Primo software as a tool for Monte Carlo simulations of intensity modulated radiotherapy: a feasibility study

Authors: Alessandro Esposito, Sofia Silva, Jorge Oliveira, Joana Lencart, João Santos

Published in: Radiation Oncology | Issue 1/2018

Login to get access

Abstract

Background

IMRT provides higher dose conformation to the target and dose sparing to surrounding tissues than 3DCRT. Monte Carlo method in Medical Physics is not a novelty to approach dosimetric problems. A new PENELOPE based code named PRIMO recently was published. The most intriguing features of PRIMO are the user-friendly approach, the stand-alone property and the built-in definition of different linear accelerators models. Nevertheless, IMRT simulations are not yet implemented.

Methods

A Varian Trilogy with a Millennium120 MLC and a Varian Novalis with 120HD MLC were studied. A RW3 multi-slab phantom was irradiated with Gafchromic films inserted between slabs. An Expression 10000XL scanner (Seiko Epson Corp., Nagano, Japan) was used to digitalize the films. PTW-Verisoft software using the global Gamma Function (2%, 2 mm) was used to compare simulated and experimental results.
The primary beam parameters were adjusted to best match reference data previously obtained in a water phantom. Static MLC simulations were performed to validate the MLC models in use. Two Dynamic IMRT preliminary tests were performed with leaves moving with constant and variable speed. A further test of an in phantom delivery of a real IMRT field allowed simulating a clinical-like MLC modulation.

Results

Simulated PDD, X- and Y-profiles in reference conditions showed respectively 100.0%, 100.0% and 99.4% of Gamma points < 1 (2%, 2 mm). Static MLC simulations showed 100.0% of Gamma points < 1 with the 120HD MLC and 99.1% with the Millennium compared with the scanned images.
The fixed speed test showed 99.5 and 98.9% of Gamma points < 1 respectively with two different MLC configuration-sampling algorithms when the 120HD MLC was used. The higher modulation MLC motion simulation showed 99.1% of Gamma points < 1 with respect to the experimental. This result depends on the number of the fields to reproduce the MLC motion, as well as calculation time. The clinical-like simulation showed 96.2% of Gamma points < 1 using the same analysis conditions.

