The influence of two forms of chlorhexidine on the accuracy of contemporary electronic apex locators

Authors: Ewa Marek, Ryta Łagocka, Katarzyna Kot, Krzysztof Woźniak, Mariusz Lipski

Published in: BMC Oral Health | Issue 1/2020

Login to get access

Abstract

Background

Accurate determination of working length (WL) is crucial for the success of endodontic therapy. The aim of this study was to determine the influence of 2% chlorhexidine digluconate solution, 2% chlorhexidine digluconate gel and 2% hypochlorite solution on the accuracy of two devices: the Raypex 5 and the ApexDal.

Methods

Twenty-nine single-rooted human teeth were used in this study. The crowns were cut horizontally and embedded in an alginate mass. In each tooth, six endodontic measurements were made using two electronic apex locators (EALs): a Raypex 5 and an ApexDal. For each EAL, measurements were taken with the following products: 2% chlorhexidine solution (CHX-S group), 2% chlorhexidine gel (CHX-G group) and 2% NaOCl (NaOCl group). After performing an endodontic measurement, the endodontic instruments were stabilized with flow resin composite. Afterwards, the roots were removed from the alginate mass, and the apical one-third of each root was cut lengthways to recover the canal system. Last, the distance between the file tip and the apical foramen was measured under a microscope at 60 x magnification.

Results

Statistically significant differences were found between CHX-S and NaOCl and CHX-G and NaOCl, but no significant differences were detected between CHX-S and CHX-G during the testing of both devices. No statistically significant differences were observed between the Raypex 5 and ApexDal for all intracanal media tested.

