Published in: BMC Pregnancy and Childbirth 1/2016

Open Access 01-12-2016 | Research article

Fetal heart rate monitoring of short term variation (STV): a methodological observational study

Authors: Stina Wretler, Malin Holzmann, Sophie Graner, Pelle Lindqvist, Susanne Falck, Lennart Nordström

Published in: BMC Pregnancy and Childbirth | Issue 1/2016

Login to get access

Abstract

Background

Cardiotocography (CTG) has high sensitivity, but less specificity in detection of fetal hypoxia. There is need for adjunctive methods easy to apply during labor. Low fetal heart rate short term variation (STV) is predictive for hypoxia during the antenatal period. The objectives of our study were to methodologically evaluate monitoring of STV during labor and to compare two different monitors (Sonicaid™ and EDAN™) for antenatal use.

Methods

A prospective observational study at the obstetric department, Karolinska University hospital, Stockholm (between September 2011 and April 2015). In 100 women of ≥ 36 weeks gestation, STV values were calculated during active labor. In a subset of 20 women we compared STV values between internal and external signal acquisition. Additionally we compared antenatal monitoring with two different monitors in another 20 women.

Results

Median STV in 100 fetuses monitored with scalp electrode during labor (EDAN™) was 7.1 msec (range 1.3–25.9) with no difference between early (3–6 cm) and late (7–10 cm) labor (7.1 vs 6.8 msec; p = 0.80). STV calculated from scalp electrode signals were positively correlated with delta-STV (STV internal –external) (R = 0.70; p < 0.01). No significant differences were found between Sonicaid™ and EDAN™ in antenatal external monitoring of STV (median difference 0.9 msec, Spearman Rank Correlation Sonicaid vs delta-STV; R = 0.35; p = 0.14).

