Published in: EJNMMI Research 1/2013

Open Access 01-12-2013 | Original research

Assessment of acquisition protocols for routine imaging of Y-90 using PET/CT

Authors: Thomas Carlier, Thomas Eugène, Caroline Bodet-Milin, Etienne Garin, Catherine Ansquer, Caroline Rousseau, Ludovic Ferrer, Jacques Barbet, Frédéric Schoenahl, Françoise Kraeber-Bodéré

Published in: EJNMMI Research | Issue 1/2013

Login to get access

Abstract

Background

Despite the early theoretical prediction of the 0+-0+ transition of 90Zr, 90Y-PET underwent only recently a growing interest for the development of imaging radioembolization of liver tumors. The aim of this work was to determine the minimum detectable activity (MDA) of 90Y by PET imaging and the impact of time-of-flight (TOF) reconstruction on detectability and quantitative accuracy according to the lesion size.

Methods

The study was conducted using a Siemens Biograph® mCT with a 22 cm large axial field of view. An IEC torso-shaped phantom containing five coplanar spheres was uniformly filled to achieve sphere-to-background ratios of 40:1. The phantom was imaged nine times in 14 days over 30 min. Sinograms were reconstructed with and without TOF information. A contrast-to-noise ratio (CNR) index was calculated using the Rose criterion, taking partial volume effects into account. The impact of reconstruction parameters on quantification accuracy, detectability, and spatial localization of the signal was investigated. Finally, six patients with hepatocellular carcinoma and four patients included in different 90Y-based radioimmunotherapy protocols were enrolled for the evaluation of the imaging parameters in a clinical situation.

Results

The highest CNR was achieved with one iteration for both TOF and non-TOF reconstructions. The MDA, however, was found to be lower with TOF than with non-TOF reconstruction. There was no gain by adding TOF information in terms of CNR for concentrations higher than 2 to 3 MBq mL−1, except for infra-centimetric lesions. Recovered activity was highly underestimated when a single iteration or non-TOF reconstruction was used (10% to 150% less depending on the lesion size). The MDA was estimated at 1 MBq mL−1 for a TOF reconstruction and infra-centimetric lesions. Images from patients treated with microspheres were clinically relevant, unlike those of patients who received systemic injections of 90Y.

