Published in: Critical Care 1/2016

Open Access 01-12-2016 | Research

Manual versus Automated moNitoring Accuracy of GlucosE II (MANAGE II)

Authors: Cláudia Righy Shinotsuka, Alexandre Brasseur, David Fagnoul, Timothy So, Jean-Louis Vincent, Jean-Charles Preiser

Published in: Critical Care | Issue 1/2016

Login to get access

Abstract

Background

Intravascular continuous glucose monitoring (CGM) may facilitate glycemic control in the intensive care unit (ICU). We compared the accuracy of a CGM device (OptiScanner®) with a standard reference method.

Methods

Adult patients who had blood glucose (BG) levels >150 mg/dl and required insertion of an arterial and central venous catheter were included. The OptiScanner® was inserted into a multiple-lumen central venous catheter. Patients were treated using a dynamic-scale insulin algorithm to achieve BG values between 80 and 150 mg/dl. The BG values measured by the OptiScanner® were plotted against BG values measured using a reference analyzer. The correlation between the BG values measured using the two methods and the clinical relevance of any differences were assessed using the coefficient of determination (r 2) and the Clarke error grid, respectively; bias was assessed by the mean absolute relative difference (MARD). Three different standards of glucose monitoring were used to assess accuracy. Glycemic control was assessed using the time in range (TIR). Six indices of glycemic variability were calculated.

Results

The analysis included 929 paired samples from 88 patients, monitored for a total of 2584 hours. Reference BG values ranged between 60 and 484 mg/dl. The r 2 value was 0.89. The percentage of BG values within zones A and B of the Clarke error grid was 99.9%; the MARD was 7.7%. Using the ISO 15197 standard and Food and Drug Administration and consensus standards, respectively, 80.4% of measurements were within 15 mg/dl and 88.2% within 15% of reference values, 40% of measurements were within 7 mg/dl and 72.5% within 10% of reference values, and 65.2% of measurements were within 10 mg/dl and 82.7% within 12.5% of reference values. The TIR was slightly lower with the OptiScanner® than with the reference method. The J-index, standard deviation and maximal glucose change were the indices of glycemic variability least affected by the measurement device.

Conclusions

Based on the MARD, the performance of the OptiScanner® is adequate for use in ICU patients. Because recent standards for accuracy were not met, the OptiScanner® should not be used as a sole monitor. The assessment of glycemic variability is influenced by the time interval between BG determinations.

