Published in: International Journal of Implant Dentistry 1/2017

Open Access 01-12-2017 | Technical Advances Article

Customized SmartPeg for measurement of resonance frequency of mini dental implants

Authors: Jagjit Singh Dhaliwal, Rubens F. Albuquerque Jr, Ali Fakhry, Sukhbir Kaur, Jocelyne S. Feine

Published in: International Journal of Implant Dentistry | Issue 1/2017

Login to get access

Abstract

Background

One-piece narrow diameter implants (NDIs) have been recommended as “Single-tooth replacements in the anterior zones, single posterior, multiple-unit fixed dental prosthesis (FDP), edentulous jaws to be rehabilitated with FDP, and edentulous jaws rehabilitation with overdentures in situations with reduced mesiodistal space or reduced ridge width.” (ITI consensus 2013). Since NDIs can be immediately loaded, it is important to be able to carry out stability testing. We developed and validated a customized SmartPeg for this type of implant to measure the Implant Stability Quotient (ISQ). The ISQ of mini dental implants (MDIs) was measured and compared with the stability of standard and in a rabbit model.

Objective

The aim of the study is to test the feasibility of a customized SmartPeg for resonance frequency measurement of single-piece mini dental implants and to compare primary stability of a standard and the mini dental implant (3M™ESPE™ MDI) in a rabbit model after 6 weeks of healing.

Methods

Eight New Zealand white rabbits were used for the study. The protocol was approved by the McGill University Animal Ethics Review Board. Sixteen 3M™ESPE™ MDI and equal number of standard implants (Ankylos® Friadent, Dentsply) were inserted into the tibia/femur of the rabbits and compared. Each rabbit randomly received two 3M™ESPE™ MDI and two Ankylos® implants in each leg. ISQ values were measured with the help of an Osstell ISQ device using custom-made SmartPegs for the MDIs and implant-specific SmartPegs™ (Osstell) for the Ankylos®. Measurements were obtained both immediately following implant placement surgery and after a 6-week healing period. Each reading was taken thrice and their average compared using Wilcoxon matched pairs signed-rank tests.

Results

The median ISQ and interquartile range (IQR) values were 53.3 (8.3) at insertion and 60.5 (5.5) at 6 weeks for the 3M™ESPE™MDI and, respectively, 58.5 (4.75) and 65.5 (9.3) for the Ankylos® implant. These values also indicate that both types of implants achieved primary and secondary stability, and this is supported by histological data. ISQ values of both 3M™ESPE™ MDI and Ankylos® increased significantly from the time of insertion to 6 weeks post-insertion (p < 0.05).

