Artefacts in 1.5 Tesla and 3 Tesla cardiovascular magnetic resonance imaging in patients with leadless cardiac pacemakers

Authors: Daniel Kiblboeck, Christian Reiter, Juergen Kammler, Pierre Schmit, Hermann Blessberger, Joerg Kellermair, Franz Fellner, Clemens Steinwender

Published in: Journal of Cardiovascular Magnetic Resonance | Issue 1/2018

Login to get access

Abstract

Background

There are limited data on patients with leadless cardiac pacemakers (LCP) undergoing magnetic resonance imaging. The aim of this prospective, single-center, observational study was to evaluate artefacts on cardiovascular magnetic resonance (CMR) images in patients with LCP.

Methods

Fifteen patients with Micra™ LCP, implanted at least 6 weeks prior to CMR scan, were enrolled and underwent either 1.5 Tesla or 3 Tesla CMR imaging. Artefacts were categorized into grade 1 (excellent image quality), grade 2 (good), grade 3 (poor) and grade 4 (non-diagnostic) for each myocardial segment. One patient was excluded because of an incomplete CMR investigation due to claustrophobia.

Results

LCP caused an arc-shaped artefact (0.99 ± 0.16 cm2) at the right ventricular (RV) apex. Of 224 analyzed myocardial segments of the left ventricle (LV) 158 (70.5%) were affected by grade 1, 27 (12.1%) by grade 2, 17 (7.6%) by grade 3 and 22 (9.8%) by grade 4 artefacts. The artefact burden of grade 3 and 4 artefacts was significantly higher in the 3 Tesla group (3 Tesla vs 1.5 Tesla: 3.7 ± 1.6 vs 1.9 ± 1.4 myocardial segments per patient, p = 0.03). A high artefact burden was particularly observed in the mid anteroseptal, inferoseptal and apical septal myocardial segments of the LV and in the mid and apical segments of the RV. Quantification of LV function and assessment of valves were feasible in all patients. We did not observe any clinical or device-related adverse events.

