Skip to main content
Top
Published in: Critical Care 5/2007

Open Access 01-10-2007 | Research

Interference by new-generation mobile phones on critical care medical equipment

Authors: Erik Jan van Lieshout, Sabine N van der Veer, Reinout Hensbroek, Johanna C Korevaar, Margreeth B Vroom, Marcus J Schultz

Published in: Critical Care | Issue 5/2007

Login to get access

Abstract

Introduction

The aim of the study was to assess and classify incidents of electromagnetic interference (EMI) by second-generation and third-generation mobile phones on critical care medical equipment.

Methods

EMI was assessed with two General Packet Radio Service (GPRS) signals (900 MHz, 2 W, two different time-slot occupations) and one Universal Mobile Telecommunications System (UMTS) signal (1,947.2 MHz, 0.2 W), corresponding to maximal transmit performance of mobile phones in daily practice, generated under controlled conditions in the proximity of 61 medical devices. Incidents of EMI were classified in accordance with an adjusted critical care event scale.

Results

A total of 61 medical devices in 17 categories (27 different manufacturers) were tested and demonstrated 48 incidents in 26 devices (43%); 16 (33%) were classified as hazardous, 20 (42%) as significant and 12 (25%) as light. The GPRS-1 signal induced the most EMI incidents (41%), the GRPS-2 signal induced fewer (25%) and the UMTS signal induced the least (13%; P < 0.001). The median distance between antenna and medical device for EMI incidents was 3 cm (range 0.1 to 500 cm). One hazardous incident occurred beyond 100 cm (in a ventilator with GRPS-1 signal at 300 cm).

