Open Access 01-12-2015 | Research

Comparative dosimetric findings using accelerated partial breast irradiation across five catheter subtypes

Authors: Zaker Rana, Nadim M. Nasr, Huaying Ji, Virginia Lorio, Stephanie Akbari, Molly Sebastian, Mami Martin, Robert L. Hong

Published in: Radiation Oncology | Issue 1/2015

Login to get access

Abstract

Purpose

Accelerated partial breast irradiation (APBI) with balloon and strut adjusted volume implants (SAVI) show promising results with excellent tumor control and minimal toxicity. Knowing the factors that contribute to a high skin dose, rib dose, and D95 coverage may reduce toxicity, improve tumor control, and help properly predict patient outcomes following APBI.

Methods and materials

A retrospective analysis of 594 patients treated with brachytherapy based APBI at a single institution from May 2008 to September 2014 was grouped by applicator subtype. Patients were treated to a total of 34 Gy (3.4 Gy x 10 fractions over 5 days delivered BID) targeting a planning target volume (PTV) 1.0 cm beyond the lumpectomy cavity using a high dose rate source.

Results

SAVI devices had the lowest statistically significant values of DmaxSkin (81.00 ± 29.83), highest values of D90 (101.50 ± 3.66), and D95 (96.09 ± 4.55). SAVI-mini devices had the lowest statistically significant values of DmaxRib (77.66 ± 32.92) and smallest V150 (18.01 ± 3.39). Multi-lumen balloons were able to obtain the smallest V200 (5.89 ± 2.21). Strut-based applicators were more likely to achieve a DmaxSkin and a DmaxRib less than or equal to 100 %. The effect of PTV on V150 showed a strong positive relationship (p < .001). PTV and DmaxSkin showed a weak negative relationship in multi-lumen applicators (p = .016) and SAVI-mini devices (p < .001). PTV and DmaxRib showed a weak negative relationship in multi-lumen applicators (p = .009), SAVI devices (p < .001), and SAVI-mini devices (p < .001).

