Use of a beta microprobe system to measure arterial input function in PET via an arteriovenous shunt in rats

Authors: Geoff Warnock, Mohamed-Ali Bahri, David Goblet, Fabrice Giacomelli, Christian Lemaire, Joel Aerts, Alain Seret, Xavier Langlois, Andre Luxen, Alain Plenevaux

Published in: EJNMMI Research | Issue 1/2011

Login to get access

Abstract

Background

Kinetic modeling of physiological function using imaging techniques requires the accurate measurement of the time-activity curve of the tracer in plasma, known as the arterial input function (IF). The measurement of IF can be achieved through manual blood sampling, the use of small counting systems such as beta microprobes, or by derivation from PET images. Previous studies using beta microprobe systems to continuously measure IF have suffered from high background counts.

Methods

In the present study, a light-insensitive beta microprobe with a temporal resolution of up to 1 s was used in combination with a pump-driven femoral arteriovenous shunt to measure IF in rats. The shunt apparatus was designed such that the placement of the beta microprobe was highly reproducible. The probe-derived IF was compared to that obtained from manual sampling at 5-s intervals and IF derived from a left ventricle VOI in a dynamic PET image of the heart.

Results

Probe-derived IFs were very well matched to that obtained by "gold standard" manual blood sampling, but with an increased temporal resolution of up to 1 s. The area under the curve (AUC) ratio between probe- and manually derived IFs was 1.07 ± 0.05 with a coefficient of variation of 0.04. However, image-derived IFs were significantly underestimated compared to the manually sampled IFs, with an AUC ratio of 0.76 ± 0.24 with a coefficient of variation of 0.32.