Conclusions

Median intrapartum STV was 7.1 msec. Significant differences were found between internal and external signal acquisition, a finding that suggests further intrapartum studies to be analysed separately depending upon type of signal acquisition. Antenatal external monitoring with Sonicaid™ and EDAN™ indicates that the devices perform equally well in the identification of acidemic fetuses. Further studies are needed to assess the clinical value of intrapartum STV.
Literature
1.
go back to reference Holzmann M, Nordstrom L. Follow-up national survey (Sweden) of routines for intrapartum fetal surveillance. Acta Obstet Gynecol Scand. 2010;89(5):712–4.CrossRefPubMed Holzmann M, Nordstrom L. Follow-up national survey (Sweden) of routines for intrapartum fetal surveillance. Acta Obstet Gynecol Scand. 2010;89(5):712–4.CrossRefPubMed
2.
go back to reference Alfirevic Z, Devane D, Gyte GM. Continuous cardiotocography (CTG) as a form of electronic fetal monitoring (EFM) for fetal assessment during labour. Cochrane Database Syst Rev. 2013;5:CD006066.PubMed Alfirevic Z, Devane D, Gyte GM. Continuous cardiotocography (CTG) as a form of electronic fetal monitoring (EFM) for fetal assessment during labour. Cochrane Database Syst Rev. 2013;5:CD006066.PubMed
3.
go back to reference Vintzileos AM, Nochimson DJ, Guzman ER, Knuppel RA, Lake M, Schifrin BS. Intrapartum electronic fetal heart rate monitoring versus intermittent auscultation: a meta-analysis. Obstet Gynecol. 1995;85(1):149–55.CrossRefPubMed Vintzileos AM, Nochimson DJ, Guzman ER, Knuppel RA, Lake M, Schifrin BS. Intrapartum electronic fetal heart rate monitoring versus intermittent auscultation: a meta-analysis. Obstet Gynecol. 1995;85(1):149–55.CrossRefPubMed
4.
go back to reference Nelson KB, Dambrosia JM, Ting TY, Grether JK. Uncertain value of electronic fetal monitoring in predicting cerebral palsy. N Engl J Med. 1996;334(10):613–8.CrossRefPubMed Nelson KB, Dambrosia JM, Ting TY, Grether JK. Uncertain value of electronic fetal monitoring in predicting cerebral palsy. N Engl J Med. 1996;334(10):613–8.CrossRefPubMed
5.
go back to reference Di Tommaso M, Seravalli V, Cordisco A, Consorti G, Mecacci F, Rizzello F. Comparison of five classification systems for interpreting electronic fetal monitoring in predicting neonatal status at birth. J Matern Fetal Neonatal Med. 2013;26(5):487–90.CrossRefPubMed Di Tommaso M, Seravalli V, Cordisco A, Consorti G, Mecacci F, Rizzello F. Comparison of five classification systems for interpreting electronic fetal monitoring in predicting neonatal status at birth. J Matern Fetal Neonatal Med. 2013;26(5):487–90.CrossRefPubMed
6.
go back to reference Schiermeier S, Pildner von Steinburg S, Thieme A, Reinhard J, Daumer M, Scholz M, Hatzmann W, Schneider KT. Sensitivity and specificity of intrapartum computerised FIGO criteria for cardiotocography and fetal scalp pH during labour: multicentre, observational study. BJOG. 2008;115(12):1557–63.CrossRefPubMed Schiermeier S, Pildner von Steinburg S, Thieme A, Reinhard J, Daumer M, Scholz M, Hatzmann W, Schneider KT. Sensitivity and specificity of intrapartum computerised FIGO criteria for cardiotocography and fetal scalp pH during labour: multicentre, observational study. BJOG. 2008;115(12):1557–63.CrossRefPubMed
7.
go back to reference Palomaki O, Luukkaala T, Luoto R, Tuimala R. Intrapartum cardiotocography -- the dilemma of interpretational variation. J Perinat Med. 2006;34(4):298–302.PubMed Palomaki O, Luukkaala T, Luoto R, Tuimala R. Intrapartum cardiotocography -- the dilemma of interpretational variation. J Perinat Med. 2006;34(4):298–302.PubMed
8.
go back to reference Beaulieu MD, Fabia J, Leduc B, Brisson J, Bastide A, Blouin D, Gauthier RJ, Lalonde A. The reproducibility of intrapartum cardiotocogram assessments. Can Med Assoc J. 1982;127(3):214–6.PubMedPubMedCentral Beaulieu MD, Fabia J, Leduc B, Brisson J, Bastide A, Blouin D, Gauthier RJ, Lalonde A. The reproducibility of intrapartum cardiotocogram assessments. Can Med Assoc J. 1982;127(3):214–6.PubMedPubMedCentral
