Should the tip-apex distance (TAD) rule be modified for the proximal femoral nail antirotation (PFNA)? A retrospective study

Authors: Andrej N Nikoloski, Anthony L Osbrough, Piers J Yates

Published in: Journal of Orthopaedic Surgery and Research | Issue 1/2013

Login to get access

Abstract

Background

Unstable proximal femoral fractures are common and challenging for the orthopaedic surgeon. Often, these are treated with intramedullary nails. The most common mode of failure of any device to treat these fractures is cut-out. The Synthes proximal femoral nail antirotation (PFNA) is unique because it is the only proximal femoral intramedullary nail which employs a helical blade in lieu of a lag screw. The optimal tip-apex distance is 25 mm or less for a dynamic hip screw. The optimal blade tip placement is not known for the PFNA.

Aim

The aim of this study is to determine if the traditional tip-apex distance rule (<25 mm) applies to the PFNA.

Method

A retrospective study of all proximal femoral fractures treated with the PFNA in Western Australian public teaching hospitals between August 2006 and October 2007 was performed. Cases were identified from company and theatre implant use records. Patient demographic data was obtained from hospital records. Fractures were classified according to Arbeitsgemeinschaft für Osteosynthesefragen/Association for the Study of Internal Fixation. Fracture reduction, distal locking type and blade position within the head (tip-apex distance and Cleveland zone) were recorded from the intraoperative and immediate postoperative radiographs. Postoperative radiographs obtained in the routine treatment of patients were studied for review looking primarily for cut-out. Clinical outcomes were measured with the Oxford hip score.

Results

One hundred eighty-eight PFNAs were implanted during the study period, with 178 cases included in this study. Ninety-seven patients could be followed up clinically. There were 18 surgical implant-related failures (19%). The single most common mode of failure was cut-out in six cases (6.2%). Three cut-outs (two medial perforation and one varus collapse) occurred with tip-apex distance (TAD) less than 20 mm. There was no cut-out in cases where the TAD was from 20–30 mm. There were three implant-related failures (nail fracture, missed nail and loose locking screw), four implant-related femoral fractures, two non-unions, two delayed unions and one loss of reduction.

