Skip to main content
Top
Published in: Critical Care 1/2010

Open Access 01-02-2010 | Research

PEEP titration guided by ventilation homogeneity: a feasibility study using electrical impedance tomography

Authors: Zhanqi Zhao, Daniel Steinmann, Inéz Frerichs, Josef Guttmann, Knut Möller

Published in: Critical Care | Issue 1/2010

Login to get access

Abstract

Introduction

Lung protective ventilation requires low tidal volume and suitable positive end-expiratory pressure (PEEP). To date, few methods have been accepted for clinical use to set the appropriate PEEP. The aim of this study was to test the feasibility of PEEP titration guided by ventilation homogeneity using the global inhomogeneity (GI) index based on electrical impedance tomography (EIT) images.

Methods

In a retrospective study, 10 anesthetized patients with healthy lungs mechanically ventilated under volume-controlled mode were investigated. Ventilation distribution was monitored by EIT. A standardized incremental PEEP trial (PEEP from 0 to 28 mbar, 2 mbar per step) was conducted. During the PEEP trial, "optimal" PEEP level for each patient was determined when the air was most homogeneously distributed in the lung, indicated by the lowest GI index value. Two published methods for setting PEEP were included for comparison based on the maximum global dynamic compliance and the intra-tidal compliance-volume curve.

Results

No significant differences in the results were observed between the GI index method (12.2 ± 4.6 mbar) and the dynamic compliance method (11.4 ± 2.3 mbar, P > 0.6), or between the GI index and the compliance-volume curve method (12.2 ± 4.9 mbar, P > 0.6).

