Validation of transpulmonary thermodilution variables in hemodynamically stable patients with heart diseases

Authors: Matthias Peter Hilty, Daniel Peter Franzen, Christophe Wyss, Patric Biaggi, Marco Maggiorini

Published in: Annals of Intensive Care | Issue 1/2017

Login to get access

Abstract

Background

Transpulmonary thermodilution is recommended in the treatment of critically ill patients presenting with complex shock. However, so far it has not been validated in hemodynamically stable patients with heart disease.

Methods

We assessed the validity of cardiac output, global end-diastolic volume index (GEDVI), an established marker of preload thought to reflect the volume of all four heart chambers, global ejection fraction (GEF) and cardiac function index (CFI) as variables of cardiac function, and extravascular lung water index (EVLWI) as indicator of pulmonary edema in 29 patients undergoing elective left and right heart catheterization including left ventricular angiography with stable coronary heart disease and normal cardiac function (controls, n = 11), moderate-to-severe aortic valve stenosis (AS, n = 10), or dilated cardiomyopathy (DCM, n = 8).

Results

Cardiac output was similar in controls, AS, and DCM, with good correlation between transpulmonary thermodilution and pulmonary artery catheter using the Fick method (r = 0.69, p < 0.0001). Left ventricular end-diastolic volume was normal in controls and AS, but significantly higher in DCM (104 ± 37 vs 135 ± 63 vs 234 ± 24 ml, p < 0.01). GEDVI did not differentiate between patients with normal and patients with enlarged left ventricular end-diastolic volume (848 ± 128 vs 882 ± 213 ml m−2, p = 0.60). No difference in GEF and CFI was found between patients with normal and patients with reduced left ventricular ejection fraction. Patients with AS but not DCM had higher EVLWI than controls (9 ± 2 vs 12 ± 4 vs 11 ± 3 ml kg−1, p = 0.04), while there was only a trend in pulmonary artery occlusion pressure (8 ± 3 vs 10 ± 5 vs 14 ± 7 mmHg, p = 0.05).

