Bronchodilator delivery with metered-dose inhaler during mechanical ventilation

Authors: Dimitris Georgopoulos, Eleni Mouloudi, Eumorfia Kondili, Maria Klimathianaki

Published in: Critical Care | Issue 4/2000

Login to get access

Abstract

The delivery of bronchodilators with metered-dose inhaler (MDI) in mechanically ventilated patients has attracted considerable interest in recent years. This is because the use of the MDI has several advantages over the nebulizer, such as reduced cost, ease of administration, less personnel time, reliability of dosing and a lower risk of contamination. A spacer device is fundamental in order to demonstrate the efficacy of the bronchodilatory therapy delivered by MDI. Provided that the technique of administration is appropriate, MDIs are as effective as nebulizers, despite a significantly lower dose of bronchodilator given by the MDI.
Literature
1.
go back to reference Boucher BA, Kuhl DA, Coffey BC, Fabian TC: Drug use in a trauma intensive care unit. Am J Hosp Pharm 1990, 47: 805-810.PubMed Boucher BA, Kuhl DA, Coffey BC, Fabian TC: Drug use in a trauma intensive care unit. Am J Hosp Pharm 1990, 47: 805-810.PubMed
2.
go back to reference Dhand R, Tobin MJ: Inhaled bronchodilator therapy in mechanically ventilated patients. Am J Respir Crit Care Med 1997, 156: 3-10.CrossRefPubMed Dhand R, Tobin MJ: Inhaled bronchodilator therapy in mechanically ventilated patients. Am J Respir Crit Care Med 1997, 156: 3-10.CrossRefPubMed
3.
go back to reference Dhand R, Tobin MJ: Bronhodilator delivery with metered-dose inhalers in mechanically-ventilated patients. Eur Respir J 1996, 9: 585-595. 10.1183/09031936.96.09030585CrossRefPubMed Dhand R, Tobin MJ: Bronhodilator delivery with metered-dose inhalers in mechanically-ventilated patients. Eur Respir J 1996, 9: 585-595. 10.1183/09031936.96.09030585CrossRefPubMed
4.
go back to reference Wright PE, Carmichael LC, Bernard GR: Effect of bronchodilators on lung mechanics in the acute respiratory distress syndrome (ARDS). Chest 1994, 106: 1517-1523.CrossRefPubMed Wright PE, Carmichael LC, Bernard GR: Effect of bronchodilators on lung mechanics in the acute respiratory distress syndrome (ARDS). Chest 1994, 106: 1517-1523.CrossRefPubMed
5.
go back to reference Georgopoulos D, Brochard L: Ventilatory strategies in acute exacerbations of chronic obstructive pulmonary disease. Eur Respir Monit 1998, 8: 12-44. Georgopoulos D, Brochard L: Ventilatory strategies in acute exacerbations of chronic obstructive pulmonary disease. Eur Respir Monit 1998, 8: 12-44.
6.
go back to reference Georgopoulos D, Burchardi H: Ventilatory strategies in adult patients with status asthmaticus. Eur Respir Monit 1998, 8: 45-83. Georgopoulos D, Burchardi H: Ventilatory strategies in adult patients with status asthmaticus. Eur Respir Monit 1998, 8: 45-83.
7.
go back to reference Summer W, Elston R, Tharpe L, Nelson S, Haponic EF: Aerosol bronchodilator delivery methods: relative impact on pulmonary function and cost of respiratory care. Arch Intern Mer 1989, 149: 618-623. 10.1001/archinte.149.3.618CrossRef Summer W, Elston R, Tharpe L, Nelson S, Haponic EF: Aerosol bronchodilator delivery methods: relative impact on pulmonary function and cost of respiratory care. Arch Intern Mer 1989, 149: 618-623. 10.1001/archinte.149.3.618CrossRef
8.
go back to reference Bowton DL, Goldsmoth WM, Haponic EF: Substitution of metered-dose inhalers for hand nebulizers: success and cost-saving in a large, acute care hospital. Chest 1998, 101: 305-308.CrossRef Bowton DL, Goldsmoth WM, Haponic EF: Substitution of metered-dose inhalers for hand nebulizers: success and cost-saving in a large, acute care hospital. Chest 1998, 101: 305-308.CrossRef
9.
go back to reference Alvine GF, Rodgers P, Fitzsimmons KM, Ahrens RC: Disposable jet nebulizers; how reliable are they? Chest 1992, 101: 316-319.CrossRefPubMed Alvine GF, Rodgers P, Fitzsimmons KM, Ahrens RC: Disposable jet nebulizers; how reliable are they? Chest 1992, 101: 316-319.CrossRefPubMed
10.