Conclusions

According to the results, it is feasible and reasonable to use the GI index to select the PEEP level with respect to ventilation homogeneity. The GI index may provide new insights into the relationship between lung mechanics and tidal volume distribution and may be used to guide ventilator settings.
Appendix
Available only for authorised users
Literature
1.
go back to reference Rouby JJ, Lu Q, Goldstein I: Selecting the right level of positive end-expiratory pressure in patients with acute respiratory distress syndrome. Am J Respir Crit Care Med. 2002, 165: 1182-1186.CrossRefPubMed Rouby JJ, Lu Q, Goldstein I: Selecting the right level of positive end-expiratory pressure in patients with acute respiratory distress syndrome. Am J Respir Crit Care Med. 2002, 165: 1182-1186.CrossRefPubMed
2.
go back to reference Victorino JA, Borges JB, Okamoto VN, Matos GF, Tucci MR, Caramez MP, Tanaka H, Sipmann FS, Santos DC, Barbas CS, Carvalho CR, Amato MB: Imbalances in regional lung ventilation: a validation study on electrical impedance tomography. Am J Respir Crit Care Med. 2004, 169: 791-800. 10.1164/rccm.200301-133OC.CrossRefPubMed Victorino JA, Borges JB, Okamoto VN, Matos GF, Tucci MR, Caramez MP, Tanaka H, Sipmann FS, Santos DC, Barbas CS, Carvalho CR, Amato MB: Imbalances in regional lung ventilation: a validation study on electrical impedance tomography. Am J Respir Crit Care Med. 2004, 169: 791-800. 10.1164/rccm.200301-133OC.CrossRefPubMed
3.
go back to reference Putensen C, Wrigge H, Zinserling J: Electrical impedance tomography guided ventilation therapy. Curr Opin Crit Care. 2007, 13: 344-350. 10.1097/MCC.0b013e328136c1e2.CrossRefPubMed Putensen C, Wrigge H, Zinserling J: Electrical impedance tomography guided ventilation therapy. Curr Opin Crit Care. 2007, 13: 344-350. 10.1097/MCC.0b013e328136c1e2.CrossRefPubMed
4.
go back to reference Gattinoni L, Caironi P, Valenza F, Carlesso E: The role of CT-scan studies for the diagnosis and therapy of acute respiratory distress syndrome. Clin Chest Med. 2006, 27: 559-570. 10.1016/j.ccm.2006.06.002.CrossRefPubMed Gattinoni L, Caironi P, Valenza F, Carlesso E: The role of CT-scan studies for the diagnosis and therapy of acute respiratory distress syndrome. Clin Chest Med. 2006, 27: 559-570. 10.1016/j.ccm.2006.06.002.CrossRefPubMed
5.
go back to reference Frerichs I, Hinz J, Herrmann P, Weisser G, Hahn G, Dudykevych T, Quintel M, Hellige G: Detection of local lung air content by electrical impedance tomography compared with electron beam CT. J Appl Physiol. 2002, 93: 660-666.CrossRefPubMed Frerichs I, Hinz J, Herrmann P, Weisser G, Hahn G, Dudykevych T, Quintel M, Hellige G: Detection of local lung air content by electrical impedance tomography compared with electron beam CT. J Appl Physiol. 2002, 93: 660-666.CrossRefPubMed
6.
go back to reference Hinz J, Neumann P, Dudykevych T, Andersson LG, Wrigge H, Burchardi H, Hedenstierna G: Regional ventilation by electrical impedance tomography: a comparison with ventilation scintigraphy in pigs. Chest. 2003, 124: 314-322. 10.1378/chest.124.1.314.CrossRefPubMed Hinz J, Neumann P, Dudykevych T, Andersson LG, Wrigge H, Burchardi H, Hedenstierna G: Regional ventilation by electrical impedance tomography: a comparison with ventilation scintigraphy in pigs. Chest. 2003, 124: 314-322. 10.1378/chest.124.1.314.CrossRefPubMed
7.
go back to reference Marquis F, Coulombe N, Costa R, Gagnon H, Guardo R, Skrobik Y: Electrical impedance tomography's correlation to lung volume is not influenced by anthropometric parameters. J Clin Monit Comput. 2006, 20: 201-207. 10.1007/s10877-006-9021-4.CrossRefPubMed Marquis F, Coulombe N, Costa R, Gagnon H, Guardo R, Skrobik Y: Electrical impedance tomography's correlation to lung volume is not influenced by anthropometric parameters. J Clin Monit Comput. 2006, 20: 201-207. 10.1007/s10877-006-9021-4.CrossRefPubMed