Conclusions

Cardiac output measurement by transpulmonary thermodilution is unaffected by differences in ventricular size and outflow obstruction. However, GEDVI did not identify markedly enlarged left ventricular end-diastolic volumes, and neither GEF nor CFI reflected the increased heart chamber volumes and markedly impaired left ventricular function in patients with DCM. In contrast, EVLWI is probably a sensitive marker of subclinical pulmonary edema particularly in patients with elevated left-ventricular-filling pressure irrespective of differences in left ventricular function.
Literature
1.
go back to reference Teboul J-L, Saugel B, Cecconi M, De Backer D, Hofer CK, Monnet X, Perel A, Pinsky MR, Reuter DA, Rhodes A, Squara P, Vincent J-L, Scheeren TW. Less invasive hemodynamic monitoring in critically ill patients. Intensive Care Med. 2016;42:1350–9.CrossRefPubMed Teboul J-L, Saugel B, Cecconi M, De Backer D, Hofer CK, Monnet X, Perel A, Pinsky MR, Reuter DA, Rhodes A, Squara P, Vincent J-L, Scheeren TW. Less invasive hemodynamic monitoring in critically ill patients. Intensive Care Med. 2016;42:1350–9.CrossRefPubMed
2.
go back to reference Cecconi M, De Backer D, Antonelli M, Beale R, Bakker J, Hofer C, Jaeschke R, Mebazaa A, Pinsky MR, Teboul JL, Vincent JL, Rhodes A. Consensus on circulatory shock and hemodynamic monitoring. Task force of the European Society of Intensive Care Medicine. Intensive Care Med. 2014;40:1795–815.CrossRefPubMedPubMedCentral Cecconi M, De Backer D, Antonelli M, Beale R, Bakker J, Hofer C, Jaeschke R, Mebazaa A, Pinsky MR, Teboul JL, Vincent JL, Rhodes A. Consensus on circulatory shock and hemodynamic monitoring. Task force of the European Society of Intensive Care Medicine. Intensive Care Med. 2014;40:1795–815.CrossRefPubMedPubMedCentral
3.
go back to reference Goedje O, Seebauer T, Peyerl M, Pfeiffer UJ, Reichart B. Hemodynamic monitoring by double-indicator dilution technique in patients after orthotopic heart transplantation. Chest. 2000;118:775–81.CrossRefPubMed Goedje O, Seebauer T, Peyerl M, Pfeiffer UJ, Reichart B. Hemodynamic monitoring by double-indicator dilution technique in patients after orthotopic heart transplantation. Chest. 2000;118:775–81.CrossRefPubMed
4.
go back to reference Michard F, Alaya S, Zarka V, Bahloul M, Richard C, Teboul J-L. Global end-diastolic volume as an indicator of cardiac preload in patients with septic shock. Chest. 2003;124:1900–8.CrossRefPubMed Michard F, Alaya S, Zarka V, Bahloul M, Richard C, Teboul J-L. Global end-diastolic volume as an indicator of cardiac preload in patients with septic shock. Chest. 2003;124:1900–8.CrossRefPubMed
5.
go back to reference Monnet X, Anguel N, Osman D, Hamzaoui O, Richard C, Teboul J-L. Assessing pulmonary permeability by transpulmonary thermodilution allows differentiation of hydrostatic pulmonary edema from ALI/ARDS. Intensive Care Med. 2007;33:448–53.CrossRefPubMed Monnet X, Anguel N, Osman D, Hamzaoui O, Richard C, Teboul J-L. Assessing pulmonary permeability by transpulmonary thermodilution allows differentiation of hydrostatic pulmonary edema from ALI/ARDS. Intensive Care Med. 2007;33:448–53.CrossRefPubMed
6.
go back to reference Meybohm P, Gruenewald M, Renner J, Maracke M, Rossee S, Höcker J, Hagelstein S, Zacharowski K, Bein B. Assessment of left ventricular systolic function during acute myocardial ischemia: a comparison of transpulmonary thermodilution and transesophageal echocardiography. Minerva Anestesiol. 2011;77:132–41.PubMed Meybohm P, Gruenewald M, Renner J, Maracke M, Rossee S, Höcker J, Hagelstein S, Zacharowski K, Bein B. Assessment of left ventricular systolic function during acute myocardial ischemia: a comparison of transpulmonary thermodilution and transesophageal echocardiography. Minerva Anestesiol. 2011;77:132–41.PubMed
7.
go back to reference Jabot J, Monnet X, Bouchra L, Chemla D, Richard C, Teboul J-L. Cardiac function index provided by transpulmonary thermodilution behaves as an indicator of left ventricular systolic function. Crit Care Med. 2009;37:2913–8.CrossRefPubMed Jabot J, Monnet X, Bouchra L, Chemla D, Richard C, Teboul J-L. Cardiac function index provided by transpulmonary thermodilution behaves as an indicator of left ventricular systolic function. Crit Care Med. 2009;37:2913–8.CrossRefPubMed
8.
