Published in: BMC Musculoskeletal Disorders 1/2021

Open Access 01-12-2021 | Technical advance

A novel 3D-printed locking cage for anterior atlantoaxial fixation and fusion: case report and in vitro biomechanical evaluation

Authors: Shenglin Wang, Huijie Leng, Yinglun Tian, Nanfang Xu, Zhongjun Liu

Published in: BMC Musculoskeletal Disorders | Issue 1/2021

Login to get access

Abstract

Background

Treatment of atlantoaxial dislocation is aimed at reduction and stabilization of the atlantoaxial joint. 3D printing refers to a process where additive manufacturing is achieved under precise computer control. Literature on its utilization in anterior atlantoaxial fixation and fusion is rare. This study is the first report on a 3D-printed locking cage used in the anterior procedure for atlantoaxial dislocation.

Methods

A middle-aged male in his 40s presented with weakness and numbness of his extremities for 3 years and could only walk slowly with assistance. Imaging studies revealed severe anterior migration of C1, irreducible atlantoaxial dislocation, and severe cervical-medullary compression. A preoperative plan consisting of trans-oral soft tissue release and fixation using tailor-designed 3D-printed cages was devised. Following fluoroscopic confirmation of reduction of the atlantoaxial joints, two customized 3D-printed cages made of titanium alloy were inserted into the bilateral facet joints, which were then locked by six screws into the lateral masses of C1 and C2. The microstructure of the inserted cages was optimized for improved biomechanical stability and enhanced osseo-integration, without the need for bone grafting. In addition, a biomechanical test was performed on seven human cadaveric specimens comparing the novel implant with the conventional C1 lateral mass-C2 pedicle screw construct in three modes of motion (flexion-extension, lateral bending, axial rotation).

Results

Improvement of neurologic function in the patient was evident immediately after surgery. He was able to walk independently 1 month post-operatively. At the 12-month follow-up, coronal reconstruction of CT demonstrated properly-positioned 3D-printed cages, evidence of osseo-integration at the bone-implant interface, and no subsidence or displacement of the implant. Eighteen months out of surgery, the mJOA score improved to 15, and lateral X-ray confirmed reduction of atlanto-axial dislocation. Additionally, the new construct provided strong fixation comparable to that conferred by conventional constructs as there was no significant difference observed between the two groups in all three directions of motion.

