Optimization of the prescription isodose line for Gamma Knife radiosurgery using the shot within shot technique

Authors: Perry B. Johnson, Maria I. Monterroso, Fei Yang, Eric Mellon

Published in: Radiation Oncology | Issue 1/2017

Login to get access

Abstract

Background

This work explores how the choice of prescription isodose line (IDL) affects the dose gradient, target coverage, and treatment time for Gamma Knife radiosurgery when a smaller shot is encompassed within a larger shot at the same stereotactic coordinates (shot within shot technique).

Methods

Beam profiles for the 4, 8, and 16 mm collimator settings were extracted from the treatment planning system and characterized using Gaussian fits. The characterized data were used to create over 10,000 shot within shot configurations by systematically changing collimator weighting and choice of prescription IDL. Each configuration was quantified in terms of the dose gradient, target coverage, and beam-on time. By analyzing these configurations, it was found that there are regions of overlap in target size where a higher prescription IDL provides equivalent dose fall-off to a plan prescribed at the 50% IDL. Furthermore, the data indicate that treatment times within these regions can be reduced by up to 40%. An optimization strategy was devised to realize these gains. The strategy was tested for seven patients treated for 1–4 brain metastases (20 lesions total).

Results

For a single collimator setting, the gradient in the axial plane was steepest when prescribed to the 56–63% (4 mm), 62–70% (8 mm), and 77–84% (16 mm) IDL, respectively. Through utilization of the optimization technique, beam-on time was reduced by more than 15% in 16/20 lesions. The volume of normal brain receiving 12 Gy or above also decreased in many cases, and in only one instance increased by more than 0.5 cm3.

Conclusions

This work demonstrates that IDL optimization using the shot within shot technique can reduce treatment times without degrading treatment plan quality.
Appendix
Available only for authorised users
Literature
1.
go back to reference Paddick I, Lippitz B. A simple dose gradient measurement tool to complement the conformity index. J Neurosurg. 2006;105(Suppl 1):194–201.PubMed Paddick I, Lippitz B. A simple dose gradient measurement tool to complement the conformity index. J Neurosurg. 2006;105(Suppl 1):194–201.PubMed
2.
go back to reference Korytko T, Radivoyevitch T, Colussi V, et al. 12 Gy gamma knife radiosurgical volume is a predictor for radiation necrosis in non-avm intracranial tumors. Int J Radiat Oncol Biol Phys. 2006;64(2):419–24.CrossRefPubMed Korytko T, Radivoyevitch T, Colussi V, et al. 12 Gy gamma knife radiosurgical volume is a predictor for radiation necrosis in non-avm intracranial tumors. Int J Radiat Oncol Biol Phys. 2006;64(2):419–24.CrossRefPubMed
4.
go back to reference Novotny J, Bhatnagar J, Niranjan A, et al. Dosimetric comparison of the Leksell gamma knife Perfexion and 4C. J Neurosurg. 2008;109:8–14.PubMed Novotny J, Bhatnagar J, Niranjan A, et al. Dosimetric comparison of the Leksell gamma knife Perfexion and 4C. J Neurosurg. 2008;109:8–14.PubMed
5.
go back to reference Petti P, Larson D, Kunwar S. Use of hybrid shots in planning Perfexion gamma knife treatments for lesions close to critical structures. J Neurosurg. 2008;109:34–40.PubMed Petti P, Larson D, Kunwar S. Use of hybrid shots in planning Perfexion gamma knife treatments for lesions close to critical structures. J Neurosurg. 2008;109:34–40.PubMed
6.
go back to reference Flickinger JC, Kondziolka D, Pollock BE, et al. Complications from arteriovenous malformation radiosurgery: multivariate analysis and risk modeling. Int J Radiat Oncol Biol Phys. 1997;38:489–90.CrossRef Flickinger JC, Kondziolka D, Pollock BE, et al. Complications from arteriovenous malformation radiosurgery: multivariate analysis and risk modeling. Int J Radiat Oncol Biol Phys. 1997;38:489–90.CrossRef
7.
go back to reference Nakamura JL, Verhey LJ, Smith V, et al. Dose conformity of gamma knife radiosurgery and risk factors for complications. Int J Radiat Biol Phys. 2003;81:115–9. Nakamura JL, Verhey LJ, Smith V, et al. Dose conformity of gamma knife radiosurgery and risk factors for complications. Int J Radiat Biol Phys. 2003;81:115–9.
8.
go back to reference Nuclear Regulatory Commission. Safety precautions for remote afterloader units, teletherapy units, and gamma stereotactic radiosurgery units. 10 CFR 35.615. 2015. Nuclear Regulatory Commission. Safety precautions for remote afterloader units, teletherapy units, and gamma stereotactic radiosurgery units. 10 CFR 35.615. 2015.
9.
go back to reference Meeks S, Buatti J, Bova F, et al. Treatment planning optimization for linear accelerator radiosurgery. Int J Radiat Oncol Biol Phys. 1998;41(1):183–97.CrossRefPubMed Meeks S, Buatti J, Bova F, et al. Treatment planning optimization for linear accelerator radiosurgery. Int J Radiat Oncol Biol Phys. 1998;41(1):183–97.CrossRefPubMed
10.
go back to reference Sio T, Jang S, Lee SW, et al. Comparing gamma knife and Cyberknife in patients with brain metastases. J App Clin Med Phys. 2014;15(1):14–26.CrossRef Sio T, Jang S, Lee SW, et al. Comparing gamma knife and Cyberknife in patients with brain metastases. J App Clin Med Phys. 2014;15(1):14–26.CrossRef
11.
go back to reference Andrews DW, Scott CB, Sperduto PW, et al. Whole brain radiation therapy with or without stereotactic radiosurgery boost for patients with one to three brain metastases: phase III results of the RTOG 9508 randomised trial. Lancet. 2004;363:1665–72.CrossRefPubMed Andrews DW, Scott CB, Sperduto PW, et al. Whole brain radiation therapy with or without stereotactic radiosurgery boost for patients with one to three brain metastases: phase III results of the RTOG 9508 randomised trial. Lancet. 2004;363:1665–72.CrossRefPubMed
12.
go back to reference Shiue K, Barnett G, Suh J, et al. Using higher isodose lines for gamma knife treatment of 1 to 3 brain metastases is safe and effective. Neurosurgery. 2014;74(4):360–6.CrossRefPubMed Shiue K, Barnett G, Suh J, et al. Using higher isodose lines for gamma knife treatment of 1 to 3 brain metastases is safe and effective. Neurosurgery. 2014;74(4):360–6.CrossRefPubMed
13.
go back to reference Romano K, Trifiletti D, Garda A, et al. Choosing a prescription isodose in stereotactic radiosurgery for brain metastases: implications for local control. World Neurosurgery. 2017;98:761–7.CrossRefPubMed Romano K, Trifiletti D, Garda A, et al. Choosing a prescription isodose in stereotactic radiosurgery for brain metastases: implications for local control. World Neurosurgery. 2017;98:761–7.CrossRefPubMed
Metadata
Title
Optimization of the prescription isodose line for Gamma Knife radiosurgery using the shot within shot technique
Authors
Perry B. Johnson
Maria I. Monterroso
Fei Yang
Eric Mellon
Publication date
01-12-2017
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2017
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/s13014-017-0919-4

Other articles of this Issue 1/2017

Radiation Oncology 1/2017 Go to the issue