Open Access 01-12-2015 | Research article

Does osteoporosis reduce the primary tilting stability of cementless acetabular cups?

Authors: Christoph von Schulze Pellengahr, Lars V von Engelhardt, Bernd Wegener, Peter E Müller, Andreas Fottner, Patrick Weber, Ole Ackermann, Matthias Lahner, Wolfram Teske

Published in: BMC Musculoskeletal Disorders | Issue 1/2015

Login to get access

Abstract

Background

Cementless hip cups need sufficient primary tilting stability to achieve osseointegration. The aim of the study was to assess differences of the primary implant stability in osteoporotic bone and in bone with normal bone density. To assess the influence of different cup designs, two types of threaded and two types of press-fit cups were tested.

Methods

The maximum tilting moment for two different cementless threaded cups and two different cementless press-fit cups was determined in macerated human hip acetabuli with reduced (n=20) and normal bone density (n=20), determined using Q-CT. The tilting moments for each cup were determined five times in the group with reduced bone density and five times in the group with normal bone density, and the respective average values were calculated.

Results

The mean maximum extrusion force of the threaded cup Zintra was 5670.5 N (max. tilting moment 141.8 Nm) in bone with normal density and.5748.3 N (max. tilting moment 143.7 Nm) in osteoporotic bone. For the Hofer Imhof (HI) threaded cup it was 7681.5 N (192.0 Nm) in bone with normal density and 6828.9 N (max. tilting moment 170.7 Nm) in the group with osteoporotic bone. The mean maximum extrusion force of the macro-textured press-fit cup Metallsockel CL was 3824.6 N (max. tilting moment 95.6 Nm) in bone with normal and 2246.2 N (max. tilting moment 56.2 Nm) in osteoporotic bone. For the Monoblock it was 1303.8 N (max. tilting moment 32.6 Nm) in normal and 1317 N (max. tilting moment 32.9 Nm) in osteoporotic bone. There was no significance. A reduction of the maximum tilting moment in osteoporotic bone of the ESKA press-fit cup Metallsockel CL was noticed.