Conclusion

The PFNA is a suitable fixation device for the treatment of unstable proximal femoral fractures. There were still a relatively large number of cut-outs, and the tip-apex distance in the failures showed a bimodal distribution, not like previously demonstrated with dynamic hip screw. We propose that the helical blade behaves differently to a screw, and placement too close to the subchondral bone may lead to penetration through the head.
Appendix
Available only for authorised users
Literature
1.
go back to reference Parker MJ, Handoll HH: Gamma and other cephalocondylic intramedullary nails versus extramedullary implants for extracapsular hip fractures. Cochrane Library Database System Rev. 2010, 9: CD000093-doi:10.1002/14651858.CD000093.pub5 Parker MJ, Handoll HH: Gamma and other cephalocondylic intramedullary nails versus extramedullary implants for extracapsular hip fractures. Cochrane Library Database System Rev. 2010, 9: CD000093-doi:10.1002/14651858.CD000093.pub5
2.
go back to reference Klinger HM, Baums MH, Eckert M, Neugebauer R: A comparative study of unstable per- and intertrochanteric femoral fractures treated with dynamic hip screw (DHS) and trochanteric butt-press plate vs. proximal femoral nail (PFN). Zentralblatt fur Chirurgie. 2005, 130 (4): 301-306. 10.1055/s-2005-836784.CrossRefPubMed Klinger HM, Baums MH, Eckert M, Neugebauer R: A comparative study of unstable per- and intertrochanteric femoral fractures treated with dynamic hip screw (DHS) and trochanteric butt-press plate vs. proximal femoral nail (PFN). Zentralblatt fur Chirurgie. 2005, 130 (4): 301-306. 10.1055/s-2005-836784.CrossRefPubMed
3.
go back to reference Windolf J, Hollander DA, Hakimi M: Pitfalls and complications in the use of the proximal femoral nail. Langenbecks Archiv Surg. 2005, 390: 59-65. 10.1007/s00423-004-0466-y.CrossRef Windolf J, Hollander DA, Hakimi M: Pitfalls and complications in the use of the proximal femoral nail. Langenbecks Archiv Surg. 2005, 390: 59-65. 10.1007/s00423-004-0466-y.CrossRef
4.
go back to reference Sommers MB, Roth C, Hall H, Kam BC, Ehmke LW, Krieg JC, Madey SM, Bottlang M: A laboratory model to evaluate cutout resistance of implants for pertrochanteric fracture fixation. J Orthop Trauma. 2004, 18: 361-368. 10.1097/00005131-200407000-00006.CrossRefPubMed Sommers MB, Roth C, Hall H, Kam BC, Ehmke LW, Krieg JC, Madey SM, Bottlang M: A laboratory model to evaluate cutout resistance of implants for pertrochanteric fracture fixation. J Orthop Trauma. 2004, 18: 361-368. 10.1097/00005131-200407000-00006.CrossRefPubMed
5.
go back to reference Simmermacher RKJ, Ljungqvist J, Bail H, Hockertz T, Vochteloo AJ, Ochs U, Werken C, AO - PFNA studygroup: The new proximal femoral nail antirotation (PFNA) in daily practice: results of a multicentre clinical study. Injury. 2008, 39 (8): 932-939. 10.1016/j.injury.2008.02.005.CrossRefPubMed Simmermacher RKJ, Ljungqvist J, Bail H, Hockertz T, Vochteloo AJ, Ochs U, Werken C, AO - PFNA studygroup: The new proximal femoral nail antirotation (PFNA) in daily practice: results of a multicentre clinical study. Injury. 2008, 39 (8): 932-939. 10.1016/j.injury.2008.02.005.CrossRefPubMed
6.
go back to reference PFNA: Leading the way to optimal stability: Synthes. Original instruments and implants of the association for the study of internal fixation. AO/ASIF. Technique guide. Stratec Med. 2004, 1-44. PFNA: Leading the way to optimal stability: Synthes. Original instruments and implants of the association for the study of internal fixation. AO/ASIF. Technique guide. Stratec Med. 2004, 1-44.
7.
go back to reference Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM: The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am. 1995, 77 (7): 1058-1064.PubMed Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM: The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am. 1995, 77 (7): 1058-1064.PubMed
8.
go back to reference Cleveland M, Thompson F, Wilson H, Ishizuka T: A ten-year analysis of intertrochanteric fractures of the femur. J Bone Joint Surg Br. 1959, 41-A: 1399-1408. Cleveland M, Thompson F, Wilson H, Ishizuka T: A ten-year analysis of intertrochanteric fractures of the femur. J Bone Joint Surg Br. 1959, 41-A: 1399-1408.
9.
go back to reference Baumgaertner MR, Curtin SL, Lindskog DM: Intramedullary versus extramedullary fixation for the treatment of intertrochanteric hip fractures. Clin Orthop Relat Res. 1998, 348: 87-94.CrossRefPubMed Baumgaertner MR, Curtin SL, Lindskog DM: Intramedullary versus extramedullary fixation for the treatment of intertrochanteric hip fractures. Clin Orthop Relat Res. 1998, 348: 87-94.CrossRefPubMed
10.
