Electrical activity of the diaphragm (EAdi) as a monitoring parameter in difficult weaning from respirator: a pilot study

Authors: Jürgen Barwing, Cristina Pedroni, Ulrike Olgemöller, Michael Quintel, Onnen Moerer

Published in: Critical Care | Issue 4/2013

Login to get access

Abstract

Introduction

A reliable prediction of successful weaning from respiratory support may be crucial for the overall outcome of the critically ill patient. The electrical activity of the diaphragm (EAdi) allows one to monitor the patients’ respiratory drive and their ability to meet the increased respiratory demand. In this pilot study, we compared the EAdi with conventional parameters of weaning failure, such as the ratio of respiratory rate to tidal volume.

Methods

We studied 18 mechanically ventilated patients considered difficult to wean. For a spontaneous breathing trial (SBT), the patients were disconnected from the ventilator and given oxygen through a T-piece. The SBT was evaluated by using standard criteria.

Results

Twelve patients completed the SBT successfully, and six failed. The EAdi was significantly different in the two groups. We found an early increase in EAdi in the failing patients that was more pronounced than in any of the patients who successfully passed the SBT. Changes in EAdi predicted an SBT failure earlier than did conventional parameters.

Conclusions

EAdi monitoring adds valuable information during weaning from the ventilator and may help to identify patients who are not ready for discontinuation of respiratory support.
Appendix
Available only for authorised users
Literature
1.
go back to reference Yang KL, Tobin MJ: A prospective study of indexes predicting the outcome of trials of weaning from mechanical ventilation. N Engl J Med 1991, 324: 1445-1450. 10.1056/NEJM199105233242101PubMedCrossRef Yang KL, Tobin MJ: A prospective study of indexes predicting the outcome of trials of weaning from mechanical ventilation. N Engl J Med 1991, 324: 1445-1450. 10.1056/NEJM199105233242101PubMedCrossRef
2.
go back to reference Berg KM, Lang J, Salciccioli JD, Bak E, Cocchi MN, Gautam S, Donnin MW: The Rapid Shallow Breathing Index as a predictor of failure of noninvasive ventilation for patients with acute respiratory failure. Respir Care 2012, 57: 1548-1554. 10.4187/respcare.01597PubMedCrossRef Berg KM, Lang J, Salciccioli JD, Bak E, Cocchi MN, Gautam S, Donnin MW: The Rapid Shallow Breathing Index as a predictor of failure of noninvasive ventilation for patients with acute respiratory failure. Respir Care 2012, 57: 1548-1554. 10.4187/respcare.01597PubMedCrossRef
3.
go back to reference Sinderby C, Navalesi P, Beck J, Skrobik Y, Comtois N, Friberg S, Gottfried SB, Lindström L: Neural control of mechanical ventilation in respiratory failure. Nat Med 1999, 5: 1433-1436. 10.1038/71012PubMedCrossRef Sinderby C, Navalesi P, Beck J, Skrobik Y, Comtois N, Friberg S, Gottfried SB, Lindström L: Neural control of mechanical ventilation in respiratory failure. Nat Med 1999, 5: 1433-1436. 10.1038/71012PubMedCrossRef
4.
go back to reference Beck J, Reilly M, Grasselli G, Mirabella L, Slutsky AS, Dunn MS, Sinderby C: Patient-ventilator interaction during neurally adjusted ventilatory assist in low birth weight infants. Pediatr Res 2009, 65: 663-668. 10.1203/PDR.0b013e31819e72abPubMedPubMedCentralCrossRef Beck J, Reilly M, Grasselli G, Mirabella L, Slutsky AS, Dunn MS, Sinderby C: Patient-ventilator interaction during neurally adjusted ventilatory assist in low birth weight infants. Pediatr Res 2009, 65: 663-668. 10.1203/PDR.0b013e31819e72abPubMedPubMedCentralCrossRef