go back to reference Sakka SG, Rühl CC, Pfeiffer UJ, Beale R, McLuckie A, Reinhart K, Meier-Hellmann A. Assessment of cardiac preload and extravascular lung water by single transpulmonary thermodilution. Intensive Care Med. 2000;26:180–7.CrossRefPubMed Sakka SG, Rühl CC, Pfeiffer UJ, Beale R, McLuckie A, Reinhart K, Meier-Hellmann A. Assessment of cardiac preload and extravascular lung water by single transpulmonary thermodilution. Intensive Care Med. 2000;26:180–7.CrossRefPubMed
9.
go back to reference Hofer CK, Furrer L, Matter-Ensner S, Maloigne M, Klaghofer R, Genoni M, Zollinger A. Volumetric preload measurement by thermodilution: a comparison with transoesophageal echocardiography. Br J Anaesth. 2005;94:748–55.CrossRefPubMed Hofer CK, Furrer L, Matter-Ensner S, Maloigne M, Klaghofer R, Genoni M, Zollinger A. Volumetric preload measurement by thermodilution: a comparison with transoesophageal echocardiography. Br J Anaesth. 2005;94:748–55.CrossRefPubMed
10.
go back to reference Marik PE. Techniques for assessment of intravascular volume in critically ill patients. J Intensive Care Med. 2009;24:329–37.CrossRefPubMed Marik PE. Techniques for assessment of intravascular volume in critically ill patients. J Intensive Care Med. 2009;24:329–37.CrossRefPubMed
11.
go back to reference Ritter S, Rudiger A, Maggiorini M. Transpulmonary thermodilution-derived cardiac function index identifies cardiac dysfunction in acute heart failure and septic patients: an observational study. Crit Care Lond Engl. 2009;13:R133.CrossRef Ritter S, Rudiger A, Maggiorini M. Transpulmonary thermodilution-derived cardiac function index identifies cardiac dysfunction in acute heart failure and septic patients: an observational study. Crit Care Lond Engl. 2009;13:R133.CrossRef
12.
go back to reference Lang RM, Badano LP, Mor-Avi V, Afilalo J, Armstrong A, Ernande L, Flachskampf FA, Foster E, Goldstein SA, Kuznetsova T, Lancellotti P, Muraru D, Picard MH, Rietzschel ER, Rudski L, Spencer KT, Tsang W, Voigt J-U. Recommendations for cardiac chamber quantification by echocardiography in adults: an update from the American Society of Echocardiography and the European Association of Cardiovascular Imaging. J Am Soc Echocardiogr. 2015;28(1–39):e14. Lang RM, Badano LP, Mor-Avi V, Afilalo J, Armstrong A, Ernande L, Flachskampf FA, Foster E, Goldstein SA, Kuznetsova T, Lancellotti P, Muraru D, Picard MH, Rietzschel ER, Rudski L, Spencer KT, Tsang W, Voigt J-U. Recommendations for cardiac chamber quantification by echocardiography in adults: an update from the American Society of Echocardiography and the European Association of Cardiovascular Imaging. J Am Soc Echocardiogr. 2015;28(1–39):e14.
13.
go back to reference Nagueh SF, Appleton CP, Gillebert TC, Marino PN, Oh JK, Smiseth OA, Waggoner AD, Flachskampf FA, Pellikka PA, Evangelista A. Recommendations for the evaluation of left ventricular diastolic function by echocardiography. J Am Soc Echocardiogr. 2009;22:107–33.CrossRefPubMed Nagueh SF, Appleton CP, Gillebert TC, Marino PN, Oh JK, Smiseth OA, Waggoner AD, Flachskampf FA, Pellikka PA, Evangelista A. Recommendations for the evaluation of left ventricular diastolic function by echocardiography. J Am Soc Echocardiogr. 2009;22:107–33.CrossRefPubMed
14.
go back to reference Levey AS, Stevens LA, Schmid CH, Zhang YL, Castro AF, Feldman HI, Kusek JW, Eggers P, Van Lente F, Greene T, Coresh J. CKD-EPI (Chronic Kidney Disease Epidemiology Collaboration). A new equation to estimate glomerular filtration rate. Ann Intern Med. 2009;150:604–12.CrossRefPubMedPubMedCentral Levey AS, Stevens LA, Schmid CH, Zhang YL, Castro AF, Feldman HI, Kusek JW, Eggers P, Van Lente F, Greene T, Coresh J. CKD-EPI (Chronic Kidney Disease Epidemiology Collaboration). A new equation to estimate glomerular filtration rate. Ann Intern Med. 2009;150:604–12.CrossRefPubMedPubMedCentral
15.
16.
go back to reference Prontera C, Emdin M, Zucchelli GC, Ripoli A, Passino C, Clerico A. Analytical performance and diagnostic accuracy of a fully-automated electrochemiluminescent assay for the N-terminal fragment of the pro-peptide of brain natriuretic peptide in patients with cardiomyopathy: comparison with immunoradiometric assay methods for brain natriuretic peptide and atrial natriuretic peptide. Clin Chem Lab Med. 2004;42:37–44.CrossRefPubMed Prontera C, Emdin M, Zucchelli GC, Ripoli A, Passino C, Clerico A. Analytical performance and diagnostic accuracy of a fully-automated electrochemiluminescent assay for the N-terminal fragment of the pro-peptide of brain natriuretic peptide in patients with cardiomyopathy: comparison with immunoradiometric assay methods for brain natriuretic peptide and atrial natriuretic peptide. Clin Chem Lab Med. 2004;42:37–44.CrossRefPubMed
