Published in: Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine 1/2017

Open Access 01-12-2017 | Original research

Performance of point-of-care international normalized ratio measurement to diagnose trauma-induced coagulopathy

Authors: Thomas Mistral, Yvonnick Boué, Jean-Luc Bosson, Pauline Manhes, Jules Greze, Julien Brun, Pierre Albaladejo, Jean-François Payen, Pierre Bouzat

Published in: Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine | Issue 1/2017

Login to get access

Abstract

Background

Trauma-induced coagulopathy (TIC) is a common feature after severe trauma. Detection of TIC is based upon classic coagulation tests including international normalized ratio (INR) value. Point-of-care (POC) devices have been developed to rapidly measure INR at the bedside on whole blood. The aim of the study was to test the precision of the Coagucheck® XS Pro device for INR measurement at hospital admission after severe trauma.

Methods

We conducted a prospective observational study in a French level I trauma center. From January 2015 to May 2016, 98 patients with a suspicion of a post-traumatic acute hemorrhage had POC-INR measurement on whole blood concomitantly to classic laboratory INR determination (lab-INR) on plasma at hospital admission. The agreement between the two methods in sorting three predefined categories of INR (normal coagulation, moderate TIC and severe TIC) was evaluated using the Cohen’s kappa test with a quadratic weighting. The correlation between POC-INR and lab-INR was measured using the Pearson’s coefficient. We also performed a Bland and Altman analysis.

Results

The agreement between the lab-INR and the POC-INR was moderate (Kappa = 0.45 [95% CI 0.36–0.50]) and the correlation between the two measurements was also weak (Pearson’s coefficient = 0.44 [95% CI 0.27–0.59]). Using a Bland and Altman analysis, the mean difference (bias) for INR was 0.22 [95% CI 0.02–0.42], and the standard deviation (precision) of the difference was 1.01.

Discussion/conclusion

POC Coagucheck® XS Pro device is not reliable to measure bedside INR. Its moderate agreement with lab-INR weakens the usefulness of such device after severe trauma.