Conclusions

The numerical model of the Varian Trilogy and Novalis in the PRIMO software was validated. The algorithms to simulate MLC motion were considered reliable. A clinical-like procedure was successfully simulated.
Literature
1.
2.
go back to reference Burman C, et al. Planning, delivery, and quality assurance of intensity-modulated radiotherapy using dynamic multileaf collimator: A strategy for large-scale implementation for the treatment of carcinoma of the prostate. Int J Radiat Oncol Biol Phys. 1997;39:863–73.CrossRefPubMed Burman C, et al. Planning, delivery, and quality assurance of intensity-modulated radiotherapy using dynamic multileaf collimator: A strategy for large-scale implementation for the treatment of carcinoma of the prostate. Int J Radiat Oncol Biol Phys. 1997;39:863–73.CrossRefPubMed
3.
go back to reference Webb S. Intensity-modulated radiation therapy: Dynamic mlc (dmlc) therapy, multisegment therapy and tomo- therapy. an example of qa in dmlc therapy. Strahlenther Onkol. 1998;174:8–12.PubMed Webb S. Intensity-modulated radiation therapy: Dynamic mlc (dmlc) therapy, multisegment therapy and tomo- therapy. an example of qa in dmlc therapy. Strahlenther Onkol. 1998;174:8–12.PubMed
4.
go back to reference Boyer AL, Yu CX. Intensity-modulated radiation therapy with dynamic multileaf collimators. Semin Radiat Oncol. 1999;9:48–59.CrossRefPubMed Boyer AL, Yu CX. Intensity-modulated radiation therapy with dynamic multileaf collimators. Semin Radiat Oncol. 1999;9:48–59.CrossRefPubMed
5.
go back to reference Grein E, et al. An investigation of a new amorphous silicon electronic portal imaging device for transit dosimetry. Med Phys. 2002;29:2262–8.CrossRefPubMed Grein E, et al. An investigation of a new amorphous silicon electronic portal imaging device for transit dosimetry. Med Phys. 2002;29:2262–8.CrossRefPubMed
6.
go back to reference Greer PB, Popescu CC. Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapy. Med Phys. 2003;30:1618–27.CrossRefPubMed Greer PB, Popescu CC. Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapy. Med Phys. 2003;30:1618–27.CrossRefPubMed
7.
go back to reference Iori M, et al. Imat-sim: A new method for the clinical dosimetry of intensity-modulated arc therapy (imat). Med Phys. 2007;34:2759–73.CrossRefPubMed Iori M, et al. Imat-sim: A new method for the clinical dosimetry of intensity-modulated arc therapy (imat). Med Phys. 2007;34:2759–73.CrossRefPubMed
8.
go back to reference Pallotta S, et al. Design and implementation of a water phantom for imrt, arc therapy, and tomotherapy dose distribution measurements. Med Phys. 2007;34:3724–31.CrossRefPubMed Pallotta S, et al. Design and implementation of a water phantom for imrt, arc therapy, and tomotherapy dose distribution measurements. Med Phys. 2007;34:3724–31.CrossRefPubMed
9.
go back to reference Dong L, et al. Patient-specific point dose measurement for imrt monitor unit verification. Int J Radiat Oncol Biol Phys. 2003;56:867–77.CrossRefPubMed Dong L, et al. Patient-specific point dose measurement for imrt monitor unit verification. Int J Radiat Oncol Biol Phys. 2003;56:867–77.CrossRefPubMed
10.
go back to reference Ma CM, et al. A quality assurance phantom for imrt dose verification. Phys Med Biol. 2003;48:561–72.CrossRefPubMed Ma CM, et al. A quality assurance phantom for imrt dose verification. Phys Med Biol. 2003;48:561–72.CrossRefPubMed
11.
go back to reference Dobler B, et al. Hybrid plan verification for intensity-modulated radiation therapy (imrt) using the 2d ionization chamber array i’mrt matrixx–a feasibility study. Phys Med Biol. 2010;55:39–55.CrossRef Dobler B, et al. Hybrid plan verification for intensity-modulated radiation therapy (imrt) using the 2d ionization chamber array i’mrt matrixx–a feasibility study. Phys Med Biol. 2010;55:39–55.CrossRef
12.
go back to reference Bedford JL, et al. Evaluation of the Delta(4) phantom for IMRT and VMAT verification. Phys Med Biol. 2009;54:167–76.CrossRef Bedford JL, et al. Evaluation of the Delta(4) phantom for IMRT and VMAT verification. Phys Med Biol. 2009;54:167–76.CrossRef
13.
go back to reference Pethoukova AL, et al. The ArcCHECK diode array for dosimetric verification of HybridArc. Phys Med Biol. 2011;56:5411–28.CrossRef Pethoukova AL, et al. The ArcCHECK diode array for dosimetric verification of HybridArc. Phys Med Biol. 2011;56:5411–28.CrossRef
14.
go back to reference Stathakis S, et al. Characterization of a novel 2D array dosimeter for patient-specific quality assurance with volumetric arc therapy. Med Phys. 2013;40:071731.CrossRefPubMed Stathakis S, et al. Characterization of a novel 2D array dosimeter for patient-specific quality assurance with volumetric arc therapy. Med Phys. 2013;40:071731.CrossRefPubMed
15.
go back to reference Chetty I, et al. Report of the aapm task group no. 105: issues associated with clinical implementation of monte carlo-based photon and electron external beam treatment planning. Med Phys. 2007;34:4818–53.CrossRefPubMed Chetty I, et al. Report of the aapm task group no. 105: issues associated with clinical implementation of monte carlo-based photon and electron external beam treatment planning. Med Phys. 2007;34:4818–53.CrossRefPubMed
16.
