Published in: Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine 1/2014

Open Access 01-12-2014 | Original research

Noninvasive continuous versus intermittent arterial pressure monitoring: evaluation of the vascular unloading technique (CNAP device) in the emergency department

Authors: Julia Y Wagner, Julia S Prantner, Agnes S Meidert, Alexander Hapfelmeier, Roland M Schmid, Bernd Saugel

Published in: Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine | Issue 1/2014

Login to get access

Abstract

Background

Monitoring cardiovascular function in acutely ill patients in the emergency department (ED) is of paramount importance. Arterial pressure (AP) is usually monitored using intermittent oscillometric measurements with an upper arm cuff. The vascular unloading technique (VUT) allows continuous noninvasive AP monitoring. In this study, we compare continuous AP measurements obtained by VUT with intermittent oscillometric AP measurements in ED patients. In addition, we aimed to investigate whether continuous noninvasive AP monitoring allows detection of relevant hypotensive episodes that might be missed with intermittent AP monitoring.

Methods

In a German university hospital, 130 ED patients who required AP monitoring were analyzed in this prospective method comparison study. Continuous AP monitoring was performed using VUT (CNAP technology; CNSystems Medizintechnik AG, Graz, Austria) over a 2-hour period. The oscillometric AP values were recorded simultaneously every 15 minutes for the comparison of both methods. For statistical evaluation, Bland-Altman plots accounting for repeated AP measurements per individual were used.

Results

The mean difference (±standard deviation) between AP measurements obtained by VUT and oscillometric AP measurements was -5 mmHg (±22 mmHg) for systolic AP (SAP), -2 mmHg (±15 mmHg) for diastolic AP (DAP), and -6 mmHg (±16 mmHg) for mean AP (MAP), respectively. In the interval between two oscillometric measurements, the VUT device detected hypotensive episodes (≥4 minutes) defined as either SAP <90 mmHg or MAP <65 mmHg in 30 patients and 16 patients, respectively. In 11 (SAP <90 mmHg) and 6 (MAP <65 mmHg) of these patients, hypotension was also detected by the subsequent intermittent oscillometric AP measurement.