Trial registration

Clinicaltrials.gov NCT01720381. Registered 31 October 2012.
Literature
1.
go back to reference Gunst J, Van den Berghe G. Blood glucose control in the ICU: don't throw out the baby with the bathwater! Intensive Care Med. 2016;42:1478–81. Gunst J, Van den Berghe G. Blood glucose control in the ICU: don't throw out the baby with the bathwater! Intensive Care Med. 2016;42:1478–81.
2.
go back to reference Marik PE. Tight glycemic control in acutely ill patients: low evidence of benefit, high evidence of harm! Intensive Care Med. 2016;42:1475–77. Marik PE. Tight glycemic control in acutely ill patients: low evidence of benefit, high evidence of harm! Intensive Care Med. 2016;42:1475–77.
3.
go back to reference Mesotten D, Preiser JC, Kosiborod M. Glucose management in critically ill adults and children. Lancet Diabetes Endocrinol. 2015;3:723–33.CrossRefPubMed Mesotten D, Preiser JC, Kosiborod M. Glucose management in critically ill adults and children. Lancet Diabetes Endocrinol. 2015;3:723–33.CrossRefPubMed
4.
go back to reference Marik PE, Preiser JC. Toward understanding tight glycemic control in the ICU: a systematic review and metaanalysis. Chest. 2010;137:544–51.CrossRefPubMed Marik PE, Preiser JC. Toward understanding tight glycemic control in the ICU: a systematic review and metaanalysis. Chest. 2010;137:544–51.CrossRefPubMed
5.
go back to reference Van den Berghe G, Wouters P, Weekers F, Verwaest C, Bruyninckx F, Schetz M, et al. Intensive insulin therapy in the critically ill patients. N Engl J Med. 2001;345:1359–67.CrossRefPubMed Van den Berghe G, Wouters P, Weekers F, Verwaest C, Bruyninckx F, Schetz M, et al. Intensive insulin therapy in the critically ill patients. N Engl J Med. 2001;345:1359–67.CrossRefPubMed
6.
go back to reference Finfer S, Chittock DR, Su SY, Blair D, Foster D, Dhingra V, et al. Intensive versus conventional glucose control in critically ill patients. N Engl J Med. 2009;360:1283–97.CrossRefPubMed Finfer S, Chittock DR, Su SY, Blair D, Foster D, Dhingra V, et al. Intensive versus conventional glucose control in critically ill patients. N Engl J Med. 2009;360:1283–97.CrossRefPubMed
7.
go back to reference Preiser JC, Devos P, Ruiz-Santana S, Mélot C, Annane D, Groeneveld J, et al. A prospective randomised multi-centre controlled trial on tight glucose control by intensive insulin therapy in adult intensive care units: the Glucontrol study. Intensive Care Med. 2009;35:1738–48.CrossRefPubMed Preiser JC, Devos P, Ruiz-Santana S, Mélot C, Annane D, Groeneveld J, et al. A prospective randomised multi-centre controlled trial on tight glucose control by intensive insulin therapy in adult intensive care units: the Glucontrol study. Intensive Care Med. 2009;35:1738–48.CrossRefPubMed
8.
go back to reference Brunkhorst FM, Engel C, Bloos F, Meier-Hellmann A, Ragaller M, Weiler N, et al. Intensive insulin therapy and pentastarch resuscitation in severe sepsis. N Engl J Med. 2008;358:125–39.CrossRefPubMed Brunkhorst FM, Engel C, Bloos F, Meier-Hellmann A, Ragaller M, Weiler N, et al. Intensive insulin therapy and pentastarch resuscitation in severe sepsis. N Engl J Med. 2008;358:125–39.CrossRefPubMed
9.
go back to reference Schultz MJ, Harmsen RE, Spronk PE. Clinical review: Strict or loose glycemic control in critically ill patients—implementing best available evidence from randomized controlled trials. Crit Care. 2010;14:223.CrossRefPubMedPubMedCentral Schultz MJ, Harmsen RE, Spronk PE. Clinical review: Strict or loose glycemic control in critically ill patients—implementing best available evidence from randomized controlled trials. Crit Care. 2010;14:223.CrossRefPubMedPubMedCentral
10.
go back to reference Finfer S, Wernerman J, Preiser JC, Cass T, Desaive T, Hovorka R, et al. Clinical review: Consensus recommendations on measurement of blood glucose and reporting glycemic control in critically ill adults. Crit Care. 2013;17:229.CrossRefPubMedPubMedCentral Finfer S, Wernerman J, Preiser JC, Cass T, Desaive T, Hovorka R, et al. Clinical review: Consensus recommendations on measurement of blood glucose and reporting glycemic control in critically ill adults. Crit Care. 2013;17:229.CrossRefPubMedPubMedCentral
11.
