Published in: BMC Musculoskeletal Disorders 1/2018

Open Access 01-12-2018 | Research article

Elasticity standard values of the Achilles tendon assessed with acoustic radiation force impulse elastography on healthy volunteers: a cross section study

Published in: BMC Musculoskeletal Disorders | Issue 1/2018

Login to get access

Abstract

Background

To determine normal values for acoustic radiation force impulse (ARFI) shear wave elastography of the Achilles tendon.

Methods

Using the VTIQ mode with the Acuson S3000™ (Siemens Healthcare, Erlangen, Germany), we measured the shear wave velocity (SWV) in m/s and the diameter in mm of both Achilles tendons in 182 healthy volunteers. The tendon was displayed in a sagittal view with a relaxed tendon. The parameters were tested for correlations with the anthropometric data of the subjects, between the genders and different age groups, as well as information obtained from the history, such as smoking and sporting activities.

Results

Using a sagittal acoustic window, we determined a mean SWV of 9.09 ± 0.71 m/s for the left Achilles tendon and 9.17 ± 0.61 m/s for the right. There was a significant difference between the results for the right and left side (p < 0.05). The diameter on the left was 4.7 ± 0.9 mm. On the right, it was 4.8 ± 0.9 mm. Likewise there was a significant difference between the results for the diameter of the left and right side (p < 0.05). Neither gender, body mass index (BMI) nor age had a significant effect on either of the measured parameters (p > 0.05). The same goes for the consumption of tobacco and alcohol (p > 0.05).

Conclusions

Age, gender, BMI, smoking or the consumption of alcohol did not affect either the elasticity or the diameter of the Achilles tendon.
Literature
1.
5.
go back to reference Trottmann M, Rübenthaler J, Marcon J, Stief CG, Reiser MF, Clevert DA. Differences of standard values of supersonic shear imaging and ARFI technique - in-vivo study of testicular tissue. Clin Hemorheol Microcirc. 2016; https://doi.org/10.3233/CH-168039. Trottmann M, Rübenthaler J, Marcon J, Stief CG, Reiser MF, Clevert DA. Differences of standard values of supersonic shear imaging and ARFI technique - in-vivo study of testicular tissue. Clin Hemorheol Microcirc. 2016; https://​doi.​org/​10.​3233/​CH-168039.
10.
go back to reference Gerber L, Buslau A, Reichhardt A, et al. Assessment of liver fibrosis with supersonic shear wave elastography (Aixplorer™) in comparison to the transient elastography (Fibroscan®) and acoustic radiation force impulse (ARFI)-imaging in patients with chronic liver disease. Ultraschall Med - Eur J Ultrasound. 2013;34(S 01):PS7_01. https://doi.org/10.1055/s-0033-1355078. Gerber L, Buslau A, Reichhardt A, et al. Assessment of liver fibrosis with supersonic shear wave elastography (Aixplorer™) in comparison to the transient elastography (Fibroscan®) and acoustic radiation force impulse (ARFI)-imaging in patients with chronic liver disease. Ultraschall Med - Eur J Ultrasound. 2013;34(S 01):PS7_01. https://​doi.​org/​10.​1055/​s-0033-1355078.
16.
go back to reference Madaras EI, Perez J, Sobel BE, Mottley JG, Miller JG. Anisotropy of the ultrasonic backscatter of myocardial tissue: II. Measurements in vivo. J Acoust Soc Am. 1988;83(2):762–9.CrossRefPubMed Madaras EI, Perez J, Sobel BE, Mottley JG, Miller JG. Anisotropy of the ultrasonic backscatter of myocardial tissue: II. Measurements in vivo. J Acoust Soc Am. 1988;83(2):762–9.CrossRefPubMed
17.
go back to reference Holland MR, Lewis SH, Hall CS, et al. Effects of tissue anisotropy on the spectral characteristics of ultrasonic backscatter measured with a clinical imaging system. Ultrason Imaging. 1998;20(3):178–90.CrossRefPubMed Holland MR, Lewis SH, Hall CS, et al. Effects of tissue anisotropy on the spectral characteristics of ultrasonic backscatter measured with a clinical imaging system. Ultrason Imaging. 1998;20(3):178–90.CrossRefPubMed
18.
go back to reference Ruan Z, Zhao B, Qi H, et al. Elasticity of healthy Achilles tendon decreases with the increase of age as determined by acoustic radiation force impulse imaging. Int J Clin Exp Med. 2015;8(1):1043–50.PubMedPubMedCentral Ruan Z, Zhao B, Qi H, et al. Elasticity of healthy Achilles tendon decreases with the increase of age as determined by acoustic radiation force impulse imaging. Int J Clin Exp Med. 2015;8(1):1043–50.PubMedPubMedCentral
Metadata
Title
Elasticity standard values of the Achilles tendon assessed with acoustic radiation force impulse elastography on healthy volunteers: a cross section study
Publication date
01-12-2018
Published in
BMC Musculoskeletal Disorders / Issue 1/2018
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-018-2056-0

Other articles of this Issue 1/2018

BMC Musculoskeletal Disorders 1/2018 Go to the issue