go back to reference Lenich A, Mayr E, Ruter A, Möckl C, Füchtmeier B: First results with the trochanter fixation nail (TFN): a report on 120 cases. Arch Orthop Trauma Surg. 2006, 126: 706-712. 10.1007/s00402-006-0117-6.CrossRefPubMed Lenich A, Mayr E, Ruter A, Möckl C, Füchtmeier B: First results with the trochanter fixation nail (TFN): a report on 120 cases. Arch Orthop Trauma Surg. 2006, 126: 706-712. 10.1007/s00402-006-0117-6.CrossRefPubMed
11.
go back to reference Saudan M, Lübbeke A, Sadowski C, Riand N, Stern R, Hoffmeyer P: Pertrochanteric fractures: is there an advantage to an intramedullary nail? A randomized, prospective study of 206 patients comparing the dynamic hip screw and proximal femoral nail. J Orthop Trauma. 2002, 16 (6): 386-393. 10.1097/00005131-200207000-00004.CrossRefPubMed Saudan M, Lübbeke A, Sadowski C, Riand N, Stern R, Hoffmeyer P: Pertrochanteric fractures: is there an advantage to an intramedullary nail? A randomized, prospective study of 206 patients comparing the dynamic hip screw and proximal femoral nail. J Orthop Trauma. 2002, 16 (6): 386-393. 10.1097/00005131-200207000-00004.CrossRefPubMed
12.
go back to reference Butt MS, Krikler SJ, Nafie S, Ali MS: Comparison of dynamic hip screw and gamma nail: a prospective, randomized, controlled trial. Injury. 1995, 26: 615-618. 10.1016/0020-1383(95)00126-T.CrossRefPubMed Butt MS, Krikler SJ, Nafie S, Ali MS: Comparison of dynamic hip screw and gamma nail: a prospective, randomized, controlled trial. Injury. 1995, 26: 615-618. 10.1016/0020-1383(95)00126-T.CrossRefPubMed
13.
go back to reference Zou J, Xu Y, Yang H: A comparison of proximal femoral nail antirotation and dynamic hip screw devices in trochanteric fractures. J Int Med Res. 2009, 37 (4): 1057-1064. 10.1177/147323000903700410.CrossRefPubMed Zou J, Xu Y, Yang H: A comparison of proximal femoral nail antirotation and dynamic hip screw devices in trochanteric fractures. J Int Med Res. 2009, 37 (4): 1057-1064. 10.1177/147323000903700410.CrossRefPubMed
14.
go back to reference Bridle SH, Patel AD, Bircher M, Calvert P: Fixation of intertrochanteric fractures of the femur. A randomised prospective comparison of the gamma nail and the dynamic hip screw. J Bone Joint Surg Br. 1991, 73: 330-334.PubMed Bridle SH, Patel AD, Bircher M, Calvert P: Fixation of intertrochanteric fractures of the femur. A randomised prospective comparison of the gamma nail and the dynamic hip screw. J Bone Joint Surg Br. 1991, 73: 330-334.PubMed
15.
go back to reference Jones HW, Johnston P, Parker M: Are short femoral nails superior to the sliding hip screw? A meta-analysis of 24 studies involving 3279 fractures. Int Orthop. 2006, 30: 69-78. 10.1007/s00264-005-0028-0.PubMedCentralCrossRefPubMed Jones HW, Johnston P, Parker M: Are short femoral nails superior to the sliding hip screw? A meta-analysis of 24 studies involving 3279 fractures. Int Orthop. 2006, 30: 69-78. 10.1007/s00264-005-0028-0.PubMedCentralCrossRefPubMed
16.
go back to reference Gardner MJ, Stephen M, Briggs A, Kopjar B, Helfet DL, Lorich DG: Radiographic outcomes of intertrochanteric hip fractures treated with the trochanteric fixation nail. Injury. 2007, 38: 1189-1196. 10.1016/j.injury.2007.03.014.CrossRefPubMed Gardner MJ, Stephen M, Briggs A, Kopjar B, Helfet DL, Lorich DG: Radiographic outcomes of intertrochanteric hip fractures treated with the trochanteric fixation nail. Injury. 2007, 38: 1189-1196. 10.1016/j.injury.2007.03.014.CrossRefPubMed
17.
go back to reference Mereddy P, Kamath S, Ramakrishnan M, Malik H, Donnachie N: The AO/ASIF proximal femoral nail antirotation (PFNA): a new design for the treatment of unstable proximal femoral fractures. Injury. 2009, 40 (4): 428-432. 10.1016/j.injury.2008.10.014.CrossRefPubMed Mereddy P, Kamath S, Ramakrishnan M, Malik H, Donnachie N: The AO/ASIF proximal femoral nail antirotation (PFNA): a new design for the treatment of unstable proximal femoral fractures. Injury. 2009, 40 (4): 428-432. 10.1016/j.injury.2008.10.014.CrossRefPubMed
18.
go back to reference Brunner A, Jockel JA, Babst R: The PFNA proximal femur nail in treatment of unstable proximal femur fractures—3 cases of postoperative perforation of the helical blade into the hip joint. J Orthop Trauma. 2008, 22 (10): 731-736. 10.1097/BOT.0b013e3181893b1b.CrossRefPubMed Brunner A, Jockel JA, Babst R: The PFNA proximal femur nail in treatment of unstable proximal femur fractures—3 cases of postoperative perforation of the helical blade into the hip joint. J Orthop Trauma. 2008, 22 (10): 731-736. 10.1097/BOT.0b013e3181893b1b.CrossRefPubMed
19.
