Examination of inter-rater and intra-rater reliability during retentive force measurement of different clasps using the developed small-sized retentive force measurement device

Authors: Hitoshi Akiyama, Maiko Sakamoto, Ryoichi Akama, Jun Takei

Published in: BMC Oral Health | Issue 1/2020

Login to get access

Abstract

Background

The design process of a removable partial denture (RPD) consists of rests, major connectors, minor connectors, denture base, and retainer. The abutment tooth contour determines the location of the retention portion of the clasp at the retention areas. The load capacity of the clasp depends on various factors such as type, the position of a clasp, tooth position, clasp length, and pullout location. As a general rule, the amount of retention required to dislodge the RPD from the supporting structure should always be the minimum necessary to resist reasonable dislodging forces. Excessive force from the clasps may cause many problems. Although there are many reports on the retentive force of cast clasps using large devices, it has not been possible to measure it using simple devices until now.

Method

The purpose of the present study is to develop a small-sized retentive force measurement device that can easily measure the retentive force of a clasp used for an RPD. It is to examine inter-rater and intra-rater reliability. A small-sized retentive force measurement device that can be easily measured in the laboratory has been developed. Using commercially available hard plaster cast, a skilled dental technician has made 10 types of cast clasps used in clinical practice using conventional techniques. Three assessors measured the retentive force of 10 types of cast clasps. To confirm the reliability, the intra-class correlation coefficients ICC (1,1) and ICC (1,3) of the 3 assessors were calculated, and the reliability within the assessor was examined. The inter-class correlation coefficients ICC (3,1) and ICC (3,3) were calculated, and the reliability of the assessors was examined.

Result

The intra-class correlation coefficients of 3 assessors are as follows: assessor 1 has ICC (1,1) = 0.971, ICC (1,3) = 0.990, assessor 2 has ICC (1,1) = 0.967, ICC (1,3) = 0.989, assessor 3 has ICC (1,1) = 0.962, ICC (1,3) = 0.987. The inter-class correlation coefficients of 3 assessors are as follows: ICC (3,1) = 0.993, ICC (3,3) = 0.998. From the evaluation standard of the intraclass correlation coefficients of reliability value by ICC, it was evaluated as almost perfect and high reproducibility was confirmed.

