Published in: BMC Ophthalmology 1/2017

Open Access 01-12-2017 | Research article

Effects of diurnal, lighting, and angle-of-incidence variation on anterior segment optical coherence tomography (AS-OCT) angle metrics

Authors: Handan Akil, Anna Dastiridou, Kenneth Marion, Brian A. Francis, Vikas Chopra

Published in: BMC Ophthalmology | Issue 1/2017

Login to get access

Abstract

Background

First reported study to assess the effect of diurnal variation on anterior chamber angle measurements, as well as, to re-test the effects of lighting and angle-of-incidence variation on anterior chamber angle (ACA) measurements acquired by time-domain anterior segment optical coherence tomography (AS-OCT).

Methods

A total of 30 eyes from 15 healthy, normal subjects underwent anterior chamber imaging using a Visante time-domain AS-OCT according to an IRB-approved protocol. For each eye, the inferior angle was imaged twice in the morning (8 am – 10 am) and then again in the afternoon (3 pm – 5 pm), under light meter-controlled conditions with ambient room lighting ‘ON’ and lights ‘OFF’, and at 5° angle of incidence increments. The ACA metrics measured for each eye were: angle opening distance (AOD, measured 500 and 750 μm anterior from scleral spur), the trabecular-iris-space area (TISA, measured 500 and 750 μm anterior from scleral spur), and scleral spur angle. Measurements were performed by masked, certified Reading Center graders using the Visante’s Internal Measurement Tool. Differences in measurements between morning and afternoon, lighting variations, and angle of incidence were compared.

Results

Mean age of the participants was 31.2 years (range 23–58). Anterior chamber angle metrics did not differ significantly from morning to afternoon imaging, or when the angle of incidence was offset by 5° in either direction away from the inferior angle 6 o’clock position. (p-value 0.13-0.93). Angle metrics at the inferior corneal limbus, 6 o’clock position (IC270), with room lighting ‘OFF’, showed a significant decrease (p < 0.05) compared to room lighting ‘ON’.