go back to reference Strauss E, Frank J, Lee J, Kummer F, Tejwani N: Helical blade versus sliding hip screw for treatment of unstable intertrochanteric hip fractures: a biomechanical evaluation. Injury. 2006, 37 (10): 984-989. 10.1016/j.injury.2006.06.008.CrossRefPubMed Strauss E, Frank J, Lee J, Kummer F, Tejwani N: Helical blade versus sliding hip screw for treatment of unstable intertrochanteric hip fractures: a biomechanical evaluation. Injury. 2006, 37 (10): 984-989. 10.1016/j.injury.2006.06.008.CrossRefPubMed
20.
go back to reference Penzkofer J, Mendel T, Bauer C, Brehme K: Treatment results of pertrochanteric and subtrochanteric femoral fractures: a retrospective comparison of PFN and PFNA. Der Unfallchirurg. 2009, 112 (8): 699-705. 10.1007/s00113-009-1642-x.CrossRefPubMed Penzkofer J, Mendel T, Bauer C, Brehme K: Treatment results of pertrochanteric and subtrochanteric femoral fractures: a retrospective comparison of PFN and PFNA. Der Unfallchirurg. 2009, 112 (8): 699-705. 10.1007/s00113-009-1642-x.CrossRefPubMed
21.
go back to reference Wild M, Jungbkuth P, Thelen S, Laffrée Q, Gehrmann S, Betsch M, Windolf J, Hakimi M: The dynamics of proximal femoral nails: a clinical comparison between PFNA and Targon PF. Orthop. 2010, 33 (8): 10- Wild M, Jungbkuth P, Thelen S, Laffrée Q, Gehrmann S, Betsch M, Windolf J, Hakimi M: The dynamics of proximal femoral nails: a clinical comparison between PFNA and Targon PF. Orthop. 2010, 33 (8): 10-
22.
go back to reference Takigami I, Matsumoto K, Ohara A, Yamanaka K, Naganawa T, Ohashi M, Date K, Shimizu K: Treatment of trochanteric fractures with the PFNA (proximal femoral nail antirotation) nail system - report of early results. Bull New York Univ Hosp Joint Dis. 2008, 66 (4): 276-279. Takigami I, Matsumoto K, Ohara A, Yamanaka K, Naganawa T, Ohashi M, Date K, Shimizu K: Treatment of trochanteric fractures with the PFNA (proximal femoral nail antirotation) nail system - report of early results. Bull New York Univ Hosp Joint Dis. 2008, 66 (4): 276-279.
23.
go back to reference Liu Y, Tao R, Liu F, Wang Y, Zhou Z, Cao Y, Wang H: Mid-term outcomes after intramedullary fixation of peritrochanteric femoral fractures using the new proximal femoral nail antirotation (PFNA). Injury. 2010, 41: 810-817. 10.1016/j.injury.2010.03.020.CrossRefPubMed Liu Y, Tao R, Liu F, Wang Y, Zhou Z, Cao Y, Wang H: Mid-term outcomes after intramedullary fixation of peritrochanteric femoral fractures using the new proximal femoral nail antirotation (PFNA). Injury. 2010, 41: 810-817. 10.1016/j.injury.2010.03.020.CrossRefPubMed
24.
go back to reference Pu JS, Liu L, Wang GL, Fang Y, Yang TF: Results of the proximal femoral nail anti-rotation (PFNA) in elderly Chinese patients. Int Orthop. 2009, 33 (5): 1441-1444. 10.1007/s00264-009-0776-3.PubMedCentralCrossRefPubMed Pu JS, Liu L, Wang GL, Fang Y, Yang TF: Results of the proximal femoral nail anti-rotation (PFNA) in elderly Chinese patients. Int Orthop. 2009, 33 (5): 1441-1444. 10.1007/s00264-009-0776-3.PubMedCentralCrossRefPubMed
25.
go back to reference Born C, Karich B, Bauer C, von Oldenburg G, Augat P: Hip screw migration testing: first results for hip screws and helical blades utilizing a new oscillating test method. J Orthop Res. 2011, 29 (5): 760-766. 10.1002/jor.21236.CrossRefPubMed Born C, Karich B, Bauer C, von Oldenburg G, Augat P: Hip screw migration testing: first results for hip screws and helical blades utilizing a new oscillating test method. J Orthop Res. 2011, 29 (5): 760-766. 10.1002/jor.21236.CrossRefPubMed
26.
go back to reference Xu Y, Geng D, Yang H, Wang M, Zhu G: Treatment of unstable proximal femoral fractures: comparison of the proximal femoral nail antirotation and gamma nail 3. Orthop. 2010, 33 (7): 473- Xu Y, Geng D, Yang H, Wang M, Zhu G: Treatment of unstable proximal femoral fractures: comparison of the proximal femoral nail antirotation and gamma nail 3. Orthop. 2010, 33 (7): 473-
27.
go back to reference Zhou JQ, Chang SM: Failure of PFNA: helical blade perforation and tip-apex distance. Injury. 2012, 43 (7): 1227-1228. 10.1016/j.injury.2011.10.024.CrossRefPubMed Zhou JQ, Chang SM: Failure of PFNA: helical blade perforation and tip-apex distance. Injury. 2012, 43 (7): 1227-1228. 10.1016/j.injury.2011.10.024.CrossRefPubMed
Metadata
Title
Should the tip-apex distance (TAD) rule be modified for the proximal femoral nail antirotation (PFNA)? A retrospective study
Authors
Andrej N Nikoloski
Anthony L Osbrough
Piers J Yates
Publication date
01-12-2013
Publisher
BioMed Central
Published in
Journal of Orthopaedic Surgery and Research / Issue 1/2013
Electronic ISSN: 1749-799X
DOI
https://doi.org/10.1186/1749-799X-8-35

Other articles of this Issue 1/2013

Journal of Orthopaedic Surgery and Research 1/2013 Go to the issue