Conclusion

The EALs Raypex 5 and ApexDal had higher accuracy in the anatomical foramen of the root containing chlorhexidine in the gel or in the solution form than in the canal containing sodium hypochlorite.
Literature
1.
go back to reference Ziegler AA, Altenburger MJ, Wrbas KT. In vivo comparison of working length determination with electronic apex locators. Int Endod J. 2007;40:133–8.PubMedCrossRef Ziegler AA, Altenburger MJ, Wrbas KT. In vivo comparison of working length determination with electronic apex locators. Int Endod J. 2007;40:133–8.PubMedCrossRef
2.
go back to reference Stein TJ, Corcoran JF. Noioniozing method of locating the apical constriction (minor foramen) in root canals. Oral Surg Oral Med Oral Pathol. 1991;71:96–9.PubMedCrossRef Stein TJ, Corcoran JF. Noioniozing method of locating the apical constriction (minor foramen) in root canals. Oral Surg Oral Med Oral Pathol. 1991;71:96–9.PubMedCrossRef
3.
go back to reference Basmadijan-Charles CL, Farge P, Bourgeois DM, Lebrun T. Factors influencing the long-term results of endodontic treatment: a review of the literature. Int Dent J. 2002;52:81–6.CrossRef Basmadijan-Charles CL, Farge P, Bourgeois DM, Lebrun T. Factors influencing the long-term results of endodontic treatment: a review of the literature. Int Dent J. 2002;52:81–6.CrossRef
4.
go back to reference Kang JA, Kim SK. Accuracies of seven different apex locators under various conditions. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008;106:e57–62.PubMedCrossRef Kang JA, Kim SK. Accuracies of seven different apex locators under various conditions. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008;106:e57–62.PubMedCrossRef
5.
go back to reference Plotino G, Grande NM, Brigsnte L, Lesti B, Somma F. Ex vivo accuracy of three electronic apex locators root ZX, elements diagnostic unit, apex locator and ProPex. Int Endod J. 2006;39:408–14.PubMedCrossRef Plotino G, Grande NM, Brigsnte L, Lesti B, Somma F. Ex vivo accuracy of three electronic apex locators root ZX, elements diagnostic unit, apex locator and ProPex. Int Endod J. 2006;39:408–14.PubMedCrossRef
6.
go back to reference Meares WA, Steiman HR. The influence of sodium hypochlorite irrigation on the accuracy of the root ZX electronic apex locator. J Endod. 2002;28:595–8.PubMedCrossRef Meares WA, Steiman HR. The influence of sodium hypochlorite irrigation on the accuracy of the root ZX electronic apex locator. J Endod. 2002;28:595–8.PubMedCrossRef
7.
go back to reference Kim E, Marmo M. An in vivo comparison of working length determination by only root ZX apex locator versus combining root ZX apex locator with radiographs using a new impression technique. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008;105:e79–83.PubMedCrossRef Kim E, Marmo M. An in vivo comparison of working length determination by only root ZX apex locator versus combining root ZX apex locator with radiographs using a new impression technique. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008;105:e79–83.PubMedCrossRef
8.
go back to reference Ózsezer E, Inan U, Aydin U. In vivo evaluation of ProPex electronic apex locator. J Endod. 2007;33:974–7.PubMedCrossRef Ózsezer E, Inan U, Aydin U. In vivo evaluation of ProPex electronic apex locator. J Endod. 2007;33:974–7.PubMedCrossRef
9.
go back to reference ElAyouti A, Weiger R, Löst C. Frequency of overinstrumentation with an acceptable radiographic working length. J Endod. 2001;27:49–52.PubMedCrossRef ElAyouti A, Weiger R, Löst C. Frequency of overinstrumentation with an acceptable radiographic working length. J Endod. 2001;27:49–52.PubMedCrossRef
10.
go back to reference ElAyouti A, Weiger R, Löst C. The ability of root ZX apex locator to reduce the frequency of overestimated radiographic working length. J Endod. 2002;28:116–9.PubMedCrossRef ElAyouti A, Weiger R, Löst C. The ability of root ZX apex locator to reduce the frequency of overestimated radiographic working length. J Endod. 2002;28:116–9.PubMedCrossRef
