Skip to main content
Top
Published in: BMC Anesthesiology 1/2018

Open Access 01-12-2018 | Research article

Laryngeal mask airway Unique™ position in paediatric patients undergoing magnetic resonance imaging (MRI): prospective observational study

Authors: Jozef Klučka, Jan Šenkyřík, Jarmila Skotáková, Roman Štoudek, Michaela Ťoukalková, Ivo Křikava, Lukáš Mareček, Tomáš Pavlík, Alena Štouračová, Petr Štourač

Published in: BMC Anesthesiology | Issue 1/2018

Login to get access

Abstract

Background

Laryngeal mask UNIQUE® (LMAU) is supraglottic airway device with good clinical performance and low failure rate. Little is known about the ideal position of the LMAU on the magnetic resonance imaging (MRI) and whether radiological malposition can be associated with clinical performance (audible leak) in children. The primary aim of the study was to evaluate incidence of the radiologic malposition of the LMAU according to size. The secondary outcome was the clinical performance and associated complications (1st attempt success rate, audible leak) in LMAUs in correct position vs. radiologically misplaced LMAUs.

Methods

In prospective observational study, all paediatric patients undergoing MRI of the brain under general anaesthesia with the LMAU were included (1.9.2016–16.5.2017). The radiologically correct position: LMAU in hypopharynx, proximal cuff opposite to the C1 or C2 and distance A (proximal cuff end and aditus laryngis) ≤ distance B (distal cuff end and aditus laryngis). Malposition A: LMAU outside the hypopharynx. Malposition B: proximal cuff outside C1-C2. Malposition C: distance A ≥ distance B. We measured distances on the MRI image. Malposition incidence between LMAU sizes and first attempt success rate in trainees and consultant groups was compared using Fisher exact test, difference in incidence of malpositions using McNemar test and difference in leakage according to radiological position using two-sample binomial test.

Results

Overall 202 paediatric patients were included. The incidence of radiologically defined malposition was 26.2% (n = 53). Laryngeal mask was successfully inserted on the 1st attempt in 91.1% (n = 184) cases. Audible leak was detected in 3.5% (n = 7) patients. The radiologically defined malposition was present in 42.9% (n = 3) cases with audible leak. The rate of associated complications was 1.5% (n = 3): laryngospasm, desaturation, cough. In 4.0% (n = 8) the LMAU was soiled from blood.
Higher incidence of radiological malposition was in LMAU 1.0, 1.5 and LMAU 3, 4 compared to LMAU 2 or LMAU 2.5 (p < 0.001).

Conclusion

Malposition was not associated with impaired clinical performance (audible leak, complications) of the LMAU or the need for alternative airway management.

