Published in: EJNMMI Research 1/2013

Open Access 01-12-2013 | Original research

Optimized dose regimen for whole-body FDG-PET imaging

Published in: EJNMMI Research | Issue 1/2013

Login to get access

Abstract

Background

The European Association of Nuclear Medicine procedure guidelines for whole-body fluorodeoxyglucose positron-emission tomography (FDG-PET) scanning prescribe a dose proportional to the patient’s body mass. However, clinical practice shows degraded image quality in obese patients indicating that using an FDG dose proportional to body mass does not overcome size-related degradation of the image quality. The aim of this study was to optimize the administered FDG dose as a function of the patient’s body mass or a different patient-dependent parameter, providing whole-body FDG-PET images of a more constant quality.

Methods

Using a linear relation between administered dose and body mass, FDG-PET imaging was performed on two PET/computed tomography scanners (Biograph TruePoint and Biograph mCT, Siemens). Image quality was assessed by the signal-to-noise ratio (SNR) in the liver in 102 patients with a body mass of 46 to 130 kg. Moreover, the best correlating patient-dependent parameter was derived, and an optimized FDG dose regimen was determined. This optimized dose regimen was validated on the Biograph TruePoint system in 42 new patients. Furthermore, this relation was verified by a simulation study, in which patients with different body masses were simulated with cylindrical phantoms.

Results

As expected, both PET systems showed a significant decrease in SNR with increasing patient’s body mass when using a linear dosage. When image quality was fitted to the patient-dependent parameters, the fit with the patient’s body mass had the highest R2. The optimized dose regimen was found to be A new = c/t × m2, where m is the body mass, t is the acquisition time per bed position and c is a constant (depending on scanner type). Using this relation, SNR no longer varied with the patient’s body mass. This quadratic relation between dose and body mass was confirmed by the simulation study.