go back to reference Hamill RJ, Houston ED, Georghiu PR, Wright CE, Koza MA, Cadle RM, Goeptert PA, Lewis DA, Zenon GJ, Clarridge JE: An outbreak of Burkholderia (formerly Pseudomonas ) cepacia respiratory track colonisation and infection associated with nebulized albuterol therapy. Ann Intern Med 1995, 122: 762-766.CrossRefPubMed Hamill RJ, Houston ED, Georghiu PR, Wright CE, Koza MA, Cadle RM, Goeptert PA, Lewis DA, Zenon GJ, Clarridge JE: An outbreak of Burkholderia (formerly Pseudomonas ) cepacia respiratory track colonisation and infection associated with nebulized albuterol therapy. Ann Intern Med 1995, 122: 762-766.CrossRefPubMed
11.
go back to reference Beaty CD, Ritz Rh, Benson Ms: Continuous in-line nebulizers complicate pressure support ventilation. Chest 1989, 96: 1360-1363.CrossRefPubMed Beaty CD, Ritz Rh, Benson Ms: Continuous in-line nebulizers complicate pressure support ventilation. Chest 1989, 96: 1360-1363.CrossRefPubMed
12.
go back to reference Hess D: Aerosol bronchodilator delivery during mechanical ventilation nebulizer or inhaler? Chest 1991, 100: 1103-1104.CrossRefPubMed Hess D: Aerosol bronchodilator delivery during mechanical ventilation nebulizer or inhaler? Chest 1991, 100: 1103-1104.CrossRefPubMed
13.
go back to reference Manthous CA, Hall JB, Schmidt GA, Wood LD: Metered-dose inhaler versus nebulized albuterol in mechanically ventilated patients. Am Rev Respir Dis 1993, 148: 1567-1570.CrossRefPubMed Manthous CA, Hall JB, Schmidt GA, Wood LD: Metered-dose inhaler versus nebulized albuterol in mechanically ventilated patients. Am Rev Respir Dis 1993, 148: 1567-1570.CrossRefPubMed
14.
go back to reference Fink JB, Dhand R, Duarte AG, Jenne JW, Tobin MJ: Deposition of aerosol from metered-dose inhaler during mechanical ventilation: an in vitro model. Am J Respir Crit Care Med 1996, 154: 382-387.CrossRefPubMed Fink JB, Dhand R, Duarte AG, Jenne JW, Tobin MJ: Deposition of aerosol from metered-dose inhaler during mechanical ventilation: an in vitro model. Am J Respir Crit Care Med 1996, 154: 382-387.CrossRefPubMed
15.
go back to reference Crogan SJ, Bishop MJ: Delivery efficiency of metered-dose aerosols given via endotracheal tube. Anesthesiology 1989, 70: 1008-1010.CrossRefPubMed Crogan SJ, Bishop MJ: Delivery efficiency of metered-dose aerosols given via endotracheal tube. Anesthesiology 1989, 70: 1008-1010.CrossRefPubMed
16.
go back to reference Fuller HD, Dolovich MB, Postmituck G, Wong Pack W, Newhouse MT: Pressurized aerosol versus jet aerosol delivery to mechanically-ventilated patients: comparison of dose to the lungs. Am Rev Respir Dis 1990, 141: 440-444.CrossRefPubMed Fuller HD, Dolovich MB, Postmituck G, Wong Pack W, Newhouse MT: Pressurized aerosol versus jet aerosol delivery to mechanically-ventilated patients: comparison of dose to the lungs. Am Rev Respir Dis 1990, 141: 440-444.CrossRefPubMed
17.
go back to reference Ebert J, Adams AB, Green-Eide B: An evaluation of MDI spacers and adapters: their effect on the respirable volume of medication. Respir Care 1992, 37: 862-868.PubMed Ebert J, Adams AB, Green-Eide B: An evaluation of MDI spacers and adapters: their effect on the respirable volume of medication. Respir Care 1992, 37: 862-868.PubMed
18.
go back to reference Fuller HD, Dolovich MB, Chambers C, Newhouse MT: Aerosol delivery during mechanical ventilation: a predictive in vitro model. J Aerosol Med 1992, 5: 251-259.CrossRef Fuller HD, Dolovich MB, Chambers C, Newhouse MT: Aerosol delivery during mechanical ventilation: a predictive in vitro model. J Aerosol Med 1992, 5: 251-259.CrossRef
19.
