Published in: BMC Ophthalmology 1/2018

Open Access 01-12-2018 | Case report

Eye damage due to cosmetic ultrasound treatment: a case report

Authors: Yuanyuan Chen, Zhongyu Shi, Yin Shen

Published in: BMC Ophthalmology | Issue 1/2018

Login to get access

Abstract

Background

Rejuvenation of aging eyelids is one of cosmetic changes to the individual to create the appearance of youth. Tightening treatment of eyelid by ultrasonic heat could possibly develop acute eye injury, including acute increase of IOP, cataract and rarely myopia.

Case presentation

A case report of rejuvenation tightening treatment caused eye injury with 6 months’ follow-up. All examinations were performed at a university teaching hospital. A healthy 32-year-old Asian woman had pain, photophobia and blurred vision in the right eye after rejuvenation tightening eye brow treatment. Intraocular pressure (IOP) was 31 mmHg in the right eye. Tyndall phenomena were observed. Visual acuity of the right eye dropped to 20/200 (from 20/20), with best-corrected visual acuities (BCVAs) 20/20. An iris pigment detachment was found. Neuro-ophthalmic examination was relative afferent pupillary defect (RAPD) positive with pericentral scotoma in the right eye, indicating optic nerve damage. In the optical quality analysis system (OQAS) exam, the objective scatter index (OSI) was 1.0 in the right eye and 0.7 in the left. Clearing additional plus lens power was difficult for this patient, indicating accommodation spasm in the right eye.

Conclusions

Rejuvenation with intense-focused ultrasound (IFUS) could cause heat injury, leads to acute increase of IOP. Heat damage in zonular fibers could cause accommodation spasm and myopia. Eye injuries caused by IFUS has been seldom reported. We recommend that cosmetic treatment in the eye area should be highly aware of side effect.
Appendix
Available only for authorised users
Literature
1.
go back to reference Gliklich RE, White WM, Slayton MH, Barthe PG, Makin IR. Clinical pilot study of intense ultrasound therapy to deep dermal facial skin and subcutaneous tissues. Arch Facial Plast Surg. 2007;9(2):88–95.CrossRefPubMed Gliklich RE, White WM, Slayton MH, Barthe PG, Makin IR. Clinical pilot study of intense ultrasound therapy to deep dermal facial skin and subcutaneous tissues. Arch Facial Plast Surg. 2007;9(2):88–95.CrossRefPubMed
2.
go back to reference McMurray CJ, Burley CD, Elder MJ. Clear lens extraction for the treatment of persistent accommodative spasm after head trauma. J Cataract Refract Surg. 2004;30(12):2629–31.CrossRefPubMed McMurray CJ, Burley CD, Elder MJ. Clear lens extraction for the treatment of persistent accommodative spasm after head trauma. J Cataract Refract Surg. 2004;30(12):2629–31.CrossRefPubMed
3.
go back to reference Pak CS, Lee YK, Jeong JH, Kim JH, Seo JD, Heo CY. Safety and efficacy of ulthera in the rejuvenation of aging lower eyelids: a pivotal clinical trial. Aesthet Plast Surg. 2014;38(5):861–8.CrossRef Pak CS, Lee YK, Jeong JH, Kim JH, Seo JD, Heo CY. Safety and efficacy of ulthera in the rejuvenation of aging lower eyelids: a pivotal clinical trial. Aesthet Plast Surg. 2014;38(5):861–8.CrossRef
4.
go back to reference Dubinsky TJ, Cuevas C, Dighe MK, Kolokythas O, Hwang JH. High-intensity focused ultrasound: current potential and oncologic applications. Am J Roentgenol. 2008;190(1):191–9.CrossRef Dubinsky TJ, Cuevas C, Dighe MK, Kolokythas O, Hwang JH. High-intensity focused ultrasound: current potential and oncologic applications. Am J Roentgenol. 2008;190(1):191–9.CrossRef
5.
go back to reference CHUA SH, Ang P, Khoo LS, GOH CL. Nonablative infrared skin tightening in type IV to V Asian skin: a prospective clinical study. Dermatol Surg. 2007;33(2):146–51.PubMed CHUA SH, Ang P, Khoo LS, GOH CL. Nonablative infrared skin tightening in type IV to V Asian skin: a prospective clinical study. Dermatol Surg. 2007;33(2):146–51.PubMed
6.
go back to reference Suh D-H, Oh Y-J, Lee S-J, Rho J-H, Song K-Y, Kim N-I, Shin M-K. A intense-focused ultrasound tightening for the treatment of infraorbital laxity. J Cosmet Laser Ther. 2012;14(6):290–5.CrossRefPubMed Suh D-H, Oh Y-J, Lee S-J, Rho J-H, Song K-Y, Kim N-I, Shin M-K. A intense-focused ultrasound tightening for the treatment of infraorbital laxity. J Cosmet Laser Ther. 2012;14(6):290–5.CrossRefPubMed
7.
go back to reference Lee HS, Jang WS, Cha YJ, Choi YH, Tak Y, Hwang E, Kim BJ, Kim MN. Multiple pass ultrasound tightening of skin laxity of the lower face and neck. Dermatol Surg. 2012;38(1):20–7.CrossRefPubMed Lee HS, Jang WS, Cha YJ, Choi YH, Tak Y, Hwang E, Kim BJ, Kim MN. Multiple pass ultrasound tightening of skin laxity of the lower face and neck. Dermatol Surg. 2012;38(1):20–7.CrossRefPubMed
9.
go back to reference Kyung Jung S, Yang SW, Soo Kim M, Kim C. Corneal stromal damage through the eyelid after tightening using intense focused ultrasound. Can J Ophthalmol. 2015;50(4):e54–7.CrossRefPubMed Kyung Jung S, Yang SW, Soo Kim M, Kim C. Corneal stromal damage through the eyelid after tightening using intense focused ultrasound. Can J Ophthalmol. 2015;50(4):e54–7.CrossRefPubMed
Metadata
Title
Eye damage due to cosmetic ultrasound treatment: a case report
Authors
Yuanyuan Chen
Zhongyu Shi
Yin Shen
Publication date
01-12-2018
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2018
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-018-0891-2

Other articles of this Issue 1/2018

BMC Ophthalmology 1/2018 Go to the issue