Conclusions

The novel implant represents a new option in the treatment of irreducible atlantoaxial dislocation. It can provide strong anterior support for solid fixation and fusion with a low profile and a microstructure that obviates the need for bone grafting.
Appendix
Available only for authorised users
Literature
1.
go back to reference Wang S, et al. Novel surgical classification and treatment strategy for atlantoaxial dislocations. Spine (Phila Pa 1976). 2013;38(21):E1348–56.CrossRef Wang S, et al. Novel surgical classification and treatment strategy for atlantoaxial dislocations. Spine (Phila Pa 1976). 2013;38(21):E1348–56.CrossRef
2.
go back to reference Wu X, et al. Surgical strategies for the treatment of os odontoideum with atlantoaxial dislocation. J Neurosurg Spine. 2018;28(2):131–9.CrossRef Wu X, et al. Surgical strategies for the treatment of os odontoideum with atlantoaxial dislocation. J Neurosurg Spine. 2018;28(2):131–9.CrossRef
3.
go back to reference Guan J, et al. Effectiveness of posterior reduction and fixation in atlantoaxial dislocation: a retrospective cohort study of 135 patients with a treatment algorithm proposal. Eur Spine J. 2019. Guan J, et al. Effectiveness of posterior reduction and fixation in atlantoaxial dislocation: a retrospective cohort study of 135 patients with a treatment algorithm proposal. Eur Spine J. 2019.
4.
go back to reference Wang S, et al. Atlantoaxial rotatory fixed dislocation: report on a series of 32 pediatric cases. Spine (Phila Pa 1976). 2016;41(12):E725–32.CrossRef Wang S, et al. Atlantoaxial rotatory fixed dislocation: report on a series of 32 pediatric cases. Spine (Phila Pa 1976). 2016;41(12):E725–32.CrossRef
5.
go back to reference Wang C, et al. Open reduction of irreducible atlantoaxial dislocation by transoral anterior atlantoaxial release and posterior internal fixation. Spine (Phila Pa 1976). 2006;31(11):E306–13.CrossRef Wang C, et al. Open reduction of irreducible atlantoaxial dislocation by transoral anterior atlantoaxial release and posterior internal fixation. Spine (Phila Pa 1976). 2006;31(11):E306–13.CrossRef
6.
go back to reference Yin QS, et al. Transoral atlantoaxial reduction plate fixation for irreducible atlantoaxial dislocation. Chin J Traumatol. 2006;9(1):14–20.PubMed Yin QS, et al. Transoral atlantoaxial reduction plate fixation for irreducible atlantoaxial dislocation. Chin J Traumatol. 2006;9(1):14–20.PubMed
7.
go back to reference Cai H, et al. 3D Printing in Spine Surgery. Adv Exp Med Biol. 2018;1093:345–59.CrossRef Cai H, et al. 3D Printing in Spine Surgery. Adv Exp Med Biol. 2018;1093:345–59.CrossRef
8.
go back to reference Xu N, et al. Reconstruction of the upper cervical spine using a personalized 3D-printed vertebral body in an adolescent with Ewing sarcoma. Spine (Phila Pa 1976). 2016;41(1):E50–4.CrossRef Xu N, et al. Reconstruction of the upper cervical spine using a personalized 3D-printed vertebral body in an adolescent with Ewing sarcoma. Spine (Phila Pa 1976). 2016;41(1):E50–4.CrossRef
9.
go back to reference Tetreault L, et al. The modified Japanese Orthopaedic association scale: establishing criteria for mild, moderate and severe impairment in patients with degenerative cervical myelopathy. Eur Spine J. 2017;26(1):78–84.CrossRef Tetreault L, et al. The modified Japanese Orthopaedic association scale: establishing criteria for mild, moderate and severe impairment in patients with degenerative cervical myelopathy. Eur Spine J. 2017;26(1):78–84.CrossRef
10.
go back to reference Wang S, et al. Radiographic evaluation of the technique for C1 lateral mass and C2 pedicle screw fixation in three hundred nineteen cases. Spine (Phila Pa 1976). 2011;36(1):3–8.CrossRef Wang S, et al. Radiographic evaluation of the technique for C1 lateral mass and C2 pedicle screw fixation in three hundred nineteen cases. Spine (Phila Pa 1976). 2011;36(1):3–8.CrossRef
11.
go back to reference Yang J, et al. In vivo study of a self-stabilizing artificial vertebral body fabricated by electron beam melting. Spine (Phila Pa 1976). 2014;39(8):E486–92.CrossRef Yang J, et al. In vivo study of a self-stabilizing artificial vertebral body fabricated by electron beam melting. Spine (Phila Pa 1976). 2014;39(8):E486–92.CrossRef
12.
go back to reference Phan K, et al. Application of a 3D custom printed patient specific spinal implant for C1/2 arthrodesis. J Spine Surg. 2016;2(4):314–8.CrossRef Phan K, et al. Application of a 3D custom printed patient specific spinal implant for C1/2 arthrodesis. J Spine Surg. 2016;2(4):314–8.CrossRef
13.
go back to reference Du JY, et al. Biomechanical analysis of screw constructs for atlantoaxial fixation in cadavers: a systematic review and meta-analysis. J Neurosurg Spine. 2015;22(2):151–61.CrossRef Du JY, et al. Biomechanical analysis of screw constructs for atlantoaxial fixation in cadavers: a systematic review and meta-analysis. J Neurosurg Spine. 2015;22(2):151–61.CrossRef
14.
go back to reference Chun DH, et al. Biomechanical comparison of four different Atlantoaxial posterior fixation constructs in adults: a finite element study. Spine (Phila Pa 1976). 2018;43(15):E891–7.CrossRef Chun DH, et al. Biomechanical comparison of four different Atlantoaxial posterior fixation constructs in adults: a finite element study. Spine (Phila Pa 1976). 2018;43(15):E891–7.CrossRef
15.
go back to reference Liu S, et al. Biomechanical evaluation of C1 lateral mass and C2 translaminar bicortical screws in atlantoaxial fixation: an in vitro human cadaveric study. Spine J. 2018;18(4):674–81.CrossRef Liu S, et al. Biomechanical evaluation of C1 lateral mass and C2 translaminar bicortical screws in atlantoaxial fixation: an in vitro human cadaveric study. Spine J. 2018;18(4):674–81.CrossRef
16.
go back to reference Zhang B, et al. Biomechanical comparison of modified TARP technique versus modified Goel technique for the treatment of basilar invagination: a finite element analysis. Spine (Phila Pa 1976). 2016;41(8):E459–66.CrossRef Zhang B, et al. Biomechanical comparison of modified TARP technique versus modified Goel technique for the treatment of basilar invagination: a finite element analysis. Spine (Phila Pa 1976). 2016;41(8):E459–66.CrossRef
Metadata
Title
A novel 3D-printed locking cage for anterior atlantoaxial fixation and fusion: case report and in vitro biomechanical evaluation
Authors
Shenglin Wang
Huijie Leng
Yinglun Tian
Nanfang Xu
Zhongjun Liu
Publication date
01-12-2021
Publisher
BioMed Central
Published in
BMC Musculoskeletal Disorders / Issue 1/2021
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-021-03987-2

Other articles of this Issue 1/2021

BMC Musculoskeletal Disorders 1/2021 Go to the issue