Accuracy of planar anteversion measurements using anteroposterior radiographs

Authors: Gwang Chul Lee, Sang Hong Lee, Sin Wook Kang, Hyung Seok Park, Suenghwan Jo

Published in: BMC Musculoskeletal Disorders | Issue 1/2019

Login to get access

Abstract

Background

Several methods using simple anteroposterior (AP) radiographs have been suggested for the measurement of anteversion of the cup component after total hip arthroplasty. Herein, we compared six widely used anteversion measurement methods using two different types of AP radiograph, the conventional pelvis AP and hip-centered AP radiographs, to identify the measurement method and the type of radiograph that would provide the highest accuracy and reliability.

Methods

We developed two custom-made bi-planar anteversion measurement models for the validation test. The models were designed for pelvis AP and hip-centered AP radiographs, respectively. The radiographs were acquired using the inclination angles of both models, changing from 10° to 70° at 10° increments. For each inclination angle, anteversion was changed from 0° to 30° at 5° increments. The measurements were obtained independently by two orthopedic surgeons blinded from each other’s measurements, using the methods of 1) Pradhan et al., 2) Lewinnek et al., 3) Widmer et al., 4) Liaw et al., 5) Hassan et al., and 6) Ackland et al. The measurements were repeated after 2 months. The accuracy, compared with that of the reference angle, and intra-observer and inter-observer reliabilities of each method were calculated.

Results

The highest accuracy was found when the method of Liaw et al. was used with hip-centered AP radiographs, which showed a difference of 1.37° ± 1.73 from the reference angle. Moreover, regardless of the type of radiograph, the methods by Pradhan et al., Lewinnek et al., and Liaw et al. showed excellent correlations with the reference anteversion. However, substantial differences were found when the methods by Widmer et al., Hassan et al., and Ackland et al. were used, regardless of the type of radiograph used. When anteversion was measured in an inclination between 30° and 50°, the method of Pradhan et al., when used with pelvis AP radiographs, showed the highest accuracy (1.23° ± 0.92°). We also found no significant difference in anteversions between the measurements made on pelvic and hip-centered AP radiographs. Both interobserver and intraobserver reliabilities were high for all the measurements tested.