Conclusions

VUT using the CNAP system for noninvasive continuous AP measurement shows reasonable agreement with intermittent oscillometric measurements in acutely ill ED patients. Continuous AP monitoring allows immediate recognition of clinically relevant hypotensive episodes, which are missed or only belatedly recognized with intermittent AP measurement.
Appendix
Available only for authorised users
Literature
1.
go back to reference Kumar A, Roberts D, Wood KE, Light B, Parrillo JE, Sharma S, Suppes R, Feinstein D, Zanotti S, Taiberg L, Gurka D, Kumar A, Cheang M: Duration of hypotension before initiation of effective antimicrobial therapy is the critical determinant of survival in human septic shock. Crit Care Med. 2006, 34: 1589-1596. 10.1097/01.CCM.0000217961.75225.E9.CrossRefPubMed Kumar A, Roberts D, Wood KE, Light B, Parrillo JE, Sharma S, Suppes R, Feinstein D, Zanotti S, Taiberg L, Gurka D, Kumar A, Cheang M: Duration of hypotension before initiation of effective antimicrobial therapy is the critical determinant of survival in human septic shock. Crit Care Med. 2006, 34: 1589-1596. 10.1097/01.CCM.0000217961.75225.E9.CrossRefPubMed
2.
go back to reference Frezza EE, Mezghebe H: Indications and complications of arterial catheter use in surgical or medical intensive care units: analysis of 4932 patients. Am Surg. 1998, 64: 127-131.PubMed Frezza EE, Mezghebe H: Indications and complications of arterial catheter use in surgical or medical intensive care units: analysis of 4932 patients. Am Surg. 1998, 64: 127-131.PubMed
3.
go back to reference Fortin J, Marte W, Grullenberger R, Hacker A, Habenbacher W, Heller A, Wagner C, Wach P, Skrabal F: Continuous non-invasive blood pressure monitoring using concentrically interlocking control loops. Comput Biol Med. 2006, 36: 941-957. 10.1016/j.compbiomed.2005.04.003.CrossRefPubMed Fortin J, Marte W, Grullenberger R, Hacker A, Habenbacher W, Heller A, Wagner C, Wach P, Skrabal F: Continuous non-invasive blood pressure monitoring using concentrically interlocking control loops. Comput Biol Med. 2006, 36: 941-957. 10.1016/j.compbiomed.2005.04.003.CrossRefPubMed
4.
go back to reference Biais M, Vidil L, Roullet S, Masson F, Quinart A, Revel P, Sztark F: Continuous non-invasive arterial pressure measurement: evaluation of CNAP device during vascular surgery. Ann Fr Anesth Reanim. 2010, 29: 530-535.CrossRefPubMed Biais M, Vidil L, Roullet S, Masson F, Quinart A, Revel P, Sztark F: Continuous non-invasive arterial pressure measurement: evaluation of CNAP device during vascular surgery. Ann Fr Anesth Reanim. 2010, 29: 530-535.CrossRefPubMed
5.
go back to reference Jeleazcov C, Krajinovic L, Munster T, Birkholz T, Fried R, Schuttler J, Fechner J: Precision and accuracy of a new device (CNAPTM) for continuous non-invasive arterial pressure monitoring: assessment during general anaesthesia. Br J Anaesth. 2010, 105: 264-272. 10.1093/bja/aeq143.CrossRefPubMed Jeleazcov C, Krajinovic L, Munster T, Birkholz T, Fried R, Schuttler J, Fechner J: Precision and accuracy of a new device (CNAPTM) for continuous non-invasive arterial pressure monitoring: assessment during general anaesthesia. Br J Anaesth. 2010, 105: 264-272. 10.1093/bja/aeq143.CrossRefPubMed
6.
go back to reference Hahn R, Rinosl H, Neuner M, Kettner SC: Clinical validation of a continuous non-invasive haemodynamic monitor (CNAP&trade; 500) during general anaesthesia. Br J Anaesth. 2012, 108: 581-585. 10.1093/bja/aer499.CrossRefPubMed Hahn R, Rinosl H, Neuner M, Kettner SC: Clinical validation of a continuous non-invasive haemodynamic monitor (CNAP&trade; 500) during general anaesthesia. Br J Anaesth. 2012, 108: 581-585. 10.1093/bja/aer499.CrossRefPubMed
7.
go back to reference Ilies C, Bauer M, Berg P, Rosenberg J, Hedderich J, Bein B, Hinz J, Hanss R: Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurement. Br J Anaesth. 2012, 108: 202-210. 10.1093/bja/aer394.CrossRefPubMed Ilies C, Bauer M, Berg P, Rosenberg J, Hedderich J, Bein B, Hinz J, Hanss R: Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurement. Br J Anaesth. 2012, 108: 202-210. 10.1093/bja/aer394.CrossRefPubMed
8.
go back to reference Jagadeesh AM, Singh NG, Mahankali S: A comparison of a continuous noninvasive arterial pressure (CNAP) monitor with an invasive arterial blood pressure monitor in the cardiac surgical ICU. Ann Card Anaesth. 2012, 15: 180-184. 10.4103/0971-9784.97973.CrossRefPubMed Jagadeesh AM, Singh NG, Mahankali S: A comparison of a continuous noninvasive arterial pressure (CNAP) monitor with an invasive arterial blood pressure monitor in the cardiac surgical ICU. Ann Card Anaesth. 2012, 15: 180-184. 10.4103/0971-9784.97973.CrossRefPubMed