Conclusion

CMR imaging in patients with LCP is feasible with excellent to good image quality in the majority of LV segments. The artefact burden is comparable small allowing an accurate evaluation of LV function, cardiac structures and valves. However, artefacts in the mid anteroseptal, inferoseptal and apical septal myocardial segments of the LV due to the LCP may impair or even exclude diagnostic evaluation of these segments. Artefacts on CMR images may be reduced by the use of 1.5 Tesla CMR scanners.
Literature
1.
go back to reference Penell DJ. Cardiovascular magnetic resonance. Circulation. 2010;121(5):692–705.CrossRef Penell DJ. Cardiovascular magnetic resonance. Circulation. 2010;121(5):692–705.CrossRef
2.
go back to reference Brignole M, Auricchio A, Baron-Esquivias G, Bordachar P, Boriani G, Breithardt OA, et al. 2013 ESC guidelines on cardiac pacing and cardiac resynchronization therapy. Europace. 2013;15(8):1070–118.CrossRefPubMed Brignole M, Auricchio A, Baron-Esquivias G, Bordachar P, Boriani G, Breithardt OA, et al. 2013 ESC guidelines on cardiac pacing and cardiac resynchronization therapy. Europace. 2013;15(8):1070–118.CrossRefPubMed
4.
go back to reference Russo RJ, Costa HS, Silva PD, Anderson JL, Arshad A, Biederman RW, et al. Assessing the risks associated with MRI in patients with a pacemaker or defibrillator. N Engl J Med. 2017;376(8):755–64.CrossRefPubMed Russo RJ, Costa HS, Silva PD, Anderson JL, Arshad A, Biederman RW, et al. Assessing the risks associated with MRI in patients with a pacemaker or defibrillator. N Engl J Med. 2017;376(8):755–64.CrossRefPubMed
5.
go back to reference Strom JB, Whelan JB, Shen C, Zheng SQ, Mortele KJ, Kramer DB. Safety and utility of magnetic resonance imaging in patients with cardiac implantable electronic devices. Heart Rhythm. 2017;14(8):1138–44.CrossRefPubMedPubMedCentral Strom JB, Whelan JB, Shen C, Zheng SQ, Mortele KJ, Kramer DB. Safety and utility of magnetic resonance imaging in patients with cardiac implantable electronic devices. Heart Rhythm. 2017;14(8):1138–44.CrossRefPubMedPubMedCentral
6.
go back to reference Shah AD, Patel AU, Knezevic A, Hoskins MH, Hirsh DS, Merchant FM, et al. Clinical performance of magnetic resonance imaging conditional and nonconditional cardiac implantable electronic devices. Pacing Clin Electrophysiol. 2017;40(5):467–75.CrossRefPubMed Shah AD, Patel AU, Knezevic A, Hoskins MH, Hirsh DS, Merchant FM, et al. Clinical performance of magnetic resonance imaging conditional and nonconditional cardiac implantable electronic devices. Pacing Clin Electrophysiol. 2017;40(5):467–75.CrossRefPubMed
7.
go back to reference Bertelsen L, Petersen HH, Philbert BT, Svendsen JH, Thomsen C, Vejlstrup N. Safety of magnetic resonance scanning without monitoring of patients with pacemakers. Europace. 2017;19(5):818–23.PubMed Bertelsen L, Petersen HH, Philbert BT, Svendsen JH, Thomsen C, Vejlstrup N. Safety of magnetic resonance scanning without monitoring of patients with pacemakers. Europace. 2017;19(5):818–23.PubMed
8.
go back to reference Camacho JC, Moreno CC, Shah AD, Mittal PK, Mengistu A, Lloyd MS, et al. Safety and quality of 1.5-T MRI in patients with conventional and MRI-conditional cardiac implantable electronic devices after implementation of a standardized protocol. AJR Am J Roentgenol. 2016;207(3):599–604.CrossRefPubMed Camacho JC, Moreno CC, Shah AD, Mittal PK, Mengistu A, Lloyd MS, et al. Safety and quality of 1.5-T MRI in patients with conventional and MRI-conditional cardiac implantable electronic devices after implementation of a standardized protocol. AJR Am J Roentgenol. 2016;207(3):599–604.CrossRefPubMed
9.
go back to reference Horwood L, Attili A, Luba F, Ibrahim EH, Parmar H, Stojanovska J, et al. Magnetic resonance imaging in patients with cardiac implanted electronic devices: focus on contraindications to magnetic resonance imaging protocols. Europace. 2017;19(5):812–7.PubMed Horwood L, Attili A, Luba F, Ibrahim EH, Parmar H, Stojanovska J, et al. Magnetic resonance imaging in patients with cardiac implanted electronic devices: focus on contraindications to magnetic resonance imaging protocols. Europace. 2017;19(5):812–7.PubMed
10.
go back to reference Lowe MD, Plummer CJ, Mainsty CH, Linker NJ, British Heart Rhythm Society. Safe use of MRI in people with cardiac implantable electronic devices. Heart. 2015;101(24):1950–3.CrossRefPubMed Lowe MD, Plummer CJ, Mainsty CH, Linker NJ, British Heart Rhythm Society. Safe use of MRI in people with cardiac implantable electronic devices. Heart. 2015;101(24):1950–3.CrossRefPubMed
11.
go back to reference Kypta A, Blessberger H, Hoenig S, Saleh K, Lambert T, Kammler J, et al. Clinical safety of an MRI conditional implantable cardioverter defibrillator system: a prospective monocenter ICD-magnetic resonance imaging feasibility study (MIMI). J Magn Reson Imaging. 2016;43(3):574–84.CrossRefPubMed Kypta A, Blessberger H, Hoenig S, Saleh K, Lambert T, Kammler J, et al. Clinical safety of an MRI conditional implantable cardioverter defibrillator system: a prospective monocenter ICD-magnetic resonance imaging feasibility study (MIMI). J Magn Reson Imaging. 2016;43(3):574–84.CrossRefPubMed
12.
go back to reference Tjong FV, Reddy VY. Permanent leadless cardiac pacemaker therapy: a comprehensive review. Circulation. 2017;135(15):1458–70.CrossRefPubMed Tjong FV, Reddy VY. Permanent leadless cardiac pacemaker therapy: a comprehensive review. Circulation. 2017;135(15):1458–70.CrossRefPubMed
13.
