Published in: BMC Ophthalmology 1/2017

Open Access 01-12-2017 | Research article

The effect of ocular biometric factors on the accuracy of various IOL power calculation formulas

Authors: Jinho Jeong, Han Song, Jimmy K. Lee, Roy S. Chuck, Ji-Won Kwon

Published in: BMC Ophthalmology | Issue 1/2017

Login to get access

Abstract

Background

To evaluate how differences in ocular biometry affects the Hoffer Q, Holladay 1, SRK/T, and Haigis intraocular lens power calculation formulae predictions.

Methods

This study was performed on 91 eyes of 91 patients who underwent uneventful cataract surgery. Ocular biometry values were measured using the IOL Master 500, and intraocular lens (IOL) power was calculated using the Haigis, Hoffer Q, Holladay 1, and SRK/T formulas. We calculated the expected difference (ED) of each 3rd generation formula from the Haigis formula by subtracting the predicted refraction of the Haigis formula from the predicted refraction of each 3rd generation formula. Post-operative anterior chamber depth (ACD) was measured at 1 month after surgery using the IOL master. We calculated errors of each formula by subtracting predicted from manifest refraction at post-operative 1 month. Correlation analysis was performed between ocular biometry values, formula expectation values, formula errors and absolute formula errors.

Results

Multiple regression analysis revealed that preoperative ACD was the only significant factor for ED prediction in all of the 3rd generation formulas. For mean errors, axial length and post-operative 1-month change of ACD (delta ACD) correlated significantly with the errors in all 3rd generation formulas, but not with errors of the Haigis formula. Median absolute error (MedAE) of the formulas were 0.40 D for the Hoffer Q formula, 0.37 D for the Holladay formula, 0.34 D for the SRK/T formula, and 0.41 D for the Haigis formula. The MAE of the formulas were 0.50 ± 0.47 D for the Hoffer Q formula, 0.50 ± 0.50 D for the Holladay formula, 0.47 ± 0.51 D for the SRK/T formula, and 0.50 ± 0.47 D for the Haigis formula.