9.
go back to reference Eeftinck Schattenkerk DW, van Lieshout JJ, van den Meiracker AH, Wesseling KR, Blanc S, Wieling W, van Montfrans GA, Settels JJ, Wesseling KH, Westerhof BE: Nexfin noninvasive continuous blood pressure validated against Riva-Rocci/Korotkoff. Am J Hypertens. 2009, 22: 378-383. 10.1038/ajh.2008.368.CrossRefPubMed Eeftinck Schattenkerk DW, van Lieshout JJ, van den Meiracker AH, Wesseling KR, Blanc S, Wieling W, van Montfrans GA, Settels JJ, Wesseling KH, Westerhof BE: Nexfin noninvasive continuous blood pressure validated against Riva-Rocci/Korotkoff. Am J Hypertens. 2009, 22: 378-383. 10.1038/ajh.2008.368.CrossRefPubMed
10.
go back to reference Martina JR, Westerhof BE, van Goudoever J, de Beaumont EM, Truijen J, Kim YS, Immink RV, Jobsis DA, Hollmann MW, Lahpor JR, de Mol BA, van Lieshout JJ: Noninvasive continuous arterial blood pressure monitoring with Nexfin(R). Anesthesiology. 2012, 116: 1092-1103. 10.1097/ALN.0b013e31824f94ed.CrossRefPubMed Martina JR, Westerhof BE, van Goudoever J, de Beaumont EM, Truijen J, Kim YS, Immink RV, Jobsis DA, Hollmann MW, Lahpor JR, de Mol BA, van Lieshout JJ: Noninvasive continuous arterial blood pressure monitoring with Nexfin(R). Anesthesiology. 2012, 116: 1092-1103. 10.1097/ALN.0b013e31824f94ed.CrossRefPubMed
11.
go back to reference Garnier RP, van der Spoel AG, Sibarani-Ponsen R, Markhorst DG, Boer C: Level of agreement between Nexfin non-invasive arterial pressure with invasive arterial pressure measurements in children. Br J Anaesth. 2012, 109: 609-615. 10.1093/bja/aes295.CrossRefPubMed Garnier RP, van der Spoel AG, Sibarani-Ponsen R, Markhorst DG, Boer C: Level of agreement between Nexfin non-invasive arterial pressure with invasive arterial pressure measurements in children. Br J Anaesth. 2012, 109: 609-615. 10.1093/bja/aes295.CrossRefPubMed
12.
go back to reference Janelle GM, Gravenstein N: An accuracy evaluation of the T-Line Tensymeter (continuous noninvasive blood pressure management device) versus conventional invasive radial artery monitoring in surgical patients. Anesth Analg. 2006, 102: 484-490. 10.1213/01.ane.0000194873.52453.bd.CrossRefPubMed Janelle GM, Gravenstein N: An accuracy evaluation of the T-Line Tensymeter (continuous noninvasive blood pressure management device) versus conventional invasive radial artery monitoring in surgical patients. Anesth Analg. 2006, 102: 484-490. 10.1213/01.ane.0000194873.52453.bd.CrossRefPubMed
13.
go back to reference Szmuk P, Pivalizza E, Warters RD, Ezri T, Gebhard R: An evaluation of the T-Line Tensymeter continuous noninvasive blood pressure device during induced hypotension. Anaesthesia. 2008, 63: 307-312. 10.1111/j.1365-2044.2007.05369.x.CrossRefPubMed Szmuk P, Pivalizza E, Warters RD, Ezri T, Gebhard R: An evaluation of the T-Line Tensymeter continuous noninvasive blood pressure device during induced hypotension. Anaesthesia. 2008, 63: 307-312. 10.1111/j.1365-2044.2007.05369.x.CrossRefPubMed
14.
go back to reference Dueck R, Goedje O, Clopton P: Noninvasive continuous beat-to-beat radial artery pressure via TL-200 applanation tonometry. J Clin Monit Comput. 2012, 26: 75-83. 10.1007/s10877-012-9336-2.CrossRefPubMed Dueck R, Goedje O, Clopton P: Noninvasive continuous beat-to-beat radial artery pressure via TL-200 applanation tonometry. J Clin Monit Comput. 2012, 26: 75-83. 10.1007/s10877-012-9336-2.CrossRefPubMed
15.
go back to reference Saugel B, Fassio F, Hapfelmeier A, Meidert AS, Schmid RM, Huber W: The T-Line TL-200 system for continuous non-invasive blood pressure measurement in medical intensive care unit patients. Intensive Care Med. 2012, 38: 1471-1477. 10.1007/s00134-012-2617-x.CrossRefPubMed Saugel B, Fassio F, Hapfelmeier A, Meidert AS, Schmid RM, Huber W: The T-Line TL-200 system for continuous non-invasive blood pressure measurement in medical intensive care unit patients. Intensive Care Med. 2012, 38: 1471-1477. 10.1007/s00134-012-2617-x.CrossRefPubMed
16.
