Abstract

Background

Left atrial volume is a prognostic factor in cardiac pathologies. We aimed to validate left atrial volume detection with 3D and 2D echocardiography (3DE and 2DE) by human cadaveric casts. 3DE facilitates measurement of atrial volume without geometrical assumptions or dependence on imaging angle in contrast to 2DE methods.

Methods

For method validation, six water-filled balloons were submerged in a 20-l water tank and their volumes were measured with 3DE. Seven human cadaveric left atrial casts were prepared of silicone and were transformed into ultrasound-permeable casts. Casts were imaged in the same setting, so that 3DE and 2DE of casts represented transthoracic apical view. Left ventricle analysis softwares GE 4D Auto LVQ and TomTec 4D LV-Function were used for 3DE volumetry.

Results

Balloon volumes ranged 37 to 255 ml (mean 126 ml). 3DE resulted in an excellent volumetric agreement with balloon volumes, absolute bias was − 3.7 ml (95% CI -5.9 to − 1.4). Atrial cast volumes were 38 to 94 ml (mean 56.6 ml). 3DE and 2DE volumes were excellently correlated with cast volumes (r = 0.96 to 0.99). Biases were for GE 4D LVQ -0.7 ml (95% CI -6.1 to 4.6), TomTec 4D LV-Function 3.3 ml (− 1.9 to 8.5) and 2DE 2.9 ml (− 4.0 to 9.9). 3DE resulted in lower limits of agreement and showed no volume-related bias in contrast to area-length method.

