Visual outcomes and safety after bilateral implantation of a trifocal presbyopia correcting intraocular lens in a Korean population: a prospective single-arm study

Authors: Tae-im Kim, Tae-Young Chung, Myoung Joon Kim, Kyounghwa Lee, Joon Young Hyon

Published in: BMC Ophthalmology | Issue 1/2020

Login to get access

Abstract

Background

To investigate the 3-month postoperative performance and safety after implantation of a trifocal intraocular lens (IOL) in a Korean population.

Methods

This was a clinical, prospective, multicenter, single-arm study. Forty-four subjects (88 eyes) with bilateral cataract with expected postoperative corneal astigmatism of < 1.00 diopter (D) and no ocular disease or eye condition underwent bilateral implantation of the AcrySof IQ® PanOptix IOL (TFNT00). Postoperative examination at 3 months included binocular defocus curve; binocular best corrected distance visual acuity (BCDVA); monocular/binocular uncorrected VA (UCVA) at distance (4 m), intermediate (60 cm), and near (40 cm); contrast sensitivity under photopic conditions with/without glare; and subjective outcomes, including satisfaction and spectacle independence.

Results

Binocular defocus curve at 3 months after bilateral implantation showed VA of 0.1 logMAR or better from + 0.5 D through − 2.5 D. Binocular BCDVA mean ± SD at 4 m was − 0.05 ± 0.07 logMAR. Binocular and monocular UCVA was 0.03 ± 0.1 and 0.08 ± 0.12 logMAR (4 m), − 0.00 ± 0.11 and 0.05 ± 0.13 logMAR (60 cm), and 0.03 ± 0.12 and 0.09 ± 0.13 logMAR (40 cm), respectively. Contrast sensitivity with glare was 1.67 ± 0.13, 1.91 ± 0.17, 1.54 ± 0.21, and 1.14 ± 0.20 log units at 3, 6, 12, and 18 cycles/degree, respectively. At near and intermediate distances, 84 and 77% of subjects reported good/excellent satisfaction, and 84 and 91% of subjects reported spectacle independence, respectively.

Conclusions

In a Korean population, visual performance of the trifocal TFNT00 IOL 3 months postoperatively was < 0.1 logMAR for binocular UCVA at all distances, with high subject satisfaction and spectacle independence.