go back to reference Wernerman J, Desaive T, Finfer S, Foubert L, Furnary A, Holzinger U, et al. Continuous glucose control in the ICU: report of a 2013 round table meeting. Crit Care. 2014;18:226.CrossRefPubMedPubMedCentral Wernerman J, Desaive T, Finfer S, Foubert L, Furnary A, Holzinger U, et al. Continuous glucose control in the ICU: report of a 2013 round table meeting. Crit Care. 2014;18:226.CrossRefPubMedPubMedCentral
12.
go back to reference Preiser JC. Glycemic control during critical illness. Exp Rev Endocrinol Metab. 2011;6:681–9.CrossRef Preiser JC. Glycemic control during critical illness. Exp Rev Endocrinol Metab. 2011;6:681–9.CrossRef
13.
go back to reference Knaus WA, Draper EA, Wagner DP, Zimmerman JE. APACHE II: a severity of disease classification system. Crit Care Med. 1985;13:818–29.CrossRefPubMed Knaus WA, Draper EA, Wagner DP, Zimmerman JE. APACHE II: a severity of disease classification system. Crit Care Med. 1985;13:818–29.CrossRefPubMed
14.
go back to reference Meynaar IA, Dawson L, Tangkau PL, Salm EF, Rijks L. Introduction and evaluation of a computerised insulin protocol. Intensive Care Med. 2007;33:591–6.CrossRefPubMed Meynaar IA, Dawson L, Tangkau PL, Salm EF, Rijks L. Introduction and evaluation of a computerised insulin protocol. Intensive Care Med. 2007;33:591–6.CrossRefPubMed
15.
go back to reference Clarke WL, Cox D, Gonder-Frederick LA, Carter W, Pohl SL. Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care. 1987;10:622–8.CrossRefPubMed Clarke WL, Cox D, Gonder-Frederick LA, Carter W, Pohl SL. Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care. 1987;10:622–8.CrossRefPubMed
16.
go back to reference Krinsley JS, Preiser JC. Time in blood glucose range 70 to 140 mg/dl > 80% is strongly associated with increased survival in non-diabetic critically ill adults. Crit Care. 2015;19:179.CrossRefPubMedPubMedCentral Krinsley JS, Preiser JC. Time in blood glucose range 70 to 140 mg/dl > 80% is strongly associated with increased survival in non-diabetic critically ill adults. Crit Care. 2015;19:179.CrossRefPubMedPubMedCentral
17.
go back to reference Penning S, Pretty C, Preiser JC, Shaw GM, Desaive T, Chase JG. Glucose control positively influences patient outcome: a retrospective study. J Crit Care. 2015;30:455–9.CrossRefPubMed Penning S, Pretty C, Preiser JC, Shaw GM, Desaive T, Chase JG. Glucose control positively influences patient outcome: a retrospective study. J Crit Care. 2015;30:455–9.CrossRefPubMed
18.
go back to reference Ali NA, Krinsley JK, Preiser JC. Glucose variability in critically ill patients. In: Vincent JL, editor. Yearbook of Intensive Care and Emergency Medicine. Berlin: Springer; 2010. p. 728–37. ISBN 978-3-540-92275-9. Ali NA, Krinsley JK, Preiser JC. Glucose variability in critically ill patients. In: Vincent JL, editor. Yearbook of Intensive Care and Emergency Medicine. Berlin: Springer; 2010. p. 728–37. ISBN 978-3-540-92275-9.
19.
go back to reference van Hooijdonk RT, Winters T, Fischer JC, van Dongen-Lases EC, Krinsley JS, Preiser JC, Schultz MJ. Accuracy and limitations of continuous glucose monitoring using spectroscopy in critically patients. Ann Intensive Care. 2014;4:8.CrossRefPubMedPubMedCentral van Hooijdonk RT, Winters T, Fischer JC, van Dongen-Lases EC, Krinsley JS, Preiser JC, Schultz MJ. Accuracy and limitations of continuous glucose monitoring using spectroscopy in critically patients. Ann Intensive Care. 2014;4:8.CrossRefPubMedPubMedCentral
20.
go back to reference Magarian P, Sterling B. Animal study Optiscan Plasma-generating glucose monitor accuracy demonstrated in an animal model. J Diabetes Sci Technol. 2009;3:1411–8.CrossRefPubMedPubMedCentral Magarian P, Sterling B. Animal study Optiscan Plasma-generating glucose monitor accuracy demonstrated in an animal model. J Diabetes Sci Technol. 2009;3:1411–8.CrossRefPubMedPubMedCentral
21.
go back to reference Jax T, Heise T, Nosek L, Gable J, Lim G, Calentine C. Automated near-continuous glucose monitoring measured in plasma using mid-infrared spectroscopy. J Diabetes Sci Technol. 2011;5:345–52.CrossRefPubMedPubMedCentral Jax T, Heise T, Nosek L, Gable J, Lim G, Calentine C. Automated near-continuous glucose monitoring measured in plasma using mid-infrared spectroscopy. J Diabetes Sci Technol. 2011;5:345–52.CrossRefPubMedPubMedCentral
22.
go back to reference Barassi A, Umbrello M, Ghilardi F, Damele CA, Massaccesi L, Iapichino G, et al. Evaluation of the performance of a new OptiScanner™ 5000 system for an intermittent glucose monitoring. Clin Chim Acta. 2015;438:252–4.CrossRefPubMed Barassi A, Umbrello M, Ghilardi F, Damele CA, Massaccesi L, Iapichino G, et al. Evaluation of the performance of a new OptiScanner™ 5000 system for an intermittent glucose monitoring. Clin Chim Acta. 2015;438:252–4.CrossRefPubMed