8.
go back to reference Erlandsson K, Odenstedt H, Lundin S, Stenqvist O: Positive end-expiratory pressure optimization using electric impedance tomography in morbidly obese patients during laparoscopic gastric bypass surgery. Acta Anaesthesiol Scand. 2006, 50: 833-839. 10.1111/j.1399-6576.2006.01079.x.CrossRefPubMed Erlandsson K, Odenstedt H, Lundin S, Stenqvist O: Positive end-expiratory pressure optimization using electric impedance tomography in morbidly obese patients during laparoscopic gastric bypass surgery. Acta Anaesthesiol Scand. 2006, 50: 833-839. 10.1111/j.1399-6576.2006.01079.x.CrossRefPubMed
9.
go back to reference Luepschen H, Meier T, Grossherr M, Leibecke T, Karsten J, Leonhardt S: Protective ventilation using electrical impedance tomography. Physiol Meas. 2007, 28: S247-260. 10.1088/0967-3334/28/7/S18.CrossRefPubMed Luepschen H, Meier T, Grossherr M, Leibecke T, Karsten J, Leonhardt S: Protective ventilation using electrical impedance tomography. Physiol Meas. 2007, 28: S247-260. 10.1088/0967-3334/28/7/S18.CrossRefPubMed
10.
go back to reference Zhao Z, Moller K, Steinmann D, Frerichs I, Guttmann J: Evaluation of an electrical impedance tomography-based global inhomogeneity index for pulmonary ventilation distribution. Intensive Care Med. 2009, 35: 1900-1906. 10.1007/s00134-009-1589-y.CrossRefPubMed Zhao Z, Moller K, Steinmann D, Frerichs I, Guttmann J: Evaluation of an electrical impedance tomography-based global inhomogeneity index for pulmonary ventilation distribution. Intensive Care Med. 2009, 35: 1900-1906. 10.1007/s00134-009-1589-y.CrossRefPubMed
11.
go back to reference Suarez-Sipmann F, Bohm SH, Tusman G, Pesch T, Thamm O, Reissmann H, Reske A, Magnusson A, Hedenstierna G: Use of dynamic compliance for open lung positive end-expiratory pressure titration in an experimental study. Crit Care Med. 2007, 35: 214-221. 10.1097/01.CCM.0000251131.40301.E2.CrossRefPubMed Suarez-Sipmann F, Bohm SH, Tusman G, Pesch T, Thamm O, Reissmann H, Reske A, Magnusson A, Hedenstierna G: Use of dynamic compliance for open lung positive end-expiratory pressure titration in an experimental study. Crit Care Med. 2007, 35: 214-221. 10.1097/01.CCM.0000251131.40301.E2.CrossRefPubMed
12.
go back to reference Mols G, Brandes I, Kessler V, Lichtwarck-Aschoff M, Loop T, Geiger K, Guttmann J: Volume-dependent compliance in ARDS: proposal of a new diagnostic concept. Intensive Care Med. 1999, 25: 1084-1091. 10.1007/s001340051016.CrossRefPubMed Mols G, Brandes I, Kessler V, Lichtwarck-Aschoff M, Loop T, Geiger K, Guttmann J: Volume-dependent compliance in ARDS: proposal of a new diagnostic concept. Intensive Care Med. 1999, 25: 1084-1091. 10.1007/s001340051016.CrossRefPubMed
13.
go back to reference Stahl CA, Moller K, Schumann S, Kuhlen R, Sydow M, Putensen C, Guttmann J: Dynamic versus static respiratory mechanics in acute lung injury and acute respiratory distress syndrome. Crit Care Med. 2006, 34: 2090-2098. 10.1097/01.CCM.0000227220.67613.0D.CrossRefPubMed Stahl CA, Moller K, Schumann S, Kuhlen R, Sydow M, Putensen C, Guttmann J: Dynamic versus static respiratory mechanics in acute lung injury and acute respiratory distress syndrome. Crit Care Med. 2006, 34: 2090-2098. 10.1097/01.CCM.0000227220.67613.0D.CrossRefPubMed
14.
go back to reference Zhao Z, Möller K, Steinmann D, Guttmann J: Determination of lung area in EIT images. Proc 3rd International Conference on Bioinformatics and Biomedical Engineering. 2009, Beijing, China: IEEE, 1-4. Zhao Z, Möller K, Steinmann D, Guttmann J: Determination of lung area in EIT images. Proc 3rd International Conference on Bioinformatics and Biomedical Engineering. 2009, Beijing, China: IEEE, 1-4.