11.
go back to reference Marcos-Arenal JL, Rivera EM, Caplan DJ. Evaluation the paper point technique for locating the apical foramen after canal preparation. Oral Surg. Oral Med Oral Pathol Oral Radiol Endod. 2009;108:e101–5.CrossRef Marcos-Arenal JL, Rivera EM, Caplan DJ. Evaluation the paper point technique for locating the apical foramen after canal preparation. Oral Surg. Oral Med Oral Pathol Oral Radiol Endod. 2009;108:e101–5.CrossRef
12.
go back to reference Dunlap CA, Remeikis NA, BeGole EA, Rauschenberger CR. An in vivo evaluation of an electric apex locator that uses the ratio method vital and necrotic canals. J Endod. 1998;24:48–50.PubMedCrossRef Dunlap CA, Remeikis NA, BeGole EA, Rauschenberger CR. An in vivo evaluation of an electric apex locator that uses the ratio method vital and necrotic canals. J Endod. 1998;24:48–50.PubMedCrossRef
13.
go back to reference Pratten DH, McDonald NJ. Comparison of radiographic and electronic working lengths. J Endod. 1996;22:173–6.PubMedCrossRef Pratten DH, McDonald NJ. Comparison of radiographic and electronic working lengths. J Endod. 1996;22:173–6.PubMedCrossRef
14.
go back to reference Khandewal D, Ballal NV, Sarawathi MV. Comparative evaluation of accuracy of 2 electronic apex locators with convetional radiography: an ex vivo study. J Endod. 2015;41(2):201–4.PubMedCrossRef Khandewal D, Ballal NV, Sarawathi MV. Comparative evaluation of accuracy of 2 electronic apex locators with convetional radiography: an ex vivo study. J Endod. 2015;41(2):201–4.PubMedCrossRef
15.
go back to reference Gurel MA, Helvacioglu Kivanc B, Ekici A. A comparative assessment of the accuracies of Raypex 5, Raypex 6, iPex and iPex II electronic apex locators: an in vitro study. J Istanb Univ Fac Dent. 2017;51:28–33.PubMedPubMedCentral Gurel MA, Helvacioglu Kivanc B, Ekici A. A comparative assessment of the accuracies of Raypex 5, Raypex 6, iPex and iPex II electronic apex locators: an in vitro study. J Istanb Univ Fac Dent. 2017;51:28–33.PubMedPubMedCentral
16.
go back to reference Yolagiden M, Ersahan S, Suyun G, Bilgec E, Aydin C. Comparison of four electronic apex locators in detecting working length: an ex vivo study. J Contemp Dent Pract. 2018;19:1427–33.PubMedCrossRef Yolagiden M, Ersahan S, Suyun G, Bilgec E, Aydin C. Comparison of four electronic apex locators in detecting working length: an ex vivo study. J Contemp Dent Pract. 2018;19:1427–33.PubMedCrossRef
17.
go back to reference Saraswathi V, Kedia A, Purayil TP, Ballal V, Saini A. Comparative evaluation of the accuracy of two electronic apex locators in determining the working length in teeth with simulated apical root resorption: an in vitro study. J Conserv Dent. 2016;19:402–5.PubMedPubMedCentralCrossRef Saraswathi V, Kedia A, Purayil TP, Ballal V, Saini A. Comparative evaluation of the accuracy of two electronic apex locators in determining the working length in teeth with simulated apical root resorption: an in vitro study. J Conserv Dent. 2016;19:402–5.PubMedPubMedCentralCrossRef
18.
go back to reference Marroquín BB, Fernández CC, Schmidtmann I, Willershausen B, Goldberg F. Accuracy of electronic apex locators to detect root canal perforations with inserted me-tallic posts: an ex vivo study. Head Face Med. 2014;10:57.PubMedPubMedCentralCrossRef Marroquín BB, Fernández CC, Schmidtmann I, Willershausen B, Goldberg F. Accuracy of electronic apex locators to detect root canal perforations with inserted me-tallic posts: an ex vivo study. Head Face Med. 2014;10:57.PubMedPubMedCentralCrossRef
19.
go back to reference Ding J, Gutmann JL, Fan B, Lu Y, Chen H. Investigation of Apex Locators and Related Morphological. J Endod. 2010;36:1399–403.PubMedCrossRef Ding J, Gutmann JL, Fan B, Lu Y, Chen H. Investigation of Apex Locators and Related Morphological. J Endod. 2010;36:1399–403.PubMedCrossRef
