Abstract

Background

Recent case reports have indicated significant cuff overinflation when using the standard filling volume based on the manufacturer’s recommendations in older models of laryngeal tubes. The aim of this study was to determine the minimum cuff pressure needed to perform standardized ventilation without leakage in the new, revised model of the laryngeal tube “LTS-D”.

Methods

After ethical approval, LTS-D was placed for ventilation in 60 anesthetized patients. The cuff was inflated to the recommended volume (#3: 60 ml, #4: 80 ml, and #5: 90 ml). After evaluation of the initial cuff pressure (CP), the CP was lowered in 10 cmH2O steps until a minimal cuff pressure of 30 cmH2O was achieved. The absence of an audible leak was required for a step-by-step reduction in the CP. Evacuated cuff volume, success rate, and airway injuries were documented. Data were expressed as medians (interquartile ranges [IQRs]). The comparison of CPs and cuff volumes was performed using the Mann-Whitney test.

Results

After initial inflation, the CP ranged from 105 cmH2O [90–120; #5] to 120 cmH2O [110–120; #3]. Lowering the CP to 60 cmH2O resulted in a reduced cuff volume ranging from 47 ml [44–54; #3] to 77 ml [75–82; #5] compared to the initial inflation (p < 0.001). Leakage occurred more frequently when the CP was lowered to 40 cmH2O compared to the initial inflation (44/54 [81%]; p < 0.01). Using a CP between 50 cmH2O and 60 cmH2O, a leakage rate of 3/54 (5%) was observed, compared to a rate of 11/54 (21%) when using a CP lower than 50 cmH2O. The overall success rate was 90%, and airway injury occurred in 7% of patients (4/60).

Conclusion

We found significant overinflation of the revised LTS-D using the recommended volume for initial cuff inflation. A CP of 60 cmH2O was found to be sufficient for ventilation in the majority of patients evaluated. Checking and adjusting the CP in laryngeal tubes is mandatory to avoid overinflation.