15.
go back to reference Hahn G, Frerichs I, Kleyer M, Hellige G: Local mechanics of the lung tissue determined by functional EIT. Physiol Meas. 1996, 17 (Suppl 4A): A159-166. 10.1088/0967-3334/17/4A/020.CrossRefPubMed Hahn G, Frerichs I, Kleyer M, Hellige G: Local mechanics of the lung tissue determined by functional EIT. Physiol Meas. 1996, 17 (Suppl 4A): A159-166. 10.1088/0967-3334/17/4A/020.CrossRefPubMed
16.
go back to reference Frerichs I, Hahn G, Golisch W, Kurpitz M, Burchardi H, Hellige G: Monitoring perioperative changes in distribution of pulmonary ventilation by functional electrical impedance tomography. Acta Anaesthesiol Scand. 1998, 42: 721-726. 10.1111/j.1399-6576.1998.tb05308.x.CrossRefPubMed Frerichs I, Hahn G, Golisch W, Kurpitz M, Burchardi H, Hellige G: Monitoring perioperative changes in distribution of pulmonary ventilation by functional electrical impedance tomography. Acta Anaesthesiol Scand. 1998, 42: 721-726. 10.1111/j.1399-6576.1998.tb05308.x.CrossRefPubMed
17.
go back to reference Pulletz S, van Genderingen HR, Schmitz G, Zick G, Schadler D, Scholz J, Weiler N, Frerichs I: Comparison of different methods to define regions of interest for evaluation of regional lung ventilation by EIT. Physiol Meas. 2006, 27: S115-127. 10.1088/0967-3334/27/5/S10.CrossRefPubMed Pulletz S, van Genderingen HR, Schmitz G, Zick G, Schadler D, Scholz J, Weiler N, Frerichs I: Comparison of different methods to define regions of interest for evaluation of regional lung ventilation by EIT. Physiol Meas. 2006, 27: S115-127. 10.1088/0967-3334/27/5/S10.CrossRefPubMed
18.
go back to reference LaFollette R, Hojnowski K, Norton J, DiRocco J, Carney D, Nieman G: Using pressure-volume curves to set proper PEEP in acute lung injury. Nurs Crit Care. 2007, 12: 231-241. 10.1111/j.1478-5153.2007.00224.x.CrossRefPubMed LaFollette R, Hojnowski K, Norton J, DiRocco J, Carney D, Nieman G: Using pressure-volume curves to set proper PEEP in acute lung injury. Nurs Crit Care. 2007, 12: 231-241. 10.1111/j.1478-5153.2007.00224.x.CrossRefPubMed
19.
go back to reference Caramez MP, Kacmarek RM, Helmy M, Miyoshi E, Malhotra A, Amato MB, Harris RS: A comparison of methods to identify open-lung PEEP. Intensive Care Med. 2009, 35: 740-747. 10.1007/s00134-009-1412-9.PubMedCentralCrossRefPubMed Caramez MP, Kacmarek RM, Helmy M, Miyoshi E, Malhotra A, Amato MB, Harris RS: A comparison of methods to identify open-lung PEEP. Intensive Care Med. 2009, 35: 740-747. 10.1007/s00134-009-1412-9.PubMedCentralCrossRefPubMed
20.
go back to reference Iotti GA, Braschi A, Brunner JX, Smits T, Olivei M, Palo A, Veronesi R: Respiratory mechanics by least squares fitting in mechanically ventilated patients: applications during paralysis and during pressure support ventilation. Intensive Care Med. 1995, 21: 406-413. 10.1007/BF01707409.CrossRefPubMed Iotti GA, Braschi A, Brunner JX, Smits T, Olivei M, Palo A, Veronesi R: Respiratory mechanics by least squares fitting in mechanically ventilated patients: applications during paralysis and during pressure support ventilation. Intensive Care Med. 1995, 21: 406-413. 10.1007/BF01707409.CrossRefPubMed
21.
go back to reference Guttmann J, Eberhard L, Fabry B, Zappe D, Bernhard H, Lichtwarck-Aschoff M, Adolph M, Wolff G: Determination of volume-dependent respiratory system mechanics in mechanically ventilated patients using the new SLICE method. Technol Health Care. 1994, 2: 175-191.PubMed Guttmann J, Eberhard L, Fabry B, Zappe D, Bernhard H, Lichtwarck-Aschoff M, Adolph M, Wolff G: Determination of volume-dependent respiratory system mechanics in mechanically ventilated patients using the new SLICE method. Technol Health Care. 1994, 2: 175-191.PubMed