5.
go back to reference Grasselli G, Beck J, Comtois N, Brunet F, Sinderby C: Determination of patient’s contribution to a breath during mechanical ventilation [abstract]. Am J Respir Crit Care Med 2009, 179: 3070. Grasselli G, Beck J, Comtois N, Brunet F, Sinderby C: Determination of patient’s contribution to a breath during mechanical ventilation [abstract]. Am J Respir Crit Care Med 2009, 179: 3070.
6.
go back to reference Jolley CJ, Luo YM, Steier J, Reilly C, Seymour J, Lunt A, Ward K, Rafferty GF, Polkey MI, Moxham J: Neural respiratory drive in healthy subjects and in COPD. Eur Respir J 2009, 33: 289-297.PubMedCrossRef Jolley CJ, Luo YM, Steier J, Reilly C, Seymour J, Lunt A, Ward K, Rafferty GF, Polkey MI, Moxham J: Neural respiratory drive in healthy subjects and in COPD. Eur Respir J 2009, 33: 289-297.PubMedCrossRef
7.
go back to reference Beck J, Sinderby C, Lindström L, Grassino A: Influence of bipolar esophageal electrode positioning on measurements of human crural diaphragm electromyogram. J Appl Physiol 1996, 81: 1434-1449.PubMed Beck J, Sinderby C, Lindström L, Grassino A: Influence of bipolar esophageal electrode positioning on measurements of human crural diaphragm electromyogram. J Appl Physiol 1996, 81: 1434-1449.PubMed
8.
go back to reference Lourenço RV, Cherniack NS, Malm JR, Fishman AP: Nervous output from the respiratory center during obstructed breathing. J Appl Physiol 1966, 21: 527-533.PubMed Lourenço RV, Cherniack NS, Malm JR, Fishman AP: Nervous output from the respiratory center during obstructed breathing. J Appl Physiol 1966, 21: 527-533.PubMed
9.
go back to reference Petit JM, Milic-Emili G, Delhez L: Role of the diaphragm in breathing in conscious normal man: an electromyographic study. J Appl Physiol 1960, 15: 1101-1106.PubMed Petit JM, Milic-Emili G, Delhez L: Role of the diaphragm in breathing in conscious normal man: an electromyographic study. J Appl Physiol 1960, 15: 1101-1106.PubMed
10.
go back to reference Beck J, Campoccia F, Allo JC, Brander L, Brunet F, Slutsky AS, Sinderby C: Improved synchrony and respiratory unloading by neurally adjusted ventilatory assist (NAVA) in lung-injured rabbits. Pediatr Res 2007, 61: 289-294. 10.1203/01.pdr.0000257324.22406.93PubMedCrossRef Beck J, Campoccia F, Allo JC, Brander L, Brunet F, Slutsky AS, Sinderby C: Improved synchrony and respiratory unloading by neurally adjusted ventilatory assist (NAVA) in lung-injured rabbits. Pediatr Res 2007, 61: 289-294. 10.1203/01.pdr.0000257324.22406.93PubMedCrossRef
11.
go back to reference Colombo D, Cammarota G, Bergamaschi V, De Lucia M, Corte FD, Navalesi P: Physiologic response to varying levels of pressure support and neurally adjusted ventilatory assist in patients with acute respiratory failure. Intensive Care Med 2008, 34: 2010-2018. 10.1007/s00134-008-1208-3PubMedCrossRef Colombo D, Cammarota G, Bergamaschi V, De Lucia M, Corte FD, Navalesi P: Physiologic response to varying levels of pressure support and neurally adjusted ventilatory assist in patients with acute respiratory failure. Intensive Care Med 2008, 34: 2010-2018. 10.1007/s00134-008-1208-3PubMedCrossRef
12.
