High incidence of rainbow glare after femtosecond laser assisted-LASIK using the upgraded FS200 femtosecond laser

Authors: Yu Zhang, Yue-guo Chen

Published in: BMC Ophthalmology | Issue 1/2018

Login to get access

Abstract

Background

To compare the incidence of rainbow glare (RG) after femtosecond laser assisted-LASIK (FS-LASIK) using the upgraded FS200 femtosecond laser with different flap cut parameter settings.

Methods

A consecutive series of 129 patients (255 eyes) who underwent FS-LASIK for correcting myopia and/or astigmatism using upgraded WaveLight FS200 femtosecond laser with the original settings was included in group A. Another consecutive series of 129 patients (255 eyes) who underwent FS-LASIK using upgraded WaveLight FS200 femtosecond laser with flap cut parameter settings changed (decreased pulse energy, spot and line separation) was included in group B. The incidence and fading time of RG, confocal microscopic image and postoperative clinical results were compared between the two groups.

Results

There were no differences between the two groups in age, baseline refraction, excimer laser ablation depth, postoperative uncorrected visual acuity and refraction. The incidence rate of RG in group A (35/255, 13.73%) was significantly higher than that in group B (4/255, 1.57%) (P < 0.05). The median fading time was 3 months in group A and 1 month in group B (P > 0.05).The confocal microscopic images showed wider laser spot spacing in group A than group B. The incidence of RG was significantly correlated with age and grouping (P < 0.05).

