Effect of pedicle screw augmentation with a self-curing elastomeric material under cranio-caudal cyclic loading—a cadaveric biomechanical study

Authors: Werner Schmoelz, Alexander Keiler, Marko Konschake, Richard A Lindtner, Alessandro Gasbarrini

Published in: Journal of Orthopaedic Surgery and Research | Issue 1/2018

Login to get access

Abstract

Background

Pedicle screws can be augmented with polymethylmethacrylate (PMMA) cement through cannulated and fenestrated pedicle screws to improve screw anchorage. To overcome the drawbacks of PMMA, a modified augmentation technique applying a self-curing elastomeric material into a balloon-created cavity prior to screw insertion was developed and evaluated. The aim of the study was to compare the effect of the established and novel augmentation technique on pedicle screw anchorage in a biomechanical in vitro experiment.

Methods

In ten lumbar vertebral bodies, the right pedicles were instrumented with monoaxial cannulated and fenestrated pedicle screws and augmented in situ with 2 ml PMMA. The left pedicles were instrumented with monoaxial cannulated pedicle screws. Prior to left screw insertion, a balloon cavity was created and filled with 3 ml of self-curing elastomer (silicone). Each screw was subjected to a cranio-caudal cyclic load starting from − 50 to 50 N while the upper load was increased by 5 N every 100 load cycles until loosening or 11,000 cycles (600 N). After cyclic loading, a pullout test of the screws was conducted.

Results

The mean cycles to screw loosening were 9824 ± 1982 and 7401 ± 1644 for the elastomer and PMMA group, respectively (P = 0.012). The post-cycling pullout test of the loosened screws showed differences in the failure mode and failure load, with predominantly pedicle/vertebrae fractures in the PMMA group (1188.6 N ± 288.1) and screw pullout through the pedicle (671.3 N ± 332.1) in the elastomer group.