17.
go back to reference Baumgartner H, Hung J, Bermejo J, Chambers JB, Evangelista A, Griffin BP, Iung B, Otto CM, Pellikka PA, Quiñones M, American Society of Echocardiography, European Association of Echocardiography. Echocardiographic assessment of valve stenosis: EAE/ASE recommendations for clinical practice. J Am Soc Echocardiogr. 2009;22:1–23 (quiz 101–102).CrossRefPubMed Baumgartner H, Hung J, Bermejo J, Chambers JB, Evangelista A, Griffin BP, Iung B, Otto CM, Pellikka PA, Quiñones M, American Society of Echocardiography, European Association of Echocardiography. Echocardiographic assessment of valve stenosis: EAE/ASE recommendations for clinical practice. J Am Soc Echocardiogr. 2009;22:1–23 (quiz 101–102).CrossRefPubMed
18.
go back to reference Elliott P, Andersson B, Arbustini E, Bilinska Z, Cecchi F, Charron P, Dubourg O, Kühl U, Maisch B, McKenna WJ, Monserrat L, Pankuweit S, Rapezzi C, Seferovic P, Tavazzi L, Keren A. Classification of the cardiomyopathies: a position statement from the European Society Of Cardiology Working Group on Myocardial and Pericardial Diseases. Eur Heart J. 2008;29:270–6.CrossRefPubMed Elliott P, Andersson B, Arbustini E, Bilinska Z, Cecchi F, Charron P, Dubourg O, Kühl U, Maisch B, McKenna WJ, Monserrat L, Pankuweit S, Rapezzi C, Seferovic P, Tavazzi L, Keren A. Classification of the cardiomyopathies: a position statement from the European Society Of Cardiology Working Group on Myocardial and Pericardial Diseases. Eur Heart J. 2008;29:270–6.CrossRefPubMed
19.
go back to reference Benjamini Y, Hochberg Y. Controlling the false discovery rate: a practical and powerful approach to multiple testing. J R Stat Soc Ser B Methodol. 1995;57:289–300. Benjamini Y, Hochberg Y. Controlling the false discovery rate: a practical and powerful approach to multiple testing. J R Stat Soc Ser B Methodol. 1995;57:289–300.
20.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet Lond Engl. 1986;1:307–10.CrossRef Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet Lond Engl. 1986;1:307–10.CrossRef
21.
go back to reference Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999;15:85–91.CrossRefPubMed Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999;15:85–91.CrossRefPubMed
22.
23.
go back to reference Wickham H. ggplot2: elegant graphics for data analysis [Internet]. 1st ed. 2009. Corr. 3rd printing 2010 edition. New York: Springer; 2010. http://ggplot2.org. Wickham H. ggplot2: elegant graphics for data analysis [Internet]. 1st ed. 2009. Corr. 3rd printing 2010 edition. New York: Springer; 2010. http://​ggplot2.​org.
24.
go back to reference Hamilton WF, Walker Moore J, Kinsman JM, Spurling RG. Simultaneous determination of the pulmonary and systemic circulation times in man and of a figure related to the cardiac output. Am J Physiol. 1928;84:338–44. Hamilton WF, Walker Moore J, Kinsman JM, Spurling RG. Simultaneous determination of the pulmonary and systemic circulation times in man and of a figure related to the cardiac output. Am J Physiol. 1928;84:338–44.
25.
go back to reference Rasmussen P, Widmer M, Hilty MP, Hug M, Sørensen H, Ogoh S, Sato K, Secher NH, Maggiorini M, Lundby C. Thermodilution-determined Internal Jugular Venous Flow. Med Sci Sports Exerc. 2016. Rasmussen P, Widmer M, Hilty MP, Hug M, Sørensen H, Ogoh S, Sato K, Secher NH, Maggiorini M, Lundby C. Thermodilution-determined Internal Jugular Venous Flow. Med Sci Sports Exerc. 2016.
26.
go back to reference Forrester JS, Ganz W, Diamond G, McHugh T, Chonette DW, Swan HJ. Thermodilution cardiac output determination with a single flow-directed catheter. Am Heart J. 1972;83:306–11.CrossRefPubMed Forrester JS, Ganz W, Diamond G, McHugh T, Chonette DW, Swan HJ. Thermodilution cardiac output determination with a single flow-directed catheter. Am Heart J. 1972;83:306–11.CrossRefPubMed
27.
