Comparative evaluation of refractive outcomes after implantation of two types of intraocular lenses with different diopter intervals (0.25 diopter versus 0.50 diopter)

Authors: Minjung Kim, Youngsub Eom, Jong Suk Song, Hyo Myung Kim

Published in: BMC Ophthalmology | Issue 1/2018

Login to get access

Abstract

Background

Intraocular lenses (IOLs) with different diopter (D) intervals may have different tolerance, and may provide different accuracy of refractive outcome after cataract surgery. The aim of the study is to compare the accuracy of refractive outcome after implantation of IOLs with different D intervals after cataract surgery.

Methods

A total of 80 eyes from 40 patients who underwent phacoemulsification with implantation of a 0.50 D interval Akreos AO IOL in one eye and a 0.25 D interval Softec HD™ IOL in the other eye were enrolled. The percentages of eyes with refractive prediction error within ±0.50 D at one month after surgery were compared. To evaluate the effect of the dioptric errors of the IOL itself on refractive prediction error, the percentage of eyes with refractive prediction error within ±0.25 D of the IOL with a standard deviation (SD) of ±0.40 D was compared with that of the IOL with a SD of ±0.11 D through Monte Carlo simulations.

Results

In this clinical study, the percentage of eyes with refractive prediction error within ±0.50 D by the Haigis formula in the Softec HD™ group (85.0%) was significantly greater than that in the Akreos AO group (57.5%; P = 0.027). In Monte Carlo simulations, all percentages of eyes with refractive prediction error within ±0.25 D by the Haigis and SRK/T formulas in the Softec HD™ group were significantly greater than those in the Akreos AO group.

Conclusions

The IOL with a 0.25 D interval was more accurate than the IOL with a 0.50 D interval in predicting refractive outcome after cataract surgery.