Conclusion

Critical care equipment is vulnerable to EMI by new-generation wireless telecommunication technologies with median distances of about 3 cm. The policy to keep mobile phones '1 meter' from the critical care bedside in combination with easily accessed areas of unrestricted use still seems warranted.
Appendix
Available only for authorised users
Literature
2.
go back to reference International Organization for Standardization (ISO): Health Informatics – Use of Mobile Wireless Communication and Computing Technology in Healthcare Facilities. Geneva: ISO; 2005. [Report no. ISO/TR 21730:2005.] International Organization for Standardization (ISO): Health Informatics – Use of Mobile Wireless Communication and Computing Technology in Healthcare Facilities. Geneva: ISO; 2005. [Report no. ISO/TR 21730:2005.]
3.
go back to reference Lawrentschuk N, Bolton DM: Mobile phone interference with medical equipment and its clinical relevance: a systematic review. Med J Aust 2004, 181: 145-149.PubMed Lawrentschuk N, Bolton DM: Mobile phone interference with medical equipment and its clinical relevance: a systematic review. Med J Aust 2004, 181: 145-149.PubMed
4.
go back to reference Tri JL, Severson RP, Hyberger LK, Hayes DL: Use of cellular telephones in the hospital environment. Mayo Clin Proc 2007, 82: 282-285.CrossRefPubMed Tri JL, Severson RP, Hyberger LK, Hayes DL: Use of cellular telephones in the hospital environment. Mayo Clin Proc 2007, 82: 282-285.CrossRefPubMed
5.
go back to reference Shaw CI, Kacmarek RM, Hampton RL, Riggi V, El Masry A, Cooper JB, Hurford WE: Cellular phone interference with the operation of mechanical ventilators. Crit Care Med 2004, 32: 928-931. 10.1097/01.CCM.0000120061.01431.DBCrossRefPubMed Shaw CI, Kacmarek RM, Hampton RL, Riggi V, El Masry A, Cooper JB, Hurford WE: Cellular phone interference with the operation of mechanical ventilators. Crit Care Med 2004, 32: 928-931. 10.1097/01.CCM.0000120061.01431.DBCrossRefPubMed
6.
go back to reference Barbaro V, Bartolini P, Benassi M, Di Nallo AM, Reali L, Valsecchi S: Electromagnetic interference by GSM cellular phones and UHF radios with intensive-care and operating-room ventilators. Biomed Instrum Technol 2000, 34: 361-369.PubMed Barbaro V, Bartolini P, Benassi M, Di Nallo AM, Reali L, Valsecchi S: Electromagnetic interference by GSM cellular phones and UHF radios with intensive-care and operating-room ventilators. Biomed Instrum Technol 2000, 34: 361-369.PubMed
7.
go back to reference Irnich WE, Tobisch R: Mobile phones in hospitals. Biomed Instrum Technol 1999, 33: 28-34.PubMed Irnich WE, Tobisch R: Mobile phones in hospitals. Biomed Instrum Technol 1999, 33: 28-34.PubMed
8.
go back to reference Soto RG, Chu LF, Goldman JM, Rampil IJ, Ruskin KJ: Communication in critical care environments: mobile telephones improve patient care. Anesth Analg 2006, 102: 535-541. 10.1213/01.ane.0000194506.79408.79CrossRefPubMed Soto RG, Chu LF, Goldman JM, Rampil IJ, Ruskin KJ: Communication in critical care environments: mobile telephones improve patient care. Anesth Analg 2006, 102: 535-541. 10.1213/01.ane.0000194506.79408.79CrossRefPubMed
9.
go back to reference Hahn IH, Schnadower D, Dakin RJ, Nelson LS: Cellular phone interference as a cause of acute epinephrine poisoning. Ann Emerg Med 2005, 46: 298-299. 10.1016/j.annemergmed.2005.03.026CrossRefPubMed Hahn IH, Schnadower D, Dakin RJ, Nelson LS: Cellular phone interference as a cause of acute epinephrine poisoning. Ann Emerg Med 2005, 46: 298-299. 10.1016/j.annemergmed.2005.03.026CrossRefPubMed
10.
go back to reference Anonymous: Wireless communication devices and electromagnetic interference. ECRI's updated recommendations. Health Devices 2001, 30: 403-409. Anonymous: Wireless communication devices and electromagnetic interference. ECRI's updated recommendations. Health Devices 2001, 30: 403-409.
11.
go back to reference Lapinsky SE, Easty AC: Electromagnetic interference in critical care. J Crit Care 2006, 21: 267-270. 10.1016/j.jcrc.2006.03.010CrossRefPubMed Lapinsky SE, Easty AC: Electromagnetic interference in critical care. J Crit Care 2006, 21: 267-270. 10.1016/j.jcrc.2006.03.010CrossRefPubMed
12.
go back to reference Wallin MK, Marve T, Hakansson PK: Modern wireless telecommunication technologies and their electromagnetic compatibility with life-supporting equipment. Anesth Analg 2005, 101: 1393-1400. 10.1213/01.ANE.0000180216.83554.00CrossRefPubMed Wallin MK, Marve T, Hakansson PK: Modern wireless telecommunication technologies and their electromagnetic compatibility with life-supporting equipment. Anesth Analg 2005, 101: 1393-1400. 10.1213/01.ANE.0000180216.83554.00CrossRefPubMed
13.
go back to reference Ettelt S, Nolte E, McKee M, Haugen OA, Karlberg I, Klazinga N, Ricciardi W, Teperi J: Evidence-based policy? The use of mobile phones in hospital. J Public Health (Oxf) 2006, 28: 299-303. 10.1093/pubmed/fdl067CrossRef Ettelt S, Nolte E, McKee M, Haugen OA, Karlberg I, Klazinga N, Ricciardi W, Teperi J: Evidence-based policy? The use of mobile phones in hospital. J Public Health (Oxf) 2006, 28: 299-303. 10.1093/pubmed/fdl067CrossRef
14.
go back to reference Institute of Electrical and Electronics Engineers: American National Standard Recommended Practice for On-site ad hoc Test Method for Estimating Radiated Electromagnetic Immunity of Medical Devices to Specific Radio-frequency Transmitters (Standard C63.18). Piscataway, NJ: IEEE; 1997. Institute of Electrical and Electronics Engineers: American National Standard Recommended Practice for On-site ad hoc Test Method for Estimating Radiated Electromagnetic Immunity of Medical Devices to Specific Radio-frequency Transmitters (Standard C63.18). Piscataway, NJ: IEEE; 1997.
15.
go back to reference Kivlahan C, Sangster W, Nelson K, Buddenbaum J, Lobenstein K: Developing a comprehensive electronic adverse event reporting system in an academic health center. Jt Comm J Qual Improv 2002, 28: 583-594.PubMed Kivlahan C, Sangster W, Nelson K, Buddenbaum J, Lobenstein K: Developing a comprehensive electronic adverse event reporting system in an academic health center. Jt Comm J Qual Improv 2002, 28: 583-594.PubMed
16.
go back to reference Imhoff M: Everybody on the phone? Anesth Analg 2006, 102: 533-534. 10.1213/01.ane.0000194505.53207.35CrossRefPubMed Imhoff M: Everybody on the phone? Anesth Analg 2006, 102: 533-534. 10.1213/01.ane.0000194505.53207.35CrossRefPubMed
17.
go back to reference Lönn S, Forssén U, Vecchia P, Ahlbom A, Feychting M: Output power levels from mobile phones in different geographical areas; implications for exposure assessment. Occup Environ Med 2004, 61: 769-772. 10.1136/oem.2003.012567PubMedCentralCrossRefPubMed Lönn S, Forssén U, Vecchia P, Ahlbom A, Feychting M: Output power levels from mobile phones in different geographical areas; implications for exposure assessment. Occup Environ Med 2004, 61: 769-772. 10.1136/oem.2003.012567PubMedCentralCrossRefPubMed
18.
go back to reference IEC: Medical Electrical Equipment. Part 1–2: General Requirements for Safety – Collateral Standard: Electromagnetic Compatibility – Requirements and Tests. Geneva: International Electrotechnical Commission; 2004. [Report no. IEC 60601-1-2:2001+A1:2004.] IEC: Medical Electrical Equipment. Part 1–2: General Requirements for Safety – Collateral Standard: Electromagnetic Compatibility – Requirements and Tests. Geneva: International Electrotechnical Commission; 2004. [Report no. IEC 60601-1-2:2001+A1:2004.]
Metadata
Title
Interference by new-generation mobile phones on critical care medical equipment
Authors
Erik Jan van Lieshout
Sabine N van der Veer
Reinout Hensbroek
Johanna C Korevaar
Margreeth B Vroom
Marcus J Schultz
Publication date
01-10-2007
Publisher
BioMed Central
Published in
Critical Care / Issue 5/2007
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc6115

Other articles of this Issue 5/2007

Critical Care 5/2007 Go to the issue