Conclusions

Only one iteration and TOF were necessary to achieve an MDA around 1 MBq mL−1 and the most accurate localization of lesions. For precise quantification, at least three iterations gave the best performance, using TOF reconstruction and keeping an MDA of roughly 1 MBq mL−1. One and three iterations were mandatory to prevent false positive results for quantitative analysis of clinical data.
Appendix
Available only for authorised users
Literature
1.
go back to reference Morschhauser F, Kraeber-Bodéré F, Wegener WA, Harousseau JL, Petillon MO, Huglo D, Trümper LH, Meller J, Pfreundschuh M, Kirsch CM, Naumann R, Kropp J, Horne H, Teoh N, Le Gouill S, Bodet-milin C, Chatal F, Goldenberg DM: High rates of durable responses with anti-CD22 fractionated radioimmunotherapy: results of a multicenter, phase I/II study in non-Hodgkin's lymphoma. J Clin Oncol 2010, 28: 3709–3716. 10.1200/JCO.2009.27.7863CrossRefPubMed Morschhauser F, Kraeber-Bodéré F, Wegener WA, Harousseau JL, Petillon MO, Huglo D, Trümper LH, Meller J, Pfreundschuh M, Kirsch CM, Naumann R, Kropp J, Horne H, Teoh N, Le Gouill S, Bodet-milin C, Chatal F, Goldenberg DM: High rates of durable responses with anti-CD22 fractionated radioimmunotherapy: results of a multicenter, phase I/II study in non-Hodgkin's lymphoma. J Clin Oncol 2010, 28: 3709–3716. 10.1200/JCO.2009.27.7863CrossRefPubMed
2.
go back to reference Illidge TM: Radioimmunotherapy of lymphoma: a treatment approach ahead of its time or past its sell-by date? J Clin Oncol 2010, 28: 2944–2946. 10.1200/JCO.2009.26.8748CrossRefPubMed Illidge TM: Radioimmunotherapy of lymphoma: a treatment approach ahead of its time or past its sell-by date? J Clin Oncol 2010, 28: 2944–2946. 10.1200/JCO.2009.26.8748CrossRefPubMed
3.
go back to reference Bodei L, Cremonesi M, Grana C, Rocca P, Bartolomei M, Chinol M, Paganelli G: Receptor radionuclide therapy with 90Y-[DOTA]0-Tyr3-octreotide (90Y-DOTATOC) in neuroendocrine tumours. Eur J Nucl Med Mol Imaging 2004, 31: 1038–1046.CrossRefPubMed Bodei L, Cremonesi M, Grana C, Rocca P, Bartolomei M, Chinol M, Paganelli G: Receptor radionuclide therapy with 90Y-[DOTA]0-Tyr3-octreotide (90Y-DOTATOC) in neuroendocrine tumours. Eur J Nucl Med Mol Imaging 2004, 31: 1038–1046.CrossRefPubMed
4.
go back to reference Ansquer C, Kraeber-Bodéré F, Chatal JF: Current status and perspectives in peptide receptor radiation therapy. Cur Pharm Des 2009, 15: 2453–2462. 10.2174/138161209788682262CrossRef Ansquer C, Kraeber-Bodéré F, Chatal JF: Current status and perspectives in peptide receptor radiation therapy. Cur Pharm Des 2009, 15: 2453–2462. 10.2174/138161209788682262CrossRef
5.
go back to reference Salem R, Thurston KG: Radioembolization with 90Yttrium microspheres: a state-of-the-art brachytherapy treatment for primary and secondary liver malignancies: part 1—technical and methodologic considerations. J Vasc Interv Radiol 2006, 17: 1251–1278. 10.1097/01.RVI.0000233785.75257.9ACrossRefPubMed Salem R, Thurston KG: Radioembolization with 90Yttrium microspheres: a state-of-the-art brachytherapy treatment for primary and secondary liver malignancies: part 1—technical and methodologic considerations. J Vasc Interv Radiol 2006, 17: 1251–1278. 10.1097/01.RVI.0000233785.75257.9ACrossRefPubMed
6.
go back to reference Bult W, Vente MAD, Zonnenberg BA, Van Het Schip AD, Nijsen JFW: Microsphere radioembolization of liver malignancies: current developments. Q J Nucl Med Mol Imaging 2009, 53: 325–335.PubMed Bult W, Vente MAD, Zonnenberg BA, Van Het Schip AD, Nijsen JFW: Microsphere radioembolization of liver malignancies: current developments. Q J Nucl Med Mol Imaging 2009, 53: 325–335.PubMed
7.