Conclusions

IF derived from the beta microprobe accurately represented the IF as measured by blood sampling, was reproducible, and was more accurate than an image-derived technique. The use of the shunt removed problems of tissue-background activity, and the use of a light-tight probe with minimal gamma sensitivity refined the system. The probe/shunt apparatus can be used in both microprobe and PET studies.
Appendix
Available only for authorised users
Literature
1.
go back to reference Diehl KH, Hull R, Morton D, Pfister R, Rabemampianina Y, Smith D, Vidal JM, van de Vorstenbosch C: A good practice guide to the administration of substances and removal of blood, including routes and volumes. J Appl Toxicol 2001, 21: 15–23. 10.1002/jat.727PubMedCrossRef Diehl KH, Hull R, Morton D, Pfister R, Rabemampianina Y, Smith D, Vidal JM, van de Vorstenbosch C: A good practice guide to the administration of substances and removal of blood, including routes and volumes. J Appl Toxicol 2001, 21: 15–23. 10.1002/jat.727PubMedCrossRef
2.
go back to reference Sharp TL, Dence CS, Engelbach JA, Herrero P, Gropler RJ, Welch MJ: Techniques necessary for multiple tracer quantitative small-animal imaging studies. Nucl Med Biol 2005, 32: 875–884. 10.1016/j.nucmedbio.2005.05.010PubMedCrossRef Sharp TL, Dence CS, Engelbach JA, Herrero P, Gropler RJ, Welch MJ: Techniques necessary for multiple tracer quantitative small-animal imaging studies. Nucl Med Biol 2005, 32: 875–884. 10.1016/j.nucmedbio.2005.05.010PubMedCrossRef
3.
go back to reference Wu HM, Sui G, Lee CC, Prins ML, Ladno W, Lin HD, Yu AS, Phelps ME, Huang SC: In vivo quantitation of glucose metabolism in mice using small-animal PET and a microfluidic device. J Nucl Med 2007, 48: 837–845. 10.2967/jnumed.106.038182PubMedCrossRef Wu HM, Sui G, Lee CC, Prins ML, Ladno W, Lin HD, Yu AS, Phelps ME, Huang SC: In vivo quantitation of glucose metabolism in mice using small-animal PET and a microfluidic device. J Nucl Med 2007, 48: 837–845. 10.2967/jnumed.106.038182PubMedCrossRef
4.
go back to reference Kudomi N, Choi E, Yamamoto S, Watabe H, Min Kim K, Shidahara M, Ogawa M, Teramoto N, Sakamoto E, H I: Development of a GSO Detector Assembly for a Continuous Blood Sampling System. IEEE Trans Nucl Sci 2003, 50: 70–73. 10.1109/TNS.2002.807869CrossRef Kudomi N, Choi E, Yamamoto S, Watabe H, Min Kim K, Shidahara M, Ogawa M, Teramoto N, Sakamoto E, H I: Development of a GSO Detector Assembly for a Continuous Blood Sampling System. IEEE Trans Nucl Sci 2003, 50: 70–73. 10.1109/TNS.2002.807869CrossRef
5.
go back to reference Lapointe D, Cadorette J, Rodrigue S, Rouleau D, R L: A microvolumetric blood counter/sampler for metabolic PET studies in small animals. IEEE Trans Nucl Sci 1998, 45: 2195–2199. 10.1109/23.708343CrossRef Lapointe D, Cadorette J, Rodrigue S, Rouleau D, R L: A microvolumetric blood counter/sampler for metabolic PET studies in small animals. IEEE Trans Nucl Sci 1998, 45: 2195–2199. 10.1109/23.708343CrossRef
6.
go back to reference Boellaard R, van Lingen A, van Balen SC, Hoving BG, Lammertsma AA: Characteristics of a new fully programmable blood sampling device for monitoring blood radioactivity during PET. Eur J Nucl Med 2001, 28: 81–89. 10.1007/s002590000405PubMedCrossRef Boellaard R, van Lingen A, van Balen SC, Hoving BG, Lammertsma AA: Characteristics of a new fully programmable blood sampling device for monitoring blood radioactivity during PET. Eur J Nucl Med 2001, 28: 81–89. 10.1007/s002590000405PubMedCrossRef
7.