Trial registration

ClinicalTrials.gov NCT02300337. Registered: 20 November 2014.
Appendix
Available only for authorised users
Literature
1.
go back to reference Genzwurker H, Finteis T, Hinkelbein J, et al. First clinical experiences with the new LTS. A laryngeal tube with an oesophageal drain. Anaesthesist. 2003;52:697–702.CrossRefPubMed Genzwurker H, Finteis T, Hinkelbein J, et al. First clinical experiences with the new LTS. A laryngeal tube with an oesophageal drain. Anaesthesist. 2003;52:697–702.CrossRefPubMed
2.
go back to reference Mandal NG. A new device has to be safe and reliable too. Anaesth. 2001;56:382–3.CrossRef Mandal NG. A new device has to be safe and reliable too. Anaesth. 2001;56:382–3.CrossRef
3.
go back to reference Agro F, Cataldo R, Alfano A, Galli B. A new prototype for airway management in an emergency: the Laryngeal Tube. Resuscitation. 1999;41:284–6.CrossRefPubMed Agro F, Cataldo R, Alfano A, Galli B. A new prototype for airway management in an emergency: the Laryngeal Tube. Resuscitation. 1999;41:284–6.CrossRefPubMed
4.
go back to reference Soar J, Nolan JP, Bottiger BW, et al. European Resuscitation Council Guidelines for Resuscitation 2015: Section 3. Adult advanced life support. Resuscitation. 2015;95:100–47.CrossRefPubMed Soar J, Nolan JP, Bottiger BW, et al. European Resuscitation Council Guidelines for Resuscitation 2015: Section 3. Adult advanced life support. Resuscitation. 2015;95:100–47.CrossRefPubMed
5.
go back to reference Ulrich-Pur H, Hrska F, Krafft P, et al. Comparison of mucosal pressures induced by cuffs of different airway devices. Anesthesiol. 2006;104:933–8.CrossRef Ulrich-Pur H, Hrska F, Krafft P, et al. Comparison of mucosal pressures induced by cuffs of different airway devices. Anesthesiol. 2006;104:933–8.CrossRef
6.
go back to reference Burgard G, Mollhoff T, Prien T. The effect of laryngeal mask cuff pressure on postoperative sore throat incidence. J Clin Anesth. 1996;8:198–201.CrossRefPubMed Burgard G, Mollhoff T, Prien T. The effect of laryngeal mask cuff pressure on postoperative sore throat incidence. J Clin Anesth. 1996;8:198–201.CrossRefPubMed
7.
go back to reference Nott MR, Noble PD, Parmar M. Reducing the incidence of sore throat with the laryngeal mask airway. Eur J Anaesthesiol. 1998;15:153–7.CrossRefPubMed Nott MR, Noble PD, Parmar M. Reducing the incidence of sore throat with the laryngeal mask airway. Eur J Anaesthesiol. 1998;15:153–7.CrossRefPubMed
8.
go back to reference Keller C, Brimacombe J, Boehler M, et al. The influence of cuff volume and anatomic location on pharyngeal, esophageal, and tracheal mucosal pressures with the esophageal tracheal combitube. Anesthesiol. 2002;96:1074–7.CrossRef Keller C, Brimacombe J, Boehler M, et al. The influence of cuff volume and anatomic location on pharyngeal, esophageal, and tracheal mucosal pressures with the esophageal tracheal combitube. Anesthesiol. 2002;96:1074–7.CrossRef
9.
go back to reference Keller C, Brimacombe J, Kleinsasser A, et al. Pharyngeal mucosal pressures with the laryngeal tube airway versus ProSeal laryngeal mask airway. Anasthesiol Intensivmed Notfallmed Schmerzther. 2003;38:393–6.CrossRefPubMed Keller C, Brimacombe J, Kleinsasser A, et al. Pharyngeal mucosal pressures with the laryngeal tube airway versus ProSeal laryngeal mask airway. Anasthesiol Intensivmed Notfallmed Schmerzther. 2003;38:393–6.CrossRefPubMed
10.
go back to reference Brimacombe J, Keller C, Puhringer F. Pharyngeal mucosal pressure and perfusion: a fiberoptic evaluation of the posterior pharynx in anesthetized adult patients with a modified cuffed oropharyngeal airway. Anesthesiol. 1999;91:1661–5.CrossRef Brimacombe J, Keller C, Puhringer F. Pharyngeal mucosal pressure and perfusion: a fiberoptic evaluation of the posterior pharynx in anesthetized adult patients with a modified cuffed oropharyngeal airway. Anesthesiol. 1999;91:1661–5.CrossRef
11.