Conclusions

We conclude that measurement of human cadaveric left atrial cast volumes by 3DE is in excellent agreement with true cast volumes.
Literature
3.
4.
go back to reference Ristow B, Ali S, Whooley MA, Schiller NB. Usefulness of left atrial volume index to predict heart failure hospitalization and mortality in ambulatory patients with coronary heart disease and comparison to left ventricular ejection fraction (from the heart and soul study). Am J Cardiol. 2008. https://doi.org/10.1016/j.amjcard.2008.02.099. Ristow B, Ali S, Whooley MA, Schiller NB. Usefulness of left atrial volume index to predict heart failure hospitalization and mortality in ambulatory patients with coronary heart disease and comparison to left ventricular ejection fraction (from the heart and soul study). Am J Cardiol. 2008. https://​doi.​org/​10.​1016/​j.​amjcard.​2008.​02.​099.
5.
go back to reference Gulati A, Ismail TF, Jabbour A, Ismail NA, Morarji K, Ali A, et al. Clinical utility and prognostic value of left atrial volume assessment by cardiovascular magnetic resonance in non-ischaemic dilated cardiomyopathy. Eur J Heart Fail. 2013. https://doi.org/10.1093/eurjhf/hft019. Gulati A, Ismail TF, Jabbour A, Ismail NA, Morarji K, Ali A, et al. Clinical utility and prognostic value of left atrial volume assessment by cardiovascular magnetic resonance in non-ischaemic dilated cardiomyopathy. Eur J Heart Fail. 2013. https://​doi.​org/​10.​1093/​eurjhf/​hft019.
9.
go back to reference Lang RM, Badano LP, Mor-Avi V, Afilalo J, Armstrong A, Ernande L, et al. Recommendations for cardiac chamber quantification by echocardiography in adults: an update from the American Society of Echocardiography and the European Association of Cardiovascular Imaging. J Am Soc Echocardiogr. 2015. https://doi.org/10.1016/j.echo.2014.10.003. Lang RM, Badano LP, Mor-Avi V, Afilalo J, Armstrong A, Ernande L, et al. Recommendations for cardiac chamber quantification by echocardiography in adults: an update from the American Society of Echocardiography and the European Association of Cardiovascular Imaging. J Am Soc Echocardiogr. 2015. https://​doi.​org/​10.​1016/​j.​echo.​2014.​10.​003.
10.
go back to reference Donal E, Lip GYH, Galderisi M, Goette A, Shah D, Marwan M, et al. EACVI/EHRA expert consensus document on the role of multi-modality imaging for the evaluation of patients with atrial fibrillation. Eur Heart J Cardiovasc Imaging. 2016. https://doi.org/10.1093/ehjci/jev354. Donal E, Lip GYH, Galderisi M, Goette A, Shah D, Marwan M, et al. EACVI/EHRA expert consensus document on the role of multi-modality imaging for the evaluation of patients with atrial fibrillation. Eur Heart J Cardiovasc Imaging. 2016. https://​doi.​org/​10.​1093/​ehjci/​jev354.
11.
go back to reference Kühl JT, Lønborg J, Fuchs A, Andersen MJ, Vejlstrup N, Kelbæk H, et al. Assessment of left atrial volume and function: a comparative study between echocardiography, magnetic resonance imaging and multi slice computed tomography. Int J Cardiovasc Imaging. 2012. https://doi.org/10.1007/s10554-011-9930-2. Kühl JT, Lønborg J, Fuchs A, Andersen MJ, Vejlstrup N, Kelbæk H, et al. Assessment of left atrial volume and function: a comparative study between echocardiography, magnetic resonance imaging and multi slice computed tomography. Int J Cardiovasc Imaging. 2012. https://​doi.​org/​10.​1007/​s10554-011-9930-2.
14.
go back to reference Artang R, Migrino RQ, Harmann L, Bowers M, Woods TD. Left atrial volume measurement with automated border detection by 3-dimensional echocardiography: comparison with magnetic resonance imaging. Cardiovasc Ultrasound. 2009. https://doi.org/10.1186/1476-7120-7-16. Artang R, Migrino RQ, Harmann L, Bowers M, Woods TD. Left atrial volume measurement with automated border detection by 3-dimensional echocardiography: comparison with magnetic resonance imaging. Cardiovasc Ultrasound. 2009. https://​doi.​org/​10.​1186/​1476-7120-7-16.
15.
go back to reference Perez de Isla L, Feltes G, Moreno J, Martinez W, Saltijeral A, de Agustin JA, et al. Quantification of left atrial volumes using three-dimensional wall motion tracking echocardiographic technology: comparison with cardiac magnetic resonance. Eur Hear J - Cardiovasc Imaging. 2014. https://doi.org/10.1093/ehjci/jeu001. Perez de Isla L, Feltes G, Moreno J, Martinez W, Saltijeral A, de Agustin JA, et al. Quantification of left atrial volumes using three-dimensional wall motion tracking echocardiographic technology: comparison with cardiac magnetic resonance. Eur Hear J - Cardiovasc Imaging. 2014. https://​doi.​org/​10.​1093/​ehjci/​jeu001.
17.
20.
go back to reference Kebed K, Kruse E, Addetia K, Ciszek B, Thykattil M, Guile B, et al. Atrial-focused views improve the accuracy of two-dimensional echocardiographic measurements of the left and right atrial volumes: a contribution to the increase in normal values in the guidelines update. Int J Cardiovasc Imaging. 2016. https://doi.org/10.1007/s10554-016-0988-8. Kebed K, Kruse E, Addetia K, Ciszek B, Thykattil M, Guile B, et al. Atrial-focused views improve the accuracy of two-dimensional echocardiographic measurements of the left and right atrial volumes: a contribution to the increase in normal values in the guidelines update. Int J Cardiovasc Imaging. 2016. https://​doi.​org/​10.​1007/​s10554-016-0988-8.
22.
go back to reference Rabbat MG, Wilber D, Thomas K, Malick O, Bashir A, Agrawal A, et al. Left atrial volume assessment in atrial fibrillation using multimodality imaging: a comparison of echocardiography, invasive three-dimensional CARTO and cardiac magnetic resonance imaging. Int J Cardiovasc Imaging. 2015. https://doi.org/10.1007/s10554-015-0641-y. Rabbat MG, Wilber D, Thomas K, Malick O, Bashir A, Agrawal A, et al. Left atrial volume assessment in atrial fibrillation using multimodality imaging: a comparison of echocardiography, invasive three-dimensional CARTO and cardiac magnetic resonance imaging. Int J Cardiovasc Imaging. 2015. https://​doi.​org/​10.​1007/​s10554-015-0641-y.
Metadata
Title
Validation of 3D echocardiographic volume detection of left atrium by human cadaveric casts
Authors
Jouni K. Kuusisto
Vesa M. Järvinen
Juha P. Sinisalo
Publication date
01-12-2018
Publisher
BioMed Central
Published in
BMC Medical Imaging / Issue 1/2018
Electronic ISSN: 1471-2342
DOI
https://doi.org/10.1186/s12880-018-0282-4

Other articles of this Issue 1/2018

BMC Medical Imaging 1/2018 Go to the issue