Conclusions

The new custom-made SmartPeg is suitable for measuring the Implant Stability Quotient of 3M™ESPE™MDIs. The primary stability of 3M™ESPE™MDIs is similar to the primary stability attained by standard implants in the rabbit tibia.
Literature
1.
go back to reference Branemark PI, Adell R, Breine U, Hansson BO, Lindstrom J, Ohlsson A. Intra-osseous anchorage of dental prostheses. I. Experimental studies. Scand J Plast Reconstr Surg. 1969;3(2):81–100. Epub 1969/01/01.CrossRefPubMed Branemark PI, Adell R, Breine U, Hansson BO, Lindstrom J, Ohlsson A. Intra-osseous anchorage of dental prostheses. I. Experimental studies. Scand J Plast Reconstr Surg. 1969;3(2):81–100. Epub 1969/01/01.CrossRefPubMed
2.
go back to reference Meredith N. Assessment of implant stability as a prognostic determinant. Int J Prosthodont. 1998;11(5):491–501. Epub 1999/01/29.PubMed Meredith N. Assessment of implant stability as a prognostic determinant. Int J Prosthodont. 1998;11(5):491–501. Epub 1999/01/29.PubMed
3.
go back to reference Lioubavina-Hack N, Lang NP, Karring T. Significance of primary stability for osseointegration of dental implants. Clin Oral Implants Res. 2006;17(3):244–50. Epub 2006/05/05.CrossRefPubMed Lioubavina-Hack N, Lang NP, Karring T. Significance of primary stability for osseointegration of dental implants. Clin Oral Implants Res. 2006;17(3):244–50. Epub 2006/05/05.CrossRefPubMed
4.
go back to reference Albrektsson T, Dahl E, Enbom L, Engevall S, Engquist B, Eriksson AR, et al. Osseointegrated oral implants. A Swedish multicenter study of 8139 consecutively inserted nobelpharma implants. J Periodontol. 1988;59(5):287–96. Epub 1988/05/01.CrossRefPubMed Albrektsson T, Dahl E, Enbom L, Engevall S, Engquist B, Eriksson AR, et al. Osseointegrated oral implants. A Swedish multicenter study of 8139 consecutively inserted nobelpharma implants. J Periodontol. 1988;59(5):287–96. Epub 1988/05/01.CrossRefPubMed
5.
go back to reference Bornstein MM, Valderrama P, Jones AA, Wilson TG, Seibl R, Cochran DL. Bone apposition around two different sandblasted and acid-etched titanium implant surfaces: a histomorphometric study in canine mandibles. Clin Oral Implants Res. 2008;19(3):233–41. Epub 2008/01/08.CrossRefPubMed Bornstein MM, Valderrama P, Jones AA, Wilson TG, Seibl R, Cochran DL. Bone apposition around two different sandblasted and acid-etched titanium implant surfaces: a histomorphometric study in canine mandibles. Clin Oral Implants Res. 2008;19(3):233–41. Epub 2008/01/08.CrossRefPubMed
6.
go back to reference Bulard RA, Vance JB. Multi-clinic evaluation using mini-dental implants for long-term denture stabilization: a preliminary biometric evaluation. Compend Contin Educ Dent. 2005;26(12):892–7. Epub 2006/01/05.PubMed Bulard RA, Vance JB. Multi-clinic evaluation using mini-dental implants for long-term denture stabilization: a preliminary biometric evaluation. Compend Contin Educ Dent. 2005;26(12):892–7. Epub 2006/01/05.PubMed
7.
go back to reference Diedrich P, Fritz U, Kinzinger G, Angelakis J. Movement of periodontally affected teeth after guided tissue regeneration (GTR)—an experimental pilot study in animals. J Orofac Orthop. 2003;64(3):214–27. Epub 2003/07/02.CrossRefPubMed Diedrich P, Fritz U, Kinzinger G, Angelakis J. Movement of periodontally affected teeth after guided tissue regeneration (GTR)—an experimental pilot study in animals. J Orofac Orthop. 2003;64(3):214–27. Epub 2003/07/02.CrossRefPubMed
8.
go back to reference Buchter A, Kleinheinz J, Wiesmann HP, Jayaranan M, Joos U, Meyer U. Interface reaction at dental implants inserted in condensed bone. Clin Oral Implants Res. 2005;16(5):509–17. Epub 2005/09/17.CrossRefPubMed Buchter A, Kleinheinz J, Wiesmann HP, Jayaranan M, Joos U, Meyer U. Interface reaction at dental implants inserted in condensed bone. Clin Oral Implants Res. 2005;16(5):509–17. Epub 2005/09/17.CrossRefPubMed
9.