Conclusion

The developed small-sized retentive force measurement device has reproducibility within and between the assessors.
Literature
1.
go back to reference Marie A, Keeling A, Hyde TP, Nattress BR, Pavitt S, Murphy RJ, Shary TJ, Dillon S, Osnes C, Wood DJ. Deformation and retentive force following invitro cyclic fatigue of cobalt-chrome and aryl ketone polymer (AKP) clasps. Dent Mater. 2019;35:e113–e21.CrossRef Marie A, Keeling A, Hyde TP, Nattress BR, Pavitt S, Murphy RJ, Shary TJ, Dillon S, Osnes C, Wood DJ. Deformation and retentive force following invitro cyclic fatigue of cobalt-chrome and aryl ketone polymer (AKP) clasps. Dent Mater. 2019;35:e113–e21.CrossRef
2.
go back to reference Kratochvil FJ. Partial removable prosthodontics. 1st ed: WB.Saunders Co; 1988. Kratochvil FJ. Partial removable prosthodontics. 1st ed: WB.Saunders Co; 1988.
3.
go back to reference Renner RP, Boucher LJ. Removable partial dentures: Chicago:Quintessence Pub.Co Inc; 1987. Renner RP, Boucher LJ. Removable partial dentures: Chicago:Quintessence Pub.Co Inc; 1987.
4.
go back to reference Mourshed B, Al-Sabri FA, Qaed NA, Alaizari N, Al-Shamiri HM, Alfaqih A. Effect of clasp type and pullout location on clasp retention in different environment: in vitro study. Eur J Dent. 2017;11:216–20.CrossRef Mourshed B, Al-Sabri FA, Qaed NA, Alaizari N, Al-Shamiri HM, Alfaqih A. Effect of clasp type and pullout location on clasp retention in different environment: in vitro study. Eur J Dent. 2017;11:216–20.CrossRef
5.
go back to reference Turner JW, Radford DR, Sherriff M. Flexural properties and surface finishing of acetal resin denture clasps. J Prosthodont. 1999;8:188–95.CrossRef Turner JW, Radford DR, Sherriff M. Flexural properties and surface finishing of acetal resin denture clasps. J Prosthodont. 1999;8:188–95.CrossRef
6.
go back to reference Reddy JC, Chintapatla SB, Srikakula NK, Juturu RK, Paidi SK, Tedlapu SK, Mannava P, Khatoon R. Comparison of retention of clasps made of different materials using three-dimensional finite element analysis. J Clin Diagn Res. 2016;10:ZC13–6.PubMedPubMedCentral Reddy JC, Chintapatla SB, Srikakula NK, Juturu RK, Paidi SK, Tedlapu SK, Mannava P, Khatoon R. Comparison of retention of clasps made of different materials using three-dimensional finite element analysis. J Clin Diagn Res. 2016;10:ZC13–6.PubMedPubMedCentral
7.
go back to reference Kim D, Park C, Yi Y, Cho L. Comparison of cast Ti-Ni alloy clasp retention with conventional removable partial denture clasps. J Prosthet Dent. 2004;91:374–82.CrossRef Kim D, Park C, Yi Y, Cho L. Comparison of cast Ti-Ni alloy clasp retention with conventional removable partial denture clasps. J Prosthet Dent. 2004;91:374–82.CrossRef
8.
go back to reference Helal MA, Baraka OA, Sanad ME, Al-Khiary Y, Ludwig K, Kern M. Effect of clasp design on retention at different intervals using different abutment materials and in a simulated oral condition. J Prosthodont. 2014;23:140–5.CrossRef Helal MA, Baraka OA, Sanad ME, Al-Khiary Y, Ludwig K, Kern M. Effect of clasp design on retention at different intervals using different abutment materials and in a simulated oral condition. J Prosthodont. 2014;23:140–5.CrossRef
9.
go back to reference Cheng H, Xu M, Zhang H, Wu W, Zheng M, Li X. Cyclic fatigue properties of cobalt-chromium alloy clasps for partial removable dental prostheses. J Prosthet Dent. 2010;104:389–96.CrossRef Cheng H, Xu M, Zhang H, Wu W, Zheng M, Li X. Cyclic fatigue properties of cobalt-chromium alloy clasps for partial removable dental prostheses. J Prosthet Dent. 2010;104:389–96.CrossRef
10.
go back to reference Meenakshi A, Gupta R, Bharti V, Sriramaprabu G, Prabhakar R. An evaluation of retentive ability and deformation of acetal resin and cobalt-chromium clasps. J Clin Diagn Res. 2016;10:ZC37–41.PubMedPubMedCentral Meenakshi A, Gupta R, Bharti V, Sriramaprabu G, Prabhakar R. An evaluation of retentive ability and deformation of acetal resin and cobalt-chromium clasps. J Clin Diagn Res. 2016;10:ZC37–41.PubMedPubMedCentral
11.
go back to reference Firtell DN, Grisius RJ. Retention of obturator-removable partial dentures: a comparison of buccal and lingual retention. J Prosthet Dent. 1980;43:212–7.CrossRef Firtell DN, Grisius RJ. Retention of obturator-removable partial dentures: a comparison of buccal and lingual retention. J Prosthet Dent. 1980;43:212–7.CrossRef
12.
go back to reference Rodrigues RS, Ribeiro RF, de Mattos MC, Bezzon OL. Comparative study of circumferential clasp retention force for titanium and cobalt-chromium removable partial dentures. J Prosthet Dent. 2002;88:290–6.CrossRef Rodrigues RS, Ribeiro RF, de Mattos MC, Bezzon OL. Comparative study of circumferential clasp retention force for titanium and cobalt-chromium removable partial dentures. J Prosthet Dent. 2002;88:290–6.CrossRef
13.
go back to reference Epprecht A, Zeltner M, Benic G, Özcan M. A strain gauge analysis comparing 4-unit veneered zirconium dioxide implant-borne fixed dental prosthesis on engaging and non-engaging abutments before and after torque application. Clin Exp Dent Res. 2018;4:13–8.CrossRef Epprecht A, Zeltner M, Benic G, Özcan M. A strain gauge analysis comparing 4-unit veneered zirconium dioxide implant-borne fixed dental prosthesis on engaging and non-engaging abutments before and after torque application. Clin Exp Dent Res. 2018;4:13–8.CrossRef
14.