Trial registration

Current Controlled Trials KCT0002192, Retrospectively registered (Date of registration: 6 January 2017).
Literature
1.
go back to reference Olsen T. Calculation of intraocular lens power: a review. Acta Ophthalmol. 2007;85(5):472–85.CrossRef Olsen T. Calculation of intraocular lens power: a review. Acta Ophthalmol. 2007;85(5):472–85.CrossRef
2.
go back to reference Olsen T. Improved accuracy of intraocular lens power calculation with the Zeiss IOLMaster. Acta Ophthalmol. 2007;85(1):84–7.CrossRef Olsen T. Improved accuracy of intraocular lens power calculation with the Zeiss IOLMaster. Acta Ophthalmol. 2007;85(1):84–7.CrossRef
3.
go back to reference Mamalis N. Intraocular lens power accuracy: how are we doing? J Cataract Refract Surg. 2003;29(1):1–3.CrossRefPubMed Mamalis N. Intraocular lens power accuracy: how are we doing? J Cataract Refract Surg. 2003;29(1):1–3.CrossRefPubMed
4.
go back to reference Norrby S. Sources of error in intraocular lens power calculation. J Cataract Refract Surg. 2008;34(3):368–76.CrossRefPubMed Norrby S. Sources of error in intraocular lens power calculation. J Cataract Refract Surg. 2008;34(3):368–76.CrossRefPubMed
6.
go back to reference Olsen T. Sources of error in intraocular lens power calculation. J Cataract Refract Surg. 1992;18(2):125–9.CrossRefPubMed Olsen T. Sources of error in intraocular lens power calculation. J Cataract Refract Surg. 1992;18(2):125–9.CrossRefPubMed
7.
go back to reference Norrby NS, Grossman LW, Geraghty EP, Kreiner CF, Mihori M, Patel AS, Portney V, Silberman DM. Accuracy in determining intraocular lens dioptric power assessed by interlaboratory tests. J Cataract Refract Surg. 1996;22(7):983–93.CrossRefPubMed Norrby NS, Grossman LW, Geraghty EP, Kreiner CF, Mihori M, Patel AS, Portney V, Silberman DM. Accuracy in determining intraocular lens dioptric power assessed by interlaboratory tests. J Cataract Refract Surg. 1996;22(7):983–93.CrossRefPubMed
9.
go back to reference Eom Y, Kang SY, Song JS, Kim HM. Use of corneal power-specific constants to improve the accuracy of the SRK/T formula. Ophthalmology. 2013;120(3):477–81.CrossRefPubMed Eom Y, Kang SY, Song JS, Kim HM. Use of corneal power-specific constants to improve the accuracy of the SRK/T formula. Ophthalmology. 2013;120(3):477–81.CrossRefPubMed
10.
go back to reference Eom Y, Song JS, Kim YY, Kim HM. Comparison of SRK/T and Haigis formulas for predicting corneal astigmatism correction with toric intraocular lenses. J Cataract Refract Surg. 2015;41(8):1650–7.CrossRefPubMed Eom Y, Song JS, Kim YY, Kim HM. Comparison of SRK/T and Haigis formulas for predicting corneal astigmatism correction with toric intraocular lenses. J Cataract Refract Surg. 2015;41(8):1650–7.CrossRefPubMed
11.
go back to reference Eom Y, Kang SY, Song JS, Kim YY, Kim HM. Comparison of Hoffer Q and Haigis formulae for intraocular lens power calculation according to the anterior chamber depth in short eyes. Am J Ophthalmol. 2014;157(4):818–24. e812.CrossRefPubMed Eom Y, Kang SY, Song JS, Kim YY, Kim HM. Comparison of Hoffer Q and Haigis formulae for intraocular lens power calculation according to the anterior chamber depth in short eyes. Am J Ophthalmol. 2014;157(4):818–24. e812.CrossRefPubMed
13.
go back to reference Brown DC, Gills JP, Trattler WB, Newsom TH, Weinstock RJ, Sanders DR. Prospective multicenter trial assessing effectiveness, refractive predictability and safety of a new aberration free, bi-aspheric intraocular lens. Cont Lens Anterior Eye. 2011;34(4):188–92.CrossRefPubMed Brown DC, Gills JP, Trattler WB, Newsom TH, Weinstock RJ, Sanders DR. Prospective multicenter trial assessing effectiveness, refractive predictability and safety of a new aberration free, bi-aspheric intraocular lens. Cont Lens Anterior Eye. 2011;34(4):188–92.CrossRefPubMed
14.
go back to reference Choi Y, Eom Y, Song JS, Kim HM. Influence of corneal power on intraocular lens power of the second eye in the SRK/T formula in bilateral cataract surgery. BMC Ophthalmol. 2017;17(1):261.CrossRefPubMedPubMedCentral Choi Y, Eom Y, Song JS, Kim HM. Influence of corneal power on intraocular lens power of the second eye in the SRK/T formula in bilateral cataract surgery. BMC Ophthalmol. 2017;17(1):261.CrossRefPubMedPubMedCentral
15.
go back to reference Li Y, Bao FJ. Interocular symmetry analysis of bilateral eyes. J Med Eng Technol. 2014;38(4):179–87.CrossRefPubMed Li Y, Bao FJ. Interocular symmetry analysis of bilateral eyes. J Med Eng Technol. 2014;38(4):179–87.CrossRefPubMed
16.
go back to reference Petternel V, Menapace R, Findl O, Kiss B, Wirtitsch M, Rainer G, Drexler W. Effect of optic edge design and haptic angulation on postoperative intraocular lens position change. J Cataract Refract Surg. 2004;30(1):52–7.CrossRefPubMed Petternel V, Menapace R, Findl O, Kiss B, Wirtitsch M, Rainer G, Drexler W. Effect of optic edge design and haptic angulation on postoperative intraocular lens position change. J Cataract Refract Surg. 2004;30(1):52–7.CrossRefPubMed
17.
go back to reference Eom Y, Kang SY, Song JS, Kim HM. Comparison of the actual amount of axial movement of 3 aspheric intraocular lenses using anterior segment optical coherence tomography. J Cataract Refract Surg. 2013;39(10):1528–33.CrossRefPubMed Eom Y, Kang SY, Song JS, Kim HM. Comparison of the actual amount of axial movement of 3 aspheric intraocular lenses using anterior segment optical coherence tomography. J Cataract Refract Surg. 2013;39(10):1528–33.CrossRefPubMed
18.
go back to reference Lim SJ, Kang SJ, Kim HB, Apple DJ. Ideal size of an intraocular lens for capsular bag fixation. J Cataract Refract Surg. 1998;24(3):397–402.CrossRefPubMed Lim SJ, Kang SJ, Kim HB, Apple DJ. Ideal size of an intraocular lens for capsular bag fixation. J Cataract Refract Surg. 1998;24(3):397–402.CrossRefPubMed
Metadata
Title
Comparative evaluation of refractive outcomes after implantation of two types of intraocular lenses with different diopter intervals (0.25 diopter versus 0.50 diopter)
Authors
Minjung Kim
Youngsub Eom
Jong Suk Song
Hyo Myung Kim
Publication date
01-12-2018
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2018
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-018-0840-0

Other articles of this Issue 1/2018

BMC Ophthalmology 1/2018 Go to the issue