go back to reference Convert L, Morin-Brassard G, Cadorette J, Archambault M, Bentourkia M, Lecomte R: A new tool for molecular imaging: the microvolumetric beta blood counter. J Nucl Med 2007, 48: 1197–1206. 10.2967/jnumed.107.042606PubMedCrossRef Convert L, Morin-Brassard G, Cadorette J, Archambault M, Bentourkia M, Lecomte R: A new tool for molecular imaging: the microvolumetric beta blood counter. J Nucl Med 2007, 48: 1197–1206. 10.2967/jnumed.107.042606PubMedCrossRef
8.
go back to reference Reymond J-M, Guez D, Kerhoas S, Mangeot P, Boisgard R, Jan S, Tavitian B, Trebossen R: Development of an instrument for time-activity curve measurements during PET imaging of rodents. Nuc Instr Methods Phys Res A 2007, 571: 358–361. 10.1016/j.nima.2006.11.019CrossRef Reymond J-M, Guez D, Kerhoas S, Mangeot P, Boisgard R, Jan S, Tavitian B, Trebossen R: Development of an instrument for time-activity curve measurements during PET imaging of rodents. Nuc Instr Methods Phys Res A 2007, 571: 358–361. 10.1016/j.nima.2006.11.019CrossRef
9.
go back to reference Maramraju S, Stoll S, Woody C, Schlyer D, Schiffer W, Lee D, Dewey S, P V: A LSO [beta] microprobe for measuring input functions for quantitative small animal PET. Nuc Instr Methods Phys Res A 2007, 571: 407–410. 10.1016/j.nima.2006.10.121CrossRef Maramraju S, Stoll S, Woody C, Schlyer D, Schiffer W, Lee D, Dewey S, P V: A LSO [beta] microprobe for measuring input functions for quantitative small animal PET. Nuc Instr Methods Phys Res A 2007, 571: 407–410. 10.1016/j.nima.2006.10.121CrossRef
10.
go back to reference Laforest R, Sharp TL, Engelbach JA, Fettig NM, Herrero P, Kim J, Lewis JS, Rowland DJ, Tai YC, Welch MJ: Measurement of input functions in rodents: challenges and solutions. Nucl Med Biol 2005, 32: 679–685. 10.1016/j.nucmedbio.2005.06.012PubMedCrossRef Laforest R, Sharp TL, Engelbach JA, Fettig NM, Herrero P, Kim J, Lewis JS, Rowland DJ, Tai YC, Welch MJ: Measurement of input functions in rodents: challenges and solutions. Nucl Med Biol 2005, 32: 679–685. 10.1016/j.nucmedbio.2005.06.012PubMedCrossRef
11.
go back to reference Lee K, Fox PT, Lancaster JL, Jerabek PA: A positron-probe system for arterial input function quantification for positron emission tomography in humans. Rev Sci Instrum 2008, 79: 064301. 10.1063/1.2936880PubMedCentralPubMedCrossRef Lee K, Fox PT, Lancaster JL, Jerabek PA: A positron-probe system for arterial input function quantification for positron emission tomography in humans. Rev Sci Instrum 2008, 79: 064301. 10.1063/1.2936880PubMedCentralPubMedCrossRef
12.
go back to reference Pain F, Laniece P, Mastrippolito R, Gervais P, Hantraye P, Besret L: Arterial input function measurement without blood sampling using a beta-microprobe in rats. J Nucl Med 2004, 45: 1577–1582.PubMed Pain F, Laniece P, Mastrippolito R, Gervais P, Hantraye P, Besret L: Arterial input function measurement without blood sampling using a beta-microprobe in rats. J Nucl Med 2004, 45: 1577–1582.PubMed
13.
go back to reference Seki C, Okada H, Mori S, Kakiuchi T, Yoshikawa E, Nishiyama S, Tsukada H, Yamashita T: Application of a beta microprobe for quantification of regional cerebral blood flow with (15)O-water and PET in rhesus monkeys. Ann Nucl Med 1998, 12: 7–14. 10.1007/BF03165410PubMedCrossRef Seki C, Okada H, Mori S, Kakiuchi T, Yoshikawa E, Nishiyama S, Tsukada H, Yamashita T: Application of a beta microprobe for quantification of regional cerebral blood flow with (15)O-water and PET in rhesus monkeys. Ann Nucl Med 1998, 12: 7–14. 10.1007/BF03165410PubMedCrossRef
14.