go back to reference Kwon KR, Sachdeo A, Weber HP. Achieving immediate function with provisional prostheses after implant placement: a clinical report. J Prosthet Dent. 2005;93(6):514–7. Epub 2005/06/09.CrossRefPubMed Kwon KR, Sachdeo A, Weber HP. Achieving immediate function with provisional prostheses after implant placement: a clinical report. J Prosthet Dent. 2005;93(6):514–7. Epub 2005/06/09.CrossRefPubMed
10.
go back to reference Kokubo Y, Ohkubo C. Occlusion recording device for dental implant-supported restorations. J Prosthet Dent. 2006;95(3):262–3. Epub 2006/03/18.CrossRefPubMed Kokubo Y, Ohkubo C. Occlusion recording device for dental implant-supported restorations. J Prosthet Dent. 2006;95(3):262–3. Epub 2006/03/18.CrossRefPubMed
11.
go back to reference Ohkubo C, Kobayashi M, Suzuki Y, Sato J, Hosoi T, Kurtz KS. Evaluation of transitional implant stabilized overdentures: a case series report. J Oral Rehabil. 2006;33(6):416–22. Epub 2006/05/05.CrossRefPubMed Ohkubo C, Kobayashi M, Suzuki Y, Sato J, Hosoi T, Kurtz KS. Evaluation of transitional implant stabilized overdentures: a case series report. J Oral Rehabil. 2006;33(6):416–22. Epub 2006/05/05.CrossRefPubMed
12.
go back to reference Campelo LD, Camara JR. Flapless implant surgery: a 10-year clinical retrospective analysis. Int J Oral Maxillofac Implants. 2002;17(2):271–6. Epub 2002/04/18.PubMed Campelo LD, Camara JR. Flapless implant surgery: a 10-year clinical retrospective analysis. Int J Oral Maxillofac Implants. 2002;17(2):271–6. Epub 2002/04/18.PubMed
13.
go back to reference Shatkin TE, Oppenheimer BD, Oppenheimer AJ. Mini dental implants for long-term fixed and removable prosthetics: a retrospective analysis of 2514 implants placed over a five-year period. Compend Contin Educ Dent. 2007;28(2):92–9. quiz 100-1. Epub 2007/02/27.PubMed Shatkin TE, Oppenheimer BD, Oppenheimer AJ. Mini dental implants for long-term fixed and removable prosthetics: a retrospective analysis of 2514 implants placed over a five-year period. Compend Contin Educ Dent. 2007;28(2):92–9. quiz 100-1. Epub 2007/02/27.PubMed
14.
go back to reference Griffitts TM, Collins CP, Collins PC. Mini dental implants: an adjunct for retention, stability, and comfort for the edentulous patient. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2005;100(5):e81–4. Epub 2005/10/26.CrossRefPubMed Griffitts TM, Collins CP, Collins PC. Mini dental implants: an adjunct for retention, stability, and comfort for the edentulous patient. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2005;100(5):e81–4. Epub 2005/10/26.CrossRefPubMed
15.
go back to reference Albrektsson T, Sennerby L. State of the art in oral implants. J Clin Periodontol. 1991;18(6):474–81. Epub 1991/07/01.CrossRefPubMed Albrektsson T, Sennerby L. State of the art in oral implants. J Clin Periodontol. 1991;18(6):474–81. Epub 1991/07/01.CrossRefPubMed
16.
go back to reference Atsumi M, Park SH, Wang HL. Methods used to assess implant stability: current status. Int J Oral Maxillofac Implants. 2007;22(5):743–54. Epub 2007/11/03.PubMed Atsumi M, Park SH, Wang HL. Methods used to assess implant stability: current status. Int J Oral Maxillofac Implants. 2007;22(5):743–54. Epub 2007/11/03.PubMed
17.
go back to reference Davies JE. Mechanisms of endosseous integration. Int J Prosthodont. 1998;11(5):391–401. Epub 1999/01/29.PubMed Davies JE. Mechanisms of endosseous integration. Int J Prosthodont. 1998;11(5):391–401. Epub 1999/01/29.PubMed
18.
go back to reference Romanos GE. Bone quality and the immediate loading of implants-critical aspects based on literature, research, and clinical experience. Implant Dent. 2009;18(3):203–9. Epub 2009/06/11.CrossRefPubMed Romanos GE. Bone quality and the immediate loading of implants-critical aspects based on literature, research, and clinical experience. Implant Dent. 2009;18(3):203–9. Epub 2009/06/11.CrossRefPubMed
19.
go back to reference Cawley P, Pavlakovic B, Alleyne DN, George R, Back T, Meredith N. The design of a vibration transducer to monitor the integrity of dental implants. Proc Inst Mech Eng H J Eng Med. 1998;212(4):265–72. Epub 1998/10/14.CrossRef Cawley P, Pavlakovic B, Alleyne DN, George R, Back T, Meredith N. The design of a vibration transducer to monitor the integrity of dental implants. Proc Inst Mech Eng H J Eng Med. 1998;212(4):265–72. Epub 1998/10/14.CrossRef
