Published in: Radiation Oncology 1/2018

Open Access 01-12-2018 | Research

A simple method for determining dosimetric leaf gap with cross-field dose width for rounded leaf-end multileaf collimator systems

Authors: Chih-Yuan Lin, An-Cheng Shiau, Jin-Huei Ji, Chia-Jung Lee, Ti-Hao Wang, Shu-Hui Hsu, Ji-An Liang

Published in: Radiation Oncology | Issue 1/2018

Login to get access

Abstract

Purpose

The dosimetric leaf gap (DLG) and multileaf collimator (MLC) transmission are two important systematic parameters used to model the rounded MLC leaf ends effect when commissioning an Eclipse treatment planning system (TPS). Determining the optimal DLG is a time consuming process. This study develops a simple and reliable method for determining the DLG using the cross-field dose width.

Methods and materials

A Varian TrueBeam linac with 6 MV, 10 MV, 6 MV flattening filter free (FFF) and 10 MV FFF photon beams and equipped with the 120 Millennium MLC and the Eclipse™ TPS was used in this study. Integral sliding fields and static slit MLC field doses with different gap widths were measured with an ionization chamber and GAFCHROMIC EBT3 films, respectively. Measurements were performed for different beam energies and at depths of 5 and 10 cm. DLGs were derived from a linear extrapolation to zero dose and intercepting at the gap width axis. In the ion chamber measurements method, the average MLC leaf transmission to the gap reading for each gap (RgT) were calculated with nominal and cross-field dose widths, respectively. The cross-field dose widths were determined according to the dose profile measured with EBT3 films. Additionally, the optimal DLG values were determined using plan dose measurements, as the value that produced the closest agreement between the planned and measured doses. DLGs derived from the nominal and cross-field dose width, the film measurements, and the optimal process, were obtained and compared.

Results

The DLG values are insensitive to the variations in depth (within 0.07 mm). DLGs derived from nominal gap widths showed a significantly lower values (with difference about 0.5 mm) than that from cross-field dose widths and from film measurements and from plan optimal values. The method in deriving DLGs by correcting the nominal gap widths to the cross-field dose widths has shown good agreements to the plan optimal values (with difference within 0.21 mm).