go back to reference Della Rocca G, Costa MG, Pompei L, Coccia C, Pietropaoli P. Continuous and intermittent cardiac output measurement: pulmonary artery catheter versus aortic transpulmonary technique. Br J Anaesth. 2002;88:350–6.CrossRefPubMed Della Rocca G, Costa MG, Pompei L, Coccia C, Pietropaoli P. Continuous and intermittent cardiac output measurement: pulmonary artery catheter versus aortic transpulmonary technique. Br J Anaesth. 2002;88:350–6.CrossRefPubMed
28.
go back to reference Goedje O, Hoeke K, Lichtwarck-Aschoff M, Faltchauser A, Lamm P, Reichart B. Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: comparison with pulmonary arterial thermodilution. Crit Care Med. 1999;27:2407–12.CrossRefPubMed Goedje O, Hoeke K, Lichtwarck-Aschoff M, Faltchauser A, Lamm P, Reichart B. Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: comparison with pulmonary arterial thermodilution. Crit Care Med. 1999;27:2407–12.CrossRefPubMed
29.
go back to reference Staier K, Wilhelm M, Wiesenack C, Thoma M, Keyl C. Pulmonary artery vs. transpulmonary thermodilution for the assessment of cardiac output in mitral regurgitation: a prospective observational study. Eur J Anaesthesiol. 2012;29:431–7.CrossRefPubMed Staier K, Wilhelm M, Wiesenack C, Thoma M, Keyl C. Pulmonary artery vs. transpulmonary thermodilution for the assessment of cardiac output in mitral regurgitation: a prospective observational study. Eur J Anaesthesiol. 2012;29:431–7.CrossRefPubMed
30.
go back to reference Petzoldt M, Riedel C, Braeunig J, Haas S, Goepfert MS, Treede H, Baldus S, Goetz AE, Reuter DA. Stroke volume determination using transcardiopulmonary thermodilution and arterial pulse contour analysis in severe aortic valve disease. Intensive Care Med. 2013;39:601–11.CrossRefPubMed Petzoldt M, Riedel C, Braeunig J, Haas S, Goepfert MS, Treede H, Baldus S, Goetz AE, Reuter DA. Stroke volume determination using transcardiopulmonary thermodilution and arterial pulse contour analysis in severe aortic valve disease. Intensive Care Med. 2013;39:601–11.CrossRefPubMed
31.
go back to reference Newman EV, Merrell M, Genecin A, Monge C, Milnor WR, McKeever WP. The dye dilution method for describing the central circulation. An analysis of factors shaping the time-concentration curves. Circulation. 1951;4:735–46.CrossRefPubMed Newman EV, Merrell M, Genecin A, Monge C, Milnor WR, McKeever WP. The dye dilution method for describing the central circulation. An analysis of factors shaping the time-concentration curves. Circulation. 1951;4:735–46.CrossRefPubMed
32.
go back to reference Effros RM. Lung water measurements with the mean transit time approach. J Appl Physiol Bethesda Md. 1985;1985(59):673–83. Effros RM. Lung water measurements with the mean transit time approach. J Appl Physiol Bethesda Md. 1985;1985(59):673–83.
33.
go back to reference Hofer CK, Ganter MT, Rist A, Klaghofer R, Matter-Ensner S, Zollinger A. The accuracy of preload assessment by different transesophageal echocardiographic techniques in patients undergoing cardiac surgery. J Cardiothorac Vasc Anesth. 2008;22:236–42.CrossRefPubMed Hofer CK, Ganter MT, Rist A, Klaghofer R, Matter-Ensner S, Zollinger A. The accuracy of preload assessment by different transesophageal echocardiographic techniques in patients undergoing cardiac surgery. J Cardiothorac Vasc Anesth. 2008;22:236–42.CrossRefPubMed
34.
go back to reference Wolf S, Riess A, Landscheidt JF, Lumenta CB, Friederich P, Schürer L. Global end-diastolic volume acquired by transpulmonary thermodilution depends on age and gender in awake and spontaneously breathing patients. Crit Care Lond Engl. 2009;13:R202.CrossRef Wolf S, Riess A, Landscheidt JF, Lumenta CB, Friederich P, Schürer L. Global end-diastolic volume acquired by transpulmonary thermodilution depends on age and gender in awake and spontaneously breathing patients. Crit Care Lond Engl. 2009;13:R202.CrossRef
35.
go back to reference Wisner-Euteneier A, Lichtwarck-Aschoff M, Zimmermann G, Bluemel G, Pfeiffer U. Evaluation of the cardiac function index as a new bedside indicator of cardiac performance. Clin Intensive Care. 1994;5(suppl):28. Wisner-Euteneier A, Lichtwarck-Aschoff M, Zimmermann G, Bluemel G, Pfeiffer U. Evaluation of the cardiac function index as a new bedside indicator of cardiac performance. Clin Intensive Care. 1994;5(suppl):28.
36.