Conclusions

There does not appear to be significant diurnal variation in AS-OCT parameters in normal individuals, but lighting conditions need to be strictly controlled since variation in lighting led to significant variability in AS-OCT parameters. No changes in ACA parameters were noted by varying the angle-of-incidence, which gives confidence in being able to perform longitudinal studies in approximately the same area (plus/minus 5° of original scan location).
Appendix
Available only for authorised users
Literature
2.
go back to reference Tian J, Marziliano P, Baskaran M, Wong HT, Aung T. Automatic anterior chamber angle assessment for HD-OCT images. IEEE Trans Biomed Eng. 2011;58(11):3242–9.CrossRefPubMed Tian J, Marziliano P, Baskaran M, Wong HT, Aung T. Automatic anterior chamber angle assessment for HD-OCT images. IEEE Trans Biomed Eng. 2011;58(11):3242–9.CrossRefPubMed
3.
go back to reference Liu S, Yu M, Ye C, Lam DS, Leung CK. Anterior chamber angle imaging with swept-source optical coherence tomography: an investigation on variability of angle measurement. Invest Ophthalmol Vis Sci. 2011;52(12):8598–603.CrossRefPubMed Liu S, Yu M, Ye C, Lam DS, Leung CK. Anterior chamber angle imaging with swept-source optical coherence tomography: an investigation on variability of angle measurement. Invest Ophthalmol Vis Sci. 2011;52(12):8598–603.CrossRefPubMed
4.
go back to reference Nolan WP, See JL, Chew PT, Friedman DS, Smith SD, Radhakrishan S, et al. Detection of primary angle closure using anterior segment optical coherence tomography in Asian eyes. Ophthalmology. 2007;114(1):33–9.CrossRefPubMed Nolan WP, See JL, Chew PT, Friedman DS, Smith SD, Radhakrishan S, et al. Detection of primary angle closure using anterior segment optical coherence tomography in Asian eyes. Ophthalmology. 2007;114(1):33–9.CrossRefPubMed
5.
go back to reference Leung CK, Li H, Weinreb RN, Liu J, Cheung CY, Lai RY, et al. Anterior chamber angle measurement with anterior segment optical coherence tomography: a comparison between slit lamp OCT and Visante OCT. Invest Ophthalmol Vis Sci. 2008;49(8):3469–74.CrossRefPubMed Leung CK, Li H, Weinreb RN, Liu J, Cheung CY, Lai RY, et al. Anterior chamber angle measurement with anterior segment optical coherence tomography: a comparison between slit lamp OCT and Visante OCT. Invest Ophthalmol Vis Sci. 2008;49(8):3469–74.CrossRefPubMed
6.
go back to reference Dinc UA, Oncel B, Gorgun E, Yalvac IS. Assessment of anterior chamber angle using Visante OCT, slit-lamp OCT and Pentacam. Eur J Ophthalmol. 2010;20(3):531–7.PubMed Dinc UA, Oncel B, Gorgun E, Yalvac IS. Assessment of anterior chamber angle using Visante OCT, slit-lamp OCT and Pentacam. Eur J Ophthalmol. 2010;20(3):531–7.PubMed
7.
go back to reference Wylegała E, Teper S, Nowińska AK, Milka M, Dobrowolski D. Anterior segment imaging: Fourier-domain optical coherence tomography versus time-domain optical coherence tomography. J Cataract Refract Surg. 2009;35(8):1410–4.CrossRefPubMed Wylegała E, Teper S, Nowińska AK, Milka M, Dobrowolski D. Anterior segment imaging: Fourier-domain optical coherence tomography versus time-domain optical coherence tomography. J Cataract Refract Surg. 2009;35(8):1410–4.CrossRefPubMed
8.
go back to reference Radhakrishnan S, Rollins AM, Roth JE, Yazdanfar S, Westphal V, Bardenstein DS, et al. Real-time optical coherence tomography of the anterior segment at 1310 nm. Arch Ophthalmol. 2001;119(8):1179–85.CrossRefPubMed Radhakrishnan S, Rollins AM, Roth JE, Yazdanfar S, Westphal V, Bardenstein DS, et al. Real-time optical coherence tomography of the anterior segment at 1310 nm. Arch Ophthalmol. 2001;119(8):1179–85.CrossRefPubMed
9.
go back to reference Radhakrishnan S, Goldsmith J, Huang D, Yazdanfar S, Westphal V, Bardenstein DS, et al. Comparison of optical coherence tomography and ultrasound biomicroscopy for detection of narrow anterior chamber angles. Arch Ophthalmol. 2005;123:1053–9.CrossRefPubMed Radhakrishnan S, Goldsmith J, Huang D, Yazdanfar S, Westphal V, Bardenstein DS, et al. Comparison of optical coherence tomography and ultrasound biomicroscopy for detection of narrow anterior chamber angles. Arch Ophthalmol. 2005;123:1053–9.CrossRefPubMed
10.
