Published in: BMC Medical Imaging 1/2016

Open Access 01-12-2016 | Technical advance

3D rotational fluoroscopy for intraoperative clip control in patients with intracranial aneurysms – assessment of feasibility and image quality

Authors: Thomas Westermaier, Thomas Linsenmann, György A. Homola, Mario Loehr, Christian Stetter, Nadine Willner, Ralf-Ingo Ernestus, Laszlo Solymosi, Giles H. Vince

Published in: BMC Medical Imaging | Issue 1/2016

Login to get access

Abstract

Background

Mobile 3D fluoroscopes have become increasingly available in neurosurgical operating rooms. In this series, the image quality and value of intraoperative 3D fluoroscopy with intravenous contrast agent for the evaluation of aneurysm occlusion and vessel patency after clip placement was assessed in patients who underwent surgery for intracranial aneurysms.

Materials and methods

Twelve patients were included in this retrospective analysis. Prior to surgery, a 360° rotational fluoroscopy scan was performed without contrast agent followed by another scan with 50 ml of intravenous iodine contrast agent. The image files of both scans were transferred to an Apple PowerMac® workstation, subtracted and reconstructed using OsiriX® free software. The procedure was repeated after clip placement. Both image sets were compared for assessment of aneurysm occlusion and vessel patency.

Results

Image acquisition and contrast administration caused no adverse effects. Image quality was sufficient to follow the patency of the vessels distal to the clip. Metal artifacts reduce the assessability of the immediate vicinity of the clip. Precise image subtraction and post-processing can reduce metal artifacts and make the clip-site assessable and depict larger neck-remnants.