go back to reference Saugel B, Meidert AS, Hapfelmeier A, Eyer F, Schmid RM, Huber W: Non-invasive continuous arterial pressure measurement based on radial artery tonometry in the intensive care unit: a method comparison study using the T-Line TL-200pro device. Br J Anaesth. 2013, 111: 185-190. 10.1093/bja/aet025.CrossRefPubMed Saugel B, Meidert AS, Hapfelmeier A, Eyer F, Schmid RM, Huber W: Non-invasive continuous arterial pressure measurement based on radial artery tonometry in the intensive care unit: a method comparison study using the T-Line TL-200pro device. Br J Anaesth. 2013, 111: 185-190. 10.1093/bja/aet025.CrossRefPubMed
17.
go back to reference Meidert AS, Huber W, Hapfelmeier A, Schofthaler M, Muller JN, Langwieser N, Wagner JY, Schmid RM, Saugel B: Evaluation of the radial artery applanation tonometry technology for continuous noninvasive blood pressure monitoring compared with central aortic blood pressure measurements in patients with multiple organ dysfunction syndrome. J Crit Care. 2013, 28: 908-912. 10.1016/j.jcrc.2013.06.012.CrossRefPubMed Meidert AS, Huber W, Hapfelmeier A, Schofthaler M, Muller JN, Langwieser N, Wagner JY, Schmid RM, Saugel B: Evaluation of the radial artery applanation tonometry technology for continuous noninvasive blood pressure monitoring compared with central aortic blood pressure measurements in patients with multiple organ dysfunction syndrome. J Crit Care. 2013, 28: 908-912. 10.1016/j.jcrc.2013.06.012.CrossRefPubMed
18.
go back to reference Siebig S, Rockmann F, Sabel K, Zuber-Jerger I, Dierkes C, Brunnler T, Wrede CE: Continuous non-invasive arterial pressure technique improves patient monitoring during interventional endoscopy. Int J Med Sci. 2009, 6: 37-42.PubMedCentralCrossRefPubMed Siebig S, Rockmann F, Sabel K, Zuber-Jerger I, Dierkes C, Brunnler T, Wrede CE: Continuous non-invasive arterial pressure technique improves patient monitoring during interventional endoscopy. Int J Med Sci. 2009, 6: 37-42.PubMedCentralCrossRefPubMed
19.
go back to reference Ilies C, Kiskalt H, Siedenhans D, Meybohm P, Steinfath M, Bein B, Hanss R: Detection of hypotension during Caesarean section with continuous non-invasive arterial pressure device or intermittent oscillometric arterial pressure measurement. Br J Anaesth. 2012, 109: 413-419. 10.1093/bja/aes224.CrossRefPubMed Ilies C, Kiskalt H, Siedenhans D, Meybohm P, Steinfath M, Bein B, Hanss R: Detection of hypotension during Caesarean section with continuous non-invasive arterial pressure device or intermittent oscillometric arterial pressure measurement. Br J Anaesth. 2012, 109: 413-419. 10.1093/bja/aes224.CrossRefPubMed
20.
go back to reference Penaz J, Voigt A, Teichmann W: Contribution to the continuous indirect blood pressure measurement. Z Gesamte Inn Med. 1976, 31: 1030-1033.PubMed Penaz J, Voigt A, Teichmann W: Contribution to the continuous indirect blood pressure measurement. Z Gesamte Inn Med. 1976, 31: 1030-1033.PubMed
21.
go back to reference Bland JM, Altman DG: Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007, 17: 571-582. 10.1080/10543400701329422.CrossRefPubMed Bland JM, Altman DG: Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007, 17: 571-582. 10.1080/10543400701329422.CrossRefPubMed
22.
go back to reference Nowak RM, Sen A, Garcia AJ, Wilkie H, Yang JJ, Nowak MR, Moyer ML: Noninvasive continuous or intermittent blood pressure and heart rate patient monitoring in the ED. Am J Emerg Med. 2011, 29: 782-789. 10.1016/j.ajem.2011.05.014.CrossRefPubMed Nowak RM, Sen A, Garcia AJ, Wilkie H, Yang JJ, Nowak MR, Moyer ML: Noninvasive continuous or intermittent blood pressure and heart rate patient monitoring in the ED. Am J Emerg Med. 2011, 29: 782-789. 10.1016/j.ajem.2011.05.014.CrossRefPubMed
23.
go back to reference ANSI/AAMI SP10:2002. American national standard for manual, electronic, or automated sphygmomanometers: Association for the Advancement of Medical Instrumentation. 2002 ANSI/AAMI SP10:2002. American national standard for manual, electronic, or automated sphygmomanometers: Association for the Advancement of Medical Instrumentation. 2002
Metadata
Title
Noninvasive continuous versus intermittent arterial pressure monitoring: evaluation of the vascular unloading technique (CNAP device) in the emergency department
Authors
Julia Y Wagner
Julia S Prantner
Agnes S Meidert
Alexander Hapfelmeier
Roland M Schmid
Bernd Saugel
Publication date
01-12-2014
Publisher
BioMed Central
DOI
https://doi.org/10.1186/1757-7241-22-8

Other articles of this Issue 1/2014

Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine 1/2014 Go to the issue