Trial registration

www.ClinicalTrials.gov (NCT03268746). Registered August 31, 2017.
Literature
1.
go back to reference Javitt J, Brauweiler HP, Jacobi KW, Klemen U, Kohnen S, Quentin CD, et al. Cataract extraction with multifocal intraocular lens implantation: clinical, functional, and quality-of-life outcomes. Multicenter clinical trial in Germany and Austria. J Cataract Refract Surg. 2000;26:1356–66.CrossRef Javitt J, Brauweiler HP, Jacobi KW, Klemen U, Kohnen S, Quentin CD, et al. Cataract extraction with multifocal intraocular lens implantation: clinical, functional, and quality-of-life outcomes. Multicenter clinical trial in Germany and Austria. J Cataract Refract Surg. 2000;26:1356–66.CrossRef
2.
go back to reference Vingolo EM, Grenga P, Iacobelli L, Grenga R. Visual acuity and contrast sensitivity: AcrySof ReSTOR apodized diffractive versus AcrySof SA60AT monofocal intraocular lenses. J Cataract Refract Surg. 2007;33:1244–7.CrossRef Vingolo EM, Grenga P, Iacobelli L, Grenga R. Visual acuity and contrast sensitivity: AcrySof ReSTOR apodized diffractive versus AcrySof SA60AT monofocal intraocular lenses. J Cataract Refract Surg. 2007;33:1244–7.CrossRef
3.
go back to reference Kohnen T, Allen D, Boureau C, Dublineau P, Hartmann C, Mehdorn E, et al. European multicenter study of the AcrySof ReSTOR apodized diffractive intraocular lens. Ophthalmology. 2006;113:578–84.CrossRef Kohnen T, Allen D, Boureau C, Dublineau P, Hartmann C, Mehdorn E, et al. European multicenter study of the AcrySof ReSTOR apodized diffractive intraocular lens. Ophthalmology. 2006;113:578–84.CrossRef
4.
go back to reference Alfonso JF, Fernandez-Vega L, Baamonde MB, Montes-Mico R. Prospective visual evaluation of apodized diffractive intraocular lenses. J Cataract Refract Surg. 2007;33:1235–43.CrossRef Alfonso JF, Fernandez-Vega L, Baamonde MB, Montes-Mico R. Prospective visual evaluation of apodized diffractive intraocular lenses. J Cataract Refract Surg. 2007;33:1235–43.CrossRef
5.
go back to reference Bababekova Y, Rosenfield M, Hue JE, Huang RR. Font size and viewing distance of handheld smart phones. Optom Vis Sci. 2011;88:795–7.CrossRef Bababekova Y, Rosenfield M, Hue JE, Huang RR. Font size and viewing distance of handheld smart phones. Optom Vis Sci. 2011;88:795–7.CrossRef
6.
go back to reference Occupational Safety and Health Administration. Working Safely With Video Display Terminals. OSHA 3092. US Department of Labor; 1997. Occupational Safety and Health Administration. Working Safely With Video Display Terminals. OSHA 3092. US Department of Labor; 1997.
7.
go back to reference Gatinel D, Pagnoulle C, Houbrechts Y, Gobin L. Design and qualification of a diffractive trifocal optical profile for intraocular lenses. J Cataract Refract Surg. 2011;37:2060–7.CrossRef Gatinel D, Pagnoulle C, Houbrechts Y, Gobin L. Design and qualification of a diffractive trifocal optical profile for intraocular lenses. J Cataract Refract Surg. 2011;37:2060–7.CrossRef
8.
go back to reference Carson D, Hill WE, Hong X, Karakelle M. Optical bench performance of AcrySof® IQ ReSTOR®, AT LISA® tri, and FineVision® intraocular lenses. Clin Ophthalmol. 2014;8:2105–13.PubMedPubMedCentral Carson D, Hill WE, Hong X, Karakelle M. Optical bench performance of AcrySof® IQ ReSTOR®, AT LISA® tri, and FineVision® intraocular lenses. Clin Ophthalmol. 2014;8:2105–13.PubMedPubMedCentral
9.
go back to reference Kohnen T, Titke C, Bohm M. Trifocal intraocular lens implantation to treat visual demands in various distances following lens removal. Am J Ophthalmol. 2016;161:71–7.CrossRef Kohnen T, Titke C, Bohm M. Trifocal intraocular lens implantation to treat visual demands in various distances following lens removal. Am J Ophthalmol. 2016;161:71–7.CrossRef
10.
go back to reference Charness N, Dijkstra K, Jastrzembski T, Weaver S, Champion M. Monitor viewing distance for younger and older workers. Proc Hum Factors Ergon Soc Annu Meet. 2008;52:1614–7.CrossRef Charness N, Dijkstra K, Jastrzembski T, Weaver S, Champion M. Monitor viewing distance for younger and older workers. Proc Hum Factors Ergon Soc Annu Meet. 2008;52:1614–7.CrossRef