Trial registration

NCT02869737. Registered 9 August 2016.
Literature
1.
go back to reference Chang R, Cardenas JC, Wade CE, Holcomb JB. Advances in the understanding of trauma-induced coagulopathy. Blood. 2016;128:1043–9.CrossRefPubMed Chang R, Cardenas JC, Wade CE, Holcomb JB. Advances in the understanding of trauma-induced coagulopathy. Blood. 2016;128:1043–9.CrossRefPubMed
2.
go back to reference Rossaint R, Bouillon B, Cerny V, Coats TJ, Duranteau J, Fernández-Mondéjar E, et al. The European guideline on management of major bleeding and coagulopathy following trauma: fourth Edition. Crit Care. 2016;20:100.CrossRefPubMedPubMedCentral Rossaint R, Bouillon B, Cerny V, Coats TJ, Duranteau J, Fernández-Mondéjar E, et al. The European guideline on management of major bleeding and coagulopathy following trauma: fourth Edition. Crit Care. 2016;20:100.CrossRefPubMedPubMedCentral
3.
go back to reference Frith D, Goslings JC, Gaarder C, Maegele M, Cohen MJ, Allard S, et al. Definition and drivers of acute traumatic coagulopathy: clinical and experimental investigations. J Thromb Haemost. 2010;8:1919.CrossRefPubMed Frith D, Goslings JC, Gaarder C, Maegele M, Cohen MJ, Allard S, et al. Definition and drivers of acute traumatic coagulopathy: clinical and experimental investigations. J Thromb Haemost. 2010;8:1919.CrossRefPubMed
4.
go back to reference Peltan ID, Vande Vusse LK, Maier RV, Watkins TR. An international normalized ratio-based definition of acute traumatic coagulopathy Is associated with mortality, venous thromboembolism, and multiple organ failure after injury. Crit Care Med. 2015;43:1429.CrossRefPubMedPubMedCentral Peltan ID, Vande Vusse LK, Maier RV, Watkins TR. An international normalized ratio-based definition of acute traumatic coagulopathy Is associated with mortality, venous thromboembolism, and multiple organ failure after injury. Crit Care Med. 2015;43:1429.CrossRefPubMedPubMedCentral
5.
go back to reference Callcut RA, Johannigman JA, Kadon KS, Hanseman DJ, Robinson BRH. All massive transfusion criteria are not created equal: defining the predictive value of individual transfusion triggers to better determine who benefits from blood. J Trauma. 2011;70:794.CrossRefPubMed Callcut RA, Johannigman JA, Kadon KS, Hanseman DJ, Robinson BRH. All massive transfusion criteria are not created equal: defining the predictive value of individual transfusion triggers to better determine who benefits from blood. J Trauma. 2011;70:794.CrossRefPubMed
6.
go back to reference Goodman MD, Makley AT, Hanseman DJ, Pritts TA, Robinson BRH. All the bang without the bucks: defining essential point-of-care testing for traumatic coagulopathy. J Trauma Acute Care Surg. 2015;79:117.CrossRefPubMed Goodman MD, Makley AT, Hanseman DJ, Pritts TA, Robinson BRH. All the bang without the bucks: defining essential point-of-care testing for traumatic coagulopathy. J Trauma Acute Care Surg. 2015;79:117.CrossRefPubMed
7.
go back to reference Gauss T, Hamada S, Jurcisin I, Dahmani S, Boudaoud L, Mantz J, et al. Limits of agreement between measures obtained from standard laboratory and the point-of-care device hemochron signature Elite(R) during acute haemorrhage. Br J Anaesth. 2014;112:514.CrossRefPubMed Gauss T, Hamada S, Jurcisin I, Dahmani S, Boudaoud L, Mantz J, et al. Limits of agreement between measures obtained from standard laboratory and the point-of-care device hemochron signature Elite(R) during acute haemorrhage. Br J Anaesth. 2014;112:514.CrossRefPubMed
8.
go back to reference Mitra B, O’Reilly G, Collecutt M, Cameron PA, Phillips L, Davis A. Prospective comparison of point-of-care international normalised ratio measurement versus plasma international normalised ratio for acute traumatic coagulopathy. Emerg Med Australas. 2012;24:363.CrossRefPubMed Mitra B, O’Reilly G, Collecutt M, Cameron PA, Phillips L, Davis A. Prospective comparison of point-of-care international normalised ratio measurement versus plasma international normalised ratio for acute traumatic coagulopathy. Emerg Med Australas. 2012;24:363.CrossRefPubMed
9.
go back to reference David J-S, Levrat A, Inaba K, Macabeo C, Rugeri L, Fontaine O, et al. Utility of a point-of-care device for rapid determination of prothrombin time in trauma patients: a preliminary study. J Trauma Acute Care Surg. 2012;72:703.CrossRefPubMed David J-S, Levrat A, Inaba K, Macabeo C, Rugeri L, Fontaine O, et al. Utility of a point-of-care device for rapid determination of prothrombin time in trauma patients: a preliminary study. J Trauma Acute Care Surg. 2012;72:703.CrossRefPubMed
10.
go back to reference Taborski U, Braun SL, Völler H. Analytical performance of the new coagulation monitoring system INRatio for the determination of INR compared with the coagulation monitor Coaguchek S and an established laboratory method. J Thromb Thrombolysis. 2004;18:103.CrossRefPubMed Taborski U, Braun SL, Völler H. Analytical performance of the new coagulation monitoring system INRatio for the determination of INR compared with the coagulation monitor Coaguchek S and an established laboratory method. J Thromb Thrombolysis. 2004;18:103.CrossRefPubMed