go back to reference Zhao Y, Zhao Y, Ge X. The development of a triaxial cutting force sensor based on a MEMS strain gauge. Micromachines. 2018;9:30.CrossRef Zhao Y, Zhao Y, Ge X. The development of a triaxial cutting force sensor based on a MEMS strain gauge. Micromachines. 2018;9:30.CrossRef
15.
go back to reference Carr AB, McGivney GP, Brown DT. Direct retainers, McCracken’s removable partial prosthodontics. 11th ed: Elsevier Mosby; 2000. p. 79–115. Carr AB, McGivney GP, Brown DT. Direct retainers, McCracken’s removable partial prosthodontics. 11th ed: Elsevier Mosby; 2000. p. 79–115.
16.
go back to reference Fleiss JL. Reliability of measurement, the design and analysis of clinical experiments. New York: Wiley.; 1986. p. 1–32. Fleiss JL. Reliability of measurement, the design and analysis of clinical experiments. New York: Wiley.; 1986. p. 1–32.
17.
go back to reference Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33:159–74.CrossRef Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33:159–74.CrossRef
18.
go back to reference Ahmad I, Sherriff M, Waters NE. The effect of reducing the number of clasps on removable partial denture retention. J Prosthet Dent. 1992;68:928–33.CrossRef Ahmad I, Sherriff M, Waters NE. The effect of reducing the number of clasps on removable partial denture retention. J Prosthet Dent. 1992;68:928–33.CrossRef
19.
go back to reference Frank RP, Nicholls JI. A study of the flexibility of wrought wire clasps. J Prosthet Dent. 1981;45:259–67.CrossRef Frank RP, Nicholls JI. A study of the flexibility of wrought wire clasps. J Prosthet Dent. 1981;45:259–67.CrossRef
20.
go back to reference Arda T, Arikan A. An in vitro comparison of retentive force and deformation of acetal resin and cobalt-chromium clasps. J Prosthet Dent. 2005;94:267–74.CrossRef Arda T, Arikan A. An in vitro comparison of retentive force and deformation of acetal resin and cobalt-chromium clasps. J Prosthet Dent. 2005;94:267–74.CrossRef
21.
go back to reference Bridgeman JT, Marker VA, Hummel SK, Benson BW, Pace LL. Comparison of titanium and cobalt-chromium removable partial denture clasps. J Prosthet Dent. 1997;78:187–93.CrossRef Bridgeman JT, Marker VA, Hummel SK, Benson BW, Pace LL. Comparison of titanium and cobalt-chromium removable partial denture clasps. J Prosthet Dent. 1997;78:187–93.CrossRef
22.
go back to reference Xie W, Zheng M, Wang J, Li X. The effect of build orientation on the microstructure and properties of selective laser melting Ti-6Al-4V for removable partial denture clasps. J Prosthet Dent. 2020;123:163–72.CrossRef Xie W, Zheng M, Wang J, Li X. The effect of build orientation on the microstructure and properties of selective laser melting Ti-6Al-4V for removable partial denture clasps. J Prosthet Dent. 2020;123:163–72.CrossRef
23.
go back to reference de Torres EM, de Siqueira Damasceno II, Do Amaral BA, RCS R, ADFP C, Ribeiro RF. Effect of acetyl resin retentive arms on the retentive force of circumferential clasps: an in vitro study. J Prosthodontic Res. 2012;56:216–21.CrossRef de Torres EM, de Siqueira Damasceno II, Do Amaral BA, RCS R, ADFP C, Ribeiro RF. Effect of acetyl resin retentive arms on the retentive force of circumferential clasps: an in vitro study. J Prosthodontic Res. 2012;56:216–21.CrossRef
24.
go back to reference Tse ET, Cheng LY, Luk HW, Chu FC, Chai J, Chow TW. Comparison of the retentive characteristics of cobalt-chromium and commercially pure titanium clasps using a novel method. Int J Prosthodont. 2006;19:371–2.PubMed Tse ET, Cheng LY, Luk HW, Chu FC, Chai J, Chow TW. Comparison of the retentive characteristics of cobalt-chromium and commercially pure titanium clasps using a novel method. Int J Prosthodont. 2006;19:371–2.PubMed
25.
go back to reference Bates JF. The mechanical properties of the cobalt-chromium alloys and their relation to partial denture design. Brit Dent J. 1965;119:389–96.PubMed Bates JF. The mechanical properties of the cobalt-chromium alloys and their relation to partial denture design. Brit Dent J. 1965;119:389–96.PubMed
26.
go back to reference Gapido CG, Kobayashi H, Miyakawa O, Kohno S. Fatigue resistance of cast occlusal rests using co-Cr and Ag-Pd-cu-au alloys. J Prosthet Dent. 2003;90:261–9.CrossRef Gapido CG, Kobayashi H, Miyakawa O, Kohno S. Fatigue resistance of cast occlusal rests using co-Cr and Ag-Pd-cu-au alloys. J Prosthet Dent. 2003;90:261–9.CrossRef
27.
go back to reference Bartko JJ. The intraclass correlation coefficient as a measure of reliability. Psychol Rep. 1966;19:3–11.CrossRef Bartko JJ. The intraclass correlation coefficient as a measure of reliability. Psychol Rep. 1966;19:3–11.CrossRef
28.
go back to reference Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979;86:420–8.CrossRef Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979;86:420–8.CrossRef
29.
go back to reference Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropractic Med. 2016;15:155–63.CrossRef Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropractic Med. 2016;15:155–63.CrossRef
Metadata
Title
Examination of inter-rater and intra-rater reliability during retentive force measurement of different clasps using the developed small-sized retentive force measurement device
Authors
Hitoshi Akiyama
Maiko Sakamoto
Ryoichi Akama
Jun Takei
Publication date
01-12-2020
Publisher
BioMed Central
Keyword
Denture
Published in
BMC Oral Health / Issue 1/2020
Electronic ISSN: 1472-6831
DOI
https://doi.org/10.1186/s12903-020-01215-2

Other articles of this Issue 1/2020

BMC Oral Health 1/2020 Go to the issue