11.
go back to reference Kohnen T. First implantation of a diffractive quadrafocal (trifocal) intraocular lens. J Cataract Refract Surg. 2015;41:2330–2.CrossRef Kohnen T. First implantation of a diffractive quadrafocal (trifocal) intraocular lens. J Cataract Refract Surg. 2015;41:2330–2.CrossRef
12.
go back to reference AcrySof IQ PanOptix Presbyopia-Correcting IOL. Product Information. Forth Worth: Alcon Laboratories, Inc.; 2015. AcrySof IQ PanOptix Presbyopia-Correcting IOL. Product Information. Forth Worth: Alcon Laboratories, Inc.; 2015.
13.
go back to reference Carson D, Xu Z, Alexander E, Choi M, Zhao Z, Hong X. Optical bench performance of 3 trifocal intraocular lenses. J Cataract Refract Surg. 2016;42:1361–7.CrossRef Carson D, Xu Z, Alexander E, Choi M, Zhao Z, Hong X. Optical bench performance of 3 trifocal intraocular lenses. J Cataract Refract Surg. 2016;42:1361–7.CrossRef
14.
go back to reference Gundersen KG, Potvin R. Trifocal intraocular lenses: a comparison of the visual performance and quality of vision provided by two different lens designs. Clin Ophthalmol. 2017;11:1081–7.CrossRef Gundersen KG, Potvin R. Trifocal intraocular lenses: a comparison of the visual performance and quality of vision provided by two different lens designs. Clin Ophthalmol. 2017;11:1081–7.CrossRef
15.
go back to reference Bala C, Martinez AA, Kohnen T. Multicenter visual outcomes evaluation of a new trifocal presbyopia correcting IOL – 12 month results. Presented at: American Society of Cataract and Refractive Surgery, April 13-17, 2018.. Washington, DC. Bala C, Martinez AA, Kohnen T. Multicenter visual outcomes evaluation of a new trifocal presbyopia correcting IOL – 12 month results. Presented at: American Society of Cataract and Refractive Surgery, April 13-17, 2018.. Washington, DC.
16.
go back to reference Lapid-Gortzak R, Martinez A. Multicenter clinical investigation of visual function after bilateral implantation of two presbyopia-correcting trifocal IOLs [abstract]. Presented at: European Society of Cataract and Refractive Surgeons, September 22-26, 2018. Vienna; 2018. Lapid-Gortzak R, Martinez A. Multicenter clinical investigation of visual function after bilateral implantation of two presbyopia-correcting trifocal IOLs [abstract]. Presented at: European Society of Cataract and Refractive Surgeons, September 22-26, 2018. Vienna; 2018.
17.
go back to reference Ahn JH, Kim DH, Shyn KH. Investigation of the changes in refractive surgery trends in Korea. Korean J Ophthalmol. 2018;32:8–15.CrossRef Ahn JH, Kim DH, Shyn KH. Investigation of the changes in refractive surgery trends in Korea. Korean J Ophthalmol. 2018;32:8–15.CrossRef
18.
go back to reference Lee Y. Estimation of body size and growth patterns in Korean boys. J Physiol Anthropol. 2015;34:20.CrossRef Lee Y. Estimation of body size and growth patterns in Korean boys. J Physiol Anthropol. 2015;34:20.CrossRef
20.
go back to reference Pager CK. Expectations and outcomes in cataract surgery: a prospective test of 2 models of satisfaction. Arch Ophthalmol. 2004;122:1788–92.CrossRef Pager CK. Expectations and outcomes in cataract surgery: a prospective test of 2 models of satisfaction. Arch Ophthalmol. 2004;122:1788–92.CrossRef
21.
go back to reference Monaco G, Gari M, Di Censo F, Poscia A, Ruggi G, Scialdone A. Visual performance after bilateral implantation of 2 new presbyopia-correcting intraocular lenses: trifocal versus extended range of vision. J Cataract Refract Surg. 2017;43:737–47.CrossRef Monaco G, Gari M, Di Censo F, Poscia A, Ruggi G, Scialdone A. Visual performance after bilateral implantation of 2 new presbyopia-correcting intraocular lenses: trifocal versus extended range of vision. J Cataract Refract Surg. 2017;43:737–47.CrossRef
22.
go back to reference Lee DC, Lee SY, Kim YC. An epidemiological study of the risk factors associated with myopia in young adult men in Korea. Sci Rep. 2018;8:511.CrossRef Lee DC, Lee SY, Kim YC. An epidemiological study of the risk factors associated with myopia in young adult men in Korea. Sci Rep. 2018;8:511.CrossRef