Conclusion

PTV volume is strongly correlated with V150 in all devices and V200 in strut based devices. Larger PTV volumes result in greater V150 and V200, which could help predict potential risks for hotspots and resulting toxicities in these devices. PTV volume is also weakly negatively correlated with max skin dose and max rib dose, meaning that as the PTV volumes increase one can expect slightly smaller max skin and rib doses. Strut based applicators are significantly more effective in keeping skin and rib dose constraints under 125 and 100 % when compared to any balloon based applicator.
Literature
3.
go back to reference Cuttino LW, Heffernan J, Vera R, Rosu M, Ramakrishnan VR, Arthur DW. Association between maximal skin dose and breast brachytherapy outcome: a proposal for more rigorous dosimetric constraints. Int J Radiat Oncol Biol Phys. 2011;81(3):e173–7. doi:10.1016/j.ijrobp.2010.12.023.PubMedCrossRef Cuttino LW, Heffernan J, Vera R, Rosu M, Ramakrishnan VR, Arthur DW. Association between maximal skin dose and breast brachytherapy outcome: a proposal for more rigorous dosimetric constraints. Int J Radiat Oncol Biol Phys. 2011;81(3):e173–7. doi:10.​1016/​j.​ijrobp.​2010.​12.​023.PubMedCrossRef
4.
go back to reference King TA, Bolton JS, Kuske RR, Fuhrman GM, Scroggins TG, Jiang XZ. Long-term results of wide-field brachytherapy as the sole method of radiation therapy after segmental mastectomy for T(is,1,2) breast cancer. Am J Surg. 2000;180(4):299–304.PubMedCrossRef King TA, Bolton JS, Kuske RR, Fuhrman GM, Scroggins TG, Jiang XZ. Long-term results of wide-field brachytherapy as the sole method of radiation therapy after segmental mastectomy for T(is,1,2) breast cancer. Am J Surg. 2000;180(4):299–304.PubMedCrossRef
5.
go back to reference Vicini FA, Baglan KL, Kestin LL, Mitchell C, Chen PY, Frazier RC, et al. Accelerated treatment of breast cancer. J Clin Oncol. 2001;19(7):1993–2001.PubMed Vicini FA, Baglan KL, Kestin LL, Mitchell C, Chen PY, Frazier RC, et al. Accelerated treatment of breast cancer. J Clin Oncol. 2001;19(7):1993–2001.PubMed
6.
go back to reference Polgar C, Fodor J, Major T, Nemeth G, Lovey K, Orosz Z, et al. Breast-conserving treatment with partial or whole breast irradiation for low-risk invasive breast carcinoma--5-year results of a randomized trial. Int J Radiat Oncol Biol Phys. 2007;69(3):694–702. doi:10.1016/j.ijrobp.2007.04.022.PubMedCrossRef Polgar C, Fodor J, Major T, Nemeth G, Lovey K, Orosz Z, et al. Breast-conserving treatment with partial or whole breast irradiation for low-risk invasive breast carcinoma--5-year results of a randomized trial. Int J Radiat Oncol Biol Phys. 2007;69(3):694–702. doi:10.​1016/​j.​ijrobp.​2007.​04.​022.PubMedCrossRef
7.
go back to reference Antonucci JV, Wallace M, Goldstein NS, Kestin L, Chen P, Benitez P, et al. Differences in patterns of failure in patients treated with accelerated partial breast irradiation versus whole-breast irradiation: a matched-pair analysis with 10-year follow-up. Int J Radiat Oncol Biol Phys. 2009;74(2):447–52. doi:10.1016/j.ijrobp.2008.08.025.PubMedCrossRef Antonucci JV, Wallace M, Goldstein NS, Kestin L, Chen P, Benitez P, et al. Differences in patterns of failure in patients treated with accelerated partial breast irradiation versus whole-breast irradiation: a matched-pair analysis with 10-year follow-up. Int J Radiat Oncol Biol Phys. 2009;74(2):447–52. doi:10.​1016/​j.​ijrobp.​2008.​08.​025.PubMedCrossRef
10.
go back to reference Ott OJ, Hildebrandt G, Potter R, Hammer J, Lotter M, Resch A, et al. Accelerated partial breast irradiation with multi-catheter brachytherapy: Local control, side effects and cosmetic outcome for 274 patients. Results of the German-Austrian multi-centre trial. Radiother Oncol. 2007;82(3):281–6. doi:10.1016/j.radonc.2006.08.028.PubMedCrossRef Ott OJ, Hildebrandt G, Potter R, Hammer J, Lotter M, Resch A, et al. Accelerated partial breast irradiation with multi-catheter brachytherapy: Local control, side effects and cosmetic outcome for 274 patients. Results of the German-Austrian multi-centre trial. Radiother Oncol. 2007;82(3):281–6. doi:10.​1016/​j.​radonc.​2006.​08.​028.PubMedCrossRef
11.
go back to reference Benitez PR, Keisch ME, Vicini F, Stolier A, Scroggins T, Walker A, et al. Five-year results: the initial clinical trial of MammoSite balloon brachytherapy for partial breast irradiation in early-stage breast cancer. Am J Surg. 2007;194(4):456–62. doi:10.1016/j.amjsurg.2007.06.010.PubMedCrossRef Benitez PR, Keisch ME, Vicini F, Stolier A, Scroggins T, Walker A, et al. Five-year results: the initial clinical trial of MammoSite balloon brachytherapy for partial breast irradiation in early-stage breast cancer. Am J Surg. 2007;194(4):456–62. doi:10.​1016/​j.​amjsurg.​2007.​06.​010.PubMedCrossRef