go back to reference Eschertzhuber S, Brimacombe J, Kaufmann M, et al. Directly measured mucosal pressures produced by the i-gel and Laryngeal Mask Airway Supreme in paralysed anaesthetised patients. Anaesth. 2012;67:407–10.CrossRef Eschertzhuber S, Brimacombe J, Kaufmann M, et al. Directly measured mucosal pressures produced by the i-gel and Laryngeal Mask Airway Supreme in paralysed anaesthetised patients. Anaesth. 2012;67:407–10.CrossRef
12.
go back to reference Licina A, Chambers NA, Hullett B, et al. Lower cuff pressures improve the seal of pediatric laryngeal mask airways. Paediatr Anaesth. 2008;18:952–6.CrossRefPubMed Licina A, Chambers NA, Hullett B, et al. Lower cuff pressures improve the seal of pediatric laryngeal mask airways. Paediatr Anaesth. 2008;18:952–6.CrossRefPubMed
13.
go back to reference Mihai R, Knottenbelt G, Cook TM. Evaluation of the revised laryngeal tube suction: the laryngeal tube suction II in 100 patients. Br J Anaesth. 2007;99:734–9.CrossRefPubMed Mihai R, Knottenbelt G, Cook TM. Evaluation of the revised laryngeal tube suction: the laryngeal tube suction II in 100 patients. Br J Anaesth. 2007;99:734–9.CrossRefPubMed
14.
go back to reference Russo SG, Cremer S, Galli T, Eich C, et al. Randomized comparison of the i-gel, the LMA Supreme, and the Laryngeal Tube Suction-D using clinical and fibreoptic assessments in elective patients. BMC Anesthesiol. 2012;12:18.CrossRefPubMedPubMedCentral Russo SG, Cremer S, Galli T, Eich C, et al. Randomized comparison of the i-gel, the LMA Supreme, and the Laryngeal Tube Suction-D using clinical and fibreoptic assessments in elective patients. BMC Anesthesiol. 2012;12:18.CrossRefPubMedPubMedCentral
15.
go back to reference Thee C, Serocki G, Doerges V, et al. Laryngeal tube S II, laryngeal tube S disposable, Fastrach laryngeal mask and Fastrach laryngeal mask disposable during elective surgery: a randomized controlled comparison between reusable and disposable supraglottic airway devices. Eur J Anaesthesiol. 2010;27:468–72.CrossRefPubMed Thee C, Serocki G, Doerges V, et al. Laryngeal tube S II, laryngeal tube S disposable, Fastrach laryngeal mask and Fastrach laryngeal mask disposable during elective surgery: a randomized controlled comparison between reusable and disposable supraglottic airway devices. Eur J Anaesthesiol. 2010;27:468–72.CrossRefPubMed
16.
go back to reference Genzwuerker HV, Altmayer S, Hinkelbein J, et al. Prospective randomized comparison of the new Laryngeal Tube Suction LTS II and the LMA-ProSeal for elective surgical interventions. Acta Anaesthesiol Scand. 2007;51:1373–7.CrossRefPubMed Genzwuerker HV, Altmayer S, Hinkelbein J, et al. Prospective randomized comparison of the new Laryngeal Tube Suction LTS II and the LMA-ProSeal for elective surgical interventions. Acta Anaesthesiol Scand. 2007;51:1373–7.CrossRefPubMed
17.
go back to reference Seet E, Yousaf F, Gupta S, et al. Use of manometry for laryngeal mask airway reduces postoperative pharyngolaryngeal adverse events: a prospective, randomized trial. Anesthesiol. 2010;112:652–7.CrossRef Seet E, Yousaf F, Gupta S, et al. Use of manometry for laryngeal mask airway reduces postoperative pharyngolaryngeal adverse events: a prospective, randomized trial. Anesthesiol. 2010;112:652–7.CrossRef
18.
go back to reference Esa K, Azarinah I, Muhammad M, et al. A comparison between Laryngeal Tube Suction II Airway and Proseal Laryngeal Mask Airway in laparascopic surgery. Med J Malays. 2011;66:182–6. Esa K, Azarinah I, Muhammad M, et al. A comparison between Laryngeal Tube Suction II Airway and Proseal Laryngeal Mask Airway in laparascopic surgery. Med J Malays. 2011;66:182–6.
19.
go back to reference Cavus E, Deitmer W, Francksen H, et al. Laryngeal tube S II, ProSeal laryngeal mask, and EasyTube during elective surgery: a randomized controlled comparison with the endotracheal tube in nontrained professionals. Eur J Anaesthesiol. 2009;26:730–5.CrossRefPubMed Cavus E, Deitmer W, Francksen H, et al. Laryngeal tube S II, ProSeal laryngeal mask, and EasyTube during elective surgery: a randomized controlled comparison with the endotracheal tube in nontrained professionals. Eur J Anaesthesiol. 2009;26:730–5.CrossRefPubMed
20.