20.
go back to reference Szmukler-Moncler S, Salama H, Reingewirtz Y, Dubruille JH. Timing of loading and effect of micromotion on bone-dental implant interface: review of experimental literature. J Biomed Mater Res. 1998;43(2):192–203. Epub 1998/06/10.CrossRefPubMed Szmukler-Moncler S, Salama H, Reingewirtz Y, Dubruille JH. Timing of loading and effect of micromotion on bone-dental implant interface: review of experimental literature. J Biomed Mater Res. 1998;43(2):192–203. Epub 1998/06/10.CrossRefPubMed
21.
go back to reference Wilmes B, Rademacher C, Olthoff G, Drescher D. Parameters affecting primary stability of orthodontic mini-implants. J Orofac Orthop. 2006;67(3):162–74. Epub 2006/06/01.CrossRefPubMed Wilmes B, Rademacher C, Olthoff G, Drescher D. Parameters affecting primary stability of orthodontic mini-implants. J Orofac Orthop. 2006;67(3):162–74. Epub 2006/06/01.CrossRefPubMed
22.
go back to reference Boice GW, Kraut RA. Maxillary denture retention using rare earth magnets and endosteal implants. Int J Oral Implantol. 1991;7(2):23–7. Epub 1991/01/01.PubMed Boice GW, Kraut RA. Maxillary denture retention using rare earth magnets and endosteal implants. Int J Oral Implantol. 1991;7(2):23–7. Epub 1991/01/01.PubMed
23.
go back to reference Sullivan DY, Sherwood RL, Collins TA, Krogh PH. The reverse-torque test: a clinical report. Int J Oral Maxillofac Implants. 1996;11(2):179–85. Epub 1996/03/01.PubMed Sullivan DY, Sherwood RL, Collins TA, Krogh PH. The reverse-torque test: a clinical report. Int J Oral Maxillofac Implants. 1996;11(2):179–85. Epub 1996/03/01.PubMed
24.
go back to reference Branemark R, Ohrnell LO, Nilsson P, Thomsen P. Biomechanical characterization of osseointegration during healing: an experimental in vivo study in the rat. Biomaterials. 1997;18(14):969–78. Epub 1997/07/01.CrossRefPubMed Branemark R, Ohrnell LO, Nilsson P, Thomsen P. Biomechanical characterization of osseointegration during healing: an experimental in vivo study in the rat. Biomaterials. 1997;18(14):969–78. Epub 1997/07/01.CrossRefPubMed
25.
go back to reference Branemark R, Ohrnell LO, Skalak R, Carlsson L, Branemark PI. Biomechanical characterization of osseointegration: an experimental in vivo investigation in the beagle dog. J Orthop Res. 1998;16(1):61–9. Epub 1998/05/09.CrossRefPubMed Branemark R, Ohrnell LO, Skalak R, Carlsson L, Branemark PI. Biomechanical characterization of osseointegration: an experimental in vivo investigation in the beagle dog. J Orthop Res. 1998;16(1):61–9. Epub 1998/05/09.CrossRefPubMed
26.
go back to reference Ersanli S, Karabuda C, Beck F, Leblebicioglu B. Resonance frequency analysis of one-stage dental implant stability during the osseointegration period. J Periodontol. 2005;76(7):1066–71. Epub 2005/07/16.CrossRefPubMed Ersanli S, Karabuda C, Beck F, Leblebicioglu B. Resonance frequency analysis of one-stage dental implant stability during the osseointegration period. J Periodontol. 2005;76(7):1066–71. Epub 2005/07/16.CrossRefPubMed
27.
go back to reference Bornstein MM, Chappuis V, von Arx T, Buser D. Performance of dental implants after staged sinus floor elevation procedures: 5-year results of a prospective study in partially edentulous patients. Clin Oral Implants Res. 2008;19(10):1034–43. Epub 2008/10/03.CrossRefPubMed Bornstein MM, Chappuis V, von Arx T, Buser D. Performance of dental implants after staged sinus floor elevation procedures: 5-year results of a prospective study in partially edentulous patients. Clin Oral Implants Res. 2008;19(10):1034–43. Epub 2008/10/03.CrossRefPubMed
28.
go back to reference Boronat Lopez A, Balaguer Martinez J, Lamas Pelayo J, Carrillo Garcia C, Penarrocha DM. Resonance frequency analysis of dental implant stability during the healing period. Med Oral Patol Oral Cir Bucal. 2008;13(4):E244–7. Epub 2008/04/02.PubMed Boronat Lopez A, Balaguer Martinez J, Lamas Pelayo J, Carrillo Garcia C, Penarrocha DM. Resonance frequency analysis of dental implant stability during the healing period. Med Oral Patol Oral Cir Bucal. 2008;13(4):E244–7. Epub 2008/04/02.PubMed
