01-12-2020 | Research article
Prospective assessment of plate-haptic rotationally asymmetric multifocal toric intraocular lens with near addition of + 1.5 diopters
Authors:
Tetsuro Oshika, Kazuno Negishi, Toru Noda, Hiroyuki Arai, Mikio Inamura, Yasushi Inoue, Teruyuki Miyoshi, Yoshifumi Fujita, Kazunori Miyata, Yumi Hasegawa
Published in:
BMC Ophthalmology
|
Issue 1/2020
Login to get access
Abstract
Background
To prospectively evaluate surgical results following implantation of rotationally asymmetric, plate-haptic, refractive segmented multifocal toric intraocular lenses (IOLs) with near addition of + 1.5 diopters (D) (Lentis Comfort LS-313 MF15T, Oculentis GmbH).
Methods
In 59 eyes of 41 patients, ocular examinations were conducted before and 1 day, 1 week, 1, 3, and 6 months after surgery. Uncorrected (UDVA) and corrected (CDVA) distance visual acuity, uncorrected (UIVA) and distance-corrected (DCIVA) intermediate visual acuity at 70 cm, and uncorrected (UNVA) and distance-corrected (DCNVA) near visual acuity at 30 cm were tested. A defocus curve was drawn, and the degree of disturbing photic phenomena were questioned.
Results
The IOL showed excellent rotational stability; the average absolute rotation was 1.66 ± 1.17 degrees from 1 day 1 to 6 months postoperatively, and 98.1 and 100% of eyes yielded rotation of less than 5 and 10 degrees, respectively. Postoperative distance and intermediate visual acuity were highly satisfactory; UDVA, CDVA, UIVA, and DCIVA were about 20/20, 20/16, 20/25, 20/25, respectively. Near visual acuity was suboptimal; UNVA and DCNVA were at approximately 20/60. The defocus curve analysis showed that 20/25 and 20/40 uncorrected visual acuity was attained at as close as 60 and 40 cm, respectively. Contrast sensitivity was within a normal range, and subjective photic phenomena were minimum.
Conclusions
The refractive segmented, rotationally asymmetric multifocal toric IOLs with + 1.5 D near addition showed superb rotational stability and highly satisfactory distance and intermediate vision. Contrast sensitivity was high and incidence of photic symptoms was very low.