go back to reference Ettorre GM, Santoro R, Puoti C, Sciuto R, Carpanese L, Antonini M, Antonucci G, Maini CL, Miglioresi L, Vennarecci G: Short-term follow-up of radioembolization with yttrium-90 microspheres before liver transplantation: new perspectives in advances hepatocellular carcinoma. Transplantation 2010, 90: 930–931.CrossRefPubMed Ettorre GM, Santoro R, Puoti C, Sciuto R, Carpanese L, Antonini M, Antonucci G, Maini CL, Miglioresi L, Vennarecci G: Short-term follow-up of radioembolization with yttrium-90 microspheres before liver transplantation: new perspectives in advances hepatocellular carcinoma. Transplantation 2010, 90: 930–931.CrossRefPubMed
8.
go back to reference Carrasquillo JA, White JD, Paik CH, Raubitschek A, Le N, Rotman M, Brechbiel MW, Gansow OA, Top LE, Perentesis P, Reynolds JC, Nelson DL, Waldmann TA: Similarities and differences in 111In- and 90Y-labeled 1B4M-DTPA antiTac monoclonal antibody distribution. J Nucl Med 1999, 40: 268–276.PubMed Carrasquillo JA, White JD, Paik CH, Raubitschek A, Le N, Rotman M, Brechbiel MW, Gansow OA, Top LE, Perentesis P, Reynolds JC, Nelson DL, Waldmann TA: Similarities and differences in 111In- and 90Y-labeled 1B4M-DTPA antiTac monoclonal antibody distribution. J Nucl Med 1999, 40: 268–276.PubMed
9.
go back to reference Minarik D, Sjogreen GK, Ljungberg M: Evaluation of quantitative (90)Y SPECT based on experimental phantom studies. Phys Med Biol 2008, 53: 5689–5703. 10.1088/0031-9155/53/20/008CrossRefPubMed Minarik D, Sjogreen GK, Ljungberg M: Evaluation of quantitative (90)Y SPECT based on experimental phantom studies. Phys Med Biol 2008, 53: 5689–5703. 10.1088/0031-9155/53/20/008CrossRefPubMed
10.
go back to reference Rault E, Vandenberghe S, Staelens S, Van Holen R, Lemahieu I: Optimization of 90Y bremsstrahlung image reconstruction using multiple energy window subsets [abstract]. J Nucl Med 2008, 49: 399P. 10.2967/jnumed.107.033530CrossRef Rault E, Vandenberghe S, Staelens S, Van Holen R, Lemahieu I: Optimization of 90Y bremsstrahlung image reconstruction using multiple energy window subsets [abstract]. J Nucl Med 2008, 49: 399P. 10.2967/jnumed.107.033530CrossRef
11.
go back to reference Walrand S, Hesse M, Demonceau G, Pauwels S, Jamar F: Yttrium-90-labeled microsphere tracking during liver selective internal radiotherapy by bremsstrahlung pinhole SPECT: feasibility study and evaluation in an abdominal phantom. EJNMMI Research 2011, 1: 32–46. 10.1186/2191-219X-1-32CrossRefPubMed Walrand S, Hesse M, Demonceau G, Pauwels S, Jamar F: Yttrium-90-labeled microsphere tracking during liver selective internal radiotherapy by bremsstrahlung pinhole SPECT: feasibility study and evaluation in an abdominal phantom. EJNMMI Research 2011, 1: 32–46. 10.1186/2191-219X-1-32CrossRefPubMed
12.
go back to reference Ford K: Predicted 0+ level of Zr90. Phys Rev 1955, 98: 1516. 10.1103/PhysRev.98.1516CrossRef Ford K: Predicted 0+ level of Zr90. Phys Rev 1955, 98: 1516. 10.1103/PhysRev.98.1516CrossRef
13.
go back to reference Selwyn RG, Nickles RJ, Thomadsen BR, DeWerd LA, Micka JA: A new internal pair production branching ratio of 90Y: the development of a non-destructive assay for 90 Y and 90 Sr. Appl Radiat Isot 2007, 65: 318–327. 10.1016/j.apradiso.2006.08.009CrossRefPubMed Selwyn RG, Nickles RJ, Thomadsen BR, DeWerd LA, Micka JA: A new internal pair production branching ratio of 90Y: the development of a non-destructive assay for 90 Y and 90 Sr. Appl Radiat Isot 2007, 65: 318–327. 10.1016/j.apradiso.2006.08.009CrossRefPubMed
14.