29.
go back to reference Quesada-Garcia MP, Prados-Sanchez E, Olmedo-Gaya MV, Munoz-Soto E, Gonzalez-Rodriguez MP, Valllecillo-Capilla M. Measurement of dental implant stability by resonance frequency analysis: a review of the literature. Med Oral Patol Oral Cir Bucal. 2009;14(10):e538–46. Epub 2009/08/15.CrossRefPubMed Quesada-Garcia MP, Prados-Sanchez E, Olmedo-Gaya MV, Munoz-Soto E, Gonzalez-Rodriguez MP, Valllecillo-Capilla M. Measurement of dental implant stability by resonance frequency analysis: a review of the literature. Med Oral Patol Oral Cir Bucal. 2009;14(10):e538–46. Epub 2009/08/15.CrossRefPubMed
30.
go back to reference Guler AU, Duran I, Yucel AC, Ozkan P. Effects of air-polishing powders on color stability of composite resins. J Appl Oral Sci. 2011;19(5):505–10. Epub 2011/09/17.PubMedPubMedCentral Guler AU, Duran I, Yucel AC, Ozkan P. Effects of air-polishing powders on color stability of composite resins. J Appl Oral Sci. 2011;19(5):505–10. Epub 2011/09/17.PubMedPubMedCentral
31.
go back to reference Friberg B, Sennerby L, Meredith N, Lekholm U. A comparison between cutting torque and resonance frequency measurements of maxillary implants. A 20-month clinical study. Int J Oral Maxillofac Surg. 1999;28(4):297–303. Epub 1999/07/23.CrossRefPubMed Friberg B, Sennerby L, Meredith N, Lekholm U. A comparison between cutting torque and resonance frequency measurements of maxillary implants. A 20-month clinical study. Int J Oral Maxillofac Surg. 1999;28(4):297–303. Epub 1999/07/23.CrossRefPubMed
32.
go back to reference Barewal RM, Oates TW, Meredith N, Cochran DL. Resonance frequency measurement of implant stability in vivo on implants with a sandblasted and acid-etched surface. Int J Oral Maxillofac Implants. 2003;18(5):641–51. Epub 2003/10/29.PubMed Barewal RM, Oates TW, Meredith N, Cochran DL. Resonance frequency measurement of implant stability in vivo on implants with a sandblasted and acid-etched surface. Int J Oral Maxillofac Implants. 2003;18(5):641–51. Epub 2003/10/29.PubMed
33.
go back to reference Glauser R, Sennerby L, Meredith N, Ree A, Lundgren A, Gottlow J, et al. Resonance frequency analysis of implants subjected to immediate or early functional occlusal loading. Successful vs. failing implants. Clin Oral Implants Res. 2004;15(4):428–34. Epub 2004/07/14.CrossRefPubMed Glauser R, Sennerby L, Meredith N, Ree A, Lundgren A, Gottlow J, et al. Resonance frequency analysis of implants subjected to immediate or early functional occlusal loading. Successful vs. failing implants. Clin Oral Implants Res. 2004;15(4):428–34. Epub 2004/07/14.CrossRefPubMed
34.
go back to reference Sjostrom M, Lundgren S, Nilson H, Sennerby L. Monitoring of implant stability in grafted bone using resonance frequency analysis. A clinical study from implant placement to 6 months of loading. Int J Oral Maxillofac Surg. 2005;34(1):45–51. Epub 2004/12/25.PubMed Sjostrom M, Lundgren S, Nilson H, Sennerby L. Monitoring of implant stability in grafted bone using resonance frequency analysis. A clinical study from implant placement to 6 months of loading. Int J Oral Maxillofac Surg. 2005;34(1):45–51. Epub 2004/12/25.PubMed
35.
go back to reference Gallucci GO, Belser UC, Bernard JP, Magne P. Modeling and characterization of the CEJ for optimization of esthetic implant design. Int J Periodontics Restorative Dent. 2004;24(1):19–29. Epub 2004/02/27.PubMed Gallucci GO, Belser UC, Bernard JP, Magne P. Modeling and characterization of the CEJ for optimization of esthetic implant design. Int J Periodontics Restorative Dent. 2004;24(1):19–29. Epub 2004/02/27.PubMed
36.
go back to reference Meredith N, Book K, Friberg B, Jemt T, Sennerby L. Resonance frequency measurements of implant stability in vivo. A cross-sectional and longitudinal study of resonance frequency measurements on implants in the edentulous and partially dentate maxilla. Clin Oral Implants Res. 1997;8(3):226–33. Epub 1997/06/01.CrossRefPubMed Meredith N, Book K, Friberg B, Jemt T, Sennerby L. Resonance frequency measurements of implant stability in vivo. A cross-sectional and longitudinal study of resonance frequency measurements on implants in the edentulous and partially dentate maxilla. Clin Oral Implants Res. 1997;8(3):226–33. Epub 1997/06/01.CrossRefPubMed