23.
go back to reference Han SB, Jang J, Yang HK, Hwang JM, Park SK. Prevalence and risk factors of myopia in adult Korean population: Korea national health and nutrition examination survey 2013-2014 (KNHANES VI). PLoS One. 2019;14:e0211204.CrossRef Han SB, Jang J, Yang HK, Hwang JM, Park SK. Prevalence and risk factors of myopia in adult Korean population: Korea national health and nutrition examination survey 2013-2014 (KNHANES VI). PLoS One. 2019;14:e0211204.CrossRef
24.
go back to reference Lim DH, Han J, Chung TY, Kang S, Yim HW. Epidemiologic survey Committee of the Korean Ophthalmologic Society. The high prevalence of myopia in Korean children with influence of parental refractive errors: the 2008-2012 Korean National Health and nutrition examination survey. PLoS One. 2018;13:e0207690.CrossRef Lim DH, Han J, Chung TY, Kang S, Yim HW. Epidemiologic survey Committee of the Korean Ophthalmologic Society. The high prevalence of myopia in Korean children with influence of parental refractive errors: the 2008-2012 Korean National Health and nutrition examination survey. PLoS One. 2018;13:e0207690.CrossRef
25.
go back to reference Lee JH, Jee D, Kwon JW, Lee WK. Prevalence and risk factors for myopia in a rural Korean population. Invest Ophthalmol Vis Sci. 2013;54:5466–71.CrossRef Lee JH, Jee D, Kwon JW, Lee WK. Prevalence and risk factors for myopia in a rural Korean population. Invest Ophthalmol Vis Sci. 2013;54:5466–71.CrossRef
26.
go back to reference Ruiz-Mesa R, Abengozar-Vela A, Ruiz-Santos M. A comparative study of the visual outcomes between a new trifocal and an extended depth of focus intraocular lens. Eur J Ophthalmol. 2018;28:182–7.CrossRef Ruiz-Mesa R, Abengozar-Vela A, Ruiz-Santos M. A comparative study of the visual outcomes between a new trifocal and an extended depth of focus intraocular lens. Eur J Ophthalmol. 2018;28:182–7.CrossRef
27.
go back to reference Yu S, Kim YI, Ha SW, Lee GJ, Lee KW, Park YJ. Comparison of the visual outcomes after cataract surgery with implantation of a bifocal and trifocal diffractive intraocular lens. J Korean Ophthalmol Soc. 2016;57:405–12.CrossRef Yu S, Kim YI, Ha SW, Lee GJ, Lee KW, Park YJ. Comparison of the visual outcomes after cataract surgery with implantation of a bifocal and trifocal diffractive intraocular lens. J Korean Ophthalmol Soc. 2016;57:405–12.CrossRef
28.
go back to reference Kwon YK, Kim HK, Lee JH. Clinical outcomes of diffractive trifocal intraocular lens in both eyes: a 6-month follow-up. J Korean Ophthalmol Soc. 2015;56:1331–7.CrossRef Kwon YK, Kim HK, Lee JH. Clinical outcomes of diffractive trifocal intraocular lens in both eyes: a 6-month follow-up. J Korean Ophthalmol Soc. 2015;56:1331–7.CrossRef
29.
go back to reference de Silva SR, Evans JR, Kirthi V, Ziaei M, Leyland M. Multifocal versus monofocal intraocular lenses after cataract extraction. Cochrane Database Syst Rev. 2016;12:CD003169.PubMed de Silva SR, Evans JR, Kirthi V, Ziaei M, Leyland M. Multifocal versus monofocal intraocular lenses after cataract extraction. Cochrane Database Syst Rev. 2016;12:CD003169.PubMed
30.
go back to reference Koefoed VF, Baste V, Roumes C, Hovding G. Contrast sensitivity measured by two different test methods in healthy, young adults with normal visual acuity. Acta Ophthalmol. 2015;93:154–61.CrossRef Koefoed VF, Baste V, Roumes C, Hovding G. Contrast sensitivity measured by two different test methods in healthy, young adults with normal visual acuity. Acta Ophthalmol. 2015;93:154–61.CrossRef
Metadata
Title
Visual outcomes and safety after bilateral implantation of a trifocal presbyopia correcting intraocular lens in a Korean population: a prospective single-arm study
Authors
Tae-im Kim
Tae-Young Chung
Myoung Joon Kim
Kyounghwa Lee
Joon Young Hyon
Publication date
01-12-2020
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2020
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-020-01549-z

Other articles of this Issue 1/2020

BMC Ophthalmology 1/2020 Go to the issue