11.
go back to reference Hemkens LG, Hilden KM, Hartschen S, Kaiser T, Didjurgeit U, Hansen R, et al. A randomized trial comparing INR monitoring devices in patients with anticoagulation self-management: evaluation of a novel error-grid approach. J Thromb Thrombolysis. 2008;26:22.CrossRefPubMed Hemkens LG, Hilden KM, Hartschen S, Kaiser T, Didjurgeit U, Hansen R, et al. A randomized trial comparing INR monitoring devices in patients with anticoagulation self-management: evaluation of a novel error-grid approach. J Thromb Thrombolysis. 2008;26:22.CrossRefPubMed
12.
go back to reference Torreiro EG, Fernández EG, Rodríguez RM, López CV, Núñez JB. Comparative study of accuracy and clinical agreement of the CoaguChek XS portable device versus standard laboratory practice in unexperienced patients. Thromb Haemost. 2009;101:969.PubMed Torreiro EG, Fernández EG, Rodríguez RM, López CV, Núñez JB. Comparative study of accuracy and clinical agreement of the CoaguChek XS portable device versus standard laboratory practice in unexperienced patients. Thromb Haemost. 2009;101:969.PubMed
13.
go back to reference Christensen TD, Larsen TB. Precision and accuracy of point-of-care testing coagulometers used for self-testing and self-management of oral anticoagulation therapy. J Thromb Haemost. 2012;10:251.CrossRefPubMed Christensen TD, Larsen TB. Precision and accuracy of point-of-care testing coagulometers used for self-testing and self-management of oral anticoagulation therapy. J Thromb Haemost. 2012;10:251.CrossRefPubMed
14.
go back to reference Heneghan CJ, Garcia-Alamino JM, Spencer EA, Ward AM, Perera R, Bankhead C, et al. Self-monitoring and self-management of oral anticoagulation. Cochrane Database Syst Rev. 2016;7:CD003839.PubMed Heneghan CJ, Garcia-Alamino JM, Spencer EA, Ward AM, Perera R, Bankhead C, et al. Self-monitoring and self-management of oral anticoagulation. Cochrane Database Syst Rev. 2016;7:CD003839.PubMed
15.
go back to reference Heneghan C, Ward A, Perera R, Self-Monitoring Trialist C, Bankhead C, Fuller A, et al. Self-monitoring of oral anticoagulation: systematic review and meta-analysis of individual patient data. Lancet (London, England). 2012;379:322.CrossRef Heneghan C, Ward A, Perera R, Self-Monitoring Trialist C, Bankhead C, Fuller A, et al. Self-monitoring of oral anticoagulation: systematic review and meta-analysis of individual patient data. Lancet (London, England). 2012;379:322.CrossRef
16.
go back to reference Toulon P, Ozier Y, Ankri A, Fléron M-H, Leroux G, Samama CM. Point-of-care versus central laboratory coagulation testing during haemorrhagic surgery A multicenter study. Thromb Haemost. 2009;101:394.PubMed Toulon P, Ozier Y, Ankri A, Fléron M-H, Leroux G, Samama CM. Point-of-care versus central laboratory coagulation testing during haemorrhagic surgery A multicenter study. Thromb Haemost. 2009;101:394.PubMed
17.
go back to reference Cotte J, Lacroix G, D’Aranda E, Kaiser E, Meaudre E. Management of traumatic coagulopathy during long-distance medical evacuation: utility of the Coaguchek(®) XS pro. Annales Françaises D’anesthèsie Et De Rèanimation. 2013;32:122.CrossRefPubMed Cotte J, Lacroix G, D’Aranda E, Kaiser E, Meaudre E. Management of traumatic coagulopathy during long-distance medical evacuation: utility of the Coaguchek(®) XS pro. Annales Françaises D’anesthèsie Et De Rèanimation. 2013;32:122.CrossRefPubMed
18.
go back to reference Celenza A, Skinner K. Comparison of emergency department point-of-care international normalised ratio (INR) testing with laboratory-based testing. Emerg Med J. 2011;28:136.CrossRefPubMed Celenza A, Skinner K. Comparison of emergency department point-of-care international normalised ratio (INR) testing with laboratory-based testing. Emerg Med J. 2011;28:136.CrossRefPubMed
19.
go back to reference Beynon C, Erk AG, Potzy A, Mohr S, Popp E. Point of care coagulometry in prehospital emergency care: an observational study. Scand J Trauma Resusc Emerg Med. 2015;23:58.CrossRefPubMedPubMedCentral Beynon C, Erk AG, Potzy A, Mohr S, Popp E. Point of care coagulometry in prehospital emergency care: an observational study. Scand J Trauma Resusc Emerg Med. 2015;23:58.CrossRefPubMedPubMedCentral
Metadata
Title
Performance of point-of-care international normalized ratio measurement to diagnose trauma-induced coagulopathy
Authors
Thomas Mistral
Yvonnick Boué
Jean-Luc Bosson
Pauline Manhes
Jules Greze
Julien Brun
Pierre Albaladejo
Jean-François Payen
Pierre Bouzat
Publication date
01-12-2017
Publisher
BioMed Central
DOI
https://doi.org/10.1186/s13049-017-0404-y

Other articles of this Issue 1/2017

Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine 1/2017 Go to the issue