23.
go back to reference Umbrello M, Salice V, Spanu P, Formenti P, Barassi A, Melzi d'Eril GV, et al. Performance assessment of a glucose control protocol in septic patients with an automated intermittent plasma glucose monitoring device. Clin Nutr. 2014;33:867–71.CrossRefPubMed Umbrello M, Salice V, Spanu P, Formenti P, Barassi A, Melzi d'Eril GV, et al. Performance assessment of a glucose control protocol in septic patients with an automated intermittent plasma glucose monitoring device. Clin Nutr. 2014;33:867–71.CrossRefPubMed
24.
go back to reference Van Herpe T, De Moor B, Van den Berghe G, Mesotten D. Modeling of effect of glucose sensor errors on insulin dosage and glucose bolus computed by LOGIC-Insulin. Clin Chem. 2014;60:1510–8.CrossRefPubMed Van Herpe T, De Moor B, Van den Berghe G, Mesotten D. Modeling of effect of glucose sensor errors on insulin dosage and glucose bolus computed by LOGIC-Insulin. Clin Chem. 2014;60:1510–8.CrossRefPubMed
25.
go back to reference Wilinska ME, Hovorka R. Glucose control in the intensive care unit by use of continuous glucose monitoring: what level of measurement error is acceptable? Clin Chem. 2014;60:1500–9.CrossRefPubMed Wilinska ME, Hovorka R. Glucose control in the intensive care unit by use of continuous glucose monitoring: what level of measurement error is acceptable? Clin Chem. 2014;60:1500–9.CrossRefPubMed
26.
go back to reference Finfer S, Liu B, Chittock DR, Norton R, Myburgh JA, McArthur C, Mitchell I, et al. Hypoglycemia and risk of death in critically ill patients. N Engl J Med. 2012;367:1108–18.CrossRefPubMed Finfer S, Liu B, Chittock DR, Norton R, Myburgh JA, McArthur C, Mitchell I, et al. Hypoglycemia and risk of death in critically ill patients. N Engl J Med. 2012;367:1108–18.CrossRefPubMed
27.
go back to reference Foubert LA, Lecomte PV, Nobels FR, Gulino AM, De Decker KH. Accuracy of a feasibility version of an intravenous continuous glucose monitor in volunteers with diabetes and hospitalized patients. Diabetes Technol Ther. 2014;16:858–66.CrossRefPubMed Foubert LA, Lecomte PV, Nobels FR, Gulino AM, De Decker KH. Accuracy of a feasibility version of an intravenous continuous glucose monitor in volunteers with diabetes and hospitalized patients. Diabetes Technol Ther. 2014;16:858–66.CrossRefPubMed
28.
go back to reference Crane BC, Barwell NP, Gopal P, Gopichand M, Higgs T, James TD, et al. The Development of a continuous intravascular glucose monitoring sensor. J Diabetes Sci Technol. 2015;9:751–61.CrossRefPubMedPubMedCentral Crane BC, Barwell NP, Gopal P, Gopichand M, Higgs T, James TD, et al. The Development of a continuous intravascular glucose monitoring sensor. J Diabetes Sci Technol. 2015;9:751–61.CrossRefPubMedPubMedCentral
29.
go back to reference Strasma PJ, Finfer S, Flower O, Hipszer B, Kosiborod M, Macken LJ, et al. Use of an intravascular fluorescent continuous glucose sensor in ICU patients. J Diabetes Sci Technol. 2015;9:762–70.CrossRefPubMedPubMedCentral Strasma PJ, Finfer S, Flower O, Hipszer B, Kosiborod M, Macken LJ, et al. Use of an intravascular fluorescent continuous glucose sensor in ICU patients. J Diabetes Sci Technol. 2015;9:762–70.CrossRefPubMedPubMedCentral
30.
go back to reference Schierenbeck F, Nijsten MW, Franco-Cereceda A, Liska J. Introducing intravascular microdialysis for continuous lactate monitoring in patients undergoing cardiac surgery: a prospective observational study. Crit Care. 2014;18:R56.CrossRefPubMedPubMedCentral Schierenbeck F, Nijsten MW, Franco-Cereceda A, Liska J. Introducing intravascular microdialysis for continuous lactate monitoring in patients undergoing cardiac surgery: a prospective observational study. Crit Care. 2014;18:R56.CrossRefPubMedPubMedCentral
Metadata
Title
Manual versus Automated moNitoring Accuracy of GlucosE II (MANAGE II)
Authors
Cláudia Righy Shinotsuka
Alexandre Brasseur
David Fagnoul
Timothy So
Jean-Louis Vincent
Jean-Charles Preiser
Publication date
01-12-2016
Publisher
BioMed Central
Published in
Critical Care / Issue 1/2016
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/s13054-016-1547-3

Other articles of this Issue 1/2016

Critical Care 1/2016 Go to the issue