Conclusion

This technique quickly supplies images at adequate quality to evaluate distal vessel patency after aneurysm clipping. Significant aneurysm remnants may be depicted as well. As it does not require visual control of all vessels that are supposed to be evaluated intraoperatively, this technique may be complementary to other intraoperative tools like indocyanine green videoangiography and micro-Doppler, especially for the assessment of larger aneurysms. At the momentary state of this technology, it cannot replace postoperative conventional angiography. However, 3D fluoroscopy and image post-processing are young technologies. Further technical developments are likely to result in improved image quality.
Literature
1.
go back to reference Zhang H1, Hou C, Zhou Z, Zhang H, Zhou G, Zhang G. Evaluating of small intracranial aneurysms by 64-detector CT Angiography: a comparison with 3-dimensional rotation DSA or surgical findings. J Neuroimaging. 2014;24(2):137-43. Zhang H1, Hou C, Zhou Z, Zhang H, Zhou G, Zhang G. Evaluating of small intracranial aneurysms by 64-detector CT Angiography: a comparison with 3-dimensional rotation DSA or surgical findings. J Neuroimaging. 2014;24(2):137-43.
2.
go back to reference Fischer G, Stadie A, Oertel JM. Near-infrared indocyanine green videoangiography versus microvascular Doppler sonography in aneurysm surgery. Acta Neurochir (Wien). 2010;152(9):1519–25.CrossRef Fischer G, Stadie A, Oertel JM. Near-infrared indocyanine green videoangiography versus microvascular Doppler sonography in aneurysm surgery. Acta Neurochir (Wien). 2010;152(9):1519–25.CrossRef
3.
go back to reference Siasios I, Kapsalaki EZ, Fountas KN. The role of intraoperative micro-Doppler ultrasound in verifying proper clip placement in intracranial aneurysm surgery. Neuroradiology. 2012;54(10):1109–18.CrossRefPubMed Siasios I, Kapsalaki EZ, Fountas KN. The role of intraoperative micro-Doppler ultrasound in verifying proper clip placement in intracranial aneurysm surgery. Neuroradiology. 2012;54(10):1109–18.CrossRefPubMed
4.
go back to reference Raabe A, Beck J, Gerlach R, Zimmermann M, Seifert V. Near-infrared indocyanine green video angiography: a new method for intraoperative assessment of vascular flow. Neurosurgery. 2003;52(1):132–9.PubMed Raabe A, Beck J, Gerlach R, Zimmermann M, Seifert V. Near-infrared indocyanine green video angiography: a new method for intraoperative assessment of vascular flow. Neurosurgery. 2003;52(1):132–9.PubMed
5.
go back to reference Westermaier T, Linsenmann T, Keler AF, Stetter C, Willner N, Solymosi L, Ernestus RI, Vince GH. Intraoperative cerebral angiography by intravenous contrast administration with 3-dimensional rotational fluoroscopy in patients with intracranial aneurysms: a feasibility study. Neurosurgery. 2015;11 Suppl 2:119–26.PubMed Westermaier T, Linsenmann T, Keler AF, Stetter C, Willner N, Solymosi L, Ernestus RI, Vince GH. Intraoperative cerebral angiography by intravenous contrast administration with 3-dimensional rotational fluoroscopy in patients with intracranial aneurysms: a feasibility study. Neurosurgery. 2015;11 Suppl 2:119–26.PubMed
6.
go back to reference Westermaier T, Willner N, Vince GH, Linsenmann T, Ernestus RI, Stetter C. Intraoperative 3D rotational angiography: an emergency tool for the diagnosis of intracranial aneurysms. Emerg Radiol. 2015;22(1):97–100.CrossRefPubMed Westermaier T, Willner N, Vince GH, Linsenmann T, Ernestus RI, Stetter C. Intraoperative 3D rotational angiography: an emergency tool for the diagnosis of intracranial aneurysms. Emerg Radiol. 2015;22(1):97–100.CrossRefPubMed
7.
go back to reference Smith-Bindman R, Moghadassi M, Wilson N, Nelson TR, Boone JM, Cagnon CH, Gould R, Hall DJ, Krishnam M, Lamba R, McNitt-Gray M, Seibert A, Miglioretti DL. Radiation doses in consecutive CT examinations from five university of California Medical Centers. Radiology. 2015;277(1):134–41. doi:10.1148/radiol.2015142728. Epub 2015 May 2015.CrossRefPubMed Smith-Bindman R, Moghadassi M, Wilson N, Nelson TR, Boone JM, Cagnon CH, Gould R, Hall DJ, Krishnam M, Lamba R, McNitt-Gray M, Seibert A, Miglioretti DL. Radiation doses in consecutive CT examinations from five university of California Medical Centers. Radiology. 2015;277(1):134–41. doi:10.​1148/​radiol.​2015142728. Epub 2015 May 2015.CrossRefPubMed
8.
go back to reference Geyer LL, Korner M, Harrieder A, Mueck FG, Deak Z, Wirth S, Linsenmaier U. Dose reduction in 64-row whole-body CT in multiple trauma: an optimized CT protocol with iterative image reconstruction on a gemstone-based scintillator. Br J Radiol. 2016:20160003. Geyer LL, Korner M, Harrieder A, Mueck FG, Deak Z, Wirth S, Linsenmaier U. Dose reduction in 64-row whole-body CT in multiple trauma: an optimized CT protocol with iterative image reconstruction on a gemstone-based scintillator. Br J Radiol. 2016:20160003.