go back to reference Ahn JY, Lee DS, Lee JS, Kim SK, Cheon GJ, Yeo JS, Shin SA, Chung JK, Lee MC: Quantification of regional myocardial blood flow using dynamic H2(15)O PET and factor analysis. J Nucl Med 2001, 42: 782–787.PubMed Ahn JY, Lee DS, Lee JS, Kim SK, Cheon GJ, Yeo JS, Shin SA, Chung JK, Lee MC: Quantification of regional myocardial blood flow using dynamic H2(15)O PET and factor analysis. J Nucl Med 2001, 42: 782–787.PubMed
15.
go back to reference Hermansen F, Ashburner J, Spinks TJ, Kooner JS, Camici PG, Lammertsma AA: Generation of myocardial factor images directly from the dynamic oxygen-15-water scan without use of an oxygen-15-carbon monoxide blood-pool scan. J Nucl Med 1998, 39: 1696–1702.PubMed Hermansen F, Ashburner J, Spinks TJ, Kooner JS, Camici PG, Lammertsma AA: Generation of myocardial factor images directly from the dynamic oxygen-15-water scan without use of an oxygen-15-carbon monoxide blood-pool scan. J Nucl Med 1998, 39: 1696–1702.PubMed
16.
go back to reference Iida H, Rhodes CG, de Silva R, Araujo LI, Bloomfield PM, Lammertsma AA, Jones T: Use of the left ventricular time-activity curve as a noninvasive input function in dynamic oxygen-15-water positron emission tomography. J Nucl Med 1992, 33: 1669–1677.PubMed Iida H, Rhodes CG, de Silva R, Araujo LI, Bloomfield PM, Lammertsma AA, Jones T: Use of the left ventricular time-activity curve as a noninvasive input function in dynamic oxygen-15-water positron emission tomography. J Nucl Med 1992, 33: 1669–1677.PubMed
17.
go back to reference Kim J, Herrero P, Sharp T, Laforest R, Rowland DJ, Tai YC, Lewis JS, Welch MJ: Minimally invasive method of determining blood input function from PET images in rodents. J Nucl Med 2006, 47: 330–336.PubMed Kim J, Herrero P, Sharp T, Laforest R, Rowland DJ, Tai YC, Lewis JS, Welch MJ: Minimally invasive method of determining blood input function from PET images in rodents. J Nucl Med 2006, 47: 330–336.PubMed
18.
go back to reference Mourik JE, Lubberink M, Schuitemaker A, Tolboom N, van Berckel BN, Lammertsma AA, Boellaard R: Image-derived input functions for PET brain studies. Eur J Nucl Med Mol Imaging 2009, 36: 463–471. 10.1007/s00259-008-0986-8PubMedCrossRef Mourik JE, Lubberink M, Schuitemaker A, Tolboom N, van Berckel BN, Lammertsma AA, Boellaard R: Image-derived input functions for PET brain studies. Eur J Nucl Med Mol Imaging 2009, 36: 463–471. 10.1007/s00259-008-0986-8PubMedCrossRef
19.
go back to reference Su Y, Welch MJ, Shoghi KI: The application of maximum likelihood factor analysis (MLFA) with uniqueness constraints on dynamic cardiac microPET data. Phys Med Biol 2007, 52: 2313–2334. 10.1088/0031-9155/52/8/018PubMedCrossRef Su Y, Welch MJ, Shoghi KI: The application of maximum likelihood factor analysis (MLFA) with uniqueness constraints on dynamic cardiac microPET data. Phys Med Biol 2007, 52: 2313–2334. 10.1088/0031-9155/52/8/018PubMedCrossRef
20.
go back to reference Tantawy MN, Peterson TE: Simplified [18F]FDG image-derived input function using the left ventricle, liver, and one venous blood sample. Mol Imaging 2010, 9: 76–86.PubMedCentralPubMed Tantawy MN, Peterson TE: Simplified [18F]FDG image-derived input function using the left ventricle, liver, and one venous blood sample. Mol Imaging 2010, 9: 76–86.PubMedCentralPubMed
21.
go back to reference Weinberg IN, Huang SC, Hoffman EJ, Araujo L, Nienaber C, Grover-McKay M, Dahlbom M, Schelbert H: Validation of PET-acquired input functions for cardiac studies. J Nucl Med 1988, 29: 241–247.PubMed Weinberg IN, Huang SC, Hoffman EJ, Araujo L, Nienaber C, Grover-McKay M, Dahlbom M, Schelbert H: Validation of PET-acquired input functions for cardiac studies. J Nucl Med 1988, 29: 241–247.PubMed
22.