37.
go back to reference Meredith N, Alleyne D, Cawley P. Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis. Clin Oral Implants Res. 1996;7(3):261–7. Epub 1996/09/01.CrossRefPubMed Meredith N, Alleyne D, Cawley P. Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis. Clin Oral Implants Res. 1996;7(3):261–7. Epub 1996/09/01.CrossRefPubMed
38.
go back to reference Meredith N, Shagaldi F, Alleyne D, Sennerby L, Cawley P. The application of resonance frequency measurements to study the stability of titanium implants during healing in the rabbit tibia. Clin Oral Implants Res. 1997;8(3):234–43. Epub 1997/06/01.CrossRefPubMed Meredith N, Shagaldi F, Alleyne D, Sennerby L, Cawley P. The application of resonance frequency measurements to study the stability of titanium implants during healing in the rabbit tibia. Clin Oral Implants Res. 1997;8(3):234–43. Epub 1997/06/01.CrossRefPubMed
39.
go back to reference Steigenga J, Al-Shammari K, Misch C, Nociti Jr FH, Wang HL. Effects of implant thread geometry on percentage of osseointegration and resistance to reverse torque in the tibia of rabbits. J Periodontol. 2004;75(9):1233–41. Epub 2004/11/02.CrossRefPubMed Steigenga J, Al-Shammari K, Misch C, Nociti Jr FH, Wang HL. Effects of implant thread geometry on percentage of osseointegration and resistance to reverse torque in the tibia of rabbits. J Periodontol. 2004;75(9):1233–41. Epub 2004/11/02.CrossRefPubMed
40.
go back to reference Le Guehennec L, Goyenvalle E, Lopez-Heredia MA, Weiss P, Amouriq Y, Layrolle P. Histomorphometric analysis of the osseointegration of four different implant surfaces in the femoral epiphyses of rabbits. Clin Oral Implants Res. 2008;19(11):1103–10. Epub 2008/11/06.CrossRefPubMed Le Guehennec L, Goyenvalle E, Lopez-Heredia MA, Weiss P, Amouriq Y, Layrolle P. Histomorphometric analysis of the osseointegration of four different implant surfaces in the femoral epiphyses of rabbits. Clin Oral Implants Res. 2008;19(11):1103–10. Epub 2008/11/06.CrossRefPubMed
41.
go back to reference Faeda RS, Tavares HS, Sartori R, Guastaldi AC, Marcantonio Jr E. Evaluation of titanium implants with surface modification by laser beam. Biomechanical study in rabbit tibias. Braz Oral Res. 2009;23(2):137–43. Epub 2009/08/18.CrossRefPubMed Faeda RS, Tavares HS, Sartori R, Guastaldi AC, Marcantonio Jr E. Evaluation of titanium implants with surface modification by laser beam. Biomechanical study in rabbit tibias. Braz Oral Res. 2009;23(2):137–43. Epub 2009/08/18.CrossRefPubMed
42.
go back to reference Park JW, Kim HK, Kim YJ, An CH, Hanawa T. Enhanced osteoconductivity of micro-structured titanium implants (XiVE S CELLplus) by addition of surface calcium chemistry: a histomorphometric study in the rabbit femur. Clin Oral Implants Res. 2009;20(7):684–90. Epub 2009/06/06.CrossRefPubMed Park JW, Kim HK, Kim YJ, An CH, Hanawa T. Enhanced osteoconductivity of micro-structured titanium implants (XiVE S CELLplus) by addition of surface calcium chemistry: a histomorphometric study in the rabbit femur. Clin Oral Implants Res. 2009;20(7):684–90. Epub 2009/06/06.CrossRefPubMed
43.
go back to reference Yang GL, He FM, Hu JA, Wang XX, Zhao SF. Effects of biomimetically and electrochemically deposited nano-hydroxyapatite coatings on osseointegration of porous titanium implants. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009;107(6):782–9. Epub 2009/02/10.CrossRefPubMed Yang GL, He FM, Hu JA, Wang XX, Zhao SF. Effects of biomimetically and electrochemically deposited nano-hydroxyapatite coatings on osseointegration of porous titanium implants. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009;107(6):782–9. Epub 2009/02/10.CrossRefPubMed
44.
