Open Access 01-06-2009 | Research

Electrical impedance tomography compared to positron emission tomography for the measurement of regional lung ventilation: an experimental study

Authors: JC Richard, C Pouzot, A Gros, C Tourevieille, D Lebars, F Lavenne, I Frerichs, C Guérin

Published in: Critical Care | Issue 3/2009

Login to get access

Abstract

Introduction

Electrical impedance tomography (EIT), which can assess regional lung ventilation at the bedside, has never been compared with positron-emission tomography (PET), a gold-standard to quantify regional ventilation. This experiment systematically compared both techniques in injured and non-injured lungs.

Methods

The study was performed in six mechanically ventilated female piglets. In normal lungs, tidal volume (VT) was randomly changed to 6, 8, 10 and 15 ml/kg on zero end-expiratory pressure (ZEEP), then, at VT 10 ml/kg, positive end-expiratory pressure (PEEP) was randomly changed to 5, 10 and 15 cmH2O. Afterwards, acute lung injury (ALI) was subsequently created in three animals by injecting 3 ml/kg hydrochloric acid into the trachea. Then at PEEP 5 cmH2O, VT was randomly changed to 8 and 12 ml/kg and PEEP of 10 and 15 cmH2O applied at VT 10 ml/kg. EIT and PET examinations were performed simultaneously. EIT ventilation (VTEIT) and lung volume (VL) were measured in the anterior and posterior area of each lung. On the same regions of interest, ventilation (VPET) and aerated lung volume (VAatten) were determined with PET.

Results

On ZEEP, VTEIT and VPET significantly correlated for global (VTEIT = VPET - 2E-13, R2 = 0.95, P < 0.001) and regional (VTEIT = 0.81VPET+7.65, R2 = 0.63, P < 0.001) ventilation over both conditions. For ALI condition, corresponding R2 were 0.91 and 0.73 (P < 0.01). Bias was = 0 and limits of agreement were -37.42 and +37.42 ml/min for global ventilation over both conditions. These values were 0.04 and -29.01 and +29.08 ml/min, respectively, for regional ventilation. Significant correlations were also found between VL and VAatten for global (VL = VAatten+1E-12, R2 = 0.93, P < 0.0001) and regional (VL = 0.99VAatten+0.92, R2 = 0.65, P < 0.001) volume. For ALI condition, corresponding R2 were 0.94 (P < 0.001) and 0.54 (P < 0.05). Bias was = 0 and limits of agreement ranged -38.16 and +38.16 ml for global ventilation over both conditions. These values were -0.24 and -31.96 to +31.48 ml, respectively, for regional ventilation.