go back to reference Fink JB, Dhand R, Grychowski J, Fahey PJ, Tobin MJ: Reconciling in vitro measurements of aerosol delivery from a metered-dose inhaler during mechanical ventilation and defining efficiency enhancing factors. Am J Respir Crit Care Med 1999, 159: 63-68.CrossRefPubMed Fink JB, Dhand R, Grychowski J, Fahey PJ, Tobin MJ: Reconciling in vitro measurements of aerosol delivery from a metered-dose inhaler during mechanical ventilation and defining efficiency enhancing factors. Am J Respir Crit Care Med 1999, 159: 63-68.CrossRefPubMed
20.
go back to reference Diot P, Morra L, Smaldone GC: Albuterol delivery in a model of mechanical ventilation: comparison of metered-dose inhaler and nebulizer efficiency. Am J Respir Crit Care Med 1995, 152: 1391-1394.CrossRefPubMed Diot P, Morra L, Smaldone GC: Albuterol delivery in a model of mechanical ventilation: comparison of metered-dose inhaler and nebulizer efficiency. Am J Respir Crit Care Med 1995, 152: 1391-1394.CrossRefPubMed
21.
go back to reference Rau JL, Harwood RJ, Groff JL: Evaluation of a reservoir devise for metered dose bronchodilator delivery to intubated adults: an in vitro study. Chest 1992, 102: 924-930.CrossRefPubMed Rau JL, Harwood RJ, Groff JL: Evaluation of a reservoir devise for metered dose bronchodilator delivery to intubated adults: an in vitro study. Chest 1992, 102: 924-930.CrossRefPubMed
22.
go back to reference Fuller HD, Dolovich MB, Turpie FH, Newhouse MT: Efficiency of bronchodilator aerosol delivery to the lungs from the metered dose inhaler in mechanically ventilated patients: a study comparing four different actuator devices. Chest 1994, 105: 214-218.CrossRefPubMed Fuller HD, Dolovich MB, Turpie FH, Newhouse MT: Efficiency of bronchodilator aerosol delivery to the lungs from the metered dose inhaler in mechanically ventilated patients: a study comparing four different actuator devices. Chest 1994, 105: 214-218.CrossRefPubMed
23.
go back to reference Fok TF, al-Essa M, Monkman S, Dolovich M, Girard L, Coates G, Kirpalani H: Delivery of metered dose inhaler aerosols to paralyzed and nonparalyzed rabbits. Crit Care Med 1997, 25: 140-144. 10.1097/00003246-199701000-00026CrossRefPubMed Fok TF, al-Essa M, Monkman S, Dolovich M, Girard L, Coates G, Kirpalani H: Delivery of metered dose inhaler aerosols to paralyzed and nonparalyzed rabbits. Crit Care Med 1997, 25: 140-144. 10.1097/00003246-199701000-00026CrossRefPubMed
24.
go back to reference Marik P, Hogan J, Krikorian J: A comparison of bronchodilator therapy delivered by nebulization and metered-dose inhaler in mechanically ventilated patients. Chest 1999, 115: 1653-1657. 10.1378/chest.115.6.1653CrossRefPubMed Marik P, Hogan J, Krikorian J: A comparison of bronchodilator therapy delivered by nebulization and metered-dose inhaler in mechanically ventilated patients. Chest 1999, 115: 1653-1657. 10.1378/chest.115.6.1653CrossRefPubMed
25.
go back to reference Duarte AG, Dhand R, Ralston R, Fink J, Fahey P, Tobin MJ, Jenne JW: Serum albuterol levels in mechanically ventilated patients and healthy subjects after metered-dose inhaler administration. Am J Respir Crit Care Med 1996, 154: 1658-1663.CrossRefPubMed Duarte AG, Dhand R, Ralston R, Fink J, Fahey P, Tobin MJ, Jenne JW: Serum albuterol levels in mechanically ventilated patients and healthy subjects after metered-dose inhaler administration. Am J Respir Crit Care Med 1996, 154: 1658-1663.CrossRefPubMed
26.
go back to reference Gottfried SB, Rossi A, Higgs BD, Zocchi L, Grassino A, Milic-Emili J: Respiratory mechanics in mechanically ventilated patients with respiratory failure. J Appl Physiol 1985, 58: 1849-1858.PubMed Gottfried SB, Rossi A, Higgs BD, Zocchi L, Grassino A, Milic-Emili J: Respiratory mechanics in mechanically ventilated patients with respiratory failure. J Appl Physiol 1985, 58: 1849-1858.PubMed
27.
go back to reference Bates JHT, Rossi A, Milic-Emili J: Analysis of the behaviour of the respiratory mechanics in mechanically ventilated patients with respiratory failure. J Appl Physiol 1985, 58: 1840-1848.PubMed Bates JHT, Rossi A, Milic-Emili J: Analysis of the behaviour of the respiratory mechanics in mechanically ventilated patients with respiratory failure. J Appl Physiol 1985, 58: 1840-1848.PubMed
28.
