Comparison of sodium ion levels between an arterial blood gas analyzer and an autoanalyzer in preterm infants admitted to the neonatal intensive care unit: a retrospective study

Authors: Hyunho Kim, Jin Kyu Kim, Soo Chul Cho

Published in: BMC Pediatrics | Issue 1/2016

Login to get access

Abstract

Background

The difference in sodium ion levels determined with direct and indirect methods often exceeds the permissible limit clinically. Additionally, no previous study has assessed the difference in the sodium ion levels between direct and indirect methods in premature infants. Therefore, the present study aimed to compare sodium ion levels obtained using an arterial blood gas analyzer (ABGA; direct method) and an autoanalyzer (indirect method) to determine whether they are equivalent in premature infants.

Methods

The present retrospective study included 450 preterm infants (weight, <2500 g) who were admitted to the neonatal intensive care unit (NICU) of our hospital between March 2012 and April 2014. We compared sodium ion levels in 1041 samples analyzed using an ABGA (Stat Profile® CCX Series, Nova Biomedical, Waltham, MA) and an autoanalyzer (ADVIA® 2400 Clinical Chemistry System, Siemens, Tarrytown, NY). The data were evaluated using Spearman’s correlation coefficient analysis, Bland-Altman plot, Deming regression analysis, and multivariate logistic regression analysis.

Results

The mean sodium ion levels were 134.6 ± 3.5 mmol/L using the ABGA and 138.8 ± 4.7 mmol/L using the autoanalyzer (P < 0.001). Among the 1041 samples, 957 (91.9 %) showed lower sodium ion levels with the ABGA than with the autoanalyzer and 74 (7.1 %) showed lower sodium ion levels with the autoanalyzer than with the ABGA. The incidence of hyponatremia identified using the ABGA was 51.9 % (541/1041), while the incidence of hyponatremia identified using the autoanalyzer was only 14.0 % (146/1041). The Deming regression analysis of the sodium ion levels between the ABGA and the autoanalyzer yielded the following formula: autoanalyzer Na (mmol/L) = 20.7 + (0.9 × ABGA Na [mmol/L]). In the multivariate logistic regression analysis, low plasma protein level (<4.3 g/dL) was found to be an independent risk factor for a sodium ion level difference of >4 mmol/L between the two methods (odds ratio = 2.870, P < 0.001).