9.
go back to reference Cardinale F, Miserocchi A, Moscato A, Cossu M, Castana L, Schiariti MP, Gozzo F, Pero F, Quilici L, Citterio A, Minella M, Torresin A, Lo Russo G. Talairach methodology in the multimodal imaging and robotics era. In: Scarabin JM (ed.), Stereotaxy and Epilepsy Surgery. John Libbey Eurotext. 2012;245-272. Cardinale F, Miserocchi A, Moscato A, Cossu M, Castana L, Schiariti MP, Gozzo F, Pero F, Quilici L, Citterio A, Minella M, Torresin A, Lo Russo G. Talairach methodology in the multimodal imaging and robotics era. In: Scarabin JM (ed.), Stereotaxy and Epilepsy Surgery. John Libbey Eurotext. 2012;245-272.
10.
go back to reference Westerlaan HE, Gravendeel J, Fiore D, Metzemaekers JD, Groen RJ, Mooij JJ, Oudkerk M. Multislice CT angiography in the selection of patients with ruptured intracranial aneurysms suitable for clipping or coiling. Neuroradiology. 2007;49(12):997–1007.CrossRefPubMedPubMedCentral Westerlaan HE, Gravendeel J, Fiore D, Metzemaekers JD, Groen RJ, Mooij JJ, Oudkerk M. Multislice CT angiography in the selection of patients with ruptured intracranial aneurysms suitable for clipping or coiling. Neuroradiology. 2007;49(12):997–1007.CrossRefPubMedPubMedCentral
11.
go back to reference Caplan JM, Sankey E, Yang W, Radvany MG, Colby GP, Coon AL, Tamargo RJ, Huang J. Impact of indocyanine green videoangiography on rate of clip adjustments following intraoperative angiography. Neurosurgery. 2014;75(4):437–43.CrossRefPubMed Caplan JM, Sankey E, Yang W, Radvany MG, Colby GP, Coon AL, Tamargo RJ, Huang J. Impact of indocyanine green videoangiography on rate of clip adjustments following intraoperative angiography. Neurosurgery. 2014;75(4):437–43.CrossRefPubMed
12.
go back to reference Washington CW, Zipfel GJ, Chicoine MR, Derdeyn CP, Rich KM, Moran CJ, Cross DT, Dacey RGJ. Comparing indocyanine green videoangiography to the gold standard of intraoperative digital subtraction angiography used in aneurysm surgery. J Neurosurg. 2013;118(2):420–7.CrossRefPubMed Washington CW, Zipfel GJ, Chicoine MR, Derdeyn CP, Rich KM, Moran CJ, Cross DT, Dacey RGJ. Comparing indocyanine green videoangiography to the gold standard of intraoperative digital subtraction angiography used in aneurysm surgery. J Neurosurg. 2013;118(2):420–7.CrossRefPubMed
13.
go back to reference Sun G, Ding J, Lu Y, Li M, Li L, Li GY, Zhang XP. Comparison of standard- and low-tube voltage 320-detector row volume CT angiography in detection of intracranial aneurysms with digital subtraction angiography as gold standard. Acad Radiol. 2012;19(3):281–8.CrossRefPubMed Sun G, Ding J, Lu Y, Li M, Li L, Li GY, Zhang XP. Comparison of standard- and low-tube voltage 320-detector row volume CT angiography in detection of intracranial aneurysms with digital subtraction angiography as gold standard. Acad Radiol. 2012;19(3):281–8.CrossRefPubMed
14.
go back to reference Struffert T, Deuerling-Zheng Y, Kloska S, Engelhorn T, Strother CM, Kalender WA, Kohrmann M, Schwab S, Doerfler A. Flat detector CT in the evaluation of brain parenchyma, intracranial vasculature, and cerebral blood volume: a pilot study in patients with acute symptoms of cerebral ischemia. AJNR Am J Neuroradiol. 2010;31(8):1462–9.CrossRefPubMed Struffert T, Deuerling-Zheng Y, Kloska S, Engelhorn T, Strother CM, Kalender WA, Kohrmann M, Schwab S, Doerfler A. Flat detector CT in the evaluation of brain parenchyma, intracranial vasculature, and cerebral blood volume: a pilot study in patients with acute symptoms of cerebral ischemia. AJNR Am J Neuroradiol. 2010;31(8):1462–9.CrossRefPubMed
15.
go back to reference Jeon JS, Sheen SH, Hwang GJ, Kim HC, Kwon BJ. Feasibility of intravenous flat panel detector CT angiography for intracranial arterial stenosis. AJNR Am J Neuroradiol. 2013;34(1):129–34.CrossRefPubMed Jeon JS, Sheen SH, Hwang GJ, Kim HC, Kwon BJ. Feasibility of intravenous flat panel detector CT angiography for intracranial arterial stenosis. AJNR Am J Neuroradiol. 2013;34(1):129–34.CrossRefPubMed
16.
go back to reference Kamran M, Nagaraja S, Byrne JV. C-arm flat detector computed tomography: the technique and its applications in interventional neuro-radiology. Neuroradiology. 2010;52(4):319–27.CrossRefPubMed Kamran M, Nagaraja S, Byrne JV. C-arm flat detector computed tomography: the technique and its applications in interventional neuro-radiology. Neuroradiology. 2010;52(4):319–27.CrossRefPubMed
17.