Conclusions

The methods by Pradhan et al., Liaw et al., and Lewinnek et al. may provide relatively accurate anteversion measurements with high reliability, regardless of the type of radiograph.
Literature
1.
go back to reference Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60(2):217–20.CrossRef Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60(2):217–20.CrossRef
2.
go back to reference Woo RY, Morrey BF. Dislocations after total hip arthroplasty. J Bone Joint Surg Am. 1982;64(9):1295–306.CrossRef Woo RY, Morrey BF. Dislocations after total hip arthroplasty. J Bone Joint Surg Am. 1982;64(9):1295–306.CrossRef
3.
4.
go back to reference Ritter MA. Dislocation and subluxation of the total hip replacement. Clin Orthop Relat Res. 1976;121:92–4. Ritter MA. Dislocation and subluxation of the total hip replacement. Clin Orthop Relat Res. 1976;121:92–4.
5.
go back to reference Wan Z, Dastane M, Dorr L, Malik A. Combined anteversion technique for total hip arthroplasty. J Arthroplast. 2008;467(1):119–27. Wan Z, Dastane M, Dorr L, Malik A. Combined anteversion technique for total hip arthroplasty. J Arthroplast. 2008;467(1):119–27.
6.
go back to reference Nunley RM, Keeney JA, Zhu J, Clohisy JC, Barrack RL. The reliability and variation of acetabular component anteversion measurements from cross-table lateral radiographs. J Arthroplast. 2011;26(6):84–7.CrossRef Nunley RM, Keeney JA, Zhu J, Clohisy JC, Barrack RL. The reliability and variation of acetabular component anteversion measurements from cross-table lateral radiographs. J Arthroplast. 2011;26(6):84–7.CrossRef
7.
go back to reference Eijer H, Leunig M, Mahomed MN, Ganz R. Cross-table lateral radiographs for screening of anterior femoral head-neck offset in patients with femoro-acetabular impingement. Hip Int. 2001;11(1):37–41.CrossRef Eijer H, Leunig M, Mahomed MN, Ganz R. Cross-table lateral radiographs for screening of anterior femoral head-neck offset in patients with femoro-acetabular impingement. Hip Int. 2001;11(1):37–41.CrossRef
8.
go back to reference Shin WC, Lee SM, Lee KW, Cho HJ, Lee JS, Suh KT. The reliability and accuracy of measuring anteversion of the acetabular component on plain anteroposterior and lateral radiographs after total hip arthroplasty. Bone Joint J. 2015;97-B:611–6.CrossRef Shin WC, Lee SM, Lee KW, Cho HJ, Lee JS, Suh KT. The reliability and accuracy of measuring anteversion of the acetabular component on plain anteroposterior and lateral radiographs after total hip arthroplasty. Bone Joint J. 2015;97-B:611–6.CrossRef
9.
go back to reference Arai N, Nakamura S, Matsushita T. Difference between 2 measurement methods of version angles of the acetabular component. J Arthroplast. 2007;22(5):715–20.CrossRef Arai N, Nakamura S, Matsushita T. Difference between 2 measurement methods of version angles of the acetabular component. J Arthroplast. 2007;22(5):715–20.CrossRef
10.
go back to reference Marx A, von Knoch M, Pfortner J, Wiese M, Saxler G. Misinterpretation of cup anteversion in total hip arthroplasty using planar radiography. Arch Orthop Trauma Surg. 2006;126(7):487–92.CrossRef Marx A, von Knoch M, Pfortner J, Wiese M, Saxler G. Misinterpretation of cup anteversion in total hip arthroplasty using planar radiography. Arch Orthop Trauma Surg. 2006;126(7):487–92.CrossRef
11.
go back to reference Nho JH, Lee YK, Kim HJ, Ha YC, Suh YS, Koo KH. Reliability and validity of measuring version of the acetabular component. J Bone Joint Surg Br. 2012;94(1):32–6.CrossRef Nho JH, Lee YK, Kim HJ, Ha YC, Suh YS, Koo KH. Reliability and validity of measuring version of the acetabular component. J Bone Joint Surg Br. 2012;94(1):32–6.CrossRef
12.
go back to reference Nomura T, Naito M, Nakamura Y, Ida T, Kuroda D, Kobayashi T, Sakamoto T, Seo H. An analysis of the best method for evaluating anteversion of the acetabular component after total hip replacement on plain radiographs. Bone Joint J. 2014;96-B(5):597–603.CrossRef Nomura T, Naito M, Nakamura Y, Ida T, Kuroda D, Kobayashi T, Sakamoto T, Seo H. An analysis of the best method for evaluating anteversion of the acetabular component after total hip replacement on plain radiographs. Bone Joint J. 2014;96-B(5):597–603.CrossRef
13.
go back to reference Pradhan R. Planar anteversion of the acetabular cup as determined from plain anteroposterior radiographs. J Bone Joint Surg Br. 1999;81:431–5.CrossRef Pradhan R. Planar anteversion of the acetabular cup as determined from plain anteroposterior radiographs. J Bone Joint Surg Br. 1999;81:431–5.CrossRef
14.
go back to reference Widmer KH. A simplified method to determine acetabular cup anteversion from plain radiographs. J Arthroplast. 2004;19(3):387–90.CrossRef Widmer KH. A simplified method to determine acetabular cup anteversion from plain radiographs. J Arthroplast. 2004;19(3):387–90.CrossRef
15.