Conclusion

The sodium ion levels determined using the ABGA and the autoanalyzer might not be equivalent in premature infants admitted to the NICU. Therefore, clinicians should be careful when diagnosing sodium ion imbalance in premature infants and providing treatment.
Appendix
Available only for authorised users
Literature
1.
go back to reference Kloiber LL, Winn NJ, Shaffer SG, Hassanein RS. Late hyponatremia in very-low-birth-weight infants: incidence and associated risk factors. J Am Diet Assoc. 1996;96:880–4.CrossRefPubMed Kloiber LL, Winn NJ, Shaffer SG, Hassanein RS. Late hyponatremia in very-low-birth-weight infants: incidence and associated risk factors. J Am Diet Assoc. 1996;96:880–4.CrossRefPubMed
2.
go back to reference Moritz ML, Ayus JC. Hyponatremia in preterm neonates: Not a benign condition. Pediatrics. 2009;124:e1014–6.CrossRefPubMed Moritz ML, Ayus JC. Hyponatremia in preterm neonates: Not a benign condition. Pediatrics. 2009;124:e1014–6.CrossRefPubMed
4.
go back to reference Carl A, Edward R, David E. Tietz Textbook of Clinical Chemistry and Molecular Diagnostics. 4th ed. St. Louis, Mo: Elsevier Inc.; 2006. Carl A, Edward R, David E. Tietz Textbook of Clinical Chemistry and Molecular Diagnostics. 4th ed. St. Louis, Mo: Elsevier Inc.; 2006.
5.
go back to reference Lima-Oliveira G, Lippi G, Salvagno GL, Montagnana M, Picheth G, Guidi GC. Different manufacturers of syringes: a new source of variability in blood gas, acid–base balance and related laboratory test? Clin Biochem. 2012;45:683–7.CrossRefPubMed Lima-Oliveira G, Lippi G, Salvagno GL, Montagnana M, Picheth G, Guidi GC. Different manufacturers of syringes: a new source of variability in blood gas, acid–base balance and related laboratory test? Clin Biochem. 2012;45:683–7.CrossRefPubMed
6.
go back to reference Story DA, Morimatsu H, Egi M, Bellomo R. The effect of albumin concentration on plasma sodium and chloride measurements in critically ill patients. Anesth Analg. 2007;104:893–7.CrossRefPubMed Story DA, Morimatsu H, Egi M, Bellomo R. The effect of albumin concentration on plasma sodium and chloride measurements in critically ill patients. Anesth Analg. 2007;104:893–7.CrossRefPubMed
7.
go back to reference Dimeski G, Barnett RJ. Effects of total plasma protein concentration on plasma sodium, potassium and chloride measurements by an indirect ion selective electrode measuring system. Crit Care Resusc J Australas Acad Crit Care Med. 2005;7:12–5. Dimeski G, Barnett RJ. Effects of total plasma protein concentration on plasma sodium, potassium and chloride measurements by an indirect ion selective electrode measuring system. Crit Care Resusc J Australas Acad Crit Care Med. 2005;7:12–5.
8.
go back to reference Sharma B, Sarmah D. A comparative evaluation of sodium and potassium measurements by flame photometer and by direct ISE methods. Int J Health Sci Res IJHSR. 2013;3:59–65. Sharma B, Sarmah D. A comparative evaluation of sodium and potassium measurements by flame photometer and by direct ISE methods. Int J Health Sci Res IJHSR. 2013;3:59–65.
9.
go back to reference Yip PM, Chan MK, Zielinski N, Adeli K. Heparin interference in whole blood sodium measurements in a pediatric setting. Clin Biochem. 2006;39:391–5.CrossRefPubMed Yip PM, Chan MK, Zielinski N, Adeli K. Heparin interference in whole blood sodium measurements in a pediatric setting. Clin Biochem. 2006;39:391–5.CrossRefPubMed
11.
go back to reference Kim G-H. Pseudohyponatremia: does It matter in current clinical practice? Electrolytes Blood Press E BP. 2006;4:77–82.CrossRef Kim G-H. Pseudohyponatremia: does It matter in current clinical practice? Electrolytes Blood Press E BP. 2006;4:77–82.CrossRef
12.
go back to reference Chow E, Fox N, Gama R. Effect of low serum total protein on sodium and potassium measurement by ion-selective electrodes in critically ill patients. Br J Biomed Sci. 2008;65:128–31.CrossRefPubMed Chow E, Fox N, Gama R. Effect of low serum total protein on sodium and potassium measurement by ion-selective electrodes in critically ill patients. Br J Biomed Sci. 2008;65:128–31.CrossRefPubMed
13.
go back to reference Morimatsu H, Rocktäschel J, Bellomo R, Uchino S, Goldsmith D, Gutteridge G. Comparison of point-of-care versus central laboratory measurement of electrolyte concentrations on calculations of the anion gap and the strong ion difference. Anesthesiology. 2003;98:1077–84.CrossRefPubMed Morimatsu H, Rocktäschel J, Bellomo R, Uchino S, Goldsmith D, Gutteridge G. Comparison of point-of-care versus central laboratory measurement of electrolyte concentrations on calculations of the anion gap and the strong ion difference. Anesthesiology. 2003;98:1077–84.CrossRefPubMed
14.
go back to reference Lang T, Prinsloo P, Broughton AF, Lawson N, Marenah CB. Effect of low protein concentration on serum sodium measurement: pseudohypernatraemia and pseudonormonatraemia! Ann Clin Biochem. 2002;39:66–7.CrossRefPubMed Lang T, Prinsloo P, Broughton AF, Lawson N, Marenah CB. Effect of low protein concentration on serum sodium measurement: pseudohypernatraemia and pseudonormonatraemia! Ann Clin Biochem. 2002;39:66–7.CrossRefPubMed
16.
go back to reference Robert MK, Bonita FS, Joseph WS, Nina FS, Richard EB : Nelson Textbook of Pediatrics. 20th ed, Elsevier; 2015. Robert MK, Bonita FS, Joseph WS, Nina FS, Richard EB : Nelson Textbook of Pediatrics. 20th ed, Elsevier; 2015.
17.
go back to reference Karl B, Günther B, Alexander Roithmaier MD, Martin G, Alfred P, Versmold MDHT. Body composition, nutrition and fluid balance during the first two weeks of life in preterm neonates weighing less than 1500 grams. J Pediatr. 1991;118:615–20.CrossRef Karl B, Günther B, Alexander Roithmaier MD, Martin G, Alfred P, Versmold MDHT. Body composition, nutrition and fluid balance during the first two weeks of life in preterm neonates weighing less than 1500 grams. J Pediatr. 1991;118:615–20.CrossRef
Metadata
Title
Comparison of sodium ion levels between an arterial blood gas analyzer and an autoanalyzer in preterm infants admitted to the neonatal intensive care unit: a retrospective study
Authors
Hyunho Kim
Jin Kyu Kim
Soo Chul Cho
Publication date
01-12-2016
Publisher
BioMed Central
Published in
BMC Pediatrics / Issue 1/2016
Electronic ISSN: 1471-2431
DOI
https://doi.org/10.1186/s12887-016-0636-4

Other articles of this Issue 1/2016

BMC Pediatrics 1/2016 Go to the issue