go back to reference Tuxen DV, Lane S: The effects of ventilatory pattern on hyperinflation, airway pressures and circulation in mechanical ventilation of patients with severe air-flow obstruction. Am Rev Respir Dis 1987, 136: 872-879.CrossRefPubMed Tuxen DV, Lane S: The effects of ventilatory pattern on hyperinflation, airway pressures and circulation in mechanical ventilation of patients with severe air-flow obstruction. Am Rev Respir Dis 1987, 136: 872-879.CrossRefPubMed
29.
go back to reference Georgopoulos D, Giannouli E, Patakas D: Effects of extrinsic positive end-expiratory pressure on mechanically ventilated patients with chronic obstructive pulmonary disease and dynamic hyperinflation. Intensive Care Med 1993, 19: 197-203.CrossRefPubMed Georgopoulos D, Giannouli E, Patakas D: Effects of extrinsic positive end-expiratory pressure on mechanically ventilated patients with chronic obstructive pulmonary disease and dynamic hyperinflation. Intensive Care Med 1993, 19: 197-203.CrossRefPubMed
30.
go back to reference Dawson SV, Elliott EA: Wave-speed limitation on expiratory flow-a unifying concept. Am Rev Respir Dis 1997, 140: 1-3. Dawson SV, Elliott EA: Wave-speed limitation on expiratory flow-a unifying concept. Am Rev Respir Dis 1997, 140: 1-3.
31.
go back to reference Gay PC, Patel HG, Nelson SB, Gilles B, Hubmayr RD: Metered dose inhalers for bronchodilator delivery in intubated, mechanically ventilated patients. Chest 1991, 99: 66-71.CrossRefPubMed Gay PC, Patel HG, Nelson SB, Gilles B, Hubmayr RD: Metered dose inhalers for bronchodilator delivery in intubated, mechanically ventilated patients. Chest 1991, 99: 66-71.CrossRefPubMed
32.
go back to reference Georgopoulos D, Mitrouska I, Markopoulou K, Patakas D, Anthonisen NR: Effects of breathing pattern on mechanically ventilated patients with chronic obstructive pulmonary disease and dynamic hyperinflation. Intensive Care Med 1995, 21: 880-886.CrossRefPubMed Georgopoulos D, Mitrouska I, Markopoulou K, Patakas D, Anthonisen NR: Effects of breathing pattern on mechanically ventilated patients with chronic obstructive pulmonary disease and dynamic hyperinflation. Intensive Care Med 1995, 21: 880-886.CrossRefPubMed
33.
go back to reference Tobin MJ: Monitoring respiratory mechanics in spontaneously breathing patients. In: Principles and Practice of Intensive Care Monitoring. Edited by Tobin MJ. New York: McGraw-Hill, Incorporated, Health Professions Division, 1998, 617-654. Tobin MJ: Monitoring respiratory mechanics in spontaneously breathing patients. In: Principles and Practice of Intensive Care Monitoring. Edited by Tobin MJ. New York: McGraw-Hill, Incorporated, Health Professions Division, 1998, 617-654.
34.
go back to reference Mancebo J, Amaro P, Lorino H, Lemaire F, Harf A, Brochard L: Effects of albuterol inhalation on the work of breathing during weaning from mechanical ventilation. Am Rev Respir Dis 1991, 144: 95-100.CrossRefPubMed Mancebo J, Amaro P, Lorino H, Lemaire F, Harf A, Brochard L: Effects of albuterol inhalation on the work of breathing during weaning from mechanical ventilation. Am Rev Respir Dis 1991, 144: 95-100.CrossRefPubMed
35.
go back to reference Mead J, Lindgren I, Gaensler AE: The mechanical properties of the lungs in emphysema. J Clin Invest 1956, 34: 1005-1016.CrossRef Mead J, Lindgren I, Gaensler AE: The mechanical properties of the lungs in emphysema. J Clin Invest 1956, 34: 1005-1016.CrossRef
36.
go back to reference Fernandez A, Lazaro A, Garcia A, Aragon C, Cerda E: Bronchodilators in patients with obstructive pulmonary disease on mechanical ventilation: utilisation of metered-dose inhalers. Am Rev Respir Dis 1990, 141: 164-168.CrossRefPubMed Fernandez A, Lazaro A, Garcia A, Aragon C, Cerda E: Bronchodilators in patients with obstructive pulmonary disease on mechanical ventilation: utilisation of metered-dose inhalers. Am Rev Respir Dis 1990, 141: 164-168.CrossRefPubMed
37.