go back to reference Liaw CK, Hou SM, Yang RS, Wu TY, Fuh CS. A new tool for measuring cup orientation in total hip arthroplasties from plain radiographs. Clin Orthop Relat Res. 2006;451:134–9.CrossRef Liaw CK, Hou SM, Yang RS, Wu TY, Fuh CS. A new tool for measuring cup orientation in total hip arthroplasties from plain radiographs. Clin Orthop Relat Res. 2006;451:134–9.CrossRef
16.
go back to reference Hassan DM, Johnston GH, Dust WN, Watson LG, Cassidy D. Radiographic calculation of anteversion in acetabular prostheses. J Arthroplast. 1995;10(3):369–72.CrossRef Hassan DM, Johnston GH, Dust WN, Watson LG, Cassidy D. Radiographic calculation of anteversion in acetabular prostheses. J Arthroplast. 1995;10(3):369–72.CrossRef
17.
go back to reference Ackland MK, Bourne WB, Uhthoff HK. Anteversion of the acetabular cup. Measurement of angle after total hip replacement. J Bone Joint Surg Br. 1986;68(3):409–13.CrossRef Ackland MK, Bourne WB, Uhthoff HK. Anteversion of the acetabular cup. Measurement of angle after total hip replacement. J Bone Joint Surg Br. 1986;68(3):409–13.CrossRef
18.
go back to reference Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33(1):159–74.CrossRef Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33(1):159–74.CrossRef
19.
go back to reference Coventry MB. Late dislocations in patients with Charnley total hip arthroplasty. J Bone Joint Surg Am. 1985;67(6):832–41.CrossRef Coventry MB. Late dislocations in patients with Charnley total hip arthroplasty. J Bone Joint Surg Am. 1985;67(6):832–41.CrossRef
20.
go back to reference Harris WH. Advances in surgical technique for total hip replacement: without and with osteotomy of the greater trochanter. Clin Orthop Relat Res. 1980;146:188–204. Harris WH. Advances in surgical technique for total hip replacement: without and with osteotomy of the greater trochanter. Clin Orthop Relat Res. 1980;146:188–204.
21.
go back to reference Seki M, Yuasa N, Ohkuni K. Analysis of optimal range of socket orientations in total hip arthroplasty with use of computer-aided design simulation. J Orthop Res. 1998;16(4):513–7.CrossRef Seki M, Yuasa N, Ohkuni K. Analysis of optimal range of socket orientations in total hip arthroplasty with use of computer-aided design simulation. J Orthop Res. 1998;16(4):513–7.CrossRef
22.
go back to reference Lazennec JY, Boyer P, Gorin M, Catonné Y, Rousseau MA. Acetabular anteversion with CT in supine, simulated standing, and sitting positions in a THA patient population. Clin Orthop Relat Res. 2011;469(4):1103–9.CrossRef Lazennec JY, Boyer P, Gorin M, Catonné Y, Rousseau MA. Acetabular anteversion with CT in supine, simulated standing, and sitting positions in a THA patient population. Clin Orthop Relat Res. 2011;469(4):1103–9.CrossRef
23.
go back to reference Stem ES, O’Connor MI, Kransdorf MJ, Crook J. Computed tomography analysis of acetabular anteversion and abduction. Skelet Radiol. 2006;35(6):385–9.CrossRef Stem ES, O’Connor MI, Kransdorf MJ, Crook J. Computed tomography analysis of acetabular anteversion and abduction. Skelet Radiol. 2006;35(6):385–9.CrossRef
24.
go back to reference Dorr LD, Wan Z, Malik A, Zhu J, Dastane M, Deshmane P. A comparison of surgeon estimation and computed tomographic measurement of femoral component anteversion in cementless total hip arthroplasty. J Bone Joint Surg Am. 2009;91(11):2598–604.CrossRef Dorr LD, Wan Z, Malik A, Zhu J, Dastane M, Deshmane P. A comparison of surgeon estimation and computed tomographic measurement of femoral component anteversion in cementless total hip arthroplasty. J Bone Joint Surg Am. 2009;91(11):2598–604.CrossRef
25.
go back to reference Sugano N, Noble PC, Kamaric E. A comparison of alternative methods of measuring femoral anteversion. J Comput Assist Tomogr. 1998;22(4):610–4.CrossRef Sugano N, Noble PC, Kamaric E. A comparison of alternative methods of measuring femoral anteversion. J Comput Assist Tomogr. 1998;22(4):610–4.CrossRef
26.
go back to reference Ghelman B, Kepler CK, Lyman S, Della Valle AG. CT outperforms radiography for determination of acetabular cup version after THA. Clin Orthop Relat Res. 2009;467(9):2362–70.CrossRef Ghelman B, Kepler CK, Lyman S, Della Valle AG. CT outperforms radiography for determination of acetabular cup version after THA. Clin Orthop Relat Res. 2009;467(9):2362–70.CrossRef
27.
go back to reference Lu M, Zhou YX, Du H, Zhang J, Liu J. Reliability and validity of measuring acetabular component orientation by plain anteroposterior radiographs. Clin Orthop Relat Res. 2013;471(9):2987–94.CrossRef Lu M, Zhou YX, Du H, Zhang J, Liu J. Reliability and validity of measuring acetabular component orientation by plain anteroposterior radiographs. Clin Orthop Relat Res. 2013;471(9):2987–94.CrossRef
Metadata
Title
Accuracy of planar anteversion measurements using anteroposterior radiographs
Authors
Gwang Chul Lee
Sang Hong Lee
Sin Wook Kang
Hyung Seok Park
Suenghwan Jo
Publication date
01-12-2019
Publisher
BioMed Central
Keywords
Hip-TEP
Hip-TEP
Published in
BMC Musculoskeletal Disorders / Issue 1/2019
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-019-2979-0

Other articles of this Issue 1/2019

BMC Musculoskeletal Disorders 1/2019 Go to the issue