go back to reference Nickles RJ, Roberts AD, Nye JA, Converse AK, Barnhart TE, Avila-Rodriguez MA, Sundaresan R, Dick DW, Hammas RJ, Thomadsen BR: Assaying and PET imaging of yttrium-90: 1> > 34 ppm > 0. In IEEE Nuclear Science Symposium Conference Record: October 16–22 2004; Casaccia, Italy. New York: IEEE; 2004:3412–3414.CrossRef Nickles RJ, Roberts AD, Nye JA, Converse AK, Barnhart TE, Avila-Rodriguez MA, Sundaresan R, Dick DW, Hammas RJ, Thomadsen BR: Assaying and PET imaging of yttrium-90: 1> > 34 ppm > 0. In IEEE Nuclear Science Symposium Conference Record: October 16–22 2004; Casaccia, Italy. New York: IEEE; 2004:3412–3414.CrossRef
15.
go back to reference Lhommel R, van Elmbt L, Goffette P, Van den Eynde M, Jamar F, Pauwels S, Walrand S: Feasibility of 90Y TOF PET-based dosimetry in liver metastasis therapy using SIR-spheres. Eur J Nucl Med Mol Imaging 2010, 37: 1654–1662. 10.1007/s00259-010-1470-9CrossRefPubMed Lhommel R, van Elmbt L, Goffette P, Van den Eynde M, Jamar F, Pauwels S, Walrand S: Feasibility of 90Y TOF PET-based dosimetry in liver metastasis therapy using SIR-spheres. Eur J Nucl Med Mol Imaging 2010, 37: 1654–1662. 10.1007/s00259-010-1470-9CrossRefPubMed
16.
go back to reference Gates VL, Esmail AAH, Marshall K, Spies S, Salem R: Internal pair production of 90Y permits hepatic localization of microspheres using routine PET: proof of concept. J Nucl Med 2011, 52: 72–76. 10.2967/jnumed.110.080986CrossRefPubMed Gates VL, Esmail AAH, Marshall K, Spies S, Salem R: Internal pair production of 90Y permits hepatic localization of microspheres using routine PET: proof of concept. J Nucl Med 2011, 52: 72–76. 10.2967/jnumed.110.080986CrossRefPubMed
17.
go back to reference Walrand S, Jamar F, van Elmbt L, Lhommel R, Bekonde EB, Pauwels S: 4-step renal dosimetry dependant on cortex geometry applied to 90 Y peptide receptor radiotherapy: evaluation using a fillable kidney phantom imaged by 90 Y PET. J Nucl Med 2010, 51: 1969–1973. 10.2967/jnumed.110.080093CrossRefPubMed Walrand S, Jamar F, van Elmbt L, Lhommel R, Bekonde EB, Pauwels S: 4-step renal dosimetry dependant on cortex geometry applied to 90 Y peptide receptor radiotherapy: evaluation using a fillable kidney phantom imaged by 90 Y PET. J Nucl Med 2010, 51: 1969–1973. 10.2967/jnumed.110.080093CrossRefPubMed
18.
go back to reference Baba S, Abe K, Sawamoto H, Tanabe Y, Maruoka Y, Sasaki M, Honda H: Visualization and dose estimation of 90Y-ibritumomab-tiukisetan accumulation in lymphoma patients using TOF PET/CT [abstract]. J Nucl Med 1987, 2011: 52. Baba S, Abe K, Sawamoto H, Tanabe Y, Maruoka Y, Sasaki M, Honda H: Visualization and dose estimation of 90Y-ibritumomab-tiukisetan accumulation in lymphoma patients using TOF PET/CT [abstract]. J Nucl Med 1987, 2011: 52.
19.
go back to reference van Elmbt L, Vandenberghe S, Walrand S, Pauwels S, Jamar F: Comparison of yttrium-90 quantitative imaging by TOF and non-TOF PET in a phantom of liver selective internal radiotherapy. Phys Med Biol 2011, 56: 6759–6777. 10.1088/0031-9155/56/21/001CrossRefPubMed van Elmbt L, Vandenberghe S, Walrand S, Pauwels S, Jamar F: Comparison of yttrium-90 quantitative imaging by TOF and non-TOF PET in a phantom of liver selective internal radiotherapy. Phys Med Biol 2011, 56: 6759–6777. 10.1088/0031-9155/56/21/001CrossRefPubMed
20.
go back to reference Chatal JF, Saccavini JC, Gestin JF, Thédrez P, Curtet C, Kremer M, Guerreau D, Nolibé D, Fumoleau P, Guillard Y: Biodistribution of indium-111-labeled OC 125 monoclonal antibody intraperitoneally injected into patients operated on for ovarian carcinomas. Cancer Res 1989, 49: 3087–3094.PubMed Chatal JF, Saccavini JC, Gestin JF, Thédrez P, Curtet C, Kremer M, Guerreau D, Nolibé D, Fumoleau P, Guillard Y: Biodistribution of indium-111-labeled OC 125 monoclonal antibody intraperitoneally injected into patients operated on for ovarian carcinomas. Cancer Res 1989, 49: 3087–3094.PubMed