Conclusions

The DLG values derived from the cross-field dose width method were consistent with the values derived from film measurements and from the plan optimal process. A simple and reliable method to determine DLG for rounded leaf-end MLC systems was established. This method provides a referable DLG value required during TPS commissioning.
Literature
1.
go back to reference Boyer A, Biggs P, Galvin J, Klein E, LoSasso T, Low D, et al. Basic applications of multileaf collimators. Report of the AAPM radiation therapy committee task group no. 50. Med Phys. 2001. Boyer A, Biggs P, Galvin J, Klein E, LoSasso T, Low D, et al. Basic applications of multileaf collimators. Report of the AAPM radiation therapy committee task group no. 50. Med Phys. 2001.
2.
go back to reference Vial P, Oliver L, Greer PB, Baldock C. An experimental investigation into the radiation field offset of a dynamic multileaf collimator. Phys Med Biol. 2006;51:5517–38.CrossRef Vial P, Oliver L, Greer PB, Baldock C. An experimental investigation into the radiation field offset of a dynamic multileaf collimator. Phys Med Biol. 2006;51:5517–38.CrossRef
3.
go back to reference Oliver M, Gagne I, Bush K, Zavgorodni S, Ansbacher W, Beckham W. Clinical significance of multi-leaf collimator positional errors for volumetric modulated arc therapy. Radiother Oncol. 2010;97:554–60.CrossRef Oliver M, Gagne I, Bush K, Zavgorodni S, Ansbacher W, Beckham W. Clinical significance of multi-leaf collimator positional errors for volumetric modulated arc therapy. Radiother Oncol. 2010;97:554–60.CrossRef
4.
go back to reference Kung JH, Chen GTY. Intensity modulated radiotherapy dose delivery error from radiation field offset inaccuracy. Med Phys. 2000;27:1617–22.CrossRef Kung JH, Chen GTY. Intensity modulated radiotherapy dose delivery error from radiation field offset inaccuracy. Med Phys. 2000;27:1617–22.CrossRef
5.
go back to reference LoSasso T, Chui CS, Ling CC. Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy. Med Phys. 1998;25:1919–27.CrossRef LoSasso T, Chui CS, Ling CC. Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy. Med Phys. 1998;25:1919–27.CrossRef
6.
go back to reference Rangel A, Dunscombe P. Tolerances on MLC leaf position accuracy for IMRT delivery with a dynamic MLC. Med Phys. 2009;36(7):3304–9.CrossRef Rangel A, Dunscombe P. Tolerances on MLC leaf position accuracy for IMRT delivery with a dynamic MLC. Med Phys. 2009;36(7):3304–9.CrossRef
7.
go back to reference Varian Medical Systems Inc. Eclipse Algorithms Reference Guide. Palo Alto, CA: Varian Medical Systems, Inc; 2010. Varian Medical Systems Inc. Eclipse Algorithms Reference Guide. Palo Alto, CA: Varian Medical Systems, Inc; 2010.
8.
go back to reference Kielar KN, Mok E, Hsu A, Wang L, Luxton G. Verification of dosimetric accuracy on the TrueBeam STx: rounded leaf effect of the high definition MLC. Med Phys. 2012;39(10):6360–71.CrossRef Kielar KN, Mok E, Hsu A, Wang L, Luxton G. Verification of dosimetric accuracy on the TrueBeam STx: rounded leaf effect of the high definition MLC. Med Phys. 2012;39(10):6360–71.CrossRef
9.
go back to reference Middlebrook ND, Sutherland B, Kairn T. Optimization of the dosimetric leaf gap for use in planning VMAT treatments of spine SABR cases. J Appl Clin Med Phys. 2017;18(4):133–9.CrossRef Middlebrook ND, Sutherland B, Kairn T. Optimization of the dosimetric leaf gap for use in planning VMAT treatments of spine SABR cases. J Appl Clin Med Phys. 2017;18(4):133–9.CrossRef
10.
go back to reference Kim J, Han JS, Hsia AT, Li SD, Xu ZG, Ryu S. Relationship between dosimetric leaf gap and dose calculation errors for high definition multi-leaf collimators in radiotherapy. Physics and Imaging in Radiation Oncology. 2018;5:31–6.CrossRef Kim J, Han JS, Hsia AT, Li SD, Xu ZG, Ryu S. Relationship between dosimetric leaf gap and dose calculation errors for high definition multi-leaf collimators in radiotherapy. Physics and Imaging in Radiation Oncology. 2018;5:31–6.CrossRef
11.
go back to reference Wen N, Zhao B, Kim J, Chin-Snyder K, Bellon M, Glide-Hurst C, Barton K, Chen D, Chetty IJ. IMRT and RapidArc commissioning of a TrueBeam linear accelerator using TG-119 protocol cases. J Appl Clin Med Phys. 2014;15(5):74–88.CrossRef Wen N, Zhao B, Kim J, Chin-Snyder K, Bellon M, Glide-Hurst C, Barton K, Chen D, Chetty IJ. IMRT and RapidArc commissioning of a TrueBeam linear accelerator using TG-119 protocol cases. J Appl Clin Med Phys. 2014;15(5):74–88.CrossRef
12.
go back to reference Chang Z, Wu Q, Adamson J, Ren L, Bowsher J, Yan H, Thomas A, Yin FF. Commissioning and dosimetric characteristics of TrueBeam system: composite data of three TrueBeam machines. Med Phys. 2012;39(11):6891–7018.CrossRef Chang Z, Wu Q, Adamson J, Ren L, Bowsher J, Yan H, Thomas A, Yin FF. Commissioning and dosimetric characteristics of TrueBeam system: composite data of three TrueBeam machines. Med Phys. 2012;39(11):6891–7018.CrossRef
