, at the end of dissection phase, 30 min after anhepatic phase, 30 min after reperfusion. We also evaluated the reliability of EV to track changes In SV before and after 49 fluid challenges. Finally, the SV variation (SVV) and pulse pressure variation (PPV) were tested as predictors for volume responsiveness, defined as an increase in SV ≥ 10% after 250 ml of colloid. For 112 paired SV data, the overall correlation was 0.76 and bias (limits of agreement) 0.3 (− 29 to 29) ml percentage error 62%. The EV was able to track changes in SV with a concordance rate of 97%, and a sensitivity and specificity of 93% to detect a positive fluid challenge. The AUC values (with 95% confidence intervals) for SVV and PPV were 0.68 (0.52–0.83) and 0.72 (0.57–0.86), respectively, indicating low predictive capacity in these setting. The absolute values of SV derived from EV did not agree with SV derived from TEE. However, EV was able to track the direction of changes in SV during hemodynamic management of patients undergoing liver transplantation.
Clinical trial registration: Clinicaltrials.gov Identifier: NCT03228329 prospectively Registered on 13-July-2017.
Literature
1.
go back to reference Ozier Y, Klinck JR. Anesthetic management of hepatic transplantation. Curr Opin Anaesthesiol. 2008;21:391–400.CrossRef Ozier Y, Klinck JR. Anesthetic management of hepatic transplantation. Curr Opin Anaesthesiol. 2008;21:391–400.CrossRef
2.
go back to reference Vincent JL, Weil MH. Fluid challenge revisited. Crit Care Med. 2006;34:1333–7.CrossRef Vincent JL, Weil MH. Fluid challenge revisited. Crit Care Med. 2006;34:1333–7.CrossRef
3.
go back to reference Marik PE, Levitov A, Young A, Andrews L. The use of bioreactance and carotid doppler to determine volume responsiveness and blood flow redistribution following passive leg raising in hemodynamically unstable patients. Chest. 2013;143:364–70.CrossRef Marik PE, Levitov A, Young A, Andrews L. The use of bioreactance and carotid doppler to determine volume responsiveness and blood flow redistribution following passive leg raising in hemodynamically unstable patients. Chest. 2013;143:364–70.CrossRef
4.
go back to reference Boyle M, Steel L, Flynn GM, Murgo M, Nicholson L, O’Brien M, Bihari D. Assessment of the clinical utility of an ultrasonic monitor of cardiac output (the USCOM) and agreement with thermodilution measurement. Crit Care Resusc. 2009;11:198–203.PubMed Boyle M, Steel L, Flynn GM, Murgo M, Nicholson L, O’Brien M, Bihari D. Assessment of the clinical utility of an ultrasonic monitor of cardiac output (the USCOM) and agreement with thermodilution measurement. Crit Care Resusc. 2009;11:198–203.PubMed
5.
go back to reference Mora B, Ince I, Birkenberg B, Skhirtladze K, Pernicka E, Ankersmit HJ, Dworschak M. Validation of cardiac output measurement with the LiDCOTM pulse contour system in patients with impaired left ventricular function after cardiac surgery. Anesthesia. 2011;66:675–81.CrossRef Mora B, Ince I, Birkenberg B, Skhirtladze K, Pernicka E, Ankersmit HJ, Dworschak M. Validation of cardiac output measurement with the LiDCOTM pulse contour system in patients with impaired left ventricular function after cardiac surgery. Anesthesia. 2011;66:675–81.CrossRef
6.
go back to reference Robertson AC, Eagle SS. Transesophageal echocardiography during orthotopic liver transplantation: maximizing information without the distraction. J Cardiothorac Vasc Anesth. 2014;28:141–54.CrossRef Robertson AC, Eagle SS. Transesophageal echocardiography during orthotopic liver transplantation: maximizing information without the distraction. J Cardiothorac Vasc Anesth. 2014;28:141–54.CrossRef
7.
go back to reference Critchley LA, Lee A, Ho AMH. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111:1180–92.CrossRef Critchley LA, Lee A, Ho AMH. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111:1180–92.CrossRef
8.
go back to reference Noori S, Drabu B, Soleymani S, Seri I. Continuous non-invasive cardiac output measurements in the neonate by electrical velocimetry: a comparison with echocardiography. Arch Dis Child Fetal Neonatal Ed. 2012;97:F340–3.CrossRef Noori S, Drabu B, Soleymani S, Seri I. Continuous non-invasive cardiac output measurements in the neonate by electrical velocimetry: a comparison with echocardiography. Arch Dis Child Fetal Neonatal Ed. 2012;97:F340–3.CrossRef
9.
go back to reference Martin E, Anyikam A, Ballas J, Buono K, Mantell K, Huynh-Covey T, Archer T. A validation study of electrical cardiometry in pregnant patients using transthoracic echocardiography as the reference standard. J Clin Monit Comput. 2016;30:679–86.CrossRef Martin E, Anyikam A, Ballas J, Buono K, Mantell K, Huynh-Covey T, Archer T. A validation study of electrical cardiometry in pregnant patients using transthoracic echocardiography as the reference standard. J Clin Monit Comput. 2016;30:679–86.CrossRef
10.