9.
go back to reference Schiermeier S, Westhof G, Leven A, Hatzmann H, Reinhard J. Intra- and interobserver variability of intrapartum cardiotocography: a multicenter study comparing the FIGO classification with computer analysis software. Gynecol Obstet Invest. 2011;72(3):169–73.CrossRefPubMed Schiermeier S, Westhof G, Leven A, Hatzmann H, Reinhard J. Intra- and interobserver variability of intrapartum cardiotocography: a multicenter study comparing the FIGO classification with computer analysis software. Gynecol Obstet Invest. 2011;72(3):169–73.CrossRefPubMed
10.
go back to reference Ojala K, Makikallio K, Haapsamo M, Ijas H, Tekay A. Interobserver agreement in the assessment of intrapartum automated fetal electrocardiography in singleton pregnancies. Acta Obstet Gynecol Scand. 2008;87(5):536–40.CrossRefPubMed Ojala K, Makikallio K, Haapsamo M, Ijas H, Tekay A. Interobserver agreement in the assessment of intrapartum automated fetal electrocardiography in singleton pregnancies. Acta Obstet Gynecol Scand. 2008;87(5):536–40.CrossRefPubMed
11.
go back to reference Devane D, Lalor J. Midwives’ visual interpretation of intrapartum cardiotocographs: intra- and inter-observer agreement. J Adv Nurs. 2005;52(2):133–41.CrossRefPubMed Devane D, Lalor J. Midwives’ visual interpretation of intrapartum cardiotocographs: intra- and inter-observer agreement. J Adv Nurs. 2005;52(2):133–41.CrossRefPubMed
12.
go back to reference Bernardes J, Costa-Pereira A, Ayres-de-Campos D, van Geijn HP, Pereira-Leite L. Evaluation of interobserver agreement of cardiotocograms. Int J Gynaecol Obstet. 1997;57(1):33–7.CrossRefPubMed Bernardes J, Costa-Pereira A, Ayres-de-Campos D, van Geijn HP, Pereira-Leite L. Evaluation of interobserver agreement of cardiotocograms. Int J Gynaecol Obstet. 1997;57(1):33–7.CrossRefPubMed
13.
go back to reference Keith RD, Beckley S, Garibaldi JM, Westgate JA, Ifeachor EC, Greene KR. A multicentre comparative study of 17 experts and an intelligent computer system for managing labour using the cardiotocogram. BJOG. 1995;102(9):688–700.CrossRef Keith RD, Beckley S, Garibaldi JM, Westgate JA, Ifeachor EC, Greene KR. A multicentre comparative study of 17 experts and an intelligent computer system for managing labour using the cardiotocogram. BJOG. 1995;102(9):688–700.CrossRef
14.
go back to reference Dawes G, Meir YJ, Mandruzzato GP. Computerized evaluation of fetal heart-rate patterns. J Perinat Med. 1994;22(6):491–9.CrossRefPubMed Dawes G, Meir YJ, Mandruzzato GP. Computerized evaluation of fetal heart-rate patterns. J Perinat Med. 1994;22(6):491–9.CrossRefPubMed
15.
go back to reference Dawes GS, Moulden M, Redman CW. System 8000: computerized antenatal FHR analysis. J Perinat Med. 1991;19(1–2):47–51.CrossRefPubMed Dawes GS, Moulden M, Redman CW. System 8000: computerized antenatal FHR analysis. J Perinat Med. 1991;19(1–2):47–51.CrossRefPubMed
16.
go back to reference Farmakides G, Weiner Z. Computerized analysis of the fetal heart rate. Clin Obstet Gynecol. 1995;38(1):112–20.CrossRefPubMed Farmakides G, Weiner Z. Computerized analysis of the fetal heart rate. Clin Obstet Gynecol. 1995;38(1):112–20.CrossRefPubMed
17.
go back to reference Pardey J, Moulden M, Redman CW. A computer system for the numerical analysis of nonstress tests. Am J Obstet Gynecol. 2002;186(5):1095–103.CrossRefPubMed Pardey J, Moulden M, Redman CW. A computer system for the numerical analysis of nonstress tests. Am J Obstet Gynecol. 2002;186(5):1095–103.CrossRefPubMed
18.
go back to reference Street P, Dawes GS, Moulden M, Redman CW. Short-term variation in abnormal antenatal fetal heart rate records. Am J Obstet Gynecol. 1991;165(3):515–23.CrossRefPubMed Street P, Dawes GS, Moulden M, Redman CW. Short-term variation in abnormal antenatal fetal heart rate records. Am J Obstet Gynecol. 1991;165(3):515–23.CrossRefPubMed
19.