go back to reference Wu HM, Hoh CK, Choi Y, Schelbert HR, Hawkins RA, Phelps ME, Huang SC: Factor analysis for extraction of blood time-activity curves in dynamic FDG-PET studies. J Nucl Med 1995, 36: 1714–1722.PubMed Wu HM, Hoh CK, Choi Y, Schelbert HR, Hawkins RA, Phelps ME, Huang SC: Factor analysis for extraction of blood time-activity curves in dynamic FDG-PET studies. J Nucl Med 1995, 36: 1714–1722.PubMed
23.
go back to reference Wu HM, Huang SC, Allada V, Wolfenden PJ, Schelbert HR, Phelps ME, Hoh CK: Derivation of input function from FDG-PET studies in small hearts. J Nucl Med 1996, 37: 1717–1722.PubMed Wu HM, Huang SC, Allada V, Wolfenden PJ, Schelbert HR, Phelps ME, Hoh CK: Derivation of input function from FDG-PET studies in small hearts. J Nucl Med 1996, 37: 1717–1722.PubMed
24.
go back to reference Huang SC, Wu HM, Shoghi-Jadid K, Stout DB, Chatziioannou A, Schelbert HR, Barrio JR: Investigation of a new input function validation approach for dynamic mouse microPET studies. Mol Imaging Biol 2004, 6: 34–46. 10.1016/j.mibio.2003.12.002PubMedCrossRef Huang SC, Wu HM, Shoghi-Jadid K, Stout DB, Chatziioannou A, Schelbert HR, Barrio JR: Investigation of a new input function validation approach for dynamic mouse microPET studies. Mol Imaging Biol 2004, 6: 34–46. 10.1016/j.mibio.2003.12.002PubMedCrossRef
25.
go back to reference Meyer PT, Circiumaru V, Cardi CA, Thomas DH, Bal H, Acton PD: Simplified quantification of small animal [18F]FDG PET studies using a standard arterial input function. Eur J Nucl Med Mol Imaging 2006, 33: 948–954. 10.1007/s00259-006-0121-7PubMedCrossRef Meyer PT, Circiumaru V, Cardi CA, Thomas DH, Bal H, Acton PD: Simplified quantification of small animal [18F]FDG PET studies using a standard arterial input function. Eur J Nucl Med Mol Imaging 2006, 33: 948–954. 10.1007/s00259-006-0121-7PubMedCrossRef
26.
go back to reference Ferl GZ, Zhang X, Wu HM, Huang SC: Estimation of the 18F-FDG input function in mice by use of dynamic small-animal PET and minimal blood sample data. J Nucl Med 2007, 48: 2037–2045. 10.2967/jnumed.107.041061PubMedCentralPubMedCrossRef Ferl GZ, Zhang X, Wu HM, Huang SC: Estimation of the 18F-FDG input function in mice by use of dynamic small-animal PET and minimal blood sample data. J Nucl Med 2007, 48: 2037–2045. 10.2967/jnumed.107.041061PubMedCentralPubMedCrossRef
27.
go back to reference Shoghi KI, Welch MJ: Hybrid image and blood sampling input function for quantification of small animal dynamic PET data. Nucl Med Biol 2007, 34: 989–994. 10.1016/j.nucmedbio.2007.07.010PubMedCentralPubMedCrossRef Shoghi KI, Welch MJ: Hybrid image and blood sampling input function for quantification of small animal dynamic PET data. Nucl Med Biol 2007, 34: 989–994. 10.1016/j.nucmedbio.2007.07.010PubMedCentralPubMedCrossRef
28.
go back to reference Weber B, Burger C, Biro P, Buck A: A femoral arteriovenous shunt facilitates arterial whole blood sampling in animals. Eur J Nucl Med Mol Imaging 2002, 29: 319–323. 10.1007/s00259-001-0712-2PubMedCrossRef Weber B, Burger C, Biro P, Buck A: A femoral arteriovenous shunt facilitates arterial whole blood sampling in animals. Eur J Nucl Med Mol Imaging 2002, 29: 319–323. 10.1007/s00259-001-0712-2PubMedCrossRef
29.
go back to reference Ashworth S, Ranciar A, Bloomfield PM: Development of an on-line blood detector system for PET studies in small animals. In Quantification of brain function using PET. Edited by: al Me. San Diego: Academic Press; 1996:62–66.CrossRef Ashworth S, Ranciar A, Bloomfield PM: Development of an on-line blood detector system for PET studies in small animals. In Quantification of brain function using PET. Edited by: al Me. San Diego: Academic Press; 1996:62–66.CrossRef
30.