Conclusion

Regarding ED between the third generation and Haigis formulas, preoperative ACD demonstrated the greatest influence. Calculating mean absolute errors of the formulas, all IOL formulas showed excellent and comparable accuracy. Post-operative change (delta) of ACD correlated significantly with errors of third generation formulas according to simulated ACD.
Literature
2.
go back to reference Haigis W, Lege B, Miller N, Schneider B. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graefes Arch Clin Exp Ophthalmol. 2000;238:765–73.CrossRefPubMed Haigis W, Lege B, Miller N, Schneider B. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graefes Arch Clin Exp Ophthalmol. 2000;238:765–73.CrossRefPubMed
3.
4.
go back to reference Zuberbuhler B, Morrell AJ. Errata in printed Hoffer Q formula. J Cataract Refract Surg. 2007;33:2. author reply2–3CrossRefPubMed Zuberbuhler B, Morrell AJ. Errata in printed Hoffer Q formula. J Cataract Refract Surg. 2007;33:2. author reply2–3CrossRefPubMed
5.
go back to reference Kekunnaya R, Gupta A, Sachdeva V, Rao HL, Vaddavalli PK, Om PV. Accuracy of intraocular lens power calculation formulae in children less than 2 years. Am J Ophthalmol. 2012;154:13–9. e2CrossRefPubMed Kekunnaya R, Gupta A, Sachdeva V, Rao HL, Vaddavalli PK, Om PV. Accuracy of intraocular lens power calculation formulae in children less than 2 years. Am J Ophthalmol. 2012;154:13–9. e2CrossRefPubMed
6.
go back to reference Retzlaff JA, Sanders DR, Kraff MC. Development of the SRK/T intraocular lens implant power calculation formula. J Cataract Refract Surg. 1990;16:333–40.CrossRefPubMed Retzlaff JA, Sanders DR, Kraff MC. Development of the SRK/T intraocular lens implant power calculation formula. J Cataract Refract Surg. 1990;16:333–40.CrossRefPubMed
7.
go back to reference Hoffer KJ. Accuracy of ultrasound intraocular lens calculation. Arch Ophthalmol. 1981;99:1819–23.CrossRefPubMed Hoffer KJ. Accuracy of ultrasound intraocular lens calculation. Arch Ophthalmol. 1981;99:1819–23.CrossRefPubMed
8.
go back to reference Lee AC, Qazi MA, Pepose JS. Biometry and intraocular lens power calculation. Curr Opin Ophthalmol. 2008;19:13–7.CrossRefPubMed Lee AC, Qazi MA, Pepose JS. Biometry and intraocular lens power calculation. Curr Opin Ophthalmol. 2008;19:13–7.CrossRefPubMed
9.
go back to reference Aristodemou P, Knox Cartwright NE, Sparrow JM, Johnston RL. Formula choice: Hoffer Q, Holladay 1, or SRK/T and refractive outcomes in 8108 eyes after cataract surgery with biometry by partial coherence interferometry. J Cataract Refract Surg. 2011;37:63–71.CrossRefPubMed Aristodemou P, Knox Cartwright NE, Sparrow JM, Johnston RL. Formula choice: Hoffer Q, Holladay 1, or SRK/T and refractive outcomes in 8108 eyes after cataract surgery with biometry by partial coherence interferometry. J Cataract Refract Surg. 2011;37:63–71.CrossRefPubMed
10.
go back to reference Bang S, Edell E, Yu Q, Pratzer K, Stark W. Accuracy of intraocular lens calculations using the IOLMaster in eyes with long axial length and a comparison of various formulas. Ophthalmology. 2011;118:503–6.CrossRefPubMed Bang S, Edell E, Yu Q, Pratzer K, Stark W. Accuracy of intraocular lens calculations using the IOLMaster in eyes with long axial length and a comparison of various formulas. Ophthalmology. 2011;118:503–6.CrossRefPubMed
11.
go back to reference Kucumen RB, Yenerel NM, Gorgun E, Kulacoglu DN, Dinc UA, Alimgil ML. Anterior segment optical coherence tomography measurement of anterior chamber depth and angle changes after phacoemulsification and intraocular lens implantation. J Cataract Refract Surg. 2008;34:1694–8.CrossRefPubMed Kucumen RB, Yenerel NM, Gorgun E, Kulacoglu DN, Dinc UA, Alimgil ML. Anterior segment optical coherence tomography measurement of anterior chamber depth and angle changes after phacoemulsification and intraocular lens implantation. J Cataract Refract Surg. 2008;34:1694–8.CrossRefPubMed
12.
go back to reference Gavin EA, Hammond CJ. Intraocular lens power calculation in short eyes. Eye (Lond). 2008;22:935–8.CrossRef Gavin EA, Hammond CJ. Intraocular lens power calculation in short eyes. Eye (Lond). 2008;22:935–8.CrossRef
13.
go back to reference Hoffer KJ, Aramberri J, Haigis W, Olsen T, Savini G, Shammas HJ, et al. Protocols for studies of intraocular lens formula accuracy. Am J Ophthalmol. 2015;160:403–5. e1CrossRefPubMed Hoffer KJ, Aramberri J, Haigis W, Olsen T, Savini G, Shammas HJ, et al. Protocols for studies of intraocular lens formula accuracy. Am J Ophthalmol. 2015;160:403–5. e1CrossRefPubMed
14.
go back to reference Hashemi H, Yazdani K, Mehravaran S, Fotouhi A. Anterior chamber depth measurement with a-scan ultrasonography, Orbscan II, and IOLMaster. Optom Vis Sci. 2005;82:900–4.PubMed Hashemi H, Yazdani K, Mehravaran S, Fotouhi A. Anterior chamber depth measurement with a-scan ultrasonography, Orbscan II, and IOLMaster. Optom Vis Sci. 2005;82:900–4.PubMed
15.
go back to reference MacLaren RE, Natkunarajah M, Riaz Y, Bourne RRA, Restori M, Allan BDS. Biometry and formula accuracy with intraocular lenses used for cataract surgery in extreme hyperopia. Am J Ophthalmol. 2007;143:920–31.CrossRefPubMed MacLaren RE, Natkunarajah M, Riaz Y, Bourne RRA, Restori M, Allan BDS. Biometry and formula accuracy with intraocular lenses used for cataract surgery in extreme hyperopia. Am J Ophthalmol. 2007;143:920–31.CrossRefPubMed
16.
go back to reference Sugar A, Sadri E, Dawson DG, Musch DC. Refractive stabilization after temporal phacoemulsification with foldable acrylic intraocular lens implantation. J Cataract Refract Surg. 2001;27:1741–5.CrossRefPubMed Sugar A, Sadri E, Dawson DG, Musch DC. Refractive stabilization after temporal phacoemulsification with foldable acrylic intraocular lens implantation. J Cataract Refract Surg. 2001;27:1741–5.CrossRefPubMed
17.
go back to reference Hoffer KJ. The Hoffer Q formula: a comparison of theoretic and regression formulas. J Cataract Refract Surg. 1993;19:700–12.CrossRefPubMed Hoffer KJ. The Hoffer Q formula: a comparison of theoretic and regression formulas. J Cataract Refract Surg. 1993;19:700–12.CrossRefPubMed
18.
go back to reference Ladas JG, Siddiqui AA, Devgan U, Jun AS. A 3-D “super surface” combining modern intraocular lens formulas to generate a ‘super formula’ and maximize accuracy. JAMA Ophthalmol. 2015;133:1431–6.CrossRefPubMed Ladas JG, Siddiqui AA, Devgan U, Jun AS. A 3-D “super surface” combining modern intraocular lens formulas to generate a ‘super formula’ and maximize accuracy. JAMA Ophthalmol. 2015;133:1431–6.CrossRefPubMed
Metadata
Title
The effect of ocular biometric factors on the accuracy of various IOL power calculation formulas
Authors
Jinho Jeong
Han Song
Jimmy K. Lee
Roy S. Chuck
Ji-Won Kwon
Publication date
01-12-2017
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2017
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-017-0454-y

Other articles of this Issue 1/2017

BMC Ophthalmology 1/2017 Go to the issue