go back to reference Galazios G, Tripsianis G, Tsikouras P, Koutlaki N, Liberis V. Fetal distress evaluation using and analyzing the variables of antepartum computerized cardiotocography. Arch Gynecol Obstet. 2010;281(2):229–33.CrossRefPubMed Galazios G, Tripsianis G, Tsikouras P, Koutlaki N, Liberis V. Fetal distress evaluation using and analyzing the variables of antepartum computerized cardiotocography. Arch Gynecol Obstet. 2010;281(2):229–33.CrossRefPubMed
20.
go back to reference Serra V, Bellver J, Moulden M, Redman CW. Computerized analysis of normal fetal heart rate pattern throughout gestation. Ultrasound Obstet Gynecol. 2009;34(1):74–9.CrossRefPubMed Serra V, Bellver J, Moulden M, Redman CW. Computerized analysis of normal fetal heart rate pattern throughout gestation. Ultrasound Obstet Gynecol. 2009;34(1):74–9.CrossRefPubMed
21.
go back to reference Anceschi MM, Piazze JJ, Ruozi-Berretta A, Cosmi E, Cerekja A, Maranghi L, Cosmi EV. Validity of short term variation (STV) in detection of fetal acidemia. J Perinat Med. 2003;31(3):231–6.PubMed Anceschi MM, Piazze JJ, Ruozi-Berretta A, Cosmi E, Cerekja A, Maranghi L, Cosmi EV. Validity of short term variation (STV) in detection of fetal acidemia. J Perinat Med. 2003;31(3):231–6.PubMed
22.
go back to reference Dawes GS, Moulden M, Redman CW. Short-term fetal heart rate variation, decelerations, and umbilical flow velocity waveforms before labor. Obstet Gynecol. 1992;80(4):673–8.PubMed Dawes GS, Moulden M, Redman CW. Short-term fetal heart rate variation, decelerations, and umbilical flow velocity waveforms before labor. Obstet Gynecol. 1992;80(4):673–8.PubMed
23.
go back to reference Schiermeier S, Reinhard J, Hatzmann H, Zimmermann RC, Westhof G. Fetal short time variation during labor: a non-invasive alternative to fetal scalp pH measurements? J Perinat Med. 2009;37(5):529–33.PubMed Schiermeier S, Reinhard J, Hatzmann H, Zimmermann RC, Westhof G. Fetal short time variation during labor: a non-invasive alternative to fetal scalp pH measurements? J Perinat Med. 2009;37(5):529–33.PubMed
24.
go back to reference Agrawal SK, Doucette F, Gratton R, Richardson B, Gagnon R. Intrapartum computerized fetal heart rate parameters and metabolic acidosis at birth. Obstet Gynecol. 2003;102(4):731–8.PubMed Agrawal SK, Doucette F, Gratton R, Richardson B, Gagnon R. Intrapartum computerized fetal heart rate parameters and metabolic acidosis at birth. Obstet Gynecol. 2003;102(4):731–8.PubMed
25.
go back to reference Nordström L, Waldenström U. Handläggning av normal förlossning - state of the art [Management of normal birth - state of the art]. Socialstyrelsen [The Swedish National Board of Health and Welfare]. (in Swedish) 2001-123-1. Nordström L, Waldenström U. Handläggning av normal förlossning - state of the art [Management of normal birth - state of the art]. Socialstyrelsen [The Swedish National Board of Health and Welfare]. (in Swedish) 2001-123-1.
27.
go back to reference Georgieva A, Payne SJ, Moulden M, Redman CW. Computerized intrapartum electronic fetal monitoring: analysis of the decision to deliver for fetal distress, Conference proceedings: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. 2011. p. 5888–91. Georgieva A, Payne SJ, Moulden M, Redman CW. Computerized intrapartum electronic fetal monitoring: analysis of the decision to deliver for fetal distress, Conference proceedings: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. 2011. p. 5888–91.
28.
go back to reference Steer P. Commentary on’ Antenatal cardiotocogram quality and interpretation using computers’. BJOG. 2014;121 Suppl 7:9–13.CrossRefPubMed Steer P. Commentary on’ Antenatal cardiotocogram quality and interpretation using computers’. BJOG. 2014;121 Suppl 7:9–13.CrossRefPubMed
Metadata
Title
Fetal heart rate monitoring of short term variation (STV): a methodological observational study
Authors
Stina Wretler
Malin Holzmann
Sophie Graner
Pelle Lindqvist
Susanne Falck
Lennart Nordström
Publication date
01-12-2016
Publisher
BioMed Central
Published in
BMC Pregnancy and Childbirth / Issue 1/2016
Electronic ISSN: 1471-2393
DOI
https://doi.org/10.1186/s12884-016-0845-8

Other articles of this Issue 1/2016

BMC Pregnancy and Childbirth 1/2016 Go to the issue