20.
go back to reference Erdemir A, Eldeniz AU, Ari H, Belli S, Esener T. The influence of irrigation solutions on the accuracy of the electronic apex locator facility in the tri auto ZX handpiece. Int Endod J. 2007;40:391–7.PubMedCrossRef Erdemir A, Eldeniz AU, Ari H, Belli S, Esener T. The influence of irrigation solutions on the accuracy of the electronic apex locator facility in the tri auto ZX handpiece. Int Endod J. 2007;40:391–7.PubMedCrossRef
21.
go back to reference Kuruvilla JR, Kamath MP. Antimicrobial activity of 2.5% sodium hypochlorite and 0.2% chlorhexidine gluconate separately and combined, as endodontic irrigants. J Endod. 1998;24:472–6.PubMedCrossRef Kuruvilla JR, Kamath MP. Antimicrobial activity of 2.5% sodium hypochlorite and 0.2% chlorhexidine gluconate separately and combined, as endodontic irrigants. J Endod. 1998;24:472–6.PubMedCrossRef
22.
go back to reference Mohammadia Z, Giardinoc L, Palazzid F, Asgarya S. Agonistic and Antagonistic Interactions between Chlorhexidine and Other Endodontic Agents: A Critical Review. Iran Endod J. 2015;10:1–5. Mohammadia Z, Giardinoc L, Palazzid F, Asgarya S. Agonistic and Antagonistic Interactions between Chlorhexidine and Other Endodontic Agents: A Critical Review. Iran Endod J. 2015;10:1–5.
23.
go back to reference Basrani BR, Manek S, Fillery E. Using diazotization to characterize the effect of heat orSodium hypochlorite on 2.0% Chlorhexidine. J Endod. 2009;35:1296–9.PubMedCrossRef Basrani BR, Manek S, Fillery E. Using diazotization to characterize the effect of heat orSodium hypochlorite on 2.0% Chlorhexidine. J Endod. 2009;35:1296–9.PubMedCrossRef
24.
go back to reference Khadse P, Kamra A, Banga KS. Effectiveness of various intermediate irrigants for the prevention of precipitate formed by the interaction of sodium hypochlorite and chlorhexidine-an in vitro study. Edodontology. 2014;26:309–13. Khadse P, Kamra A, Banga KS. Effectiveness of various intermediate irrigants for the prevention of precipitate formed by the interaction of sodium hypochlorite and chlorhexidine-an in vitro study. Edodontology. 2014;26:309–13.
25.
go back to reference Wang CS, Arnold RR, Trope M, Teixeira BF. Clinical efficiency of 2% chlorhexidine gel in reducing intracanal bacteria. J Endod. 2007;33:1283–9.PubMedCrossRef Wang CS, Arnold RR, Trope M, Teixeira BF. Clinical efficiency of 2% chlorhexidine gel in reducing intracanal bacteria. J Endod. 2007;33:1283–9.PubMedCrossRef
26.
go back to reference Filho M, Leonardo M, Silva L, Anibal F, Faccioli L. Inflammatory responses to different endodontic irrigating solutions. Int Endod J. 2002;35:735–9.CrossRef Filho M, Leonardo M, Silva L, Anibal F, Faccioli L. Inflammatory responses to different endodontic irrigating solutions. Int Endod J. 2002;35:735–9.CrossRef
27.
go back to reference Paquette L, Legner M, Fillery ED, Friedman S. Antibacterial efficacy of chlorhexidine gluconate intracanal medication in vivo. J Endod. 2007;33:788–95.PubMedCrossRef Paquette L, Legner M, Fillery ED, Friedman S. Antibacterial efficacy of chlorhexidine gluconate intracanal medication in vivo. J Endod. 2007;33:788–95.PubMedCrossRef
28.
go back to reference Siqueira JF, Rôças IN, Lopez HP, Magalhães FAC, Uzeda M. Elimination of Candida albicans infection of the radicular dentin by intracanal medications. J Endod. 2003;29:501–4.PubMedCrossRef Siqueira JF, Rôças IN, Lopez HP, Magalhães FAC, Uzeda M. Elimination of Candida albicans infection of the radicular dentin by intracanal medications. J Endod. 2003;29:501–4.PubMedCrossRef
29.
go back to reference Portenier I, Waltimo TD, Ørstavik D, Haapasalo M. Killing of enterococcus faecalis by MTAD and chlorhexidine digluconate with or without Cetrimide in the presence or absence of dentine powder or BSA. J Endod. 2006;32:138–41.PubMedCrossRef Portenier I, Waltimo TD, Ørstavik D, Haapasalo M. Killing of enterococcus faecalis by MTAD and chlorhexidine digluconate with or without Cetrimide in the presence or absence of dentine powder or BSA. J Endod. 2006;32:138–41.PubMedCrossRef
