Published in: BMC Ophthalmology 1/2016

Open Access 01-12-2016 | Research article

Accuracy of PlusOptix A09 distance refraction in pediatric myopia and hyperopia

Authors: Arnaud Payerols, Claudie Eliaou, Véronique Trezeguet, Max Villain, Vincent Daien

Published in: BMC Ophthalmology | Issue 1/2016

Login to get access

Abstract

Background

The PlusOptix photoscreeners (PlusOptix GmbH, Nuremberg, Germany) is used in many vision screening programs. The purpose of the present study was to further explore the accuracy of the PlusOptix A09 photoscreener in children with ametropia (myopia or hyperopia).

Methods

A total of 70 eyes (35 children) were prospectively included. Before administration with the cycloplegia treatment 1 % cyclopentolate hydrochloride, children underwent refraction measurement with the PlusOptix A09. A refraction was then performed after cycloplegia with either Retinomax hand-held or Nidek autorefractor before and after 3 years old, respectively.

Results

The median (interquartile range) age was 58 (18 to 86) months. The mean (SD) spherical equivalent differed between PlusOptix A09 and cycloplegic autorefraction (+0.54 [1.82] D vs +1.06 [2.04] D, p = 0.04). PlusOptix A09 refraction was positively correlated with cycloplegic autorefraction (r = 0.81, p < 0.001) with higher coefficient in myopic than in hyperopic children (r = 0.91, p = 0.0002 and r = 0.52, p = 0.01, respectively). The mean (SD) difference between PlusOptix A09 and cycloplegic autorefraction was higher with hyperopia than myopia (0.73 [1.34] vs 0.05 [0.66], p = 0.01). The proportion of children with < 1-D difference between cycloplegic and PlusOptix A09 refraction was 68.8 %, higher with myopia than hyperopia (90 % vs 54.5 %, p = 0.01).

Conclusion

The spherical equivalent value with non-cycloplegic PlusOptix A09 refraction is closer to that with cycloplegic autorefraction than non-cycloplegic autorefraction. The PlusOptix A09 photoscreener underestimated the hyperopia of 0.73 D and slightly overestimated myopia of 0.05 D. The PlusOptix A09 could be used for screening with higher accuracy in myopic than hyperopic children.
Literature
1.
go back to reference Cordonnier M, Kallay O. Non-cycloplegic screening for refractive errors in children with the hand-held autorefractor Retinomax: final results and comparison with non-cycloplegic photoscreening. Strabismus. 2001;9(2):59–70.CrossRefPubMed Cordonnier M, Kallay O. Non-cycloplegic screening for refractive errors in children with the hand-held autorefractor Retinomax: final results and comparison with non-cycloplegic photoscreening. Strabismus. 2001;9(2):59–70.CrossRefPubMed
2.
go back to reference Cordonnier M, Dramaix M. Screening for refractive errors in children: accuracy of the hand held refractor Retinomax to screen for astigmatism. Br J Ophthalmol. 1999;83(2):157–61.CrossRefPubMedPubMedCentral Cordonnier M, Dramaix M. Screening for refractive errors in children: accuracy of the hand held refractor Retinomax to screen for astigmatism. Br J Ophthalmol. 1999;83(2):157–61.CrossRefPubMedPubMedCentral
4.
go back to reference Moghaddam AAS, Kargozar A, Zarei-Ghanavati M, Najjaran M, Nozari V, Shakeri MT. Screening for amblyopia risk factors in pre-verbal children using the Plusoptix photoscreener: a cross-sectional population-based study. Br J Ophthalmol. 2012;96(1):83–6. doi:10.1136/bjo.2010.190405.CrossRefPubMed Moghaddam AAS, Kargozar A, Zarei-Ghanavati M, Najjaran M, Nozari V, Shakeri MT. Screening for amblyopia risk factors in pre-verbal children using the Plusoptix photoscreener: a cross-sectional population-based study. Br J Ophthalmol. 2012;96(1):83–6. doi:10.​1136/​bjo.​2010.​190405.CrossRefPubMed
5.
go back to reference Paff T, Oudesluys-Murphy AM, Wolterbeek R, et al. Screening for refractive errors in children: the plusoptiX S08 and the Retinomax K-plus2 performed by a lay screener compared to cycloplegic retinoscopy. J AAPOS Off Publ Am Assoc Pediatr Ophthalmol Strabismus. 2010;14(6):478–83. doi:10.1016/j.jaapos.2010.09.015.CrossRef Paff T, Oudesluys-Murphy AM, Wolterbeek R, et al. Screening for refractive errors in children: the plusoptiX S08 and the Retinomax K-plus2 performed by a lay screener compared to cycloplegic retinoscopy. J AAPOS Off Publ Am Assoc Pediatr Ophthalmol Strabismus. 2010;14(6):478–83. doi:10.​1016/​j.​jaapos.​2010.​09.​015.CrossRef
7.
go back to reference Ayse YK, Onder U, Suheyla K. Accuracy of Plusoptix S04 in children and teens. Can J Ophthalmol J Can Ophtalmol. 2011;46(2):153–7.CrossRef Ayse YK, Onder U, Suheyla K. Accuracy of Plusoptix S04 in children and teens. Can J Ophthalmol J Can Ophtalmol. 2011;46(2):153–7.CrossRef
8.
11.
go back to reference Matta NS, Singman EL, Silbert DI. Performance of the plusoptiX S04 photoscreener for the detection of amblyopia risk factors in children aged 3 to 5. J AAPOS Off Publ Am Assoc Pediatr Ophthalmol Strabismus. 2010;14(2):147–9. doi:10.1016/j.jaapos.2010.01.006.CrossRef Matta NS, Singman EL, Silbert DI. Performance of the plusoptiX S04 photoscreener for the detection of amblyopia risk factors in children aged 3 to 5. J AAPOS Off Publ Am Assoc Pediatr Ophthalmol Strabismus. 2010;14(2):147–9. doi:10.​1016/​j.​jaapos.​2010.​01.​006.CrossRef
13.
go back to reference Schimitzek T, Lagrèze WA. Accuracy of a new photo-refractometer in young and adult patients. Graefes Arch Clin Exp Ophthalmol Albrecht Von Graefes Arch Für Klin Exp Ophthalmol. 2005;243(7):637–45. doi:10.1007/s00417-004-1056-y.CrossRef Schimitzek T, Lagrèze WA. Accuracy of a new photo-refractometer in young and adult patients. Graefes Arch Clin Exp Ophthalmol Albrecht Von Graefes Arch Für Klin Exp Ophthalmol. 2005;243(7):637–45. doi:10.​1007/​s00417-004-1056-y.CrossRef
14.
17.
go back to reference Rajavi Z, Parsafar H, Ramezani A, Yaseri M. Is noncycloplegic photorefraction applicable for screening refractive amblyopia risk factors? J Ophthalmic Vis Res. 2012;7(1):3–9.PubMedPubMedCentral Rajavi Z, Parsafar H, Ramezani A, Yaseri M. Is noncycloplegic photorefraction applicable for screening refractive amblyopia risk factors? J Ophthalmic Vis Res. 2012;7(1):3–9.PubMedPubMedCentral
Metadata
Title
Accuracy of PlusOptix A09 distance refraction in pediatric myopia and hyperopia
Authors
Arnaud Payerols
Claudie Eliaou
Véronique Trezeguet
Max Villain
Vincent Daien
Publication date
01-12-2016
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2016
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-016-0247-8

Other articles of this Issue 1/2016

BMC Ophthalmology 1/2016 Go to the issue