Inadvertent malposition of a permanent pacemaker ventricular lead into the left ventricle which was initially missed and diagnosed two years later: a case report

Authors: Medhat F Zaher, Basem N Azab, Marc B Bogin, Soad G Bekheit

Published in: Journal of Medical Case Reports | Issue 1/2011

Login to get access

Abstract

Introduction

Inadvertent malposition of a pacemaker ventricular lead into the left ventricle is an uncommon event, and its actual incidence is probably unknown. It may be underestimated and underreported because of a possible asymptomatic course. A 12-lead electrocardiogram is important to confirm proper placement.

Case presentation

We report a case of a 60-year-old Caucasian man with a malpositioned transvenous permanent pacing lead into the left ventricle via a patent foramen ovale that was not suspected during implantation and went undiagnosed for two years without complications. The patient remained asymptomatic as he was being treated with oral anticoagulation therapy for atrial fibrillation. The decision was made to leave the pacing lead in place and continue lifelong warfarin therapy.

Conclusions

Inadvertent insertion of pacing wires into the left ventricle is a potentially dangerous complication that may happen under fluoroscopic guidance and may be overlooked by routine pacemaker interrogation. It is advisable to obtain a 12-lead electrocardiogram during or immediately after transvenous pacemaker implantation rather than use a routine pacemaker interrogation or a limited electrocardiogram.
Appendix
Available only for authorised users
Literature
1.
go back to reference Bauersfeld UK, Thakur RK, Ghani M, Yee R, Klein GJ: Malposition of transvenous pacing lead in the left ventricle: radiographic findings. AJR Am J Roentgenol. 1994, 162: 290-292.CrossRefPubMed Bauersfeld UK, Thakur RK, Ghani M, Yee R, Klein GJ: Malposition of transvenous pacing lead in the left ventricle: radiographic findings. AJR Am J Roentgenol. 1994, 162: 290-292.CrossRefPubMed
2.
go back to reference Bollinger BC, Heidenreich J: From cardiac arrhythmias. Current Diagnosis & Treatment Emergency. Edited by: Stone CK, Humphries RL. 2008, New York: Lange Medical Books/McGraw-Hill, 578-607. 6 Bollinger BC, Heidenreich J: From cardiac arrhythmias. Current Diagnosis & Treatment Emergency. Edited by: Stone CK, Humphries RL. 2008, New York: Lange Medical Books/McGraw-Hill, 578-607. 6
3.
go back to reference Schiavone WA, Castle LW, Salcedo E, Graor R: Amaurosis fugax in a patient with a left ventricular endocardial pacemaker. Pacing Clin Electrophysiol. 1984, 7: 288-292. 10.1111/j.1540-8159.1984.tb04901.x.CrossRefPubMed Schiavone WA, Castle LW, Salcedo E, Graor R: Amaurosis fugax in a patient with a left ventricular endocardial pacemaker. Pacing Clin Electrophysiol. 1984, 7: 288-292. 10.1111/j.1540-8159.1984.tb04901.x.CrossRefPubMed
4.
go back to reference Villanueva FS, Heinsimer JA, Burkman MH, Fananapazir L, Halvorsen RA, Chen JT: Echocardiographic detection of perforation of the cardiac ventricular septum by a permanent pacemaker lead. Am J Cardiol. 1987, 59: 370-371. 10.1016/0002-9149(87)90821-6.CrossRefPubMed Villanueva FS, Heinsimer JA, Burkman MH, Fananapazir L, Halvorsen RA, Chen JT: Echocardiographic detection of perforation of the cardiac ventricular septum by a permanent pacemaker lead. Am J Cardiol. 1987, 59: 370-371. 10.1016/0002-9149(87)90821-6.CrossRefPubMed
5.
go back to reference Ormond RS, Rubenfire M, Anbe DT, Drake EH: Radiographic demonstration of myocardial perforation by permanent endocardial pacemakers. Radiology. 1971, 98: 35-37.CrossRefPubMed Ormond RS, Rubenfire M, Anbe DT, Drake EH: Radiographic demonstration of myocardial perforation by permanent endocardial pacemakers. Radiology. 1971, 98: 35-37.CrossRefPubMed
6.
go back to reference Meyer JA, Millar K: Malplacement of pacemaker catheters in the coronary sinus: recognition and clinical significance. J Thorac Cardiovasc Surg. 1969, 57: 511-518.PubMed Meyer JA, Millar K: Malplacement of pacemaker catheters in the coronary sinus: recognition and clinical significance. J Thorac Cardiovasc Surg. 1969, 57: 511-518.PubMed
7.