13.
go back to reference Glide-Hurst C, Bellon M, Foster R, Altunbas C, Speiser M, Altman M, Westerly D, Wen N, Zhao B, Miften M, Chetty IJ, Solberg T. Commissioning of the Varian TrueBeam linear accelerator: a multi-institutional study. Med Phys. 2013;40(3):031719.CrossRef Glide-Hurst C, Bellon M, Foster R, Altunbas C, Speiser M, Altman M, Westerly D, Wen N, Zhao B, Miften M, Chetty IJ, Solberg T. Commissioning of the Varian TrueBeam linear accelerator: a multi-institutional study. Med Phys. 2013;40(3):031719.CrossRef
14.
go back to reference Wen N, Lu S, Kim J, Qin Y, Huang Y, Zhao B, Liu C, Chetty IJ. Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films. Radiat Oncol. 2016;11:132.CrossRef Wen N, Lu S, Kim J, Qin Y, Huang Y, Zhao B, Liu C, Chetty IJ. Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films. Radiat Oncol. 2016;11:132.CrossRef
15.
go back to reference Barbeiro AR, Ureba A, Baeza JA, Linares R, Perucha M, Jimenez-Ortega E, Velazquez S, Mateos JC, Leal A. 3D VMAT verification based on Monte Carlo log file simulation with experimental feedback from film dosimetry. PLoS One. 2016;11(11):e0166767.CrossRef Barbeiro AR, Ureba A, Baeza JA, Linares R, Perucha M, Jimenez-Ortega E, Velazquez S, Mateos JC, Leal A. 3D VMAT verification based on Monte Carlo log file simulation with experimental feedback from film dosimetry. PLoS One. 2016;11(11):e0166767.CrossRef
16.
go back to reference Li Y, Chen L, Zhu J, Liu X. The combination of the error correction methods of GAFCHROMIC EBT3 film. PLoS One. 2017;12(7):e0181958.CrossRef Li Y, Chen L, Zhu J, Liu X. The combination of the error correction methods of GAFCHROMIC EBT3 film. PLoS One. 2017;12(7):e0181958.CrossRef
17.
go back to reference Borca VC, Pasquino M, Russo G, Grosso P, Cante D, Sciacero P, Girelli G, La Porta MR, Tofani S. Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification. J Appl Clin Med Phys. 2013;14:158–71.CrossRef Borca VC, Pasquino M, Russo G, Grosso P, Cante D, Sciacero P, Girelli G, La Porta MR, Tofani S. Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification. J Appl Clin Med Phys. 2013;14:158–71.CrossRef
18.
go back to reference Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25:656–61.CrossRef Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25:656–61.CrossRef
19.
go back to reference Chandraraj V, Stathakis S, Manickam R, Esquivel E, Supe SS, Papanikolaou N. Comparison of four commercial devices for RapidArc and sliding window IMRT QA. J Appl Clin Med Phys. 2011;12:338–49.CrossRef Chandraraj V, Stathakis S, Manickam R, Esquivel E, Supe SS, Papanikolaou N. Comparison of four commercial devices for RapidArc and sliding window IMRT QA. J Appl Clin Med Phys. 2011;12:338–49.CrossRef
20.
go back to reference Li JG, Yan G, Liu C. Comparison of two commercial detector arrays for IMRT quality assurance. J Appl Clin Med Phys. 2009;10:62–74.CrossRef Li JG, Yan G, Liu C. Comparison of two commercial detector arrays for IMRT quality assurance. J Appl Clin Med Phys. 2009;10:62–74.CrossRef
21.
go back to reference Szpala S, Cao F, Kohli K. On using the dosimetric leaf gap to model the rounded leaf ends in VMAT/RapidArc plans. J Appl Clin Med Phys. 2014;15:67–84.CrossRef Szpala S, Cao F, Kohli K. On using the dosimetric leaf gap to model the rounded leaf ends in VMAT/RapidArc plans. J Appl Clin Med Phys. 2014;15:67–84.CrossRef
22.
go back to reference Kumaraswamy LK, Schmitt JD, Bailey DW, Xu ZZ, Podgorsak MB. Spatial variation of dosimetric leaf gap and its impact on dose delivery. Med Phys. 2014;41:111711.CrossRef Kumaraswamy LK, Schmitt JD, Bailey DW, Xu ZZ, Podgorsak MB. Spatial variation of dosimetric leaf gap and its impact on dose delivery. Med Phys. 2014;41:111711.CrossRef
23.
go back to reference Mullins J, DeBlois F, Syme A. Experimental characterization of the dosimetric leaf gap. Biomed Phys Eng Express. 2016;2:065013.CrossRef Mullins J, DeBlois F, Syme A. Experimental characterization of the dosimetric leaf gap. Biomed Phys Eng Express. 2016;2:065013.CrossRef
Metadata
Title
A simple method for determining dosimetric leaf gap with cross-field dose width for rounded leaf-end multileaf collimator systems
Authors
Chih-Yuan Lin
An-Cheng Shiau
Jin-Huei Ji
Chia-Jung Lee
Ti-Hao Wang
Shu-Hui Hsu
Ji-An Liang
Publication date
01-12-2018
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2018
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/s13014-018-1164-1

Other articles of this Issue 1/2018

Radiation Oncology 1/2018 Go to the issue