go back to reference Ingvar M, Eriksson L, Rogers GA, Stone-Elander S, Widen L: Rapid feasibility studies of tracers for positron emission tomography: high-resolution PET in small animals with kinetic analysis. J Cereb Blood Flow Metab 1991, 11: 926–931. 10.1038/jcbfm.1991.157PubMedCrossRef Ingvar M, Eriksson L, Rogers GA, Stone-Elander S, Widen L: Rapid feasibility studies of tracers for positron emission tomography: high-resolution PET in small animals with kinetic analysis. J Cereb Blood Flow Metab 1991, 11: 926–931. 10.1038/jcbfm.1991.157PubMedCrossRef
31.
go back to reference Weber B, Spath N, Wyss M, Wild D, Burger C, Stanley R, Buck A: Quantitative cerebral blood flow measurements in the rat using a beta-probe and H2 15O. J Cereb Blood Flow Metab 2003, 23: 1455–1460.PubMedCrossRef Weber B, Spath N, Wyss M, Wild D, Burger C, Stanley R, Buck A: Quantitative cerebral blood flow measurements in the rat using a beta-probe and H2 15O. J Cereb Blood Flow Metab 2003, 23: 1455–1460.PubMedCrossRef
32.
go back to reference Pain F, Dhenain M, Gurden H, Routier AL, Lefebvre F, Mastrippolito R, Laniece P: A method based on Monte Carlo simulations and voxelized anatomical atlases to evaluate and correct uncertainties on radiotracer accumulation quantitation in beta microprobe studies in the rat brain. Phys Med Biol 2008, 53: 5385–5404. 10.1088/0031-9155/53/19/008PubMedCrossRef Pain F, Dhenain M, Gurden H, Routier AL, Lefebvre F, Mastrippolito R, Laniece P: A method based on Monte Carlo simulations and voxelized anatomical atlases to evaluate and correct uncertainties on radiotracer accumulation quantitation in beta microprobe studies in the rat brain. Phys Med Biol 2008, 53: 5385–5404. 10.1088/0031-9155/53/19/008PubMedCrossRef
33.
go back to reference Bahri MA, Plenevaux A, Warnock G, Luxen A, Seret A: NEMA NU4–2008 image quality performance report for the microPET focus 120 and for various transmission and reconstruction methods. J Nucl Med 2009, 50: 1730–1738. 10.2967/jnumed.109.063974PubMedCrossRef Bahri MA, Plenevaux A, Warnock G, Luxen A, Seret A: NEMA NU4–2008 image quality performance report for the microPET focus 120 and for various transmission and reconstruction methods. J Nucl Med 2009, 50: 1730–1738. 10.2967/jnumed.109.063974PubMedCrossRef
34.
go back to reference Kim JS, Lee JS, Im KC, Kim SJ, Kim SY, Lee DS, Moon DH: Performance measurement of the microPET focus 120 scanner. J Nucl Med 2007, 48: 1527–1535. 10.2967/jnumed.107.040550PubMedCrossRef Kim JS, Lee JS, Im KC, Kim SJ, Kim SY, Lee DS, Moon DH: Performance measurement of the microPET focus 120 scanner. J Nucl Med 2007, 48: 1527–1535. 10.2967/jnumed.107.040550PubMedCrossRef
35.
go back to reference Tai YC, Ruangma A, Rowland D, Siegel S, Newport DF, Chow PL, Laforest R: Performance evaluation of the microPET focus: a third-generation microPET scanner dedicated to animal imaging. J Nucl Med 2005, 46: 455–463.PubMed Tai YC, Ruangma A, Rowland D, Siegel S, Newport DF, Chow PL, Laforest R: Performance evaluation of the microPET focus: a third-generation microPET scanner dedicated to animal imaging. J Nucl Med 2005, 46: 455–463.PubMed
36.
go back to reference Fang YH, Muzic RF Jr: Spillover and partial-volume correction for image-derived input functions for small-animal 18F-FDG PET studies. J Nucl Med 2008, 49: 606–614. 10.2967/jnumed.107.047613PubMedCrossRef Fang YH, Muzic RF Jr: Spillover and partial-volume correction for image-derived input functions for small-animal 18F-FDG PET studies. J Nucl Med 2008, 49: 606–614. 10.2967/jnumed.107.047613PubMedCrossRef
37.