Conclusion

The modified pedicle screw augmentation technique involving a balloon cavity creation and a self-curing elastomeric silicone resulted in a significantly improved pedicle screw anchorage under cyclic cranio-caudal loading when compared to conventional in situ PMMA augmentation.
Literature
6.
go back to reference Pfeifer BA, Krag MH, Johnson C. Repair of failed transpedicle screw fixation. A biomechanical study comparing polymethylmethacrylate, milled bone, and matchstick bone reconstruction. Spine. 1994;19:350–3.CrossRef Pfeifer BA, Krag MH, Johnson C. Repair of failed transpedicle screw fixation. A biomechanical study comparing polymethylmethacrylate, milled bone, and matchstick bone reconstruction. Spine. 1994;19:350–3.CrossRef
7.
go back to reference Sarzier JS, Evans AJ, Cahill DW. Increased pedicle screw pullout strength with vertebroplasty augmentation in osteoporotic spines. J Neurosurg. 2002;96:309–12.PubMed Sarzier JS, Evans AJ, Cahill DW. Increased pedicle screw pullout strength with vertebroplasty augmentation in osteoporotic spines. J Neurosurg. 2002;96:309–12.PubMed
9.
go back to reference McKoy BE, An YH. An injectable cementing screw for fixation in osteoporotic bone. J Biomed Mater Res. 2000;53:216–20.CrossRef McKoy BE, An YH. An injectable cementing screw for fixation in osteoporotic bone. J Biomed Mater Res. 2000;53:216–20.CrossRef
11.
go back to reference Belkoff SM, Molloy S. Temperature measurement during polymerization of polymethylmethacrylate cement used for vertebroplasty. Spine. 2003;28:1555–9.PubMed Belkoff SM, Molloy S. Temperature measurement during polymerization of polymethylmethacrylate cement used for vertebroplasty. Spine. 2003;28:1555–9.PubMed
12.
go back to reference Jefferiss CD, Lee AJ, Ling RS. Thermal aspects of self-curing polymethylmethacrylate. The Journal of Bone and Joint Surgery British. 1975;57:511–8.CrossRef Jefferiss CD, Lee AJ, Ling RS. Thermal aspects of self-curing polymethylmethacrylate. The Journal of Bone and Joint Surgery British. 1975;57:511–8.CrossRef
14.
go back to reference Freeman MA, Bradley GW, Revell PA. Observations upon the interface between bone and polymethylmethacrylate cement. The Journal of Bone and Joint Surgery British. 1982;64:489–93.CrossRef Freeman MA, Bradley GW, Revell PA. Observations upon the interface between bone and polymethylmethacrylate cement. The Journal of Bone and Joint Surgery British. 1982;64:489–93.CrossRef
16.
go back to reference Shinzato S, Nakamura T, Kokubo T, Kitamura Y. PMMA-based bioactive cement: effect of glass bead filler content and histological change with time. J Biomed Mater Res. 2002;59:225–32.CrossRef Shinzato S, Nakamura T, Kokubo T, Kitamura Y. PMMA-based bioactive cement: effect of glass bead filler content and histological change with time. J Biomed Mater Res. 2002;59:225–32.CrossRef
18.
go back to reference Ciapetti G, Granchi D, Cenni E, Savarino L, Cavedagna D, Pizzoferrato A. Cytotoxic effect of bone cements in HL-60 cells: distinction between apoptosis and necrosis. J Biomed Mater Res. 2000;52:338–45.CrossRef Ciapetti G, Granchi D, Cenni E, Savarino L, Cavedagna D, Pizzoferrato A. Cytotoxic effect of bone cements in HL-60 cells: distinction between apoptosis and necrosis. J Biomed Mater Res. 2000;52:338–45.CrossRef
19.
go back to reference Goodman SB, Huie P, Song Y, Lee K, Doshi A, Rushdieh B, Woolson S, Maloney W, Schurman D, Sibley R. Loosening and osteolysis of cemented joint arthroplasties. A biologic spectrum. Clin Orthop Relat Res. 1997;337:149–63.CrossRef Goodman SB, Huie P, Song Y, Lee K, Doshi A, Rushdieh B, Woolson S, Maloney W, Schurman D, Sibley R. Loosening and osteolysis of cemented joint arthroplasties. A biologic spectrum. Clin Orthop Relat Res. 1997;337:149–63.CrossRef
20.
go back to reference Granchi D, Stea S, Ciapetti G, Savarino L, Cavedagna D, Pizzoferrato A. In vitro effects of bone cements on the cell cycle of osteoblast-like cells. Biomaterials. 1995;16:1187–92.CrossRef Granchi D, Stea S, Ciapetti G, Savarino L, Cavedagna D, Pizzoferrato A. In vitro effects of bone cements on the cell cycle of osteoblast-like cells. Biomaterials. 1995;16:1187–92.CrossRef
26.
go back to reference ASTM. ASfTM standard specifications for acrylic bone cement, ASTM standard F451-99a. In: Annual book of ASTM standards. West Conshohocken: ASTM; 2007. ASTM. ASfTM standard specifications for acrylic bone cement, ASTM standard F451-99a. In: Annual book of ASTM standards. West Conshohocken: ASTM; 2007.