22.
go back to reference Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986, 1: 307-310.CrossRefPubMed Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986, 1: 307-310.CrossRefPubMed
23.
go back to reference Takeuchi M, Goddon S, Dolhnikoff M, Shimaoka M, Hess D, Amato MB, Kacmarek RM: Set positive end-expiratory pressure during protective ventilation affects lung injury. Anesthesiology. 2002, 97: 682-692. 10.1097/00000542-200209000-00023.CrossRefPubMed Takeuchi M, Goddon S, Dolhnikoff M, Shimaoka M, Hess D, Amato MB, Kacmarek RM: Set positive end-expiratory pressure during protective ventilation affects lung injury. Anesthesiology. 2002, 97: 682-692. 10.1097/00000542-200209000-00023.CrossRefPubMed
24.
go back to reference Harris RS, Hess DR, Venegas JG: An objective analysis of the pressure-volume curve in the acute respiratory distress syndrome. Am J Respir Crit Care Med. 2000, 161: 432-439.CrossRefPubMed Harris RS, Hess DR, Venegas JG: An objective analysis of the pressure-volume curve in the acute respiratory distress syndrome. Am J Respir Crit Care Med. 2000, 161: 432-439.CrossRefPubMed
25.
go back to reference Albaiceta GM, Taboada F, Parra D, Luyando LH, Calvo J, Menendez R, Otero J: Tomographic study of the inflection points of the pressure-volume curve in acute lung injury. Am J Respir Crit Care Med. 2004, 170: 1066-1072. 10.1164/rccm.200312-1644OC.CrossRefPubMed Albaiceta GM, Taboada F, Parra D, Luyando LH, Calvo J, Menendez R, Otero J: Tomographic study of the inflection points of the pressure-volume curve in acute lung injury. Am J Respir Crit Care Med. 2004, 170: 1066-1072. 10.1164/rccm.200312-1644OC.CrossRefPubMed
26.
go back to reference Tugrul S, Akinci O, Ozcan PE, Ince S, Esen F, Telci L, Akpir K, Cakar N: Effects of sustained inflation and postinflation positive end-expiratory pressure in acute respiratory distress syndrome: focusing on pulmonary and extrapulmonary forms. Crit Care Med. 2003, 31: 738-744. 10.1097/01.CCM.0000053554.76355.72.CrossRefPubMed Tugrul S, Akinci O, Ozcan PE, Ince S, Esen F, Telci L, Akpir K, Cakar N: Effects of sustained inflation and postinflation positive end-expiratory pressure in acute respiratory distress syndrome: focusing on pulmonary and extrapulmonary forms. Crit Care Med. 2003, 31: 738-744. 10.1097/01.CCM.0000053554.76355.72.CrossRefPubMed
27.
go back to reference Luecke T, Herrmann P, Kraincuk P, Pelosi P: Computed tomography scan assessment of lung volume and recruitment during high-frequency oscillatory ventilation. Crit Care Med. 2005, 33 (3 Suppl): S155-162. 10.1097/01.CCM.0000155916.47455.DF.CrossRefPubMed Luecke T, Herrmann P, Kraincuk P, Pelosi P: Computed tomography scan assessment of lung volume and recruitment during high-frequency oscillatory ventilation. Crit Care Med. 2005, 33 (3 Suppl): S155-162. 10.1097/01.CCM.0000155916.47455.DF.CrossRefPubMed
28.
go back to reference Borges JB, Okamoto VN, Matos GF, Caramez MP, Arantes PR, Barros F, Souza CE, Victorino JA, Kacmarek RM, Barbas CS, Carvalho CR, Amato MB: Reversibility of lung collapse and hypoxemia in early acute respiratory distress syndrome. Am J Respir Crit Care Med. 2006, 174: 268-278. 10.1164/rccm.200506-976OC.CrossRefPubMed Borges JB, Okamoto VN, Matos GF, Caramez MP, Arantes PR, Barros F, Souza CE, Victorino JA, Kacmarek RM, Barbas CS, Carvalho CR, Amato MB: Reversibility of lung collapse and hypoxemia in early acute respiratory distress syndrome. Am J Respir Crit Care Med. 2006, 174: 268-278. 10.1164/rccm.200506-976OC.CrossRefPubMed
29.