12.
go back to reference Dragun AE, Harper JL, Jenrette JM, Sinha D, Cole DJ. Predictors of cosmetic outcome following MammoSite breast brachytherapy: a single-institution experience of 100 patients with two years of follow-up. Int J Radiat Oncol Biol Phys. 2007;68(2):354–8. doi:10.1016/j.ijrobp.2006.12.014.PubMedCrossRef Dragun AE, Harper JL, Jenrette JM, Sinha D, Cole DJ. Predictors of cosmetic outcome following MammoSite breast brachytherapy: a single-institution experience of 100 patients with two years of follow-up. Int J Radiat Oncol Biol Phys. 2007;68(2):354–8. doi:10.​1016/​j.​ijrobp.​2006.​12.​014.PubMedCrossRef
13.
go back to reference Niehoff P, Polgar C, Ostertag H, Major T, Sulyok Z, Kimmig B, et al. Clinical experience with the MammoSite radiation therapy system for brachytherapy of breast cancer: results from an international phase II trial. Radiother Oncol. 2006;79(3):316–20. doi:10.1016/j.radonc.2006.05.010.PubMedCrossRef Niehoff P, Polgar C, Ostertag H, Major T, Sulyok Z, Kimmig B, et al. Clinical experience with the MammoSite radiation therapy system for brachytherapy of breast cancer: results from an international phase II trial. Radiother Oncol. 2006;79(3):316–20. doi:10.​1016/​j.​radonc.​2006.​05.​010.PubMedCrossRef
14.
go back to reference Vicini F, Beitsch PD, Quiet CA, Keleher AJ, Garcia D, Snider Jr HC, et al. Three-year analysis of treatment efficacy, cosmesis, and toxicity by the American Society of Breast Surgeons MammoSite Breast Brachytherapy Registry Trial in patients treated with accelerated partial breast irradiation (APBI). Cancer. 2008;112(4):758–66. doi:10.1002/cncr.23227.PubMedCrossRef Vicini F, Beitsch PD, Quiet CA, Keleher AJ, Garcia D, Snider Jr HC, et al. Three-year analysis of treatment efficacy, cosmesis, and toxicity by the American Society of Breast Surgeons MammoSite Breast Brachytherapy Registry Trial in patients treated with accelerated partial breast irradiation (APBI). Cancer. 2008;112(4):758–66. doi:10.​1002/​cncr.​23227.PubMedCrossRef
15.
go back to reference Khan AJ, Arthur D, Vicini F, Beitsch P, Kuerer H, Goyal S, et al. Six-year analysis of treatment-related toxicities in patients treated with accelerated partial breast irradiation on the American Society of Breast Surgeons MammoSite Breast Brachytherapy registry trial. Ann Surg Oncol. 2012;19(5):1477–83. doi:10.1245/s10434-011-2133-1.PubMedCrossRef Khan AJ, Arthur D, Vicini F, Beitsch P, Kuerer H, Goyal S, et al. Six-year analysis of treatment-related toxicities in patients treated with accelerated partial breast irradiation on the American Society of Breast Surgeons MammoSite Breast Brachytherapy registry trial. Ann Surg Oncol. 2012;19(5):1477–83. doi:10.​1245/​s10434-011-2133-1.PubMedCrossRef
16.
go back to reference Vicini F, Beitsch P, Quiet C, Gittleman M, Zannis V, Fine R, et al. Five-year analysis of treatment efficacy and cosmesis by the American Society of Breast Surgeons MammoSite Breast Brachytherapy Registry Trial in patients treated with accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys. 2011;79(3):808–17. doi:10.1016/j.ijrobp.2009.11.043.PubMedCrossRef Vicini F, Beitsch P, Quiet C, Gittleman M, Zannis V, Fine R, et al. Five-year analysis of treatment efficacy and cosmesis by the American Society of Breast Surgeons MammoSite Breast Brachytherapy Registry Trial in patients treated with accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys. 2011;79(3):808–17. doi:10.​1016/​j.​ijrobp.​2009.​11.​043.PubMedCrossRef
19.
go back to reference Kim Y, Trombetta M. Dosimetric evaluation of MammoSite® multi-lumen balloon applicator compared with contura® applicator for high dose rate (HDR) breast brachytherapy. Int J Radiat Oncol Biol Phys. 2011;81(2):902–3.CrossRef Kim Y, Trombetta M. Dosimetric evaluation of MammoSite® multi-lumen balloon applicator compared with contura® applicator for high dose rate (HDR) breast brachytherapy. Int J Radiat Oncol Biol Phys. 2011;81(2):902–3.CrossRef
20.
go back to reference Brown S, McLaughlin M, Pope K, Haile K, Hughes L, Israel PZ. Initial radiation experience evaluating early tolerance and toxicities in patients undergoing accelerated partial breast irradiation using the Contura Multi-Lumen Balloon breast brachytherapy catheter. Brachytherapy. 2009;8(2):227–33. doi:10.1016/j.brachy.2008.11.013.PubMedCrossRef Brown S, McLaughlin M, Pope K, Haile K, Hughes L, Israel PZ. Initial radiation experience evaluating early tolerance and toxicities in patients undergoing accelerated partial breast irradiation using the Contura Multi-Lumen Balloon breast brachytherapy catheter. Brachytherapy. 2009;8(2):227–33. doi:10.​1016/​j.​brachy.​2008.​11.​013.PubMedCrossRef