go back to reference Leng LZ, Rubin DG, Patsalides A, Riina HA. Fusion of intraoperative three-dimensional rotational angiography and flat-panel detector computed tomography for cerebrovascular neuronavigation. World Neurosurg. 2013;79(3-4):504–9.CrossRefPubMed Leng LZ, Rubin DG, Patsalides A, Riina HA. Fusion of intraoperative three-dimensional rotational angiography and flat-panel detector computed tomography for cerebrovascular neuronavigation. World Neurosurg. 2013;79(3-4):504–9.CrossRefPubMed
18.
go back to reference Schnell O, Morhard D, Holtmannspotter M, Reiser M, Tonn JC, Schichor C. Near-infrared indocyanine green videoangiography (ICGVA) and intraoperative computed tomography (iCT): are they complementary or competitive imaging techniques in aneurysm surgery? Acta Neurochir (Wien). 2012;154(10):1861–8.CrossRef Schnell O, Morhard D, Holtmannspotter M, Reiser M, Tonn JC, Schichor C. Near-infrared indocyanine green videoangiography (ICGVA) and intraoperative computed tomography (iCT): are they complementary or competitive imaging techniques in aneurysm surgery? Acta Neurochir (Wien). 2012;154(10):1861–8.CrossRef
19.
go back to reference Samson D, Batjer HH, Bowman G, Mootz L, Krippner WJJ, Meyer YJ, Allen BC. A clinical study of the parameters and effects of temporary arterial occlusion in the management of intracranial aneurysms. Neurosurgery. 1994;34(1):22–8.CrossRefPubMed Samson D, Batjer HH, Bowman G, Mootz L, Krippner WJJ, Meyer YJ, Allen BC. A clinical study of the parameters and effects of temporary arterial occlusion in the management of intracranial aneurysms. Neurosurgery. 1994;34(1):22–8.CrossRefPubMed
20.
go back to reference Rozeik C, Kotterer O, Preiss J, Schutz M, Dingler W, Deininger HK. Cranial CT artifacts and gantry angulation. J Comput Assist Tomogr. 1991;15(3):381–6.CrossRefPubMed Rozeik C, Kotterer O, Preiss J, Schutz M, Dingler W, Deininger HK. Cranial CT artifacts and gantry angulation. J Comput Assist Tomogr. 1991;15(3):381–6.CrossRefPubMed
21.
go back to reference Anderson GB, Steinke DE, Petruk KC, Ashforth R, Findlay JM. Computed tomographic angiography versus digital subtraction angiography for the diagnosis and early treatment of ruptured intracranial aneurysms. Neurosurgery. 1999;45(6):1315–20.CrossRefPubMed Anderson GB, Steinke DE, Petruk KC, Ashforth R, Findlay JM. Computed tomographic angiography versus digital subtraction angiography for the diagnosis and early treatment of ruptured intracranial aneurysms. Neurosurgery. 1999;45(6):1315–20.CrossRefPubMed
22.
go back to reference Pechlivanis I, Konig M, Engelhardt M, Scholz M, Heuser L, Harders A, Schmieder K. Evaluation of clip artefacts in three-dimensional computed tomography. Cent Eur Neurosurg. 2009;70(1):9–14.CrossRefPubMed Pechlivanis I, Konig M, Engelhardt M, Scholz M, Heuser L, Harders A, Schmieder K. Evaluation of clip artefacts in three-dimensional computed tomography. Cent Eur Neurosurg. 2009;70(1):9–14.CrossRefPubMed
23.
go back to reference Watanabe Y, Kashiwagi N, Yamada N, Higashi M, Fukuda T, Morikawa S, Onishi Y, Iihara K, Miyamoto S, Naito H. Subtraction 3D CT angiography with the orbital synchronized helical scan technique for the evaluation of postoperative cerebral aneurysms treated with cobalt-alloy clips. AJNR Am J Neuroradiol. 2008;29(6):1071–5.CrossRefPubMed Watanabe Y, Kashiwagi N, Yamada N, Higashi M, Fukuda T, Morikawa S, Onishi Y, Iihara K, Miyamoto S, Naito H. Subtraction 3D CT angiography with the orbital synchronized helical scan technique for the evaluation of postoperative cerebral aneurysms treated with cobalt-alloy clips. AJNR Am J Neuroradiol. 2008;29(6):1071–5.CrossRefPubMed
25.
go back to reference Paul J, Jacobi V, Farhang M, Bazrafshan B, Vogl TJ, Mbalisike EC. Radiation dose and image quality of X-ray volume imaging systems: cone-beam computed tomography, digital subtraction angiography and digital fluoroscopy. Eur Radiol. 2013;23(6):1582–93.CrossRefPubMed Paul J, Jacobi V, Farhang M, Bazrafshan B, Vogl TJ, Mbalisike EC. Radiation dose and image quality of X-ray volume imaging systems: cone-beam computed tomography, digital subtraction angiography and digital fluoroscopy. Eur Radiol. 2013;23(6):1582–93.CrossRefPubMed
Metadata
Title
3D rotational fluoroscopy for intraoperative clip control in patients with intracranial aneurysms – assessment of feasibility and image quality
Authors
Thomas Westermaier
Thomas Linsenmann
György A. Homola
Mario Loehr
Christian Stetter
Nadine Willner
Ralf-Ingo Ernestus
Laszlo Solymosi
Giles H. Vince
Publication date
01-12-2016
Publisher
BioMed Central
Published in
BMC Medical Imaging / Issue 1/2016
Electronic ISSN: 1471-2342
DOI
https://doi.org/10.1186/s12880-016-0133-0

Other articles of this Issue 1/2016

BMC Medical Imaging 1/2016 Go to the issue