go back to reference Kiehl EL, Cantillon DJ. Leadless cardiac pacing: what primary care providers and non-EP cardiologists should know. Cleve Clin J Med. 2016;83(11 Suppl 2):S24–34.CrossRefPubMed Kiehl EL, Cantillon DJ. Leadless cardiac pacing: what primary care providers and non-EP cardiologists should know. Cleve Clin J Med. 2016;83(11 Suppl 2):S24–34.CrossRefPubMed
16.
go back to reference Kypta A, Blessberger H, Kiblboeck D, Steinwender C. Three tesla cardiac magnetic resonance imaging in a patient with a leadless cardiac pacemaker system. Eur Heart J. 2017;38(34):2628.CrossRefPubMedPubMedCentral Kypta A, Blessberger H, Kiblboeck D, Steinwender C. Three tesla cardiac magnetic resonance imaging in a patient with a leadless cardiac pacemaker system. Eur Heart J. 2017;38(34):2628.CrossRefPubMedPubMedCentral
17.
go back to reference Kramer CM, Barkhausen J, Flamm SD, Kim RJ, Nagel E. Society of Cardiovascular Magnetic Resonance Board of trustees task force on standardized protocols. Standardized cardiovascular magnetic resonance (CMR) protocols 2013 update. J Cardiovasc Magn Reson. 2013;15:91.CrossRefPubMedPubMedCentral Kramer CM, Barkhausen J, Flamm SD, Kim RJ, Nagel E. Society of Cardiovascular Magnetic Resonance Board of trustees task force on standardized protocols. Standardized cardiovascular magnetic resonance (CMR) protocols 2013 update. J Cardiovasc Magn Reson. 2013;15:91.CrossRefPubMedPubMedCentral
18.
go back to reference Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. Circulation. 2002;105:539–42.CrossRefPubMed Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. Circulation. 2002;105:539–42.CrossRefPubMed
19.
go back to reference Klein-Wiele O, Garmer M, Busch M, Mateiescu S, Urbien R, Barbone G, et al. Cardiovascular magnetic resonance in patients with magnetic resonance conditional pacemaker systems at 1.5T: influence of pacemaker related artefacts on image quality including first pass perfusion, aortic and mitral valve assessment, flow measurement, short tau inversion recovery and T1-weighted imaging. Int J Cardiovasc Imaging. 2017;33(3):383–94.CrossRefPubMed Klein-Wiele O, Garmer M, Busch M, Mateiescu S, Urbien R, Barbone G, et al. Cardiovascular magnetic resonance in patients with magnetic resonance conditional pacemaker systems at 1.5T: influence of pacemaker related artefacts on image quality including first pass perfusion, aortic and mitral valve assessment, flow measurement, short tau inversion recovery and T1-weighted imaging. Int J Cardiovasc Imaging. 2017;33(3):383–94.CrossRefPubMed
20.
go back to reference Soejima K, Edmonson J, Ellingson ML, Herberg B, Wiklund C, Zhao J. Safety evaluation of a leadless pacemaker for magnetic resonance imaging use. Heart Rhythm. 2016;13(10):2056–63.CrossRefPubMed Soejima K, Edmonson J, Ellingson ML, Herberg B, Wiklund C, Zhao J. Safety evaluation of a leadless pacemaker for magnetic resonance imaging use. Heart Rhythm. 2016;13(10):2056–63.CrossRefPubMed
21.
go back to reference Sasaki T, Hansford R, Zviman MM, Kolandaivelu A, Bluemke DA, Berger RD, et al. Quantitative assessment of artifacts on cardiac magnetic resonance imaging of patients with pacemakers and implantable cardioverter-defibrillators. Circ Cardiovasc Imaging. 2011;4:662–70.CrossRefPubMedPubMedCentral Sasaki T, Hansford R, Zviman MM, Kolandaivelu A, Bluemke DA, Berger RD, et al. Quantitative assessment of artifacts on cardiac magnetic resonance imaging of patients with pacemakers and implantable cardioverter-defibrillators. Circ Cardiovasc Imaging. 2011;4:662–70.CrossRefPubMedPubMedCentral
22.
go back to reference Auricchio A, Delnoy PP, Butter C, Brachmann J, Van Erven L, Spitzer S, et al. Feasibility, safety, and short-term outcome of leadless ultrasound-based endocardial left ventricular resynchronization in heart failure patients: results of the wireless stimulation endocardially for CRT (WiSE-CRT) study. Europace. 2014;16(5):681–8.CrossRefPubMed Auricchio A, Delnoy PP, Butter C, Brachmann J, Van Erven L, Spitzer S, et al. Feasibility, safety, and short-term outcome of leadless ultrasound-based endocardial left ventricular resynchronization in heart failure patients: results of the wireless stimulation endocardially for CRT (WiSE-CRT) study. Europace. 2014;16(5):681–8.CrossRefPubMed
23.
go back to reference Tjong FV, Brouwer TF, Smeding L, Kooiman KM, de Groot JR, Ligon D, et al. Combined leadless pacemaker and subcutaneous implantable defibrillator therapy: feasibility, safety, and performance. Europace. 2016;18(11):1740–7.CrossRefPubMed Tjong FV, Brouwer TF, Smeding L, Kooiman KM, de Groot JR, Ligon D, et al. Combined leadless pacemaker and subcutaneous implantable defibrillator therapy: feasibility, safety, and performance. Europace. 2016;18(11):1740–7.CrossRefPubMed
24.
go back to reference Keller J, Neuzil P, Vymazal J, Janotka M, Brada J, Zacek R, et al. Magnetic resonance imaging in patients with subcutaneous implantable cardioverter-defibrillator. Europace. 2015;17(5):761–6.CrossRefPubMedPubMedCentral Keller J, Neuzil P, Vymazal J, Janotka M, Brada J, Zacek R, et al. Magnetic resonance imaging in patients with subcutaneous implantable cardioverter-defibrillator. Europace. 2015;17(5):761–6.CrossRefPubMedPubMedCentral
Metadata
Title
Artefacts in 1.5 Tesla and 3 Tesla cardiovascular magnetic resonance imaging in patients with leadless cardiac pacemakers
Authors
Daniel Kiblboeck
Christian Reiter
Juergen Kammler
Pierre Schmit
Hermann Blessberger
Joerg Kellermair
Franz Fellner
Clemens Steinwender
Publication date
01-12-2018
Publisher
BioMed Central
Published in
Journal of Cardiovascular Magnetic Resonance / Issue 1/2018
Electronic ISSN: 1532-429X
DOI
https://doi.org/10.1186/s12968-018-0469-4

Other articles of this Issue 1/2018

Journal of Cardiovascular Magnetic Resonance 1/2018 Go to the issue