go back to reference Brander L, Leong-Poi H, Beck J, Brunet F, Hutchison SJ, Slutsky AS, Sinderby C: Titration and implementation of neurally adjusted ventilatory assist in critically ill patients. Chest 2009, 135: 695-703. 10.1378/chest.08-1747PubMedCrossRef Brander L, Leong-Poi H, Beck J, Brunet F, Hutchison SJ, Slutsky AS, Sinderby C: Titration and implementation of neurally adjusted ventilatory assist in critically ill patients. Chest 2009, 135: 695-703. 10.1378/chest.08-1747PubMedCrossRef
13.
go back to reference Luo YM, Li RF, Jolley C, Wu HD, Steier J, Moxham J, Zhong NS: Neural respiratory drive in patients with COPD during exercise tests. Respiration 2011, 81: 294-301. 10.1159/000317136PubMedCrossRef Luo YM, Li RF, Jolley C, Wu HD, Steier J, Moxham J, Zhong NS: Neural respiratory drive in patients with COPD during exercise tests. Respiration 2011, 81: 294-301. 10.1159/000317136PubMedCrossRef
14.
go back to reference Sinderby C, Beck J, Spahija J, Weinberg J, Grassino A: Voluntary activation of the human diaphragm in health and disease. J Appl Physiol 1998, 85: 2146-2158.PubMed Sinderby C, Beck J, Spahija J, Weinberg J, Grassino A: Voluntary activation of the human diaphragm in health and disease. J Appl Physiol 1998, 85: 2146-2158.PubMed
15.
go back to reference Martin Dres M, Schmidt M, Ferre A, Mayaux J, Similowski T, Demoule A: Diaphragm electromyographic activity as a predictor of weaning failure. Intensive Care Med 2012, 38: 2017-2025. 10.1007/s00134-012-2700-3PubMedCrossRef Martin Dres M, Schmidt M, Ferre A, Mayaux J, Similowski T, Demoule A: Diaphragm electromyographic activity as a predictor of weaning failure. Intensive Care Med 2012, 38: 2017-2025. 10.1007/s00134-012-2700-3PubMedCrossRef
16.
go back to reference Wolf GK, Walsh BK, Green ML, Arnold JH: Electrical activity of the diaphragm during extubation readiness testing in critically ill children. Pediatr Crit Care Med 2011, 12: 220-224. 10.1097/PCC.0b013e3181fe28fcCrossRef Wolf GK, Walsh BK, Green ML, Arnold JH: Electrical activity of the diaphragm during extubation readiness testing in critically ill children. Pediatr Crit Care Med 2011, 12: 220-224. 10.1097/PCC.0b013e3181fe28fcCrossRef
17.
go back to reference Boles JM, Bion J, Connors A, Herridge M, Marsh B, Melot C, Pearl R, Silverman H, Stanchina M, Vieillard-Baron A, Welte T: Weaning from mechanical ventilation. Eur Respir J 2007, 29: 1033-1056. 10.1183/09031936.00010206PubMedCrossRef Boles JM, Bion J, Connors A, Herridge M, Marsh B, Melot C, Pearl R, Silverman H, Stanchina M, Vieillard-Baron A, Welte T: Weaning from mechanical ventilation. Eur Respir J 2007, 29: 1033-1056. 10.1183/09031936.00010206PubMedCrossRef
18.
go back to reference MacIntyre NR, Cook DJ, Ely EW Jr, Epstein SK, Fink JB, Heffner JE, Hess D, Hubmayer RD, Scheinhorn DJ: Evidence-based guidelines for weaning and discontinuing ventilatory support: a collective task force facilitated by the American College of Chest Physicians; the American Association for Respiratory Care; and the American College of Critical Care Medicine. Chest 2001, 120: 375-395. 10.1378/chest.120.6_suppl.375SCrossRef MacIntyre NR, Cook DJ, Ely EW Jr, Epstein SK, Fink JB, Heffner JE, Hess D, Hubmayer RD, Scheinhorn DJ: Evidence-based guidelines for weaning and discontinuing ventilatory support: a collective task force facilitated by the American College of Chest Physicians; the American Association for Respiratory Care; and the American College of Critical Care Medicine. Chest 2001, 120: 375-395. 10.1378/chest.120.6_suppl.375SCrossRef
19.