go back to reference Trepte CJC, Eichhorn V, Haas SA, Richter HP, Goepfert MS, Kubitz JC, Goetz AE, Reuter DA. Thermodilution-derived indices for assessment of left and right ventricular cardiac function in normal and impaired cardiac function. Crit Care Med. 2011;39:2106–12.CrossRefPubMed Trepte CJC, Eichhorn V, Haas SA, Richter HP, Goepfert MS, Kubitz JC, Goetz AE, Reuter DA. Thermodilution-derived indices for assessment of left and right ventricular cardiac function in normal and impaired cardiac function. Crit Care Med. 2011;39:2106–12.CrossRefPubMed
37.
go back to reference Combes A, Berneau J-B, Luyt C-E, Trouillet J-L. Estimation of left ventricular systolic function by single transpulmonary thermodilution. Intensive Care Med. 2004;30:1377–83.PubMed Combes A, Berneau J-B, Luyt C-E, Trouillet J-L. Estimation of left ventricular systolic function by single transpulmonary thermodilution. Intensive Care Med. 2004;30:1377–83.PubMed
38.
go back to reference Mihm FG, Feeley TW, Jamieson SW. Thermal dye double indicator dilution measurement of lung water in man: comparison with gravimetric measurements. Thorax. 1987;42:72–6.CrossRefPubMedPubMedCentral Mihm FG, Feeley TW, Jamieson SW. Thermal dye double indicator dilution measurement of lung water in man: comparison with gravimetric measurements. Thorax. 1987;42:72–6.CrossRefPubMedPubMedCentral
39.
go back to reference Nirmalan M, Willard TM, Edwards DJ, Little RA, Dark PM. Estimation of errors in determining intrathoracic blood volume using the single transpulmonary thermal dilution technique in hypovolemic shock. Anesthesiology. 2005;103:805–12.CrossRefPubMed Nirmalan M, Willard TM, Edwards DJ, Little RA, Dark PM. Estimation of errors in determining intrathoracic blood volume using the single transpulmonary thermal dilution technique in hypovolemic shock. Anesthesiology. 2005;103:805–12.CrossRefPubMed
40.
go back to reference Michard F. Bedside assessment of extravascular lung water by dilution methods: temptations and pitfalls. Crit Care Med. 2007;35:1186–92.CrossRefPubMed Michard F. Bedside assessment of extravascular lung water by dilution methods: temptations and pitfalls. Crit Care Med. 2007;35:1186–92.CrossRefPubMed
41.
go back to reference Sakka SG, Klein M, Reinhart K, Meier-Hellmann A. Prognostic value of extravascular lung water in critically ill patients. Chest. 2002;122:2080–6.CrossRefPubMed Sakka SG, Klein M, Reinhart K, Meier-Hellmann A. Prognostic value of extravascular lung water in critically ill patients. Chest. 2002;122:2080–6.CrossRefPubMed
42.
go back to reference Taylor BJ, Snyder EM, Richert ML, Wheatley CM, Chase SC, Olson LJ, Johnson BD. Effect of β2-adrenergic receptor stimulation on lung fluid in stable heart failure patients. J Heart Lung Transplant. 2016. Taylor BJ, Snyder EM, Richert ML, Wheatley CM, Chase SC, Olson LJ, Johnson BD. Effect of β2-adrenergic receptor stimulation on lung fluid in stable heart failure patients. J Heart Lung Transplant. 2016.
43.
go back to reference Bindels AJ, van der Hoeven JG, Meinders AE. Pulmonary artery wedge pressure and extravascular lung water in patients with acute cardiogenic pulmonary edema requiring mechanical ventilation. Am J Cardiol. 1999;84:1158–63.CrossRefPubMed Bindels AJ, van der Hoeven JG, Meinders AE. Pulmonary artery wedge pressure and extravascular lung water in patients with acute cardiogenic pulmonary edema requiring mechanical ventilation. Am J Cardiol. 1999;84:1158–63.CrossRefPubMed
44.
go back to reference Hilty MP, Franzen DP, Wyss C, Biaggi P, Maggiorini M. Validation of transpulmonary thermodilution variables in hemodynamically stable patients with heart diseases—individual patient data. Zenodo Data Repos. Hilty MP, Franzen DP, Wyss C, Biaggi P, Maggiorini M. Validation of transpulmonary thermodilution variables in hemodynamically stable patients with heart diseases—individual patient data. Zenodo Data Repos.
Metadata
Title
Validation of transpulmonary thermodilution variables in hemodynamically stable patients with heart diseases
Authors
Matthias Peter Hilty
Daniel Peter Franzen
Christophe Wyss
Patric Biaggi
Marco Maggiorini
Publication date
01-12-2017
Publisher
Springer International Publishing
Published in
Annals of Intensive Care / Issue 1/2017
Electronic ISSN: 2110-5820
DOI
https://doi.org/10.1186/s13613-017-0307-0

Other articles of this Issue 1/2017

Annals of Intensive Care 1/2017 Go to the issue