go back to reference Dhand R, Jubran A, Tobic MT: Bronchodilator delivery by metered-dose inhaler in ventilator support patients. Am J Respir Crit Care Med 1995, 151: 1827-1833.CrossRefPubMed Dhand R, Jubran A, Tobic MT: Bronchodilator delivery by metered-dose inhaler in ventilator support patients. Am J Respir Crit Care Med 1995, 151: 1827-1833.CrossRefPubMed
38.
go back to reference Dhand R, Duarte AG, Jubran A, Jenne JW, Fink JB, Fahey PJ, Tobin MJ: Dose-response to bronchodilator delivered by metered dose inhaler in ventilator-support patients. Am J Respir Crit Care Med 1996, 154: 388-393.CrossRefPubMed Dhand R, Duarte AG, Jubran A, Jenne JW, Fink JB, Fahey PJ, Tobin MJ: Dose-response to bronchodilator delivered by metered dose inhaler in ventilator-support patients. Am J Respir Crit Care Med 1996, 154: 388-393.CrossRefPubMed
39.
go back to reference Mouloudi E, Katsanoulas K, Anastasaki M, Askitopoulou E, Georgopoulos D: Bronchodilator delivery by metered-dose inhaler in mechanically ventilated COPD patients: influence of end-inspiratory pause. Eur Respir J 1998, 12: 165-169. 10.1183/09031936.98.12010165CrossRefPubMed Mouloudi E, Katsanoulas K, Anastasaki M, Askitopoulou E, Georgopoulos D: Bronchodilator delivery by metered-dose inhaler in mechanically ventilated COPD patients: influence of end-inspiratory pause. Eur Respir J 1998, 12: 165-169. 10.1183/09031936.98.12010165CrossRefPubMed
40.
go back to reference Mouloudi E, Katsanoulas K, Anastasaki M, Hoing S, Georgopoulos D: Bronchodilator delivery by metered-dose inhaler in mechanically ventilated COPD patients: influence of tidal volume. Intensive Care Med 1999, 25: 1215-1221. 10.1007/s001340051049CrossRefPubMed Mouloudi E, Katsanoulas K, Anastasaki M, Hoing S, Georgopoulos D: Bronchodilator delivery by metered-dose inhaler in mechanically ventilated COPD patients: influence of tidal volume. Intensive Care Med 1999, 25: 1215-1221. 10.1007/s001340051049CrossRefPubMed
41.
go back to reference Mouloudi E, Prinianakis G, Kondili E, Georgopoulos D: Bronchodilator delivery by metered-dose inhaler in mechanically ventilated COPD patients: influence of flow pattern. Eur Respir J 2000, in press. Mouloudi E, Prinianakis G, Kondili E, Georgopoulos D: Bronchodilator delivery by metered-dose inhaler in mechanically ventilated COPD patients: influence of flow pattern. Eur Respir J 2000, in press.
42.
go back to reference Guerin G, Chevre A, Dessirier P, Poncet T, Becquemin MH, Dequin PF, Le Guellec C, Jacques D: Inhaled fenoterol-ipratropium bromide in mechanically ventilated patients with chronic obstructive pulmonary disease. Am J Respir Crit Care Med 1999, 159: 1036-1042.CrossRefPubMed Guerin G, Chevre A, Dessirier P, Poncet T, Becquemin MH, Dequin PF, Le Guellec C, Jacques D: Inhaled fenoterol-ipratropium bromide in mechanically ventilated patients with chronic obstructive pulmonary disease. Am J Respir Crit Care Med 1999, 159: 1036-1042.CrossRefPubMed
43.
go back to reference Mouloudi E, Katsanoulas K, Hoing S, Prinianakis G, Zeaki M, Georgopoulos D: Duration of effect of bronchodilator delivered by metered dose inhaler in mechanically ventilated COPD patients. Intensive Care Med 1999, 25(Suppl 1): S129. Mouloudi E, Katsanoulas K, Hoing S, Prinianakis G, Zeaki M, Georgopoulos D: Duration of effect of bronchodilator delivered by metered dose inhaler in mechanically ventilated COPD patients. Intensive Care Med 1999, 25(Suppl 1): S129.
Metadata
Title
Bronchodilator delivery with metered-dose inhaler during mechanical ventilation
Authors
Dimitris Georgopoulos
Eleni Mouloudi
Eumorfia Kondili
Maria Klimathianaki
Publication date
01-08-2000
Publisher
BioMed Central
Published in
Critical Care / Issue 4/2000
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc698

Other articles of this Issue 4/2000

Critical Care 4/2000 Go to the issue