go back to reference Devlin JW, Boleski G, Mlynarek M, Nerenz DR, Peterson E, Jankowski M, Horst HM, Zarowitz BJ: Motor Activity Assessment Scale: a valid and reliable sedation scale for use with mechanically ventilated patients in an adult surgical intensive care unit. Crit Care Med 1999, 27: 1271-1275. 10.1097/00003246-199907000-00008PubMedCrossRef Devlin JW, Boleski G, Mlynarek M, Nerenz DR, Peterson E, Jankowski M, Horst HM, Zarowitz BJ: Motor Activity Assessment Scale: a valid and reliable sedation scale for use with mechanically ventilated patients in an adult surgical intensive care unit. Crit Care Med 1999, 27: 1271-1275. 10.1097/00003246-199907000-00008PubMedCrossRef
20.
go back to reference Barwing J, Ambold M, Linden N, Quintel M, Moerer O: Evaluation of the catheter positioning for neurally adjusted ventilatory assist. Intensive Care Med 2009, 35: 1809-1814. 10.1007/s00134-009-1587-0PubMedPubMedCentralCrossRef Barwing J, Ambold M, Linden N, Quintel M, Moerer O: Evaluation of the catheter positioning for neurally adjusted ventilatory assist. Intensive Care Med 2009, 35: 1809-1814. 10.1007/s00134-009-1587-0PubMedPubMedCentralCrossRef
21.
go back to reference Sackner MA, Watson H, Belsito AS, Feinerman D, Suarez M, Gonzalez G, Bizousky F, Krieger B: Calibration of respiratory inductive plethysmograph during natural breathing. J Appl Physiol 1989, 66: 410-420.PubMed Sackner MA, Watson H, Belsito AS, Feinerman D, Suarez M, Gonzalez G, Bizousky F, Krieger B: Calibration of respiratory inductive plethysmograph during natural breathing. J Appl Physiol 1989, 66: 410-420.PubMed
22.
go back to reference Leino K, Nunes S, Valta P, Takala J: Validation of a new respiratory inductive plethysmograph. Acta Anaesthesiol Scand 2001, 45: 104-111. 10.1034/j.1399-6576.2001.450116.xPubMedCrossRef Leino K, Nunes S, Valta P, Takala J: Validation of a new respiratory inductive plethysmograph. Acta Anaesthesiol Scand 2001, 45: 104-111. 10.1034/j.1399-6576.2001.450116.xPubMedCrossRef
23.
go back to reference Valta P, Takala J, Foster R, Weissman C, Kinney JM: Evaluation of respiratory inductive plethysmography in the measurement of breathing pattern and PEEP-induced changes in lung volume. Chest 1992, 102: 234-238. 10.1378/chest.102.1.234PubMedCrossRef Valta P, Takala J, Foster R, Weissman C, Kinney JM: Evaluation of respiratory inductive plethysmography in the measurement of breathing pattern and PEEP-induced changes in lung volume. Chest 1992, 102: 234-238. 10.1378/chest.102.1.234PubMedCrossRef
24.
go back to reference Cabello B, Thille AW, Roche-Campo F, Brochard L, Gómez FJ, Mancebo J: Physiological comparison of three spontaneous breathing trials in difficult-to-wean patients. Intensive Care Med 2010, 36: 1171-1179. 10.1007/s00134-010-1870-0PubMedCrossRef Cabello B, Thille AW, Roche-Campo F, Brochard L, Gómez FJ, Mancebo J: Physiological comparison of three spontaneous breathing trials in difficult-to-wean patients. Intensive Care Med 2010, 36: 1171-1179. 10.1007/s00134-010-1870-0PubMedCrossRef
25.
go back to reference Levine S, Nguyen T, Taylor N, Friscia ME, Budak MT, Rothenberg P, Zhu J, Sachdeva R, Sonnad S, Kaiser LR, Rubinstein NA, Powers SK, Shrager JB: Rapid disuse atrophy of diaphragm fibers in mechanically ventilated humans. N Engl J Med 2008, 358: 1327-1335. 10.1056/NEJMoa070447PubMedCrossRef Levine S, Nguyen T, Taylor N, Friscia ME, Budak MT, Rothenberg P, Zhu J, Sachdeva R, Sonnad S, Kaiser LR, Rubinstein NA, Powers SK, Shrager JB: Rapid disuse atrophy of diaphragm fibers in mechanically ventilated humans. N Engl J Med 2008, 358: 1327-1335. 10.1056/NEJMoa070447PubMedCrossRef