go back to reference Reynaert N, et al. Monte Carlo treatment planning for photon and electron beams. Rad Phys Chem. 2007;76:643–86.CrossRef Reynaert N, et al. Monte Carlo treatment planning for photon and electron beams. Rad Phys Chem. 2007;76:643–86.CrossRef
17.
go back to reference Seco J, Verhagen F. Monte Carlo Techniques in Radiation Therapy: CRC Press; 1993. Seco J, Verhagen F. Monte Carlo Techniques in Radiation Therapy: CRC Press; 1993.
18.
go back to reference Verhaegen F, Seuntjens J. Monte carlo modelling of external radiotherapy photon beams. Phys Med Biol. 2003;48:107–64.CrossRef Verhaegen F, Seuntjens J. Monte carlo modelling of external radiotherapy photon beams. Phys Med Biol. 2003;48:107–64.CrossRef
19.
go back to reference Reynaert N. Mcde: A new monte carlo dose engine for imrt with an efficient scoring method. Phys Med Biol. 2004;49:235–41.CrossRef Reynaert N. Mcde: A new monte carlo dose engine for imrt with an efficient scoring method. Phys Med Biol. 2004;49:235–41.CrossRef
20.
go back to reference Agostinelli A et al.: Geant4--a simulation toolkit. Nuclear Instruments and Methods in Physics Research, A. 2003 506, 250–303 Agostinelli A et al.: Geant4--a simulation toolkit. Nuclear Instruments and Methods in Physics Research, A. 2003 506, 250–303
21.
go back to reference Rogers DWO, et al. BEAMnrc users manual. National Research Council of Canada Report. 2005;PIRS-0509(A):509. Rogers DWO, et al. BEAMnrc users manual. National Research Council of Canada Report. 2005;PIRS-0509(A):509.
22.
go back to reference Sempau J, et al. An algorithm for Monte Carlo simulation of coupled electron-photon transport Nucl. Instrum Meth B. 1997;132:377–90.CrossRef Sempau J, et al. An algorithm for Monte Carlo simulation of coupled electron-photon transport Nucl. Instrum Meth B. 1997;132:377–90.CrossRef
23.
go back to reference Fasso A et al.: FLUKA: a multi-particle transport code. CERN-2005- 10, INFN/TC_05/11, SLAC-R-773 (2005). Fasso A et al.: FLUKA: a multi-particle transport code. CERN-2005- 10, INFN/TC_05/11, SLAC-R-773 (2005).
24.
go back to reference Brown F B MCNP—a general Monte Carlo-particle transport code, version 5. Report LA-UR-03 1987 (Los Alamos National Laboratory, Los Alamos, NM) (2003). Brown F B MCNP—a general Monte Carlo-particle transport code, version 5. Report LA-UR-03 1987 (Los Alamos National Laboratory, Los Alamos, NM) (2003).
25.
go back to reference Rodriguez L, et al. Primo: A graphical environment for the monte carlo simulation of varian and elekta linacs. Strahlentherapie und Onkologie. 2013;189:881–6.CrossRefPubMed Rodriguez L, et al. Primo: A graphical environment for the monte carlo simulation of varian and elekta linacs. Strahlentherapie und Onkologie. 2013;189:881–6.CrossRefPubMed
26.
go back to reference Ma CM, et al. Monte carlo verification of imrt dose distributions from a commercial treatment planning optimization system. Phys Med Biol. 2000;45:2483–95.CrossRefPubMed Ma CM, et al. Monte carlo verification of imrt dose distributions from a commercial treatment planning optimization system. Phys Med Biol. 2000;45:2483–95.CrossRefPubMed
27.
go back to reference Leal A, et al. Routine imrt verification by means of an automated monte carlo simulation system. Int J Radiat Oncol Biol Phys. 2003;56:58–68.CrossRefPubMed Leal A, et al. Routine imrt verification by means of an automated monte carlo simulation system. Int J Radiat Oncol Biol Phys. 2003;56:58–68.CrossRefPubMed
28.
go back to reference Seco J, et al. Head & neck imrt treatment assessed with a monte carlo dose calculation engine. Phys Med Biol. 2005;50:817–30.CrossRefPubMed Seco J, et al. Head & neck imrt treatment assessed with a monte carlo dose calculation engine. Phys Med Biol. 2005;50:817–30.CrossRefPubMed
29.
go back to reference Shih R, et al. Dynamic wedge versus physical wedges: A monte carlo study. Med Phys. 2001;28:612–9.CrossRefPubMed Shih R, et al. Dynamic wedge versus physical wedges: A monte carlo study. Med Phys. 2001;28:612–9.CrossRefPubMed
30.
go back to reference Liu HH, et al. A method of simulating dynamic multileaf collimator using monte carlo techniques for intensity-modulated radiation therapy. Phys Med Biol. 2001;46:2283–98.CrossRefPubMed Liu HH, et al. A method of simulating dynamic multileaf collimator using monte carlo techniques for intensity-modulated radiation therapy. Phys Med Biol. 2001;46:2283–98.CrossRefPubMed
31.
go back to reference Heath E, et al. A direct voxel tracking method for four-dimensional monte carlo dose calculations in deforming anatomy. Med Phys. 2006;33:434–45.CrossRefPubMed Heath E, et al. A direct voxel tracking method for four-dimensional monte carlo dose calculations in deforming anatomy. Med Phys. 2006;33:434–45.CrossRefPubMed
32.
go back to reference Low DA, et al. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25:656–61.CrossRefPubMed Low DA, et al. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25:656–61.CrossRefPubMed
Metadata
Title
Primo software as a tool for Monte Carlo simulations of intensity modulated radiotherapy: a feasibility study
Authors
Alessandro Esposito
Sofia Silva
Jorge Oliveira
Joana Lencart
João Santos
Publication date
01-12-2018
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2018
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/s13014-018-1021-2

Other articles of this Issue 1/2018

Radiation Oncology 1/2018 Go to the issue