Conclusions

Regional lung ventilation and volume were accurately measured with EIT in healthy and injured lungs and validated by simultaneous PET imaging.
Appendix
Available only for authorised users
Literature
1.
go back to reference Frerichs I, Dargaville PA, Dudykevych T, Rimensberger PC: Electrical impedance tomography: a method for monitoring regional lung aeration and tidal volume distribution? Intensive Care Med 2003,29(12):2312-2316. 10.1007/s00134-003-2029-zCrossRefPubMed Frerichs I, Dargaville PA, Dudykevych T, Rimensberger PC: Electrical impedance tomography: a method for monitoring regional lung aeration and tidal volume distribution? Intensive Care Med 2003,29(12):2312-2316. 10.1007/s00134-003-2029-zCrossRefPubMed
2.
go back to reference Victorino JA, Borges JB, Okamoto VN, Matos GF, Tucci MR, Caramez MP, Tanaka H, Sipmann FS, Santos DC, Barbas CS, Carvalho CR, Amato MB: Imbalances in regional lung ventilation: a validation study on electrical impedance tomography. Am J Respir Crit Care Med 2004,169(7):791-800. 10.1164/rccm.200301-133OCCrossRefPubMed Victorino JA, Borges JB, Okamoto VN, Matos GF, Tucci MR, Caramez MP, Tanaka H, Sipmann FS, Santos DC, Barbas CS, Carvalho CR, Amato MB: Imbalances in regional lung ventilation: a validation study on electrical impedance tomography. Am J Respir Crit Care Med 2004,169(7):791-800. 10.1164/rccm.200301-133OCCrossRefPubMed
3.
go back to reference Hinz J, Neumann P, Dudykevych T, Andersson LG, Wrigge H, Burchardi H, Hedenstierna G: Regional ventilation by electrical impedance tomography: a comparison with ventilation scintigraphy in pigs. Chest 2003,124(1):314-322. 10.1378/chest.124.1.314CrossRefPubMed Hinz J, Neumann P, Dudykevych T, Andersson LG, Wrigge H, Burchardi H, Hedenstierna G: Regional ventilation by electrical impedance tomography: a comparison with ventilation scintigraphy in pigs. Chest 2003,124(1):314-322. 10.1378/chest.124.1.314CrossRefPubMed
4.
go back to reference Richard JC, Janier M, Lavenne F, Tourvieille C, Le Bars D, Costes N, Gimenez G, Guerin C: Quantitative assessment of regional alveolar ventilation and gas volume using 13N-N2 washout and PET. J Nucl Med 2005,46(8):1375-1383.PubMed Richard JC, Janier M, Lavenne F, Tourvieille C, Le Bars D, Costes N, Gimenez G, Guerin C: Quantitative assessment of regional alveolar ventilation and gas volume using 13N-N2 washout and PET. J Nucl Med 2005,46(8):1375-1383.PubMed
5.
go back to reference Richard JC, Le Bars D, Costes N, Bregeon F, Tourvieille C, Lavenne F, Janier M, Gimenez G, Guerin C: Alveolar recruitment assessed by positron emission tomography during experimental acute lung injury. Intensive Care Med 2006,32(11):1889-1894. 10.1007/s00134-006-0331-2CrossRefPubMed Richard JC, Le Bars D, Costes N, Bregeon F, Tourvieille C, Lavenne F, Janier M, Gimenez G, Guerin C: Alveolar recruitment assessed by positron emission tomography during experimental acute lung injury. Intensive Care Med 2006,32(11):1889-1894. 10.1007/s00134-006-0331-2CrossRefPubMed
6.
go back to reference Rabbani KS, Hassan M, Kiber A: 3D object localization using EIT measurements at two levels. Physiol Meas 1996,17(3):189-199. 10.1088/0967-3334/17/3/005CrossRefPubMed Rabbani KS, Hassan M, Kiber A: 3D object localization using EIT measurements at two levels. Physiol Meas 1996,17(3):189-199. 10.1088/0967-3334/17/3/005CrossRefPubMed
7.
go back to reference Barber DC: Quantification in impedance imaging. Clin Phys Physiol Meas 1990,11(Suppl A):45-56. 10.1088/0143-0815/11/4A/306CrossRefPubMed Barber DC: Quantification in impedance imaging. Clin Phys Physiol Meas 1990,11(Suppl A):45-56. 10.1088/0143-0815/11/4A/306CrossRefPubMed
8.
go back to reference Richard JC, Bregeon F, Costes N, Bars DL, Tourvieille C, Lavenne F, Janier M, Bourdin G, Gimenez G, Guerin C: Effects of prone position and positive end-expiratory pressure on lung perfusion and ventilation. Crit Care Med 2008,36(8):2373-2380. 10.1097/CCM.0b013e31818094a9CrossRefPubMed Richard JC, Bregeon F, Costes N, Bars DL, Tourvieille C, Lavenne F, Janier M, Bourdin G, Gimenez G, Guerin C: Effects of prone position and positive end-expiratory pressure on lung perfusion and ventilation. Crit Care Med 2008,36(8):2373-2380. 10.1097/CCM.0b013e31818094a9CrossRefPubMed