Conclusion

A quadratic relation between FDG dose and the patient’s body mass is recommended. Both simulations and clinical observations confirm that image quality remains constant across patients when this quadratic dose regimen is used.
Appendix
Available only for authorised users
Literature
1.
go back to reference Boellaard R, O’Doherty MJ, Weber WA, Mottaghy FM, Lonsdale MN, Stroobants SG, Oyen WJG, Kotzerke J, Hoekstra OS, Pruim J, Marsden PK, Tatsch K, Hoekstra CJ, Visser EP, Arends B, Verzijlbergen FJ, Zijlstra JM, Comans EFI, Lammertsma AA, Paans AM, Willemsen AT, Beyer T, Bockisch A, Schaefer-Prokop C, Delbeke D, Baum RP, Chiti A, Krause BJ: FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0. Eur J Nucl Med Mol Imaging 2010,37(1):181–200. 10.1007/s00259-009-1297-4PubMedCentralCrossRefPubMed Boellaard R, O’Doherty MJ, Weber WA, Mottaghy FM, Lonsdale MN, Stroobants SG, Oyen WJG, Kotzerke J, Hoekstra OS, Pruim J, Marsden PK, Tatsch K, Hoekstra CJ, Visser EP, Arends B, Verzijlbergen FJ, Zijlstra JM, Comans EFI, Lammertsma AA, Paans AM, Willemsen AT, Beyer T, Bockisch A, Schaefer-Prokop C, Delbeke D, Baum RP, Chiti A, Krause BJ: FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0. Eur J Nucl Med Mol Imaging 2010,37(1):181–200. 10.1007/s00259-009-1297-4PubMedCentralCrossRefPubMed
2.
go back to reference Boellaard R, Oyen WJG, Hoekstra CJ, Hoekstra OS, Visser EP, Willemsen AT, Arends B, Verzijlbergen FJ, Zijlstra J, Paans AM, Comans EFI, Pruim J: The Netherlands protocol for standardisation and quantification of FDG whole body PET studies in multi-centre trials. Eur J Nucl Med Mol Imaging 2008,35(12):2320–2333. 10.1007/s00259-008-0874-2CrossRefPubMed Boellaard R, Oyen WJG, Hoekstra CJ, Hoekstra OS, Visser EP, Willemsen AT, Arends B, Verzijlbergen FJ, Zijlstra J, Paans AM, Comans EFI, Pruim J: The Netherlands protocol for standardisation and quantification of FDG whole body PET studies in multi-centre trials. Eur J Nucl Med Mol Imaging 2008,35(12):2320–2333. 10.1007/s00259-008-0874-2CrossRefPubMed
3.
go back to reference Matsumoto K, Matsuura H, Minota E, Sakamoto S, Nakamoto Y, Senda M: Evaluation of optimized injection dose and acquisition time using body mass index for three-dimensional whole-body FDG-PET. Nihon Hoshasen Gijutsu Gakkai Zasshi 2004,60(11):1564–1573.PubMed Matsumoto K, Matsuura H, Minota E, Sakamoto S, Nakamoto Y, Senda M: Evaluation of optimized injection dose and acquisition time using body mass index for three-dimensional whole-body FDG-PET. Nihon Hoshasen Gijutsu Gakkai Zasshi 2004,60(11):1564–1573.PubMed
4.
go back to reference Watson CC, Casey ME, Bendriem B, Carney JP, Townsend DW, Eberl S, Meikle S, Difilippo FP: Optimizing injected dose in clinical PET by accurately modeling the counting-rate response functions specific to individual patient scans. J Nucl Med 2005,46(11):1825–1834.PubMed Watson CC, Casey ME, Bendriem B, Carney JP, Townsend DW, Eberl S, Meikle S, Difilippo FP: Optimizing injected dose in clinical PET by accurately modeling the counting-rate response functions specific to individual patient scans. J Nucl Med 2005,46(11):1825–1834.PubMed
5.
go back to reference Everaert H, Vanhove C, Lahoutte T, Muylle K, Caveliers V, Bossuyt A, Franken PR: Optimal dose of 18F-FDG required for whole-body PET using an LSO PET camera. Eur J Nucl Med Mol Imaging 2003,30(12):1615–1619. 10.1007/s00259-003-1317-8CrossRefPubMed Everaert H, Vanhove C, Lahoutte T, Muylle K, Caveliers V, Bossuyt A, Franken PR: Optimal dose of 18F-FDG required for whole-body PET using an LSO PET camera. Eur J Nucl Med Mol Imaging 2003,30(12):1615–1619. 10.1007/s00259-003-1317-8CrossRefPubMed
6.
go back to reference Halpern BS, Dahlbom M, Auerbach MA, Schiepers C, Fueger BJ, Weber WA, Silverman DHS, Ratib O, Czernin J: Optimizing imaging protocols for overweight and obese patients: a lutetium orthosilicate PET/CT study. J Nucl Med 2005,46(4):603–607.PubMed Halpern BS, Dahlbom M, Auerbach MA, Schiepers C, Fueger BJ, Weber WA, Silverman DHS, Ratib O, Czernin J: Optimizing imaging protocols for overweight and obese patients: a lutetium orthosilicate PET/CT study. J Nucl Med 2005,46(4):603–607.PubMed
7.
go back to reference Halpern BS, Dahlbom M, Quon A, Schiepers C, Waldherr C, Silverman DH, Ratib O, Czernin J: Impact of patient weight and emission scan duration on PET/CT image quality and lesion detectability. J Nucl Med 2004,45(5):797–801.PubMed Halpern BS, Dahlbom M, Quon A, Schiepers C, Waldherr C, Silverman DH, Ratib O, Czernin J: Impact of patient weight and emission scan duration on PET/CT image quality and lesion detectability. J Nucl Med 2004,45(5):797–801.PubMed