26.
go back to reference De Jonghe B, Bastuji-Garin S, Sharshar T, Outin H, Brochard L: Does ICU-acquired paresis lengthen weaning from mechanical ventilation? Intensive Care Med 2004, 30: 1117-1121. 10.1007/s00134-004-2174-zPubMedCrossRef De Jonghe B, Bastuji-Garin S, Sharshar T, Outin H, Brochard L: Does ICU-acquired paresis lengthen weaning from mechanical ventilation? Intensive Care Med 2004, 30: 1117-1121. 10.1007/s00134-004-2174-zPubMedCrossRef
27.
go back to reference Epstein SK, Ciubotaru RL: Independent effects of etiology of failure and time to reintubation on outcome for patients failing extubation. Am J Respir Crit Care Med 1998, 158: 489-493. 10.1164/ajrccm.158.2.9711045PubMedCrossRef Epstein SK, Ciubotaru RL: Independent effects of etiology of failure and time to reintubation on outcome for patients failing extubation. Am J Respir Crit Care Med 1998, 158: 489-493. 10.1164/ajrccm.158.2.9711045PubMedCrossRef
28.
go back to reference Tobin MJ, Laghi F, Jubran A: Respiratory muscle dysfunction in mechanically ventilated patients. Mol Cell Biochem 1998, 179: 87-98. 10.1023/A:1006807904036PubMedCrossRef Tobin MJ, Laghi F, Jubran A: Respiratory muscle dysfunction in mechanically ventilated patients. Mol Cell Biochem 1998, 179: 87-98. 10.1023/A:1006807904036PubMedCrossRef
29.
go back to reference Nemer SN, Barbas CS, Caldeira JB, Cárias TC, Santos RG, Almeida LC, Azeredo LM, Noé RA, Guimarães BS, Souza PC: A new integrative weaning index of discontinuation from mechanical ventilation. Crit Care 2009, 13: 152. 10.1186/cc7882CrossRef Nemer SN, Barbas CS, Caldeira JB, Cárias TC, Santos RG, Almeida LC, Azeredo LM, Noé RA, Guimarães BS, Souza PC: A new integrative weaning index of discontinuation from mechanical ventilation. Crit Care 2009, 13: 152. 10.1186/cc7882CrossRef
30.
go back to reference El-Khatib MF, Zeineldine SM, Jamaleddine GW: Effect of pressure support ventilation and positive end expiratory pressure on the rapid shallow breathing index in intensive care unit patients. Intensive Care Med 2008, 34: 505-510. 10.1007/s00134-007-0939-xPubMedCrossRef El-Khatib MF, Zeineldine SM, Jamaleddine GW: Effect of pressure support ventilation and positive end expiratory pressure on the rapid shallow breathing index in intensive care unit patients. Intensive Care Med 2008, 34: 505-510. 10.1007/s00134-007-0939-xPubMedCrossRef
31.
go back to reference Bien MY, Shui Lin Y, Shih CH, Yang YL, Lin HW, Bai KJ, Wang JH, Ru Kou Y: Comparisons of predictive performance of breathing pattern variability measured during T-piece, automatic tube compensation, and pressure support ventilation for weaning intensive care unit patients from mechanical ventilation. Crit Care Med 2011, 39: 2253-2262. 10.1097/CCM.0b013e31822279edPubMedCrossRef Bien MY, Shui Lin Y, Shih CH, Yang YL, Lin HW, Bai KJ, Wang JH, Ru Kou Y: Comparisons of predictive performance of breathing pattern variability measured during T-piece, automatic tube compensation, and pressure support ventilation for weaning intensive care unit patients from mechanical ventilation. Crit Care Med 2011, 39: 2253-2262. 10.1097/CCM.0b013e31822279edPubMedCrossRef
32.
go back to reference Haberthur C, Mols G, Elsasser S, Bingisser R, Stocker R, Guttmann J: Extubation after breathing trials with automatic tube compensation, T-tube, or pressure support ventilation. Acta Anaesthesiol Scand 2002, 46: 973-979. 10.1034/j.1399-6576.2002.460808.xPubMedCrossRef Haberthur C, Mols G, Elsasser S, Bingisser R, Stocker R, Guttmann J: Extubation after breathing trials with automatic tube compensation, T-tube, or pressure support ventilation. Acta Anaesthesiol Scand 2002, 46: 973-979. 10.1034/j.1399-6576.2002.460808.xPubMedCrossRef