Conclusion

Results on macerated bone specimens showed no statistically significant reduction of the maximum tilting moment in specimens with osteoporotic bone density compared to normal bone, neither for threaded nor for the press-fit cups. With the limitation that the results were obtained using macerated bone, we could not detect any restrictions for the clinical indication of the examined cementless cups in osteoporotic bone.
Literature
1.
go back to reference Boutin P. Total arthroplasty of the hip by fritted aluminum prosthesis. Experimental study and 1st clinical applications. Rev Chir Orthop Reparatrice Appar Mot. 1972;58:229–46.PubMed Boutin P. Total arthroplasty of the hip by fritted aluminum prosthesis. Experimental study and 1st clinical applications. Rev Chir Orthop Reparatrice Appar Mot. 1972;58:229–46.PubMed
2.
go back to reference Charnley GJ. Anchorage of the femoral head prothesis to the shaft of the femur. J Bone Joint Surg Br. 1960;42:28.PubMed Charnley GJ. Anchorage of the femoral head prothesis to the shaft of the femur. J Bone Joint Surg Br. 1960;42:28.PubMed
3.
go back to reference Bereiter H, Burgi M, Rahn BA. The temporal behavior of the anchorage of a cement-free implanted acetabular cup in animal experiments. Orthopade. 1992;21:63–70.PubMed Bereiter H, Burgi M, Rahn BA. The temporal behavior of the anchorage of a cement-free implanted acetabular cup in animal experiments. Orthopade. 1992;21:63–70.PubMed
4.
go back to reference Albrektsson T, Branemark PI, Hansson HA, Lindstrom J. Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man. Acta Orthop Scand. 1981;52:155–70.CrossRefPubMed Albrektsson T, Branemark PI, Hansson HA, Lindstrom J. Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man. Acta Orthop Scand. 1981;52:155–70.CrossRefPubMed
6.
go back to reference Engelke K, Adams JE, Armbrecht G, Augat P, Bogado CE, Bouxsein ML, et al. Clinical use of quantitative computed tomography and peripheral quantitative computed tomography in the management of osteoporosis in adults: The 2007 ISCD Official Positions. J Clin Densitom. 2008;11 Suppl 1:123–62.CrossRefPubMed Engelke K, Adams JE, Armbrecht G, Augat P, Bogado CE, Bouxsein ML, et al. Clinical use of quantitative computed tomography and peripheral quantitative computed tomography in the management of osteoporosis in adults: The 2007 ISCD Official Positions. J Clin Densitom. 2008;11 Suppl 1:123–62.CrossRefPubMed
7.
go back to reference von Schulze PC, Bürkner A, Plitz W, Mazoochian F, Dürr HR, Müller PE, et al. Macerated human hip bone preparation as a substitute for fresh frozen preparations for testing the primary stability and frictional torques of cementless hip cups. Z Orthop. 2006;6:601–96. von Schulze PC, Bürkner A, Plitz W, Mazoochian F, Dürr HR, Müller PE, et al. Macerated human hip bone preparation as a substitute for fresh frozen preparations for testing the primary stability and frictional torques of cementless hip cups. Z Orthop. 2006;6:601–96.
8.
go back to reference Plitz W. Der Einfluß von Verformung, Spiel und Werkstoffpaarung auf das oszillierende Reibmoment bei künstlichen Hüftgelenkpfannen. PhD thesis. RWTH Aachen University; 1984. Plitz W. Der Einfluß von Verformung, Spiel und Werkstoffpaarung auf das oszillierende Reibmoment bei künstlichen Hüftgelenkpfannen. PhD thesis. RWTH Aachen University; 1984.
9.
go back to reference Hadjari MH, Hollis M, Hofmann OE, Flahiff CM, Nelson CL. Initial stability of Porus coated Acetabular Implants. Clin Orthop Relat Res. 1994;307:117–23.PubMed Hadjari MH, Hollis M, Hofmann OE, Flahiff CM, Nelson CL. Initial stability of Porus coated Acetabular Implants. Clin Orthop Relat Res. 1994;307:117–23.PubMed
10.
go back to reference Ohlin A, Balkfors B. Stability of cemented sockets after 3-14 years. J Arthroplasty. 1992;7:87–92.CrossRefPubMed Ohlin A, Balkfors B. Stability of cemented sockets after 3-14 years. J Arthroplasty. 1992;7:87–92.CrossRefPubMed
11.
go back to reference Jäger M, Küßwetter W, Rütt J, Ungethüm M, Burkhardt R. Losening of acetabular prosthesis with various methods of implantation used in cadaver-hipjoints. Z Orthop Ihre Grenzgeb. 1974;112:34–44.PubMed Jäger M, Küßwetter W, Rütt J, Ungethüm M, Burkhardt R. Losening of acetabular prosthesis with various methods of implantation used in cadaver-hipjoints. Z Orthop Ihre Grenzgeb. 1974;112:34–44.PubMed
12.
go back to reference Kuhn A, Scheller G, Schwarz M. Primary stability of cement-free press-fit acetabulum cups: In vitro displacement studies. Biomed Tech. 1999;44 Suppl 12:356–9.CrossRef Kuhn A, Scheller G, Schwarz M. Primary stability of cement-free press-fit acetabulum cups: In vitro displacement studies. Biomed Tech. 1999;44 Suppl 12:356–9.CrossRef
13.
go back to reference Olory B, Havet E, Gabrion A, Vernois J, Mertl P. Comparative in vitro assessment of the primary stability of cementless press-fit acetabular cups. Acta Orthop Belg. 2004;70 Suppl 1:31–7.PubMed Olory B, Havet E, Gabrion A, Vernois J, Mertl P. Comparative in vitro assessment of the primary stability of cementless press-fit acetabular cups. Acta Orthop Belg. 2004;70 Suppl 1:31–7.PubMed
14.
go back to reference Wetzel R, Simnacher M, Scheller G. Initial stability of press-fit acetabular cups–an in-vitro study. Biomed Tech. 2005;50 Suppl 12:400–3.CrossRef Wetzel R, Simnacher M, Scheller G. Initial stability of press-fit acetabular cups–an in-vitro study. Biomed Tech. 2005;50 Suppl 12:400–3.CrossRef
15.
go back to reference Baleani M, Fognani R, Toni A. Initial stability of a cementless acetabular cup design: experimental investigation on the effect of adding fins to the rim of the cup. Artif Organs. 2001;25:664–9.CrossRefPubMed Baleani M, Fognani R, Toni A. Initial stability of a cementless acetabular cup design: experimental investigation on the effect of adding fins to the rim of the cup. Artif Organs. 2001;25:664–9.CrossRefPubMed
16.
go back to reference Kody MH, Kabo JM, Markolf KL, Dorey FJ, Amstutz HC. Strength of initial mechanical fixation of screw ring acetabular components. Clin Orthop. 1990;257:146–53.PubMed Kody MH, Kabo JM, Markolf KL, Dorey FJ, Amstutz HC. Strength of initial mechanical fixation of screw ring acetabular components. Clin Orthop. 1990;257:146–53.PubMed
Metadata
Title
Does osteoporosis reduce the primary tilting stability of cementless acetabular cups?
Authors
Christoph von Schulze Pellengahr
Lars V von Engelhardt
Bernd Wegener
Peter E Müller
Andreas Fottner
Patrick Weber
Ole Ackermann
Matthias Lahner
Wolfram Teske
Publication date
01-12-2015
Publisher
BioMed Central
Published in
BMC Musculoskeletal Disorders / Issue 1/2015
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-015-0554-x

Other articles of this Issue 1/2015

BMC Musculoskeletal Disorders 1/2015 Go to the issue