Factors affecting the stability and repeatability of gamma camera calibration for quantitative imaging applications based on a retrospective review of clinical data

Authors: Nadège Anizan, Hao Wang, Xian C Zhou, Robert F Hobbs, Richard L Wahl, Eric C Frey

Published in: EJNMMI Research | Issue 1/2014

Login to get access

Abstract

Background

Absolute quantitative single-photon emission computed tomography (SPECT) has several important applications including monitoring tumor response after treatment and dose estimation for targeted radionuclide therapy treatment planning. Obtaining quantitative SPECT images in absolute activity units requires the use of a calibration factor, and the repeatability of this directly affects the repeatability of image quantification. This study focused on evaluating the factors affecting the repeatability of a calibration factor measured using a planar image of an in-air calibration source.

Methods

The calibration factors calculated as part of 131I-tositumomab patient dosimetry scans used in treatment planning performed over a 4-year period were retrospectively analyzed. Raw data included total counts in whole-body images of a radioactive calibration source, the activity of the source measured in a radionuclide activity meter (often referred to as a dose calibrator), and the background count rate obtained at three time points for each patient. The count rate from extrinsic flood source acquisitions and radionuclide activity meter constancy obtained on the same day as each image were also used. The data were analyzed statistically using a mixed-effects model to determine the factors affecting variations in the measured calibration factors.

Results

The global variability in the calibration factor was equal to 2.3% and was decreased by 20% to 1.8%, when the decay-corrected measurements of calibration source activity were averaged over the three time points for each patient. Camera sensitivity variation measured using a 57Co sheet source was small and had a weak relationship to calibration factor variations. When the averaged source activity was used, the main source of variance was related to preparation and measurement of the source (77%). Radionuclide activity meter constancy had a smaller but statistically significant impact on the calibration factor.

