Abstract

Background

Electronic apex locators (EALs) are modern devices used to determine the working length during root canal preparation. The newest endodontic motors provide an integrated EAL with auto-stop function to prevent instrumentation beyond the predefined working length during rotary root canal preparation. The aim of this study was to compare the accuracy of the auto-stop function of the VDW.Gold RECIPROC motor (VDW, Munich, Germany), the EndoPilot motor (Schlumbohm, Brokstedt, Germany) and the manual measurement with Raypex 6 (VDW, Munich, Germany) to detect the apical constriction (AC).

Methods

Ninety human teeth were chosen and randomly assigned to three experimental groups (30 teeth each): VDW.Gold RECIPROC motor continuous measuring (RCM), EndoPilot continuous measuring (ECM) and Raypex 6 manual measuring (RMM). When the measurement file reached the AC, the file was fixed in the tooth. The tooth was embedded in acrylic resin and the root tip was exposed, so that the histologic structure of the root canal and the file tip was visible for microscopic analysis. Afterwards, the distance of the file tip to the AC (DAC) was automatically computed with a specially developed software tool.

Results

The mean DAC were −13.18 μm (SD 88.46 μm) for RMM, −22.70 μm (SD 91.57 μm) for RCM and 18.74 μm (SD 88.11 μm) for ECM. The differences were not statistically significant (P = 0.181). The rates for instrumentation beyond the AC were not statistically different (Chi2 = 4.753, p = 0.096).

Conclusions

All measurement methods showed a high accuracy in detecting the AC. The auto-stop function of these endodontic motors is a reliable addition to the endodontic armamentarium.
Literature
1.
go back to reference Bane K, Faye B, Sarr M, Niang SO, Ndiaye D, Machtou P. Root canal shaping by single-file systems and rotary instruments: a laboratory study. Iran Endod J. 2015;10(2):135–9.PubMedPubMedCentral Bane K, Faye B, Sarr M, Niang SO, Ndiaye D, Machtou P. Root canal shaping by single-file systems and rotary instruments: a laboratory study. Iran Endod J. 2015;10(2):135–9.PubMedPubMedCentral
5.
go back to reference Stoll R, Betke K, Stachniss V. The influence of different factors on the survival of root canal fillings: a 10-year retrospective study. J Endod. 2005;31(11):783–90.CrossRefPubMed Stoll R, Betke K, Stachniss V. The influence of different factors on the survival of root canal fillings: a 10-year retrospective study. J Endod. 2005;31(11):783–90.CrossRefPubMed
7.
go back to reference Simon S, Machtou P, Adams N, Tomson P, Lumley P. Apical limit and working length in endodontics. Dent Update. 2009;36(3):146–50. 53CrossRefPubMed Simon S, Machtou P, Adams N, Tomson P, Lumley P. Apical limit and working length in endodontics. Dent Update. 2009;36(3):146–50. 53CrossRefPubMed
9.
go back to reference Al-Omari MA, Dummer PM. Canal blockage and debris extrusion with eight preparation techniques. J Endod. 1995;21(3):154–8.CrossRefPubMed Al-Omari MA, Dummer PM. Canal blockage and debris extrusion with eight preparation techniques. J Endod. 1995;21(3):154–8.CrossRefPubMed
12.
13.
go back to reference Hör D, Attin T. The accuracy of electronic working length determination. Int Endod J. 2004;37(2):125–31.CrossRef Hör D, Attin T. The accuracy of electronic working length determination. Int Endod J. 2004;37(2):125–31.CrossRef
17.
go back to reference Campbell D, Friedman S, Nguyen HQ, Kaufman A, Keila S. Apical extent of rotary canal instrumentation with an apex-locating handpiece in vitro Oral Surg Oral Med Oral Pathol Oral Radiol Endod. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1998;85(3):319–24.CrossRefPubMed Campbell D, Friedman S, Nguyen HQ, Kaufman A, Keila S. Apical extent of rotary canal instrumentation with an apex-locating handpiece in vitro Oral Surg Oral Med Oral Pathol Oral Radiol Endod. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1998;85(3):319–24.CrossRefPubMed
22.
go back to reference Schlumbohm GDI. Verfahren zum Bestimmen der Position einer in einen Zahnwurzelkanal eingeführten Messelektrode. Google Patents; 2008. Schlumbohm GDI. Verfahren zum Bestimmen der Position einer in einen Zahnwurzelkanal eingeführten Messelektrode. Google Patents; 2008.
25.
go back to reference Grimberg F, Banegas G, Chiacchio L, Zmener O. In vivo determination of root canal length: a preliminary report using the tri auto ZX apex-locating handpiece. Int Endod J. 2002;35(7):590–3.CrossRefPubMed Grimberg F, Banegas G, Chiacchio L, Zmener O. In vivo determination of root canal length: a preliminary report using the tri auto ZX apex-locating handpiece. Int Endod J. 2002;35(7):590–3.CrossRefPubMed
26.
go back to reference Kustarci A, Arslan D, Altunbas D. In vitro comparison of working length determination using three different electronic apex locators. Dent Res J (Isfahan). 2014;11(5):568–73. Kustarci A, Arslan D, Altunbas D. In vitro comparison of working length determination using three different electronic apex locators. Dent Res J (Isfahan). 2014;11(5):568–73.
Metadata
Title
The accuracy of the auto-stop function of different endodontic devices in detecting the apical constriction
Authors
David W. Christofzik
Andreas Bartols
Mahmoud Khaled
Birte Größner-Schreiber
Christof E. Dörfer
Publication date
01-12-2017
Publisher
BioMed Central
Published in
BMC Oral Health / Issue 1/2017
Electronic ISSN: 1472-6831
DOI
https://doi.org/10.1186/s12903-017-0425-y

Other articles of this Issue 1/2017

BMC Oral Health 1/2017 Go to the issue