21.
go back to reference Jakoby BW, Bercier Y, Conti M, Casey ME, Bendriem B, Townsend DW: Physical and clinical performance of the mCT time-of-flight PET/CT scanner. Phys Med Biol 2011, 56: 2375–2389. 10.1088/0031-9155/56/8/004CrossRefPubMed Jakoby BW, Bercier Y, Conti M, Casey ME, Bendriem B, Townsend DW: Physical and clinical performance of the mCT time-of-flight PET/CT scanner. Phys Med Biol 2011, 56: 2375–2389. 10.1088/0031-9155/56/8/004CrossRefPubMed
22.
go back to reference National Electrical Manufacturers Association (NEMA): Performance measurement of positron emission tomographs. In NEMA standards Publication NU 2–2007. Edited by: NEMA. Rosslyn: National Electrical Manufacturers Association; 2007:11–18. National Electrical Manufacturers Association (NEMA): Performance measurement of positron emission tomographs. In NEMA standards Publication NU 2–2007. Edited by: NEMA. Rosslyn: National Electrical Manufacturers Association; 2007:11–18.
23.
go back to reference Lhommel R, Goffette P, Van den Eynde M, Jamar F, Pauwels S, Bilbao JI, Walrand S: Yttrium-90 TOF PET scan demonstrates high–resolution biodistribution after liver SIRT. Eur J Nucl Med Mol Imaging 2009, 36: 1696. 10.1007/s00259-009-1210-1CrossRefPubMed Lhommel R, Goffette P, Van den Eynde M, Jamar F, Pauwels S, Bilbao JI, Walrand S: Yttrium-90 TOF PET scan demonstrates high–resolution biodistribution after liver SIRT. Eur J Nucl Med Mol Imaging 2009, 36: 1696. 10.1007/s00259-009-1210-1CrossRefPubMed
24.
go back to reference Panin VY, Kehren F, Michel C, Casey M: Fully 3-D PET reconstruction with system matrix derived from point source measurements. IEEE Trans Med Imag 2006, 25: 907–921.CrossRef Panin VY, Kehren F, Michel C, Casey M: Fully 3-D PET reconstruction with system matrix derived from point source measurements. IEEE Trans Med Imag 2006, 25: 907–921.CrossRef
25.
go back to reference Bao Q, Chatziioannou AF: Estimation of the minimum detectable activity of preclinical PET imaging systems with an analytical method. Med Phys 2010, 37: 6070–6083. 10.1118/1.3495817CrossRefPubMed Bao Q, Chatziioannou AF: Estimation of the minimum detectable activity of preclinical PET imaging systems with an analytical method. Med Phys 2010, 37: 6070–6083. 10.1118/1.3495817CrossRefPubMed
26.
go back to reference Goertzen AL, Suk JY, Thompson CJ: Imaging of weak-source distributions in LSO-based small-animal PET scanners. J Nucl Med 2007, 48: 1692–1698. 10.2967/jnumed.107.040584CrossRefPubMed Goertzen AL, Suk JY, Thompson CJ: Imaging of weak-source distributions in LSO-based small-animal PET scanners. J Nucl Med 2007, 48: 1692–1698. 10.2967/jnumed.107.040584CrossRefPubMed
27.
go back to reference Cherry SR, Sorenson JA, Phelps ME: Physics in Nuclear Medicine. 3rd edition. Pennsylvania: Elsevier; 2003. Cherry SR, Sorenson JA, Phelps ME: Physics in Nuclear Medicine. 3rd edition. Pennsylvania: Elsevier; 2003.
28.
go back to reference Stute S, Carlier T, Cristina K, Noblet C, Martineau A, Hutton B, Barnden L, Buvat I: Monte Carlo simulations of clinical PET and SPECT scans: impact of the input data on the simulated images. Phys Med Biol 2011, 56: 6441–6457. 10.1088/0031-9155/56/19/017CrossRefPubMed Stute S, Carlier T, Cristina K, Noblet C, Martineau A, Hutton B, Barnden L, Buvat I: Monte Carlo simulations of clinical PET and SPECT scans: impact of the input data on the simulated images. Phys Med Biol 2011, 56: 6441–6457. 10.1088/0031-9155/56/19/017CrossRefPubMed
29.