30.
go back to reference Rosenthal S, Spangberg L, Safavi K. Chlorhexidine substantivity in root canal dentin. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2004;98:488–92.PubMedCrossRef Rosenthal S, Spangberg L, Safavi K. Chlorhexidine substantivity in root canal dentin. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2004;98:488–92.PubMedCrossRef
31.
go back to reference Dametto FR, Ferraz CC, Gomes BP, Zaia AA, Teixeira FB, de Souza-Filho FJ. In vitro assessment of the immediate and prolonged antimicrobial action of chlorhexidine gel as an endodontic irrigant against enterococcus faecalis. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2005;99:768–72.PubMedCrossRef Dametto FR, Ferraz CC, Gomes BP, Zaia AA, Teixeira FB, de Souza-Filho FJ. In vitro assessment of the immediate and prolonged antimicrobial action of chlorhexidine gel as an endodontic irrigant against enterococcus faecalis. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2005;99:768–72.PubMedCrossRef
32.
go back to reference Shin HS, Yang WK, Kim M, Ko HJ, Cho KM, Park SH, Kim JW. Accuracy of root ZX in teeth with simulated root perforation in the presence of gel or liquid type endodontic irrigant. Restor Dent Endod. 2012;37:149–54.PubMedPubMedCentralCrossRef Shin HS, Yang WK, Kim M, Ko HJ, Cho KM, Park SH, Kim JW. Accuracy of root ZX in teeth with simulated root perforation in the presence of gel or liquid type endodontic irrigant. Restor Dent Endod. 2012;37:149–54.PubMedPubMedCentralCrossRef
33.
go back to reference Tinaz AC, Sevimli LS, Görgül G, Türköz G. The effects of sodium hypochlorite concentrations on the accuracy of an apex locating device. J Endod. 2002;28:160–2.PubMedCrossRef Tinaz AC, Sevimli LS, Görgül G, Türköz G. The effects of sodium hypochlorite concentrations on the accuracy of an apex locating device. J Endod. 2002;28:160–2.PubMedCrossRef
34.
go back to reference Fouad AF, Rivera EM, Krell KV. Accuracy of the Endex with variation in canal irrigants and foramen sizes. J Endod. 1983;19:63–7.CrossRef Fouad AF, Rivera EM, Krell KV. Accuracy of the Endex with variation in canal irrigants and foramen sizes. J Endod. 1983;19:63–7.CrossRef
35.
go back to reference Ebrahim AK, Yoshioka T, Kobayashi C, Suda H. The effects of size, sodium hypochlorite and blood on the accuracy of root ZX apex locator in enlarged root canals: an in vitro study. Aust Dent J. 2006;51:153–7.PubMedCrossRef Ebrahim AK, Yoshioka T, Kobayashi C, Suda H. The effects of size, sodium hypochlorite and blood on the accuracy of root ZX apex locator in enlarged root canals: an in vitro study. Aust Dent J. 2006;51:153–7.PubMedCrossRef
36.
go back to reference Akisue E, Poli de Figueiredo JA. Influence of pulp vitality on length determination by using the elements diagnostic unit and apex locator. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2007;104:e129–32.PubMedCrossRef Akisue E, Poli de Figueiredo JA. Influence of pulp vitality on length determination by using the elements diagnostic unit and apex locator. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2007;104:e129–32.PubMedCrossRef
37.
go back to reference Jenkins JA, Walker WA, Schindler WG, Flores CM. An in vitro evaluation of the accuracy of the root ZX in the presence of various irrigants. J Endod. 2001;27:209–11.PubMedCrossRef Jenkins JA, Walker WA, Schindler WG, Flores CM. An in vitro evaluation of the accuracy of the root ZX in the presence of various irrigants. J Endod. 2001;27:209–11.PubMedCrossRef
38.