Conclusions

This study indicates that calibration factors based on planar measurements have good reproducibility. The findings of this study indicate (1) the importance of accurate and precise preparation and measurement of the calibration source activity, (2) the need to carefully control background activity during calibration factor assessment and patient data acquisition, and (3) that the calibration factor and camera sensitivity were stable over time, indicating that careful but less frequent calibration is needed.
Appendix
Available only for authorised users
Literature
1.
go back to reference Wahl RL, Kroll S, Zasadny KR: Patient-specific whole-body dosimetry: principles and a simplified method for clinical implementation. J Nucl Med 1998, 39(8 Suppl):14S-20S.PubMed Wahl RL, Kroll S, Zasadny KR: Patient-specific whole-body dosimetry: principles and a simplified method for clinical implementation. J Nucl Med 1998, 39(8 Suppl):14S-20S.PubMed
2.
go back to reference Prideaux AR, Song H, Hobbs RF, He B, Frey EC, Ladenson PW, Wahl RL, Sgouros G: Three-dimensional radiobiologic dosimetry: application of radiobiologic modeling to patient-specific 3-dimensional imaging-based internal dosimetry. J Nucl Med 2007, 48: 1008–1016. 10.2967/jnumed.106.038000PubMedCentralCrossRefPubMed Prideaux AR, Song H, Hobbs RF, He B, Frey EC, Ladenson PW, Wahl RL, Sgouros G: Three-dimensional radiobiologic dosimetry: application of radiobiologic modeling to patient-specific 3-dimensional imaging-based internal dosimetry. J Nucl Med 2007, 48: 1008–1016. 10.2967/jnumed.106.038000PubMedCentralCrossRefPubMed
3.
go back to reference Wahl RL, Jacene H, Kasamon Y, Lodge MA: From RECIST to PERCIST: evolving considerations for PET response criteria in solid tumors. J Nucl Med 2009, 50: 122S-150S. 10.2967/jnumed.108.057307PubMedCentralCrossRefPubMed Wahl RL, Jacene H, Kasamon Y, Lodge MA: From RECIST to PERCIST: evolving considerations for PET response criteria in solid tumors. J Nucl Med 2009, 50: 122S-150S. 10.2967/jnumed.108.057307PubMedCentralCrossRefPubMed
4.
go back to reference He B, Frey EC: Comparison of conventional, model-based quantitative planar, and quantitative SPECT image processing methods for organ activity estimation using In-111 agents. Phys Med Biol 2006, 51: 3967–3981. 10.1088/0031-9155/51/16/006CrossRefPubMed He B, Frey EC: Comparison of conventional, model-based quantitative planar, and quantitative SPECT image processing methods for organ activity estimation using In-111 agents. Phys Med Biol 2006, 51: 3967–3981. 10.1088/0031-9155/51/16/006CrossRefPubMed
5.
go back to reference Zeintl J, Vija AH, Yahil A, Hornegger J, Kuwert T: Quantitative accuracy of clinical 99mTc SPECT/CT using ordered-subset expectation maximization with 3-dimensional resolution recovery, attenuation, and scatter correction. J Nucl Med 2010, 51: 921–928. 10.2967/jnumed.109.071571CrossRefPubMed Zeintl J, Vija AH, Yahil A, Hornegger J, Kuwert T: Quantitative accuracy of clinical 99mTc SPECT/CT using ordered-subset expectation maximization with 3-dimensional resolution recovery, attenuation, and scatter correction. J Nucl Med 2010, 51: 921–928. 10.2967/jnumed.109.071571CrossRefPubMed
6.
go back to reference Dewaraja YK, Wilderman SJ, Ljungberg M, Koral KF, Zasadny K, Kaminiski MS: Accurate dosimetry in 131I radionuclide therapy using patient-specific, 3-dimensional methods for SPECT reconstruction and absorbed dose calculation. J Nucl Med 2005, 46: 840–849.PubMedCentralPubMed Dewaraja YK, Wilderman SJ, Ljungberg M, Koral KF, Zasadny K, Kaminiski MS: Accurate dosimetry in 131I radionuclide therapy using patient-specific, 3-dimensional methods for SPECT reconstruction and absorbed dose calculation. J Nucl Med 2005, 46: 840–849.PubMedCentralPubMed
7.
go back to reference National Electrical Manufacturers Association:Performance Measurements of Gamma Cameras. NEMA Standards Publication NU 1–2007. National Electrical Manufacturers Association, Rosslyn; 2007. National Electrical Manufacturers Association:Performance Measurements of Gamma Cameras. NEMA Standards Publication NU 1–2007. National Electrical Manufacturers Association, Rosslyn; 2007.