go back to reference Bright DS, Newbury DE, Steel EB: Visibility of objects in computer simulations of noisy micrographs. J Microsc 1997, 189: 25–42.CrossRef Bright DS, Newbury DE, Steel EB: Visibility of objects in computer simulations of noisy micrographs. J Microsc 1997, 189: 25–42.CrossRef
30.
go back to reference Watson CC, Casey ME, Eriksson L, Mulnix T, Adams D, Bendriem B: NEMA NU 2 performance tests for scanners with intrinsic radioactivity. J Nucl Med 2004, 45: 822–826.PubMed Watson CC, Casey ME, Eriksson L, Mulnix T, Adams D, Bendriem B: NEMA NU 2 performance tests for scanners with intrinsic radioactivity. J Nucl Med 2004, 45: 822–826.PubMed
31.
go back to reference Selwyn R, Micka J, DeWerd L, Nickles R, Thomadsen B: Technical note: the calibration of 90 Y-labeled SIR-Spheres® using a nondestructive spectroscopy assay. Med Phys 2008, 35: 1278–1279. 10.1118/1.2889621CrossRefPubMed Selwyn R, Micka J, DeWerd L, Nickles R, Thomadsen B: Technical note: the calibration of 90 Y-labeled SIR-Spheres® using a nondestructive spectroscopy assay. Med Phys 2008, 35: 1278–1279. 10.1118/1.2889621CrossRefPubMed
32.
go back to reference Shakirin G, Braess H, Fielder F, Kunath D, Laube K, Parodi K, Priegnitz M, Enghardt W: Implementation and workflow for PET monitoring of therapeutic ion irradiation: a comparison of in-beam, in-room, and off-line techniques. Phys Med Biol 2011, 56: 1281–1298. 10.1088/0031-9155/56/5/004CrossRefPubMed Shakirin G, Braess H, Fielder F, Kunath D, Laube K, Parodi K, Priegnitz M, Enghardt W: Implementation and workflow for PET monitoring of therapeutic ion irradiation: a comparison of in-beam, in-room, and off-line techniques. Phys Med Biol 2011, 56: 1281–1298. 10.1088/0031-9155/56/5/004CrossRefPubMed
33.
go back to reference Campbell JM, Wong CO, Muzik O, Marples B, Joiner M, Burmeister J: Early dose response to yttrium-90 microsphere treatment of metastatic liver cancer by patient-specific method using single photon emission computed tomography and positron emission tomography. Int J Radiat Oncol Biol Phys 2009, 74: 313–320. 10.1016/j.ijrobp.2008.12.058CrossRefPubMed Campbell JM, Wong CO, Muzik O, Marples B, Joiner M, Burmeister J: Early dose response to yttrium-90 microsphere treatment of metastatic liver cancer by patient-specific method using single photon emission computed tomography and positron emission tomography. Int J Radiat Oncol Biol Phys 2009, 74: 313–320. 10.1016/j.ijrobp.2008.12.058CrossRefPubMed
34.
go back to reference Stute S, Benoit D, Martineau A, Rehfeld NS, Buvat I: A method for accurate modelling of the crystal response function at a crystal sub-level applied to PET reconstruction. Phys Med Biol 2011, 56: 793–809. 10.1088/0031-9155/56/3/016CrossRefPubMed Stute S, Benoit D, Martineau A, Rehfeld NS, Buvat I: A method for accurate modelling of the crystal response function at a crystal sub-level applied to PET reconstruction. Phys Med Biol 2011, 56: 793–809. 10.1088/0031-9155/56/3/016CrossRefPubMed
35.
go back to reference Watson CC: Extension of single scatter simulation to scatter correction of time-of-flight PET. IEEE Trans Nucl Sci 2007, 54: 1679–1686.CrossRef Watson CC: Extension of single scatter simulation to scatter correction of time-of-flight PET. IEEE Trans Nucl Sci 2007, 54: 1679–1686.CrossRef
Metadata
Title
Assessment of acquisition protocols for routine imaging of Y-90 using PET/CT
Authors
Thomas Carlier
Thomas Eugène
Caroline Bodet-Milin
Etienne Garin
Catherine Ansquer
Caroline Rousseau
Ludovic Ferrer
Jacques Barbet
Frédéric Schoenahl
Françoise Kraeber-Bodéré
Publication date
01-12-2013
Publisher
Springer Berlin Heidelberg
Published in
EJNMMI Research / Issue 1/2013
Electronic ISSN: 2191-219X
DOI
https://doi.org/10.1186/2191-219X-3-11

Other articles of this Issue 1/2013

EJNMMI Research 1/2013 Go to the issue