go back to reference Jain S, Kapur R. Comparative evaluation of accuracy of two electronic apex locators in the presence of various irrigants: An in vitro study. Contemp Clin Dent. 2012;3(Suppl 2):S140–5.PubMedPubMedCentralCrossRef Jain S, Kapur R. Comparative evaluation of accuracy of two electronic apex locators in the presence of various irrigants: An in vitro study. Contemp Clin Dent. 2012;3(Suppl 2):S140–5.PubMedPubMedCentralCrossRef
39.
go back to reference Khursheed I, Bansal R, Bansal T, Singh HP, Yadav M, Reddy KJ. A comparative evaluation of working length with digital radiography and third generation apex locator (ProPex) in the presence of various intracanal irrigants: an in vivo/ex vivo study. Dent Res J. 2014;11:56–60. Khursheed I, Bansal R, Bansal T, Singh HP, Yadav M, Reddy KJ. A comparative evaluation of working length with digital radiography and third generation apex locator (ProPex) in the presence of various intracanal irrigants: an in vivo/ex vivo study. Dent Res J. 2014;11:56–60.
40.
go back to reference Paras M, Manjunath V, Manjunath MK. Comparison of accuracy of two electronic apex locators in the presence of various irrigants: an in vitro study. J Conserv Dent. 2012;15:178–82.CrossRef Paras M, Manjunath V, Manjunath MK. Comparison of accuracy of two electronic apex locators in the presence of various irrigants: an in vitro study. J Conserv Dent. 2012;15:178–82.CrossRef
41.
go back to reference Kaufman A, Katz A. Realiability of root ZX apex locator texted by an in vitro model. J Endod. 1993;19:201.CrossRef Kaufman A, Katz A. Realiability of root ZX apex locator texted by an in vitro model. J Endod. 1993;19:201.CrossRef
42.
go back to reference Prasad AB, Harshit S, Aastha SA, Deepak R. An Invitro evaluation of the accuracy of two electronic apex locators to determine working length in the presence of various Irrigants. Ethiop J Health Sci. 2016;26:457–62.PubMedPubMedCentralCrossRef Prasad AB, Harshit S, Aastha SA, Deepak R. An Invitro evaluation of the accuracy of two electronic apex locators to determine working length in the presence of various Irrigants. Ethiop J Health Sci. 2016;26:457–62.PubMedPubMedCentralCrossRef
43.
go back to reference Gomes S, Oliver R, Macouzet C, Mercadé M, Roig M, Duran-Sindreu F. In vivo evaluation of the Raypex 5 by using different irrigants. J Endod. 2012;38:1075–7.PubMedCrossRef Gomes S, Oliver R, Macouzet C, Mercadé M, Roig M, Duran-Sindreu F. In vivo evaluation of the Raypex 5 by using different irrigants. J Endod. 2012;38:1075–7.PubMedCrossRef
44.
go back to reference Joesten MD, Hogg JL, Castellion ME. The word of chemistry: essentials. 4th ed. Belmont: Brooks Cole; 2004. p. 461. Joesten MD, Hogg JL, Castellion ME. The word of chemistry: essentials. 4th ed. Belmont: Brooks Cole; 2004. p. 461.
45.
go back to reference Ebrahim AK, Wadachi R, Suda H. An in vitro evaluation of the accuracy of Dentaport ZX apex locator in enlarged root canals. Aust Dent J. 2007;52:193–7.PubMedCrossRef Ebrahim AK, Wadachi R, Suda H. An in vitro evaluation of the accuracy of Dentaport ZX apex locator in enlarged root canals. Aust Dent J. 2007;52:193–7.PubMedCrossRef
46.
go back to reference Schäfer E, Bossmann K. Antimicrobial efficacy of chloroxylenol and chlorhexidine in the treatment of infected root canals. Am J Dent. 2001;14:233–7.PubMed Schäfer E, Bossmann K. Antimicrobial efficacy of chloroxylenol and chlorhexidine in the treatment of infected root canals. Am J Dent. 2001;14:233–7.PubMed
48.
go back to reference Zamany A, Safavi K, Spångberg LS. The effect of chlorhexidine as an endodontic disinfectant. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2003;96:578–81.PubMedCrossRef Zamany A, Safavi K, Spångberg LS. The effect of chlorhexidine as an endodontic disinfectant. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2003;96:578–81.PubMedCrossRef
49.
go back to reference Dammaschke T, Jung N, Harks I, Schafer E. The effect of different root canal medicaments on the elimination of enterococcus faecalis ex vivo. Eur J Dent. 2013;7:442–8.PubMedPubMedCentralCrossRef Dammaschke T, Jung N, Harks I, Schafer E. The effect of different root canal medicaments on the elimination of enterococcus faecalis ex vivo. Eur J Dent. 2013;7:442–8.PubMedPubMedCentralCrossRef