33.
go back to reference Esteban A, Alía I, Gordo F, Fernández R, Solsona JF, Vallverdú I, Macías S, Allegue JM, Blanco J, Carriedo D, León M, De La Cal MA, Taboada F, Gonzalez De Velasco J, Palazón E, Carrizosa F, Tomás R, Suarez J, Goldwasser RS: Extubation outcome after spontaneous breathing trials with T-tube or pressure support ventilation: The Spanish Lung Failure Collaborative Group. Am J Respir Crit Care Med 1997, 156: 459-465. 10.1164/ajrccm.156.2.9610109PubMedCrossRef Esteban A, Alía I, Gordo F, Fernández R, Solsona JF, Vallverdú I, Macías S, Allegue JM, Blanco J, Carriedo D, León M, De La Cal MA, Taboada F, Gonzalez De Velasco J, Palazón E, Carrizosa F, Tomás R, Suarez J, Goldwasser RS: Extubation outcome after spontaneous breathing trials with T-tube or pressure support ventilation: The Spanish Lung Failure Collaborative Group. Am J Respir Crit Care Med 1997, 156: 459-465. 10.1164/ajrccm.156.2.9610109PubMedCrossRef
34.
go back to reference Laghi F, Goyal A: Auto-PEEP in respiratory failure. Minerva Anestesiol 2012, 78: 201-221.PubMed Laghi F, Goyal A: Auto-PEEP in respiratory failure. Minerva Anestesiol 2012, 78: 201-221.PubMed
35.
go back to reference Koutsoukou A, Bekos B, Sotiropoulou C, Koulouris NG, Roussos C, Milic-Emili J: Effects of positive end-expiratory pressure on gas exchange and expiratory flow limitation in adult respiratory distress syndrome. Crit Care Med 2002, 30: 1941-1949. 10.1097/00003246-200209000-00001PubMedCrossRef Koutsoukou A, Bekos B, Sotiropoulou C, Koulouris NG, Roussos C, Milic-Emili J: Effects of positive end-expiratory pressure on gas exchange and expiratory flow limitation in adult respiratory distress syndrome. Crit Care Med 2002, 30: 1941-1949. 10.1097/00003246-200209000-00001PubMedCrossRef
36.
go back to reference Adams JA, Zabaleta IA, Stroh D, Johnson P, Sackner MA: Tidal volume measurements in newborns using respiratory inductive plethysmography. Am Rev Respir Dis 1993, 148: 585-588. 10.1164/ajrccm/148.3.585PubMedCrossRef Adams JA, Zabaleta IA, Stroh D, Johnson P, Sackner MA: Tidal volume measurements in newborns using respiratory inductive plethysmography. Am Rev Respir Dis 1993, 148: 585-588. 10.1164/ajrccm/148.3.585PubMedCrossRef
37.
go back to reference Markhorst DG, Van Gestel JP, Van Genderingen HR, Haitsma JJ, Lachmann B, Van Vught AJ: Respiratory inductive plethysmography accuracy at varying PEEP levels and degrees of acute lung injury. J Med Eng Technol 2006, 30: 166-175. 10.1080/03091900500529826PubMedCrossRef Markhorst DG, Van Gestel JP, Van Genderingen HR, Haitsma JJ, Lachmann B, Van Vught AJ: Respiratory inductive plethysmography accuracy at varying PEEP levels and degrees of acute lung injury. J Med Eng Technol 2006, 30: 166-175. 10.1080/03091900500529826PubMedCrossRef
Metadata
Title
Electrical activity of the diaphragm (EAdi) as a monitoring parameter in difficult weaning from respirator: a pilot study
Authors
Jürgen Barwing
Cristina Pedroni
Ulrike Olgemöller
Michael Quintel
Onnen Moerer
Publication date
01-08-2013
Publisher
BioMed Central
Published in
Critical Care / Issue 4/2013
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc12865

Other articles of this Issue 4/2013

Critical Care 4/2013 Go to the issue