27.
go back to reference Larsson S, Bauer TW. Use of injectable calcium phosphate cement for fracture fixation: a review. Clin Orthop Relat Res. 2002;395:23–32.CrossRef Larsson S, Bauer TW. Use of injectable calcium phosphate cement for fracture fixation: a review. Clin Orthop Relat Res. 2002;395:23–32.CrossRef
29.
go back to reference Colas A, Curtis J (2004) Silicone biomaterials: history and chemistry. Colas A, Curtis J (2004) Silicone biomaterials: history and chemistry.
30.
go back to reference Thomas X. Silicones in medical applications. In: Gleria RDJM, editor. Inorganic Polymers. New York: Nova Science Publishers (www.novapublishers.com); 2007. Thomas X. Silicones in medical applications. In: Gleria RDJM, editor. Inorganic Polymers. New York: Nova Science Publishers (www.novapublishers.com); 2007.
31.
32.
go back to reference McHanwell S, Brenner E, Chirculescu ARM, Drukker J, van Mameren H, Mazzotti G, Pais D, Paulsen F, Plaisant O, Caillaud MM, Laforet E, Riederer BM, Sanudo JR, Bueno-Lopez JL, Donate-Oliver F, Sprumont P, Teofilovski-Parapid G, BJ M. The legal and ethical framework governing Body Donation in Europe - a review of current practice and recommendations for good practice. Eur J Anat. 2008;12:1–24. McHanwell S, Brenner E, Chirculescu ARM, Drukker J, van Mameren H, Mazzotti G, Pais D, Paulsen F, Plaisant O, Caillaud MM, Laforet E, Riederer BM, Sanudo JR, Bueno-Lopez JL, Donate-Oliver F, Sprumont P, Teofilovski-Parapid G, BJ M. The legal and ethical framework governing Body Donation in Europe - a review of current practice and recommendations for good practice. Eur J Anat. 2008;12:1–24.
33.
go back to reference Riederer BM, Bolt S, Brenner E, Bueno-Lopez JL, Circulescu ARM, Davies DC, DeCaro R, Gerrits PO, McHanwell S, Pais D, Paulsen F, Sendemir E, Stabile I, BJ M. The legal and ethical framework governing Body Donation in Europe – 1st update on current practice. Eur J Anat. 2012;16:1–21. Riederer BM, Bolt S, Brenner E, Bueno-Lopez JL, Circulescu ARM, Davies DC, DeCaro R, Gerrits PO, McHanwell S, Pais D, Paulsen F, Sendemir E, Stabile I, BJ M. The legal and ethical framework governing Body Donation in Europe – 1st update on current practice. Eur J Anat. 2012;16:1–21.
36.
go back to reference Renner SM, Lim T-H, Kim W-J, Katolik L, An HS, Andersson GBJ. Augmentation of pedicle screw fixation strength using an injectable calcium phosphate cement as a function of injection timing and method. Spine. 2004;29:E212–6.CrossRef Renner SM, Lim T-H, Kim W-J, Katolik L, An HS, Andersson GBJ. Augmentation of pedicle screw fixation strength using an injectable calcium phosphate cement as a function of injection timing and method. Spine. 2004;29:E212–6.CrossRef
40.
go back to reference Rohlmann A, Bergmann G, Graichen F. Loads on an internal spinal fixation device during walking. J Biomech. 1997;30:41–7.CrossRef Rohlmann A, Bergmann G, Graichen F. Loads on an internal spinal fixation device during walking. J Biomech. 1997;30:41–7.CrossRef
41.
go back to reference Rohlmann A, Graichen F, Bergmann G. Influence of load carrying on loads in internal spinal fixators. J Biomech. 2000;33:1099–104.CrossRef Rohlmann A, Graichen F, Bergmann G. Influence of load carrying on loads in internal spinal fixators. J Biomech. 2000;33:1099–104.CrossRef
42.
go back to reference Rohlmann A, Bergmann G, Graichen F. Loads on internal spinal fixators measured in different body positions. European Spine Journal. 1999;8:354–9.CrossRef Rohlmann A, Bergmann G, Graichen F. Loads on internal spinal fixators measured in different body positions. European Spine Journal. 1999;8:354–9.CrossRef
Metadata
Title
Effect of pedicle screw augmentation with a self-curing elastomeric material under cranio-caudal cyclic loading—a cadaveric biomechanical study
Authors
Werner Schmoelz
Alexander Keiler
Marko Konschake
Richard A Lindtner
Alessandro Gasbarrini
Publication date
01-12-2018
Publisher
BioMed Central
Published in
Journal of Orthopaedic Surgery and Research / Issue 1/2018
Electronic ISSN: 1749-799X
DOI
https://doi.org/10.1186/s13018-018-0958-z

Other articles of this Issue 1/2018

Journal of Orthopaedic Surgery and Research 1/2018 Go to the issue