go back to reference Leung C, Weinreb R. Anterior chamber angle imaging with optical coherence tomography. Eye (Lond). 2011;25:261–7.CrossRef Leung C, Weinreb R. Anterior chamber angle imaging with optical coherence tomography. Eye (Lond). 2011;25:261–7.CrossRef
11.
go back to reference Akil H, Marion K, Dastiridou A, Jenkins D, Kramer B, Francis BA, et al. Identification of anterior chamber angle parameters with a portable SD-OCT device compared to a non-portable SD-OCT. Int Ophthalmol. 2017;37(1):31–7.CrossRefPubMed Akil H, Marion K, Dastiridou A, Jenkins D, Kramer B, Francis BA, et al. Identification of anterior chamber angle parameters with a portable SD-OCT device compared to a non-portable SD-OCT. Int Ophthalmol. 2017;37(1):31–7.CrossRefPubMed
12.
go back to reference Lee JR, Jeoung JW, Choi J, Jin Young C, Park KH, Kim YD. Structure-function relationships in normal and glaucomatous eyes determined by time- and spectral-domain optical coherence tomography. Invest Ophthalmol Vis Sci. 2010;51(12):6424–30.CrossRefPubMed Lee JR, Jeoung JW, Choi J, Jin Young C, Park KH, Kim YD. Structure-function relationships in normal and glaucomatous eyes determined by time- and spectral-domain optical coherence tomography. Invest Ophthalmol Vis Sci. 2010;51(12):6424–30.CrossRefPubMed
13.
go back to reference Xu L, Cao WF, Wang YX, Chen CX, Jonas JB. Anterior chamber depth and chamber angle and their associations with ocular and general parameters: the Beijing Eye study. Am J Ophthalmol. 2008;145(5):929–36.CrossRefPubMed Xu L, Cao WF, Wang YX, Chen CX, Jonas JB. Anterior chamber depth and chamber angle and their associations with ocular and general parameters: the Beijing Eye study. Am J Ophthalmol. 2008;145(5):929–36.CrossRefPubMed
14.
go back to reference Dastiridou AI, Pan X, Zhang Z, Marion KM, Francis BA, Chopra V. Comparison of physiologic versus pharmacologic mydriasis on anterior chamber angle measurements using spectral domain optical coherence tomography. J Ophthalmol. 2015;2015:845643.CrossRefPubMedPubMedCentral Dastiridou AI, Pan X, Zhang Z, Marion KM, Francis BA, Chopra V. Comparison of physiologic versus pharmacologic mydriasis on anterior chamber angle measurements using spectral domain optical coherence tomography. J Ophthalmol. 2015;2015:845643.CrossRefPubMedPubMedCentral
15.
go back to reference Wang B, Sakata LM, Friedman DS, Chan YH, He M, Lavanya R, et al. Quantitative iris parameters and association with narrow angles. Ophthalmology. 2010;117(1):11–7.CrossRefPubMed Wang B, Sakata LM, Friedman DS, Chan YH, He M, Lavanya R, et al. Quantitative iris parameters and association with narrow angles. Ophthalmology. 2010;117(1):11–7.CrossRefPubMed
16.
go back to reference Cheung CY, Liu S, Weinreb RN, Liu J, Li H, Leung DY, et al. Dynamic analysis of iris configuration with anterior segment optical coherence tomography. Invest Ophthalmol Vis Sci. 2010;51(8):4040–6.CrossRefPubMed Cheung CY, Liu S, Weinreb RN, Liu J, Li H, Leung DY, et al. Dynamic analysis of iris configuration with anterior segment optical coherence tomography. Invest Ophthalmol Vis Sci. 2010;51(8):4040–6.CrossRefPubMed
17.
go back to reference Read SA, Collins MJ, Iskander DR. Diurnal variation of axial length, intraocular pressure, and anterior eye biometrics. Invest Ophthalmol Vis Sci. 2008;49(7):2911–8.CrossRefPubMed Read SA, Collins MJ, Iskander DR. Diurnal variation of axial length, intraocular pressure, and anterior eye biometrics. Invest Ophthalmol Vis Sci. 2008;49(7):2911–8.CrossRefPubMed
18.
go back to reference Clark CV, Mapstone R. Diurnal variation in onset of acute closed angle glaucoma. Br Med J. 1986;292(6528):1106.CrossRef Clark CV, Mapstone R. Diurnal variation in onset of acute closed angle glaucoma. Br Med J. 1986;292(6528):1106.CrossRef
19.
go back to reference Khor WB, Sakata LM, Friedman DS, Narayanaswamy A, Lavanya R, Perera SA, et al. Evaluation of scanning protocols for imaging the anterior chamber angle with anterior segment-optical coherence tomography. J Glaucoma. 2010;19(6):365–8.CrossRefPubMed Khor WB, Sakata LM, Friedman DS, Narayanaswamy A, Lavanya R, Perera SA, et al. Evaluation of scanning protocols for imaging the anterior chamber angle with anterior segment-optical coherence tomography. J Glaucoma. 2010;19(6):365–8.CrossRefPubMed
20.