go back to reference Girgis K, Hamed H, Khater Y, Kacmarek RM: A decremental PEEP trial identifies the PEEP level that maintains oxygenation after lung recruitment. Respir Care. 2006, 51: 1132-1139.PubMed Girgis K, Hamed H, Khater Y, Kacmarek RM: A decremental PEEP trial identifies the PEEP level that maintains oxygenation after lung recruitment. Respir Care. 2006, 51: 1132-1139.PubMed
30.
go back to reference Grant CA, Fraser JF, Dunster KR, Schibler A: The assessment of regional lung mechanics with electrical impedance tomography: a pilot study during recruitment manoeuvres. Intensive Care Med. 2009, 35: 166-170. 10.1007/s00134-008-1324-0.CrossRefPubMed Grant CA, Fraser JF, Dunster KR, Schibler A: The assessment of regional lung mechanics with electrical impedance tomography: a pilot study during recruitment manoeuvres. Intensive Care Med. 2009, 35: 166-170. 10.1007/s00134-008-1324-0.CrossRefPubMed
31.
go back to reference Kunst PW, Bohm SH, Vazquez de Anda G, Amato MB, Lachmann B, Postmus PE, de Vries PM: Regional pressure volume curves by electrical impedance tomography in a model of acute lung injury. Crit Care Med. 2000, 28: 178-183. 10.1097/00003246-200001000-00029.CrossRefPubMed Kunst PW, Bohm SH, Vazquez de Anda G, Amato MB, Lachmann B, Postmus PE, de Vries PM: Regional pressure volume curves by electrical impedance tomography in a model of acute lung injury. Crit Care Med. 2000, 28: 178-183. 10.1097/00003246-200001000-00029.CrossRefPubMed
32.
go back to reference Meier T, Luepschen H, Karsten J, Leibecke T, Grossherr M, Gehring H, Leonhardt S: Assessment of regional lung recruitment and derecruitment during a PEEP trial based on electrical impedance tomography. Intensive Care Med. 2008, 34: 543-550. 10.1007/s00134-007-0786-9.CrossRefPubMed Meier T, Luepschen H, Karsten J, Leibecke T, Grossherr M, Gehring H, Leonhardt S: Assessment of regional lung recruitment and derecruitment during a PEEP trial based on electrical impedance tomography. Intensive Care Med. 2008, 34: 543-550. 10.1007/s00134-007-0786-9.CrossRefPubMed
33.
go back to reference Odenstedt H, Lindgren S, Olegard C, Erlandsson K, Lethvall S, Aneman A, Stenqvist O, Lundin S: Slow moderate pressure recruitment maneuver minimizes negative circulatory and lung mechanic side effects: evaluation of recruitment maneuvers using electric impedance tomography. Intensive Care Med. 2005, 31: 1706-1714. 10.1007/s00134-005-2799-6.CrossRefPubMed Odenstedt H, Lindgren S, Olegard C, Erlandsson K, Lethvall S, Aneman A, Stenqvist O, Lundin S: Slow moderate pressure recruitment maneuver minimizes negative circulatory and lung mechanic side effects: evaluation of recruitment maneuvers using electric impedance tomography. Intensive Care Med. 2005, 31: 1706-1714. 10.1007/s00134-005-2799-6.CrossRefPubMed
34.
go back to reference Frerichs I, Dargaville PA, van Genderingen H, Morel DR, Rimensberger PC: Lung volume recruitment after surfactant administration modifies spatial distribution of ventilation. Am J Respir Crit Care Med. 2006, 174: 772-779. 10.1164/rccm.200512-1942OC.CrossRefPubMed Frerichs I, Dargaville PA, van Genderingen H, Morel DR, Rimensberger PC: Lung volume recruitment after surfactant administration modifies spatial distribution of ventilation. Am J Respir Crit Care Med. 2006, 174: 772-779. 10.1164/rccm.200512-1942OC.CrossRefPubMed
Metadata
Title
PEEP titration guided by ventilation homogeneity: a feasibility study using electrical impedance tomography
Authors
Zhanqi Zhao
Daniel Steinmann
Inéz Frerichs
Josef Guttmann
Knut Möller
Publication date
01-02-2010
Publisher
BioMed Central
Published in
Critical Care / Issue 1/2010
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc8860

Other articles of this Issue 1/2010

Critical Care 1/2010 Go to the issue