go back to reference Su KH, Lee JS, Li JH, Yang YW, Liu RS, Chen JC: Partial volume correction of the microPET blood input function using ensemble learning independent component analysis. Phys Med Biol 2009, 54: 1823–1846. 10.1088/0031-9155/54/6/026PubMedCrossRef Su KH, Lee JS, Li JH, Yang YW, Liu RS, Chen JC: Partial volume correction of the microPET blood input function using ensemble learning independent component analysis. Phys Med Biol 2009, 54: 1823–1846. 10.1088/0031-9155/54/6/026PubMedCrossRef
38.
go back to reference Zanotti-Fregonara P, Fadaili el M, Maroy R, Comtat C, Souloumiac A, Jan S, Ribeiro MJ, Gaura V, Bar-Hen A, Trebossen R: Comparison of eight methods for the estimation of the image-derived input function in dynamic [(18)F]-FDG PET human brain studies. J Cereb Blood Flow Metab 2009, 29: 1825–1835. 10.1038/jcbfm.2009.93PubMedCrossRef Zanotti-Fregonara P, Fadaili el M, Maroy R, Comtat C, Souloumiac A, Jan S, Ribeiro MJ, Gaura V, Bar-Hen A, Trebossen R: Comparison of eight methods for the estimation of the image-derived input function in dynamic [(18)F]-FDG PET human brain studies. J Cereb Blood Flow Metab 2009, 29: 1825–1835. 10.1038/jcbfm.2009.93PubMedCrossRef
39.
go back to reference Votaw JR, Shulman SD: Performance evaluation of the Pico-Count flow-through detector for use in cerebral blood flow PET studies. J Nucl Med 1998, 39: 509–515.PubMed Votaw JR, Shulman SD: Performance evaluation of the Pico-Count flow-through detector for use in cerebral blood flow PET studies. J Nucl Med 1998, 39: 509–515.PubMed
40.
go back to reference Ludemann L, Sreenivasa G, Michel R, Rosner C, Plotkin M, Felix R, Wust P, Amthauer H: Corrections of arterial input function for dynamic H215O PET to assess perfusion of pelvic tumours: arterial blood sampling versus image extraction. Phys Med Biol 2006, 51: 2883–2900. 10.1088/0031-9155/51/11/014PubMedCrossRef Ludemann L, Sreenivasa G, Michel R, Rosner C, Plotkin M, Felix R, Wust P, Amthauer H: Corrections of arterial input function for dynamic H215O PET to assess perfusion of pelvic tumours: arterial blood sampling versus image extraction. Phys Med Biol 2006, 51: 2883–2900. 10.1088/0031-9155/51/11/014PubMedCrossRef
41.
go back to reference Munk OL, Keiding S, Bass L: A method to estimate dispersion in sampling catheters and to calculate dispersion-free blood time-activity curves. Med Phys 2008, 35: 3471–3481. 10.1118/1.2948391PubMedCentralPubMedCrossRef Munk OL, Keiding S, Bass L: A method to estimate dispersion in sampling catheters and to calculate dispersion-free blood time-activity curves. Med Phys 2008, 35: 3471–3481. 10.1118/1.2948391PubMedCentralPubMedCrossRef
42.
go back to reference van den Hoff J, Burchert W, Muller-Schauenburg W, Meyer GJ, Hundeshagen H: Accurate local blood flow measurements with dynamic PET: fast determination of input function delay and dispersion by multilinear minimization. J Nucl Med 1993, 34: 1770–1777.PubMed van den Hoff J, Burchert W, Muller-Schauenburg W, Meyer GJ, Hundeshagen H: Accurate local blood flow measurements with dynamic PET: fast determination of input function delay and dispersion by multilinear minimization. J Nucl Med 1993, 34: 1770–1777.PubMed
Metadata
Title
Use of a beta microprobe system to measure arterial input function in PET via an arteriovenous shunt in rats
Authors
Geoff Warnock
Mohamed-Ali Bahri
David Goblet
Fabrice Giacomelli
Christian Lemaire
Joel Aerts
Alain Seret
Xavier Langlois
Andre Luxen
Alain Plenevaux
Publication date
01-12-2011
Publisher
Springer Berlin Heidelberg
Published in
EJNMMI Research / Issue 1/2011
Electronic ISSN: 2191-219X
DOI
https://doi.org/10.1186/2191-219X-1-13

Other articles of this Issue 1/2011

EJNMMI Research 1/2011 Go to the issue