go back to reference Sakata LM, Lavanya R, Friedman DS, Aung HT, Gao H, Kumar RS, et al. Comparison of gonioscopy and anterior segment ocular coherence tomography in detecting angle closure in different quadrants of the anterior chamber angle. Ophthalmology. 2008;115(5):769–74.CrossRefPubMed Sakata LM, Lavanya R, Friedman DS, Aung HT, Gao H, Kumar RS, et al. Comparison of gonioscopy and anterior segment ocular coherence tomography in detecting angle closure in different quadrants of the anterior chamber angle. Ophthalmology. 2008;115(5):769–74.CrossRefPubMed
21.
go back to reference Pan X, Marion KM, Maram J, Zhang ZY, Francis BA, Nittala MG, et al. Reproducibility of anterior segment angle metric measurements derived from cirrus spectral domain optical coherence tomography. J Glaucoma. 2015;24(5):47–51.CrossRef Pan X, Marion KM, Maram J, Zhang ZY, Francis BA, Nittala MG, et al. Reproducibility of anterior segment angle metric measurements derived from cirrus spectral domain optical coherence tomography. J Glaucoma. 2015;24(5):47–51.CrossRef
22.
go back to reference Leung CK, Cheung CY, Li H, Dorairaj S, Yiu CK, Wong AL, et al. Dynamic analysis of dark–light changes of the anterior chamber angle with anterior segment OCT. Invest Ophthalmol Vis Sci. 2007;48(9):4116–22.CrossRefPubMed Leung CK, Cheung CY, Li H, Dorairaj S, Yiu CK, Wong AL, et al. Dynamic analysis of dark–light changes of the anterior chamber angle with anterior segment OCT. Invest Ophthalmol Vis Sci. 2007;48(9):4116–22.CrossRefPubMed
23.
go back to reference Quigley HA, Silver DM, Friedman DS, He M, Plyler RJ, Eberhart CG, et al. Iris cross-sectional area decreases with pupil dilation and its dynamic behavior is a risk factor in angle closure. J Glaucoma. 2009;18(9):173–9.CrossRefPubMed Quigley HA, Silver DM, Friedman DS, He M, Plyler RJ, Eberhart CG, et al. Iris cross-sectional area decreases with pupil dilation and its dynamic behavior is a risk factor in angle closure. J Glaucoma. 2009;18(9):173–9.CrossRefPubMed
24.
go back to reference Aptel F, Chiquet C, Beccat S, Denis P. Biometric evaluation of anterior chamber changes after physiologic pupil dilation using pentacam and anterior segment optical coherence tomography. Invest Ophthalmol Vis Sci. 2012;53(7):4005–10.CrossRefPubMed Aptel F, Chiquet C, Beccat S, Denis P. Biometric evaluation of anterior chamber changes after physiologic pupil dilation using pentacam and anterior segment optical coherence tomography. Invest Ophthalmol Vis Sci. 2012;53(7):4005–10.CrossRefPubMed
25.
go back to reference Stone RA, Quinn GE, Francis EL, Ying GS, Flitcroft DI, Parekh P, et al. Diurnal axial length fluctuations in human eyes. Invest Ophthalmol Vis Sci. 2004;45(1):63–70.CrossRefPubMed Stone RA, Quinn GE, Francis EL, Ying GS, Flitcroft DI, Parekh P, et al. Diurnal axial length fluctuations in human eyes. Invest Ophthalmol Vis Sci. 2004;45(1):63–70.CrossRefPubMed
26.
go back to reference Wilson LB, Quinn GE, Ying G, Francis EL, Schmid G, Lam A, et al. The relation of axial length and intraocular pressure fluctuations in human eyes. Invest Ophthalmol Vis Sci. 2006;47(5):1778–84.CrossRefPubMed Wilson LB, Quinn GE, Ying G, Francis EL, Schmid G, Lam A, et al. The relation of axial length and intraocular pressure fluctuations in human eyes. Invest Ophthalmol Vis Sci. 2006;47(5):1778–84.CrossRefPubMed
27.
go back to reference Pan X, Maram J, Marion K, Dastiridou A, Zhang ZY, Dominguez-Velasco MF, et al. Effect of angle of incidence on anterior chamber angle metrics from optical coherence tomography. J Glaucoma. 2016;25(1):19–23.CrossRef Pan X, Maram J, Marion K, Dastiridou A, Zhang ZY, Dominguez-Velasco MF, et al. Effect of angle of incidence on anterior chamber angle metrics from optical coherence tomography. J Glaucoma. 2016;25(1):19–23.CrossRef
Metadata
Title
Effects of diurnal, lighting, and angle-of-incidence variation on anterior segment optical coherence tomography (AS-OCT) angle metrics
Authors
Handan Akil
Anna Dastiridou
Kenneth Marion
Brian A. Francis
Vikas Chopra
Publication date
01-12-2017
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2017
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-017-0425-3

Other articles of this Issue 1/2017

BMC Ophthalmology 1/2017 Go to the issue