go back to reference Yildiz A, Esen E, Kurkcu M, Damlar I, Daglioglu K, Akova T. Effect of zoledronic acid on osseointegration of titanium implants: an experimental study in an ovariectomized rabbit model. J Oral Maxillofac Surg. 2010;68(3):515–23. Epub 2010/02/23.CrossRefPubMed Yildiz A, Esen E, Kurkcu M, Damlar I, Daglioglu K, Akova T. Effect of zoledronic acid on osseointegration of titanium implants: an experimental study in an ovariectomized rabbit model. J Oral Maxillofac Surg. 2010;68(3):515–23. Epub 2010/02/23.CrossRefPubMed
45.
go back to reference Barros RR, Novaes Jr AB, Muglia VA, Iezzi G, Piattelli A. Influence of interimplant distances and placement depth on peri-implant bone remodeling of adjacent and immediately loaded Morse cone connection implants: a histomorphometric study in dogs. Clin Oral Implants Res. 2010;21(4):371–8. Epub 2010/02/05.CrossRefPubMed Barros RR, Novaes Jr AB, Muglia VA, Iezzi G, Piattelli A. Influence of interimplant distances and placement depth on peri-implant bone remodeling of adjacent and immediately loaded Morse cone connection implants: a histomorphometric study in dogs. Clin Oral Implants Res. 2010;21(4):371–8. Epub 2010/02/05.CrossRefPubMed
46.
go back to reference Marin C, Bonfante EA, Granato R, Suzuki M, Granjeiro JM, Coelho PG. The effect of alterations on resorbable blasting media processed implant surfaces on early bone healing: a study in rabbits. Implant Dent. 2011;20(2):167–77. Epub 2011/03/31.CrossRefPubMed Marin C, Bonfante EA, Granato R, Suzuki M, Granjeiro JM, Coelho PG. The effect of alterations on resorbable blasting media processed implant surfaces on early bone healing: a study in rabbits. Implant Dent. 2011;20(2):167–77. Epub 2011/03/31.CrossRefPubMed
47.
go back to reference Green CJ, Knight J, Precious S, Simpkin S. Ketamine alone and combined with diazepam or xylazine in laboratory animals: a 10 year experience. Lab Anim. 1981;15(2):163–70. Epub 1981/04/01.CrossRefPubMed Green CJ, Knight J, Precious S, Simpkin S. Ketamine alone and combined with diazepam or xylazine in laboratory animals: a 10 year experience. Lab Anim. 1981;15(2):163–70. Epub 1981/04/01.CrossRefPubMed
48.
go back to reference Chen J, Zhang Y, Rong M, Zhao L, Jiang L, Zhang D, et al. Expression and characterization of jingzhaotoxin-34, a novel neurotoxin from the venom of the tarantula Chilobrachys jingzhao. Peptides. 2009;30(6):1042–8. Epub 2009/05/26.CrossRefPubMed Chen J, Zhang Y, Rong M, Zhao L, Jiang L, Zhang D, et al. Expression and characterization of jingzhaotoxin-34, a novel neurotoxin from the venom of the tarantula Chilobrachys jingzhao. Peptides. 2009;30(6):1042–8. Epub 2009/05/26.CrossRefPubMed
49.
go back to reference Morris HF, Ochi S, Orenstein IH, Petrazzuolo V. AICRG, Part V: factors influencing implant stability at placement and their influence on survival of Ankylos implants. J Oral Implantol. 2004;30(3):162–70. Epub 2004/07/17.CrossRefPubMed Morris HF, Ochi S, Orenstein IH, Petrazzuolo V. AICRG, Part V: factors influencing implant stability at placement and their influence on survival of Ankylos implants. J Oral Implantol. 2004;30(3):162–70. Epub 2004/07/17.CrossRefPubMed
50.
go back to reference Manresa C, Bosch M, Echeverria JJ. The comparison between implant stability quotient and bone-implant contact revisited: an experiment in Beagle dog. Clin Oral Implants Res. 2014;25(11):1213–21. Epub 2013/10/10.CrossRefPubMed Manresa C, Bosch M, Echeverria JJ. The comparison between implant stability quotient and bone-implant contact revisited: an experiment in Beagle dog. Clin Oral Implants Res. 2014;25(11):1213–21. Epub 2013/10/10.CrossRefPubMed
51.
go back to reference Alsaadi G, Quirynen M, Michiels K, Jacobs R, van Steenberghe D. A biomechanical assessment of the relation between the oral implant stability at insertion and subjective bone quality assessment. J Clin Periodontol. 2007;34(4):359–66. Epub 2007/03/24.CrossRefPubMed Alsaadi G, Quirynen M, Michiels K, Jacobs R, van Steenberghe D. A biomechanical assessment of the relation between the oral implant stability at insertion and subjective bone quality assessment. J Clin Periodontol. 2007;34(4):359–66. Epub 2007/03/24.CrossRefPubMed
52.