8.
go back to reference Masuda Y, Kondo C, Matsuo Y, Uetani M, Kusakabe K: Comparison of imaging protocols for 18F-FDG PET/CT in overweight patients: optimizing scan duration versus administered dose. J Nucl Med 2009,50(6):844–848. 10.2967/jnumed.108.060590CrossRefPubMed Masuda Y, Kondo C, Matsuo Y, Uetani M, Kusakabe K: Comparison of imaging protocols for 18F-FDG PET/CT in overweight patients: optimizing scan duration versus administered dose. J Nucl Med 2009,50(6):844–848. 10.2967/jnumed.108.060590CrossRefPubMed
9.
go back to reference Accorsi R, Karp JS, Surti S: Improved dose regimen in pediatric PET. J Nucl Med 2010,51(2):293–300. 10.2967/jnumed.109.066332CrossRefPubMed Accorsi R, Karp JS, Surti S: Improved dose regimen in pediatric PET. J Nucl Med 2010,51(2):293–300. 10.2967/jnumed.109.066332CrossRefPubMed
10.
go back to reference Conti M: State of the art and challenges of time-of-flight PET. Phys Med 2009,25(1):1–11. 10.1016/j.ejmp.2008.10.001CrossRefPubMed Conti M: State of the art and challenges of time-of-flight PET. Phys Med 2009,25(1):1–11. 10.1016/j.ejmp.2008.10.001CrossRefPubMed
11.
go back to reference Conti M: Focus on time-of-flight PET: the benefits of improved time resolution. Eur J Nucl Med Mol Imaging 2011,38(6):1147–1157. 10.1007/s00259-010-1711-yCrossRefPubMed Conti M: Focus on time-of-flight PET: the benefits of improved time resolution. Eur J Nucl Med Mol Imaging 2011,38(6):1147–1157. 10.1007/s00259-010-1711-yCrossRefPubMed
12.
go back to reference El Fakhri G, Surti S, Trott CM, Scheuermann J, Karp JS: Improvement in lesion detection with whole-body oncologic time-of-flight PET. J Nucl Med 2011,52(3):347–353. 10.2967/jnumed.110.080382PubMedCentralCrossRefPubMed El Fakhri G, Surti S, Trott CM, Scheuermann J, Karp JS: Improvement in lesion detection with whole-body oncologic time-of-flight PET. J Nucl Med 2011,52(3):347–353. 10.2967/jnumed.110.080382PubMedCentralCrossRefPubMed
13.
go back to reference Panin VY, Kehren F, Michel C, Casey M: Fully 3-D PET reconstruction with system matrix derived from point source measurements. IEEE Trans Med Imaging 2006,25(7):907–921.CrossRefPubMed Panin VY, Kehren F, Michel C, Casey M: Fully 3-D PET reconstruction with system matrix derived from point source measurements. IEEE Trans Med Imaging 2006,25(7):907–921.CrossRefPubMed
15.
go back to reference Hallynck TH, Soep HH, Thomis JA, Boelaert J, Daneels R, Dettli L: Should clearance be normalised to body surface or to lean body mass? Br J Clin Pharmacol 1981, 11: 523–526. 10.1111/j.1365-2125.1981.tb01163.xPubMedCentralCrossRefPubMed Hallynck TH, Soep HH, Thomis JA, Boelaert J, Daneels R, Dettli L: Should clearance be normalised to body surface or to lean body mass? Br J Clin Pharmacol 1981, 11: 523–526. 10.1111/j.1365-2125.1981.tb01163.xPubMedCentralCrossRefPubMed
16.
go back to reference Graham MM, Badawi RD, Wahl RL: Variations in PET/CT methodology for oncologic imaging at U.S. academic medical centers: an imaging response assessment team survey. J Nucl Med 2011,52(2):311–317. 10.2967/jnumed.109.074104PubMedCentralCrossRefPubMed Graham MM, Badawi RD, Wahl RL: Variations in PET/CT methodology for oncologic imaging at U.S. academic medical centers: an imaging response assessment team survey. J Nucl Med 2011,52(2):311–317. 10.2967/jnumed.109.074104PubMedCentralCrossRefPubMed
17.
go back to reference Beyer T, Czernin J, Freudenberg LS: Variations in clinical PET/CT operations: results of an international survey of active PET/CT users. J Nucl Med 2011,52(2):303–310. 10.2967/jnumed.110.079624CrossRefPubMed Beyer T, Czernin J, Freudenberg LS: Variations in clinical PET/CT operations: results of an international survey of active PET/CT users. J Nucl Med 2011,52(2):303–310. 10.2967/jnumed.110.079624CrossRefPubMed
18.
go back to reference Stauss J, Franzius C, Pfluger T, Juergens KU, Biassoni L, Begent J, Kluge R, Amthauer H, Voelker T, Højgaard L, Barrington S, Hain S, Lynch T, Hahn K: Guidelines for 18F-FDG PET and PET-CT imaging in paediatric oncology. Eur J Nucl Med Mol Imaging 2008,35(8):1581–1588. 10.1007/s00259-008-0826-xCrossRefPubMed Stauss J, Franzius C, Pfluger T, Juergens KU, Biassoni L, Begent J, Kluge R, Amthauer H, Voelker T, Højgaard L, Barrington S, Hain S, Lynch T, Hahn K: Guidelines for 18F-FDG PET and PET-CT imaging in paediatric oncology. Eur J Nucl Med Mol Imaging 2008,35(8):1581–1588. 10.1007/s00259-008-0826-xCrossRefPubMed
Metadata
Title
Optimized dose regimen for whole-body FDG-PET imaging
Publication date
01-12-2013
Published in
EJNMMI Research / Issue 1/2013
Electronic ISSN: 2191-219X
DOI
https://doi.org/10.1186/2191-219X-3-63