go back to reference Mazzetti H, Dussaut A, Tentori C, Dussaut E, Lazzari JO: Transarterial permanent pacing of the left ventricle. Pacing Clin Electrophysiol. 1990, 13: 588-592. 10.1111/j.1540-8159.1990.tb02073.x.CrossRefPubMed Mazzetti H, Dussaut A, Tentori C, Dussaut E, Lazzari JO: Transarterial permanent pacing of the left ventricle. Pacing Clin Electrophysiol. 1990, 13: 588-592. 10.1111/j.1540-8159.1990.tb02073.x.CrossRefPubMed
8.
go back to reference Okmen E, Erdinler I, Oguz E, Akyol A, Turek O, Cam N, Ulufer T: An electrocardiographic algorithm for determining the location of pacemaker electrode in patients with right bundle branch block configuration during permanent ventricular pacing. Angiology. 2006, 57: 623-630. 10.1177/0003319706293146.CrossRefPubMed Okmen E, Erdinler I, Oguz E, Akyol A, Turek O, Cam N, Ulufer T: An electrocardiographic algorithm for determining the location of pacemaker electrode in patients with right bundle branch block configuration during permanent ventricular pacing. Angiology. 2006, 57: 623-630. 10.1177/0003319706293146.CrossRefPubMed
9.
go back to reference Mower MM, Aranaga CE, Tabatznik B: Unusual patterns of conduction produced by pacemaker stimuli. Am Heart J. 1967, 74: 24-28. 10.1016/0002-8703(67)90036-1.CrossRefPubMed Mower MM, Aranaga CE, Tabatznik B: Unusual patterns of conduction produced by pacemaker stimuli. Am Heart J. 1967, 74: 24-28. 10.1016/0002-8703(67)90036-1.CrossRefPubMed
10.
go back to reference Sodi-Pallares D, Cadler RM: New Bases of Electrocardiography. 1956, St. Louis, MO: Mosby, 377-378. Sodi-Pallares D, Cadler RM: New Bases of Electrocardiography. 1956, St. Louis, MO: Mosby, 377-378.
11.
go back to reference Lister JW, Klotz DH, Jomain SL, Stuckey JH, Hoffman BF: Effect of pacemaker site on cardiac output and ventricular activation in dogs with complete heart block. Am J Cardiol. 1964, 14: 494-503. 10.1016/0002-9149(64)90033-5.CrossRefPubMed Lister JW, Klotz DH, Jomain SL, Stuckey JH, Hoffman BF: Effect of pacemaker site on cardiac output and ventricular activation in dogs with complete heart block. Am J Cardiol. 1964, 14: 494-503. 10.1016/0002-9149(64)90033-5.CrossRefPubMed
12.
go back to reference Konings TC, Koolbergen DR, Bouma BJ, Groenink M, Mulder BJ: Iatrogenic Perforation of the posterior mitral valve leaflet: a rare complication pacemaker lead placement. J Am Soc Echocardiogr. 2008, 21: 512.e5-512.e7. 10.1016/j.echo.2007.08.048.CrossRef Konings TC, Koolbergen DR, Bouma BJ, Groenink M, Mulder BJ: Iatrogenic Perforation of the posterior mitral valve leaflet: a rare complication pacemaker lead placement. J Am Soc Echocardiogr. 2008, 21: 512.e5-512.e7. 10.1016/j.echo.2007.08.048.CrossRef
13.
go back to reference Van Erckelens F, Sigmund M, Lambertz H, Kreis A, Reupcke C, Hanrath P: Asymptomatic left ventricular malposition of a transvenous pacemaker lead through a sinus venosus defect: follow-up over 17 years. Pacing Clin Electrophysiol. 1991, 14: 989-993. 10.1111/j.1540-8159.1991.tb04147.x.CrossRefPubMed Van Erckelens F, Sigmund M, Lambertz H, Kreis A, Reupcke C, Hanrath P: Asymptomatic left ventricular malposition of a transvenous pacemaker lead through a sinus venosus defect: follow-up over 17 years. Pacing Clin Electrophysiol. 1991, 14: 989-993. 10.1111/j.1540-8159.1991.tb04147.x.CrossRefPubMed
14.
go back to reference Van Gelder BM, Bracke FA, Oto A, Yildirir A, Haas PC, Seger JJ, Stainback RF, Botman KJ, Meijer A: Diagnosis and management of inadvertently placed pacing and ICD leads in the left ventricle: a multicenter experience and review of the literature. Pacing Clin Electrophysiol. 2000, 23: 877-883. 10.1111/j.1540-8159.2000.tb00858.x.CrossRefPubMed Van Gelder BM, Bracke FA, Oto A, Yildirir A, Haas PC, Seger JJ, Stainback RF, Botman KJ, Meijer A: Diagnosis and management of inadvertently placed pacing and ICD leads in the left ventricle: a multicenter experience and review of the literature. Pacing Clin Electrophysiol. 2000, 23: 877-883. 10.1111/j.1540-8159.2000.tb00858.x.CrossRefPubMed
Metadata
Title
Inadvertent malposition of a permanent pacemaker ventricular lead into the left ventricle which was initially missed and diagnosed two years later: a case report
Authors
Medhat F Zaher
Basem N Azab
Marc B Bogin
Soad G Bekheit
Publication date
01-12-2011
Publisher
BioMed Central
Published in
Journal of Medical Case Reports / Issue 1/2011
Electronic ISSN: 1752-1947
DOI
https://doi.org/10.1186/1752-1947-5-54

Other articles of this Issue 1/2011

Journal of Medical Case Reports 1/2011 Go to the issue