go back to reference Huwiler MA, Pjetursson BE, Bosshardt DD, Salvi GE, Lang NP. Resonance frequency analysis in relation to jawbone characteristics and during early healing of implant installation. Clin Oral Implants Res. 2007;18(3):275–80. Epub 2007/03/16.CrossRefPubMed Huwiler MA, Pjetursson BE, Bosshardt DD, Salvi GE, Lang NP. Resonance frequency analysis in relation to jawbone characteristics and during early healing of implant installation. Clin Oral Implants Res. 2007;18(3):275–80. Epub 2007/03/16.CrossRefPubMed
53.
go back to reference Zhou Y, Jiang T, Qian M, Zhang X, Wang J, Shi B, et al. Roles of bone scintigraphy and resonance frequency analysis in evaluating osseointegration of endosseous implant. Biomaterials. 2008;29(4):461–74. Epub 2007/11/07.CrossRefPubMed Zhou Y, Jiang T, Qian M, Zhang X, Wang J, Shi B, et al. Roles of bone scintigraphy and resonance frequency analysis in evaluating osseointegration of endosseous implant. Biomaterials. 2008;29(4):461–74. Epub 2007/11/07.CrossRefPubMed
54.
go back to reference Degidi M, Perrotti V, Piattelli A, Iezzi G. Mineralized bone-implant contact and implant stability quotient in 16 human implants retrieved after early healing periods: a histologic and histomorphometric evaluation. Int J Oral Maxillofac Implants. 2010;25(1):45–8. Epub 2010/03/09.PubMed Degidi M, Perrotti V, Piattelli A, Iezzi G. Mineralized bone-implant contact and implant stability quotient in 16 human implants retrieved after early healing periods: a histologic and histomorphometric evaluation. Int J Oral Maxillofac Implants. 2010;25(1):45–8. Epub 2010/03/09.PubMed
55.
go back to reference Dhaliwal J, Albuquerque R, Murshed, M, Tamimi, F, Feine, JS. A histomorphometric comparison of osseointegration with MDIs and standard implants. IADR/AADR/CADR 91st General Session, Seattle, Washington, USA, March 20–23, 2013 Dhaliwal J, Albuquerque R, Murshed, M, Tamimi, F, Feine, JS. A histomorphometric comparison of osseointegration with MDIs and standard implants. IADR/AADR/CADR 91st General Session, Seattle, Washington, USA, March 20–23, 2013
56.
go back to reference Zix J, Hug S, Kessler-Liechti G, Mericske-Stern R. Measurement of dental implant stability by resonance frequency analysis and damping capacity assessment: comparison of both techniques in a clinical trial. Int J Oral Maxillofac Implants. 2008;23(3):525–30. Epub 2008/08/15.PubMed Zix J, Hug S, Kessler-Liechti G, Mericske-Stern R. Measurement of dental implant stability by resonance frequency analysis and damping capacity assessment: comparison of both techniques in a clinical trial. Int J Oral Maxillofac Implants. 2008;23(3):525–30. Epub 2008/08/15.PubMed
57.
go back to reference Roberts WE, Smith RK, Zilberman Y, Mozsary PG, Smith RS. Osseous adaptation to continuous loading of rigid endosseous implants. Am J Orthod. 1984;86(2):95–111. Epub 1984/08/01.CrossRefPubMed Roberts WE, Smith RK, Zilberman Y, Mozsary PG, Smith RS. Osseous adaptation to continuous loading of rigid endosseous implants. Am J Orthod. 1984;86(2):95–111. Epub 1984/08/01.CrossRefPubMed
58.
go back to reference Ostman PO, Hellman M, Albrektsson T, Sennerby L. Direct loading of nobel direct and nobel perfect one-piece implants: a 1-year prospective clinical and radiographic study. Clin Oral Implants Res. 2007;18(4):409–18. Epub 2007/05/16.CrossRefPubMed Ostman PO, Hellman M, Albrektsson T, Sennerby L. Direct loading of nobel direct and nobel perfect one-piece implants: a 1-year prospective clinical and radiographic study. Clin Oral Implants Res. 2007;18(4):409–18. Epub 2007/05/16.CrossRefPubMed
Metadata
Title
Customized SmartPeg for measurement of resonance frequency of mini dental implants
Authors
Jagjit Singh Dhaliwal
Rubens F. Albuquerque Jr
Ali Fakhry
Sukhbir Kaur
Jocelyne S. Feine
Publication date
01-12-2017
Publisher
Springer Berlin Heidelberg
Published in
International Journal of Implant Dentistry / Issue 1/2017
Electronic ISSN: 2198-4034
DOI
https://doi.org/10.1186/s40729-017-0066-6

Other articles of this Issue 1/2017

International Journal of Implant Dentistry 1/2017 Go to the issue