8.
go back to reference Dewaraja YK, Wilderman SJ, Koral KF, Kaminski MS, Avram AM: Use of integrated SPECT/CT imaging for tumor dosimetry in I-131 radioimmunotherapy: a pilot patient study. Cancer Biother Radiopharm 2009, 24: 417–426. 10.1089/cbr.2008.0568PubMedCentralCrossRefPubMed Dewaraja YK, Wilderman SJ, Koral KF, Kaminski MS, Avram AM: Use of integrated SPECT/CT imaging for tumor dosimetry in I-131 radioimmunotherapy: a pilot patient study. Cancer Biother Radiopharm 2009, 24: 417–426. 10.1089/cbr.2008.0568PubMedCentralCrossRefPubMed
9.
go back to reference Ljungberg M, Sjogreen K, Liu X, Frey E, Dewaraja Y, Strand S-E: A 3-dimensional absorbed dose calculation method based on quantitative SPECT for radionuclide therapy: evaluation for 131I using Monte Carlo simulation. J Nucl Med 2002, 43: 1101–1109.PubMedCentralPubMed Ljungberg M, Sjogreen K, Liu X, Frey E, Dewaraja Y, Strand S-E: A 3-dimensional absorbed dose calculation method based on quantitative SPECT for radionuclide therapy: evaluation for 131I using Monte Carlo simulation. J Nucl Med 2002, 43: 1101–1109.PubMedCentralPubMed
10.
go back to reference He B, Du Y, Song X, Segars WP, Frey EC: A Monte Carlo and physical phantom evaluation of quantitative In-111 SPECT. Phys Med Biol 2005, 50: 4169–4185. 10.1088/0031-9155/50/17/018CrossRefPubMed He B, Du Y, Song X, Segars WP, Frey EC: A Monte Carlo and physical phantom evaluation of quantitative In-111 SPECT. Phys Med Biol 2005, 50: 4169–4185. 10.1088/0031-9155/50/17/018CrossRefPubMed
11.
go back to reference IAEA Human Health:IAEA Human Health Reports No 9: Quantitative Nuclear Medicine Imaging: Concepts, Requirements and Methods. IAEA Human Health, Vienna; 2014. IAEA Human Health:IAEA Human Health Reports No 9: Quantitative Nuclear Medicine Imaging: Concepts, Requirements and Methods. IAEA Human Health, Vienna; 2014.
12.
go back to reference Frey EC, Humm JL, Ljungberg M: Accuracy and precision of radioactivity quantification in nuclear medicine images. Semin Nucl Med 2012, 42: 208–218. 10.1053/j.semnuclmed.2011.11.003PubMedCentralCrossRefPubMed Frey EC, Humm JL, Ljungberg M: Accuracy and precision of radioactivity quantification in nuclear medicine images. Semin Nucl Med 2012, 42: 208–218. 10.1053/j.semnuclmed.2011.11.003PubMedCentralCrossRefPubMed
13.
go back to reference Wahl RL: Tositumomab and 131I therapy in non-Hodgkin’s lymphoma. J Nucl Med 2005, 46(1 suppl):128S-140S.PubMed Wahl RL: Tositumomab and 131I therapy in non-Hodgkin’s lymphoma. J Nucl Med 2005, 46(1 suppl):128S-140S.PubMed
14.
go back to reference Rao PSRS: Variance Components Estimation: Mixed Models, Methodologies and Applications. Chapman & Hall, London; 1997. Rao PSRS: Variance Components Estimation: Mixed Models, Methodologies and Applications. Chapman & Hall, London; 1997.
15.
go back to reference Carey JE, Byrne P, DeWerd L, Lieto R, Petry N: AAPM Report No. 181: The Selection, Use, Calibration, and Quality Assurance of Radionuclide Calibrators Used in Nuclear Medicine. AAPM, College Park; 2012. Carey JE, Byrne P, DeWerd L, Lieto R, Petry N: AAPM Report No. 181: The Selection, Use, Calibration, and Quality Assurance of Radionuclide Calibrators Used in Nuclear Medicine. AAPM, College Park; 2012.
16.
go back to reference Silosky M, Johnson V, Beasley C, Kappadath SC: Characterization of the count rate performance of modern gamma cameras. Med Phys 2013, 40: 032502. 10.1118/1.4792297PubMedCentralCrossRefPubMed Silosky M, Johnson V, Beasley C, Kappadath SC: Characterization of the count rate performance of modern gamma cameras. Med Phys 2013, 40: 032502. 10.1118/1.4792297PubMedCentralCrossRefPubMed
Metadata
Title
Factors affecting the stability and repeatability of gamma camera calibration for quantitative imaging applications based on a retrospective review of clinical data
Authors
Nadège Anizan
Hao Wang
Xian C Zhou
Robert F Hobbs
Richard L Wahl
Eric C Frey
Publication date
01-12-2014
Publisher
Springer Berlin Heidelberg
Published in
EJNMMI Research / Issue 1/2014
Electronic ISSN: 2191-219X
DOI
https://doi.org/10.1186/s13550-014-0067-x

Other articles of this Issue 1/2014

EJNMMI Research 1/2014 Go to the issue