23.
go back to reference Vargo JA, Verma V, Kim H, Kalash R, Heron DE, Johnson R, et al. Extended (5-year) outcomes of accelerated partial breast irradiation using MammoSite balloon brachytherapy: patterns of failure, patient selection, and dosimetric correlates for late toxicity. Int J Radiat Oncol Biol Phys. 2014;88(2):285–91. doi:10.1016/j.ijrobp.2013.05.039.PubMedCrossRef Vargo JA, Verma V, Kim H, Kalash R, Heron DE, Johnson R, et al. Extended (5-year) outcomes of accelerated partial breast irradiation using MammoSite balloon brachytherapy: patterns of failure, patient selection, and dosimetric correlates for late toxicity. Int J Radiat Oncol Biol Phys. 2014;88(2):285–91. doi:10.​1016/​j.​ijrobp.​2013.​05.​039.PubMedCrossRef
25.
27.
go back to reference Polgar C, Van Limbergen E, Potter R, Kovacs G, Polo A, Lyczek J, et al. Patient selection for accelerated partial-breast irradiation (APBI) after breast-conserving surgery: recommendations of the Groupe Europeen de Curietherapie-European Society for Therapeutic Radiology and Oncology (GEC-ESTRO) breast cancer working group based on clinical evidence (2009). Radiother Oncol. 2010;94(3):264–73. doi:10.1016/j.radonc.2010.01.014.PubMedCrossRef Polgar C, Van Limbergen E, Potter R, Kovacs G, Polo A, Lyczek J, et al. Patient selection for accelerated partial-breast irradiation (APBI) after breast-conserving surgery: recommendations of the Groupe Europeen de Curietherapie-European Society for Therapeutic Radiology and Oncology (GEC-ESTRO) breast cancer working group based on clinical evidence (2009). Radiother Oncol. 2010;94(3):264–73. doi:10.​1016/​j.​radonc.​2010.​01.​014.PubMedCrossRef
31.
go back to reference Wazer DE, Kaufman S, Cuttino L, DiPetrillo T, Arthur DW. Accelerated partial breast irradiation: an analysis of variables associated with late toxicity and long-term cosmetic outcome after high-dose-rate interstitial brachytherapy. Int J Radiat Oncol Biol Phys. 2006;64(2):489–95. doi:10.1016/j.ijrobp.2005.06.028.PubMedCrossRef Wazer DE, Kaufman S, Cuttino L, DiPetrillo T, Arthur DW. Accelerated partial breast irradiation: an analysis of variables associated with late toxicity and long-term cosmetic outcome after high-dose-rate interstitial brachytherapy. Int J Radiat Oncol Biol Phys. 2006;64(2):489–95. doi:10.​1016/​j.​ijrobp.​2005.​06.​028.PubMedCrossRef
34.
go back to reference Chao KK, Vicini FA, Wallace M, Mitchell C, Chen P, Ghilezan M, et al. Analysis of treatment efficacy, cosmesis, and toxicity using the MammoSite breast brachytherapy catheter to deliver accelerated partial-breast irradiation: the William Beaumont Hospital experience. Int J Radiat Oncol Biol Phys. 2007;69(1):32–40. doi:10.1016/j.ijrobp.2007.02.026.PubMedCrossRef Chao KK, Vicini FA, Wallace M, Mitchell C, Chen P, Ghilezan M, et al. Analysis of treatment efficacy, cosmesis, and toxicity using the MammoSite breast brachytherapy catheter to deliver accelerated partial-breast irradiation: the William Beaumont Hospital experience. Int J Radiat Oncol Biol Phys. 2007;69(1):32–40. doi:10.​1016/​j.​ijrobp.​2007.​02.​026.PubMedCrossRef
35.
go back to reference Brown S, McLaughlin M, Pope DK, Haile K, Hughes L, Israel PZ, et al. A dosimetric comparison of the Contura multilumen balloon breast brachytherapy catheter vs. the single-lumen MammoSite balloon device in patients treated with accelerated partial breast irradiation at a single institution. Brachytherapy. 2011;10(1):68–73. doi:10.1016/j.brachy.2010.01.001.PubMedCrossRef Brown S, McLaughlin M, Pope DK, Haile K, Hughes L, Israel PZ, et al. A dosimetric comparison of the Contura multilumen balloon breast brachytherapy catheter vs. the single-lumen MammoSite balloon device in patients treated with accelerated partial breast irradiation at a single institution. Brachytherapy. 2011;10(1):68–73. doi:10.​1016/​j.​brachy.​2010.​01.​001.PubMedCrossRef
36.
go back to reference Patel R, Wong G, Tran L, et al. Multi-lumen hybrid breast brachytherapy: Strut versus balloon—Dosimetric results in first 100 patients. Brachytherapy. 2010;9:S32–S3.CrossRef Patel R, Wong G, Tran L, et al. Multi-lumen hybrid breast brachytherapy: Strut versus balloon—Dosimetric results in first 100 patients. Brachytherapy. 2010;9:S32–S3.CrossRef
Metadata
Title
Comparative dosimetric findings using accelerated partial breast irradiation across five catheter subtypes
Authors
Zaker Rana
Nadim M. Nasr
Huaying Ji
Virginia Lorio
Stephanie Akbari
Molly Sebastian
Mami Martin
Robert L. Hong
Publication date
01-12-2015
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2015
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/s13014-015-0468-7

Other articles of this Issue 1/2015

Radiation Oncology 1/2015 Go to the issue