Trial registration

Clinicaltrials.​gov  (NCT02940652) Registered 18 October 18 2016.
Literature
1.
go back to reference Goldmann K. Recent developments in airway management of the paediatric patient. Curr Opin Anaesthesiol. 2006;19:278–84.CrossRef Goldmann K. Recent developments in airway management of the paediatric patient. Curr Opin Anaesthesiol. 2006;19:278–84.CrossRef
2.
go back to reference Luce V, Harkouk H, Brasher C, et al. Supraglottic airway devices vs tracheal intubation in children: a quantitative metaanalysis of respiratory complications. Paediatr Anaesth. 2014;24:1088–98.CrossRef Luce V, Harkouk H, Brasher C, et al. Supraglottic airway devices vs tracheal intubation in children: a quantitative metaanalysis of respiratory complications. Paediatr Anaesth. 2014;24:1088–98.CrossRef
3.
go back to reference Lopez-Gil M, Brimacombe J, Alvarez M. Safety and efficacy of the laryngeal mask airway. A prospective survey of 1400 children. Anaesthesia. 1996;51:969–72.CrossRef Lopez-Gil M, Brimacombe J, Alvarez M. Safety and efficacy of the laryngeal mask airway. A prospective survey of 1400 children. Anaesthesia. 1996;51:969–72.CrossRef
4.
go back to reference Lopez-Gil M, Brimacombe J, Cebrian J, Arranz J. Laryngeal mask airway in pediatric practice: a prospective study of skill acquisition by anesthesia residents. Anesthesiology. 1996;84:807–11.CrossRef Lopez-Gil M, Brimacombe J, Cebrian J, Arranz J. Laryngeal mask airway in pediatric practice: a prospective study of skill acquisition by anesthesia residents. Anesthesiology. 1996;84:807–11.CrossRef
5.
go back to reference Jagannathan N, Ramsey MA, White MC, Sohn L. An update on newer pediatric supraglottic airways with recommendations for clinical use. Paediatr Anaesth. 2015;25(4):334–45.CrossRef Jagannathan N, Ramsey MA, White MC, Sohn L. An update on newer pediatric supraglottic airways with recommendations for clinical use. Paediatr Anaesth. 2015;25(4):334–45.CrossRef
6.
go back to reference Verghese C, Berlet J, Kapila A, Pollard R. Clinical assessment of the single use laryngeal mask airway—the LMA-unique. Br J Anaesth. 1998;80(5):677–9.CrossRef Verghese C, Berlet J, Kapila A, Pollard R. Clinical assessment of the single use laryngeal mask airway—the LMA-unique. Br J Anaesth. 1998;80(5):677–9.CrossRef
7.
go back to reference Brimacombe J, von Goedecke A, Keller C, Brimacombe L, Brimacombe M. The laryngeal mask airway unique versus the soft seal laryngeal mask: a randomized, crossover study in paralyzed, anesthetized patients. Anesth Analg. 2004;99(5):1560–3; table of contents. Brimacombe J, von Goedecke A, Keller C, Brimacombe L, Brimacombe M. The laryngeal mask airway unique versus the soft seal laryngeal mask: a randomized, crossover study in paralyzed, anesthetized patients. Anesth Analg. 2004;99(5):1560–3; table of contents.
8.
go back to reference Goudsouzian NG, Denman W, Cleveland R, Shorten G. Radiologic localization of the laryngeal mask airway in children. Anesthesiology. 1992;77:1085-9.CrossRef Goudsouzian NG, Denman W, Cleveland R, Shorten G. Radiologic localization of the laryngeal mask airway in children. Anesthesiology. 1992;77:1085-9.CrossRef
9.
go back to reference Brain AI. Laryngeal mask misplacement--causes, consequences and solutions. Anaesthesia. 1992;47:531-2.CrossRef Brain AI. Laryngeal mask misplacement--causes, consequences and solutions. Anaesthesia. 1992;47:531-2.CrossRef
10.
go back to reference Monclus E, Garcés A, De Jose Maria B, Artés D, Mabrock M. Study of the adjustment of the Ambu laryngeal mask under magnetic resonance imaging. Paediatr Anaesth. 2007;17:1182-6.CrossRef Monclus E, Garcés A, De Jose Maria B, Artés D, Mabrock M. Study of the adjustment of the Ambu laryngeal mask under magnetic resonance imaging. Paediatr Anaesth. 2007;17:1182-6.CrossRef
11.
go back to reference Vialet R, Nau A, Chaumoître K, Martin C. Effects of head posture on the oral, pharyngeal and laryngeal axis alignment in infants and young children by magnetic resonance imaging. Paediatr Anaesth. 2008;18:525–31.CrossRef Vialet R, Nau A, Chaumoître K, Martin C. Effects of head posture on the oral, pharyngeal and laryngeal axis alignment in infants and young children by magnetic resonance imaging. Paediatr Anaesth. 2008;18:525–31.CrossRef
14.
go back to reference Brain AI. The laryngeal mask--a new concept in airway management. Br J Anaesth. 1983;55:801-5.CrossRef Brain AI. The laryngeal mask--a new concept in airway management. Br J Anaesth. 1983;55:801-5.CrossRef
Metadata
Title
Laryngeal mask airway Unique™ position in paediatric patients undergoing magnetic resonance imaging (MRI): prospective observational study
Authors
Jozef Klučka
Jan Šenkyřík
Jarmila Skotáková
Roman Štoudek
Michaela Ťoukalková
Ivo Křikava
Lukáš Mareček
Tomáš Pavlík
Alena Štouračová
Petr Štourač
Publication date
01-12-2018
Publisher
BioMed Central
Published in
BMC Anesthesiology / Issue 1/2018
Electronic ISSN: 1471-2253
DOI
https://doi.org/10.1186/s12871-018-0617-2

Other articles of this Issue 1/2018

BMC Anesthesiology 1/2018 Go to the issue