Conclusions

The upgraded FS200 femtosecond laser with original flap cut parameter settings could increase the incidence of RG. The narrower grating size and lower pulse energy could ameliorate this side effect.
Literature
2.
go back to reference Farjo AA, Sugar A, Schallorn SC, Majmudar PA, Tanzer DJ, Trattler WB, et al. Femtosecond lasers for LASIK flap creation: a report by the American Academy of ophthalmology. Ophthalmology. 2013;120:e5-e20. Farjo AA, Sugar A, Schallorn SC, Majmudar PA, Tanzer DJ, Trattler WB, et al. Femtosecond lasers for LASIK flap creation: a report by the American Academy of ophthalmology. Ophthalmology. 2013;120:e5-e20.
3.
go back to reference Lubatschowski H. Overview of commercially available femtosecond lasers in refractive surgery. J Refract Surg. 2008;24:S102–7.PubMed Lubatschowski H. Overview of commercially available femtosecond lasers in refractive surgery. J Refract Surg. 2008;24:S102–7.PubMed
4.
go back to reference Kymionis GD, Kontadakis GA, Naoumidi I, Kankariya VP, Panagopoulou S, Manousaki A, et al. Comparative study of stromal bed of LASIK flaps created with femtosecond lasers (IntraLase FS150, WaveLight FS200) and mechanical microkeratome. Br J Ophthalmol. 2014;98:133–7.CrossRefPubMed Kymionis GD, Kontadakis GA, Naoumidi I, Kankariya VP, Panagopoulou S, Manousaki A, et al. Comparative study of stromal bed of LASIK flaps created with femtosecond lasers (IntraLase FS150, WaveLight FS200) and mechanical microkeratome. Br J Ophthalmol. 2014;98:133–7.CrossRefPubMed
5.
go back to reference Chung SH, Mazur E. Surgical applications of femtosecond laser. J Biophotonics. 2009;210:557–72.CrossRef Chung SH, Mazur E. Surgical applications of femtosecond laser. J Biophotonics. 2009;210:557–72.CrossRef
6.
go back to reference Kurtz RM, Liu X, Elner VM, Squier JA. Du D, Mourou GA. Photodisruption in the human cornea as a function of laser pulse width. J Refract Surg. 1997;13:653–8.PubMed Kurtz RM, Liu X, Elner VM, Squier JA. Du D, Mourou GA. Photodisruption in the human cornea as a function of laser pulse width. J Refract Surg. 1997;13:653–8.PubMed
7.
go back to reference Reinstein DZ, Archer TJ, Gobbe M, Johnson N. Accuracy and reproducibility of Artemis central flap thickness and visual outcomes of LASIK with the Carl Zeiss Meditec VisuMax femtosecond laser and MEL80 excimer laser platforms. J Refract Surg. 2010;26:107–19.CrossRefPubMed Reinstein DZ, Archer TJ, Gobbe M, Johnson N. Accuracy and reproducibility of Artemis central flap thickness and visual outcomes of LASIK with the Carl Zeiss Meditec VisuMax femtosecond laser and MEL80 excimer laser platforms. J Refract Surg. 2010;26:107–19.CrossRefPubMed
8.
go back to reference Zhang Y, Chen YG, Xia YJ. Comparison of corneal flap morphology using AS-OCT in LASIK with the WaveLight FS200 femtosecond laser versus a mechanical microkeratome. J Refract Surg. 2013;29:320–4.CrossRefPubMed Zhang Y, Chen YG, Xia YJ. Comparison of corneal flap morphology using AS-OCT in LASIK with the WaveLight FS200 femtosecond laser versus a mechanical microkeratome. J Refract Surg. 2013;29:320–4.CrossRefPubMed
9.
go back to reference Krueger RR, Thornton IL, Xu M, Bor Z, van den Berg TJ. Rainbow glare as an optical side effect of IntraLASIK. Ophthalmology. 2008;115:1187–95.CrossRefPubMed Krueger RR, Thornton IL, Xu M, Bor Z, van den Berg TJ. Rainbow glare as an optical side effect of IntraLASIK. Ophthalmology. 2008;115:1187–95.CrossRefPubMed
10.
go back to reference Bamba S, Rocha KM, Ramos-Esteban JC, Krueger RR. Incidence of rainbow glare after laser in situ keratomileusis flap creation with a 60 kHz femtosecond laser. J Cataract Refract Surg. 2009;35(6):1082.CrossRefPubMed Bamba S, Rocha KM, Ramos-Esteban JC, Krueger RR. Incidence of rainbow glare after laser in situ keratomileusis flap creation with a 60 kHz femtosecond laser. J Cataract Refract Surg. 2009;35(6):1082.CrossRefPubMed
11.
go back to reference Kamm A, Tünnermann A, Merker M, Kamm A, Tünnermann A, Nolte S. Optical side-effects of fs-laser treatment in refractive surgery investigated by means of a model eye. Biomedical Optics Express. 2013;4:220–9.CrossRefPubMedPubMedCentral Kamm A, Tünnermann A, Merker M, Kamm A, Tünnermann A, Nolte S. Optical side-effects of fs-laser treatment in refractive surgery investigated by means of a model eye. Biomedical Optics Express. 2013;4:220–9.CrossRefPubMedPubMedCentral
12.
go back to reference Mochen M, Wüllner C, Krause J, Klafke M, Donitzky C, Seiler T. Technical aspects of the WaveLight FS200 femtosecond laser. J Refract Surg. 2010;26:833–40.CrossRef Mochen M, Wüllner C, Krause J, Klafke M, Donitzky C, Seiler T. Technical aspects of the WaveLight FS200 femtosecond laser. J Refract Surg. 2010;26:833–40.CrossRef
13.
go back to reference Kanellopoulos AJ, Asimellis G. Refractive and keratometric stability in high myopic LASIK with high-frequency femtosecond and excimer lasers. J Refract Surg. 2013;29:832–7.CrossRefPubMed Kanellopoulos AJ, Asimellis G. Refractive and keratometric stability in high myopic LASIK with high-frequency femtosecond and excimer lasers. J Refract Surg. 2013;29:832–7.CrossRefPubMed
14.
go back to reference Kanellopoulos AJ, Asimellis G. Digital analysis of flap parameter accuracy and objective assessment of opaque bubble layer in femtosecond laser-assisted LASIK: a novel technique. Clin Ophthalmol. 2013;7:343–51.CrossRefPubMedPubMedCentral Kanellopoulos AJ, Asimellis G. Digital analysis of flap parameter accuracy and objective assessment of opaque bubble layer in femtosecond laser-assisted LASIK: a novel technique. Clin Ophthalmol. 2013;7:343–51.CrossRefPubMedPubMedCentral
15.
go back to reference Gatinel D, Saad A, Guilbert E, Rouger H. Unilateral rainbow glare after uncomplicated Femto-LASIK using the FS-200 femtosecond laser. J Refract Surg. 2013;29:498–501.CrossRefPubMed Gatinel D, Saad A, Guilbert E, Rouger H. Unilateral rainbow glare after uncomplicated Femto-LASIK using the FS-200 femtosecond laser. J Refract Surg. 2013;29:498–501.CrossRefPubMed
16.
go back to reference Gatinel D, Saad A, Guilbert E, Rouger H. Simultaneous correction of unilateral rainbow glare and residual astigmatism by undersurface flap photoablation after femtosecond laser-assisted LASIK. J Refract Surg. 2015;31:406–10.CrossRefPubMed Gatinel D, Saad A, Guilbert E, Rouger H. Simultaneous correction of unilateral rainbow glare and residual astigmatism by undersurface flap photoablation after femtosecond laser-assisted LASIK. J Refract Surg. 2015;31:406–10.CrossRefPubMed
17.
go back to reference Moshirfar M, Desautels JD, Quist TS, Skanchy DF, Williams MT, Wallace RT. Rainbow glare after laser-assisted in situ keratomileusis: a review of literature. Clin Ophthalmol. 2016;10:2245–9.CrossRefPubMedPubMedCentral Moshirfar M, Desautels JD, Quist TS, Skanchy DF, Williams MT, Wallace RT. Rainbow glare after laser-assisted in situ keratomileusis: a review of literature. Clin Ophthalmol. 2016;10:2245–9.CrossRefPubMedPubMedCentral
Metadata
Title
High incidence of rainbow glare after femtosecond laser assisted-LASIK using the upgraded FS200 femtosecond laser
Authors
Yu Zhang
Yue-guo Chen
Publication date
01-12-2018
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2018
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-018-0734-1

Other articles of this Issue 1/2018

BMC Ophthalmology 1/2018 Go to the issue