9.
go back to reference Meier T, Luepschen H, Karsten J, Leibecke T, Grossherr M, Gehring H, Leonhardt S: Assessment of regional lung recruitment and derecruitment during a PEEP trial based on electrical impedance tomography. Intensive Care Med 2008,34(3):543-550. 10.1007/s00134-007-0786-9CrossRefPubMed Meier T, Luepschen H, Karsten J, Leibecke T, Grossherr M, Gehring H, Leonhardt S: Assessment of regional lung recruitment and derecruitment during a PEEP trial based on electrical impedance tomography. Intensive Care Med 2008,34(3):543-550. 10.1007/s00134-007-0786-9CrossRefPubMed
10.
go back to reference Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clincial measurement. Lancet 1986, 1: 307-310.CrossRefPubMed Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clincial measurement. Lancet 1986, 1: 307-310.CrossRefPubMed
11.
go back to reference Frerichs I, Hinz J, Herrmann P, Weisser G, Hahn G, Dudykevych T, Quintel M, Hellige G: Detection of local lung air content by electrical impedance tomography compared with electron beam CT. J Appl Physiol 2002,93(2):660-666.CrossRefPubMed Frerichs I, Hinz J, Herrmann P, Weisser G, Hahn G, Dudykevych T, Quintel M, Hellige G: Detection of local lung air content by electrical impedance tomography compared with electron beam CT. J Appl Physiol 2002,93(2):660-666.CrossRefPubMed
12.
go back to reference Wrigge H, Zinserling J, Muders T, Varelmann D, Gunther U, Groeben C, Magnusson A, Hedenstierna G, Putensen C: Electrical impedance tomography compared with thoracic computed tomography during a slow inflation maneuver in experimental models of lung injury. Crit Care Med 2008,36(3):903-909. 10.1097/CCM.0B013E3181652EDDCrossRefPubMed Wrigge H, Zinserling J, Muders T, Varelmann D, Gunther U, Groeben C, Magnusson A, Hedenstierna G, Putensen C: Electrical impedance tomography compared with thoracic computed tomography during a slow inflation maneuver in experimental models of lung injury. Crit Care Med 2008,36(3):903-909. 10.1097/CCM.0B013E3181652EDDCrossRefPubMed
13.
go back to reference Gattinoni L, Caironi P, Cressoni M, Chiumello D, Ranieri VM, Quintel M, Russo S, Patroniti N, Cornejo R, Bugedo G: Lung recruitment in patients with the acute respiratory distress syndrome. N Engl J Med 2006,354(17):1775-1786. 10.1056/NEJMoa052052CrossRefPubMed Gattinoni L, Caironi P, Cressoni M, Chiumello D, Ranieri VM, Quintel M, Russo S, Patroniti N, Cornejo R, Bugedo G: Lung recruitment in patients with the acute respiratory distress syndrome. N Engl J Med 2006,354(17):1775-1786. 10.1056/NEJMoa052052CrossRefPubMed
14.
go back to reference Cressoni M, Caironi P, Polli F, Carlesso E, Chiumello D, Cadringher P, Quintel M, Ranieri VM, Bugedo G, Gattinoni L: Anatomical and functional intrapulmonary shunt in acute respiratory distress syndrome. Crit Care Med 2008, 36: 669-675. 10.1097/01.CCM.0000300276.12074.E1CrossRefPubMed Cressoni M, Caironi P, Polli F, Carlesso E, Chiumello D, Cadringher P, Quintel M, Ranieri VM, Bugedo G, Gattinoni L: Anatomical and functional intrapulmonary shunt in acute respiratory distress syndrome. Crit Care Med 2008, 36: 669-675. 10.1097/01.CCM.0000300276.12074.E1CrossRefPubMed
15.
go back to reference Frerichs I, Hinz J, Herrmann P, Weisser G, Hahn G, Quintel M, Hellige G: Regional lung perfusion as determined by electrical impedance tomography in comparison with electron beam CT imaging. IEEE Trans Med Imaging 2002,21(6):646-652. 10.1109/TMI.2002.800585CrossRefPubMed Frerichs I, Hinz J, Herrmann P, Weisser G, Hahn G, Quintel M, Hellige G: Regional lung perfusion as determined by electrical impedance tomography in comparison with electron beam CT imaging. IEEE Trans Med Imaging 2002,21(6):646-652. 10.1109/TMI.2002.800585CrossRefPubMed
Metadata
Title
Electrical impedance tomography compared to positron emission tomography for the measurement of regional lung ventilation: an experimental study
Authors
JC Richard
C Pouzot
A Gros
C Tourevieille
D Lebars
F Lavenne
I Frerichs
C Guérin
Publication date
01-06-2009
Publisher
BioMed Central
Published in
Critical Care / Issue 3/2009
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc7900

Other articles of this Issue 3/2009

Critical Care 3/2009 Go to the issue