go back to reference Kette F, Reffo I, Giordani G, et al. The use of laryngeal tube by nurses in out-of-hospital emergencies: preliminary experience. Resuscitation. 2005;66:21–5.CrossRefPubMed Kette F, Reffo I, Giordani G, et al. The use of laryngeal tube by nurses in out-of-hospital emergencies: preliminary experience. Resuscitation. 2005;66:21–5.CrossRefPubMed
21.
go back to reference Heuer JF, Barwing J, Eich C, et al. Initial ventilation through laryngeal tube instead of face mask in out-of-hospital cardiopulmonary arrest is effective and safe. Eur J Emerg Med. 2010;17:10–5.CrossRefPubMed Heuer JF, Barwing J, Eich C, et al. Initial ventilation through laryngeal tube instead of face mask in out-of-hospital cardiopulmonary arrest is effective and safe. Eur J Emerg Med. 2010;17:10–5.CrossRefPubMed
22.
go back to reference Maignan M, Koch FX, Kraemer M, et al. Impact of laryngeal tube use on chest compression fraction during out-of-hospital cardiac arrest. A prospective alternate month study. Resuscitation. 2015;93:113–7.CrossRefPubMed Maignan M, Koch FX, Kraemer M, et al. Impact of laryngeal tube use on chest compression fraction during out-of-hospital cardiac arrest. A prospective alternate month study. Resuscitation. 2015;93:113–7.CrossRefPubMed
23.
go back to reference Townsend R, Brimacombe J, Keller C, et al. Jaw thrust as a predictor of insertion conditions for the proseal laryngeal mask airway. Middle East J Anaesthesiol. 2009;20:59–62.PubMed Townsend R, Brimacombe J, Keller C, et al. Jaw thrust as a predictor of insertion conditions for the proseal laryngeal mask airway. Middle East J Anaesthesiol. 2009;20:59–62.PubMed
24.
go back to reference Kurola J, Paakkonen H, Kettunen T, et al. Feasibility of written instructions in airway management training of laryngeal tube. Scand J Trauma Resusc Emerg Med. 2011;19:56.CrossRefPubMedPubMedCentral Kurola J, Paakkonen H, Kettunen T, et al. Feasibility of written instructions in airway management training of laryngeal tube. Scand J Trauma Resusc Emerg Med. 2011;19:56.CrossRefPubMedPubMedCentral
25.
go back to reference Langlois C, Pean D, Testa S, et al. The LTD laryngeal tube: a new single-use airway device. Ann Fr Anesth Reanim. 2007;26:197–201.CrossRefPubMed Langlois C, Pean D, Testa S, et al. The LTD laryngeal tube: a new single-use airway device. Ann Fr Anesth Reanim. 2007;26:197–201.CrossRefPubMed
26.
go back to reference Zhang L, Seet E, Mehta V, et al. Oropharyngeal leak pressure with the laryngeal mask airway Supreme at different intracuff pressures: a randomized controlled trial. Can J Anaesth. 2011;58:624–9.CrossRefPubMed Zhang L, Seet E, Mehta V, et al. Oropharyngeal leak pressure with the laryngeal mask airway Supreme at different intracuff pressures: a randomized controlled trial. Can J Anaesth. 2011;58:624–9.CrossRefPubMed
27.
go back to reference Wong JG, Heaney M, Chambers NA, et al. Impact of laryngeal mask airway cuff pressures on the incidence of sore throat in children. Paediatr Anaesth. 2009;19:464–9.CrossRefPubMed Wong JG, Heaney M, Chambers NA, et al. Impact of laryngeal mask airway cuff pressures on the incidence of sore throat in children. Paediatr Anaesth. 2009;19:464–9.CrossRefPubMed
28.
go back to reference Cattano D, Ferrario L, Patel CB, et al. Laryngeal Tube Suction-D, Combitube and Proseal Laryngeal Mask Airway: randomized clinical trial. J Anesthesiol Clin Sci. 2012;2049–9752:1–8. Cattano D, Ferrario L, Patel CB, et al. Laryngeal Tube Suction-D, Combitube and Proseal Laryngeal Mask Airway: randomized clinical trial. J Anesthesiol Clin Sci. 2012;2049–9752:1–8.
Metadata
Title
Evaluation of the optimal cuff volume and cuff pressure of the revised laryngeal tube “LTS-D” in surgical patients
Authors
Marc Kriege
Christian Alflen
Johannes Eisel
Thomas Ott
Tim Piepho
Ruediger R. Noppens
Publication date
01-12-2017
Publisher
BioMed Central
Published in
BMC Anesthesiology / Issue 1/2017
Electronic ISSN: 1471-2253
DOI
https://doi.org/10.1186/s12871-017-0308-4

Other articles of this Issue 1/2017

BMC Anesthesiology 1/2017 Go to the issue