go back to reference Sasaki K, Mutoh T, Mutoh T, Kawashima R, Tsubone H. Electrical velocimetry for noninvasive cardiac output and stroke volume variation measurements in dogs undergoing cardiovascular surgery. Vet Anaesth Analg. 2017;44:7–16.CrossRef Sasaki K, Mutoh T, Mutoh T, Kawashima R, Tsubone H. Electrical velocimetry for noninvasive cardiac output and stroke volume variation measurements in dogs undergoing cardiovascular surgery. Vet Anaesth Analg. 2017;44:7–16.CrossRef
11.
go back to reference Mukhtar A, Aboulfetouh F, Obayah G, Salah M, Emam M, Khater Y, Akram R, Hoballah A, Bahaa M, Elmeteini M, Hamza A. The safety of modern hydroxyethyl starch in living donor liver transplantation: a comparison with human albumin. Anesth Analg. 2009;109:924–30.CrossRef Mukhtar A, Aboulfetouh F, Obayah G, Salah M, Emam M, Khater Y, Akram R, Hoballah A, Bahaa M, Elmeteini M, Hamza A. The safety of modern hydroxyethyl starch in living donor liver transplantation: a comparison with human albumin. Anesth Analg. 2009;109:924–30.CrossRef
12.
go back to reference Bataille B, Bertuit M, Mora M, Mazerolles M, Cocquet P, Masson B, et al. Comparison of esCCO and transthoracic echocardiography for non-invasive measurement of cardiac output intensive care. Br J Anaesth. 2012;109:879–86.CrossRef Bataille B, Bertuit M, Mora M, Mazerolles M, Cocquet P, Masson B, et al. Comparison of esCCO and transthoracic echocardiography for non-invasive measurement of cardiac output intensive care. Br J Anaesth. 2012;109:879–86.CrossRef
13.
go back to reference Peyton PJ, Chong SW. Bias and precision statistics: should we still adhere to the 30% benchmark for cardiac output monitor validation studies? Anesthesiology. 2011;114:1245–6.CrossRef Peyton PJ, Chong SW. Bias and precision statistics: should we still adhere to the 30% benchmark for cardiac output monitor validation studies? Anesthesiology. 2011;114:1245–6.CrossRef
14.
go back to reference Liu YH, Dhakal BP, Keesakul C, Kacmarek RM, Lewis GD, Jiang Y. Continuous non-invasive cardiac output monitoring during exercise: validation of electrical cardiometry with Fick and thermodilution methods. Br J Anaesth. 2016;117:129–31.CrossRef Liu YH, Dhakal BP, Keesakul C, Kacmarek RM, Lewis GD, Jiang Y. Continuous non-invasive cardiac output monitoring during exercise: validation of electrical cardiometry with Fick and thermodilution methods. Br J Anaesth. 2016;117:129–31.CrossRef
15.
go back to reference Mukhtar A, Dabbous H. Modulation of splanchnic circulation: role in perioperative management of liver transplant patients. World J Gastroenterol. 2016;22:1582–92.CrossRef Mukhtar A, Dabbous H. Modulation of splanchnic circulation: role in perioperative management of liver transplant patients. World J Gastroenterol. 2016;22:1582–92.CrossRef
16.
go back to reference Sasaki K, Mutoh T, Mutoh T, Kawashima R, Tsubone H. Electrical velocimetry for noninvasive cardiac output and stroke volume variation measurements in dogs undergoing cardiovascular surgery. Vet Anaesth Analg. 2017;44:7–16.CrossRef Sasaki K, Mutoh T, Mutoh T, Kawashima R, Tsubone H. Electrical velocimetry for noninvasive cardiac output and stroke volume variation measurements in dogs undergoing cardiovascular surgery. Vet Anaesth Analg. 2017;44:7–16.CrossRef
17.
go back to reference Gouvêa G, Diaz R, Auler L, Toledo R, Martinho JM. Evaluation of the pulse pressure variation index as a predictor of fluid responsiveness during orthotopic liver transplantation. Br J Anaesth. 2009;103:238–43.CrossRef Gouvêa G, Diaz R, Auler L, Toledo R, Martinho JM. Evaluation of the pulse pressure variation index as a predictor of fluid responsiveness during orthotopic liver transplantation. Br J Anaesth. 2009;103:238–43.CrossRef
Metadata
Title
Validation of electrical velocimetry in resuscitation of patients undergoing liver transplantation. Observational study
Authors
Ahmed M. Mukhtar
Mohamed Elayashy
Amr H. Sayed
Gihan M. Obaya
Akram A. Eladawy
Mai A. Ali
Hisham M. Dahab
Dina Z. Khalaf
Mostafa A. Mohamed
Amr H. Elfouly
Gad M. Behairy
Amr A. Abdelaal
Publication date
01-04-2020
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 2/2020
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-019-00313-z

Other articles of this Issue 2/2020

Journal of Clinical Monitoring and Computing 2/2020 Go to the issue