50.
go back to reference Dammaschke T, Schneider U, Stratmann U, Yoo JM, Schäfer E. Effect of root canal dressings on the regeneration of inflamed periapical tissue. Acta Odontol Scand. 2005;63:143–52.PubMedCrossRef Dammaschke T, Schneider U, Stratmann U, Yoo JM, Schäfer E. Effect of root canal dressings on the regeneration of inflamed periapical tissue. Acta Odontol Scand. 2005;63:143–52.PubMedCrossRef
51.
go back to reference Oliveira TN, Vivacqua-Gomes N, Bernardes RA, Vivan RR, MAH D, Vasconcelos BC. Determination of the accuracy of 5 electronic apex locators in the function of different employment protocols. J Endod. 2017;43:1663–7.PubMedCrossRef Oliveira TN, Vivacqua-Gomes N, Bernardes RA, Vivan RR, MAH D, Vasconcelos BC. Determination of the accuracy of 5 electronic apex locators in the function of different employment protocols. J Endod. 2017;43:1663–7.PubMedCrossRef
52.
go back to reference Saatchi M, Aminozarbian MG, Noormohammadi H, Baghaei B. Influence of blood on the accuracy of Raypex 5 and root ZX electronic foramen locators: an in vivo study. Braz Dent J. 2016;27:336–9.PubMedCrossRef Saatchi M, Aminozarbian MG, Noormohammadi H, Baghaei B. Influence of blood on the accuracy of Raypex 5 and root ZX electronic foramen locators: an in vivo study. Braz Dent J. 2016;27:336–9.PubMedCrossRef
53.
go back to reference Al-Hadlaq SM. Effect of chloroform, orange solvent and eucalyptol on the accuracy of four electronic apex locators. Aust Endod J. 2013;39:112–5.PubMedCrossRef Al-Hadlaq SM. Effect of chloroform, orange solvent and eucalyptol on the accuracy of four electronic apex locators. Aust Endod J. 2013;39:112–5.PubMedCrossRef
55.
go back to reference Steffen H, Splieth CH, Behr K. Comparison of measurements obtaining with hand files or the canal leader attached to electronic apex locators: an in vitro study. Int Endod J. 1999;32:103–7.PubMedCrossRef Steffen H, Splieth CH, Behr K. Comparison of measurements obtaining with hand files or the canal leader attached to electronic apex locators: an in vitro study. Int Endod J. 1999;32:103–7.PubMedCrossRef
56.
go back to reference Luigi Cianconi L, Angott V, Felici R, Conte G, Mancin M. Accuracy of Three Electronic Apex Locators Compared with Digital Radiography: An Ex Vivo Study. J Endod. 2010;36:2003–7.PubMedCrossRef Luigi Cianconi L, Angott V, Felici R, Conte G, Mancin M. Accuracy of Three Electronic Apex Locators Compared with Digital Radiography: An Ex Vivo Study. J Endod. 2010;36:2003–7.PubMedCrossRef
57.
go back to reference Mancini M, Palopoli P, Iorio L, Conte G, Ciancon L. Accuracy of an Electronic Apex Locator in the Retreatment of Teeth Obturated with Plastic or Cross-linked Gutta-percha Carrier-based Materials: An Ex Vivo Study. J Endod. 2014;40:2061–5.PubMedCrossRef Mancini M, Palopoli P, Iorio L, Conte G, Ciancon L. Accuracy of an Electronic Apex Locator in the Retreatment of Teeth Obturated with Plastic or Cross-linked Gutta-percha Carrier-based Materials: An Ex Vivo Study. J Endod. 2014;40:2061–5.PubMedCrossRef
58.
go back to reference Kaufman AY, Keila S, Yoshpe M. Accuracy of new apex locator: an in vitro study. Int Endod J. 2002;35:186–92.PubMedCrossRef Kaufman AY, Keila S, Yoshpe M. Accuracy of new apex locator: an in vitro study. Int Endod J. 2002;35:186–92.PubMedCrossRef
Metadata
Title
The influence of two forms of chlorhexidine on the accuracy of contemporary electronic apex locators
Authors
Ewa Marek
Ryta Łagocka
Katarzyna Kot
Krzysztof Woźniak
Mariusz Lipski
Publication date
01-12-2020
Publisher
BioMed Central
Published in
BMC Oral Health / Issue 1/2020
Electronic ISSN: 1472-6831
DOI
https://doi.org/10.1186/s12903-019-0994-z

Other articles of this Issue 1/2020

BMC Oral Health 1/2020 Go to the issue