Open Access 01-12-2014 | Research

Optimal set of grid size and angular increment for practical dose calculation using the dynamic conformal arc technique: a systematic evaluation of the dosimetric effects in lung stereotactic body radiation therapy

Authors: Ji-Yeon Park, Siyong Kim, Hae-Jin Park, Jeong-Woo Lee, Yeon-Sil Kim, Tae-Suk Suh

Published in: Radiation Oncology | Issue 1/2014

Login to get access

Abstract

Purpose

To recommend the optimal plan parameter set of grid size and angular increment for dose calculations in treatment planning for lung stereotactic body radiation therapy (SBRT) using dynamic conformal arc therapy (DCAT) considering both accuracy and computational efficiency.

Materials and methods

Dose variations with varying grid sizes (2, 3, and 4 mm) and angular increments (2°, 4°, 6°, and 10°) were analyzed in a thorax phantom for 3 spherical target volumes and in 9 patient cases. A 2-mm grid size and 2° angular increment are assumed sufficient to serve as reference values. The dosimetric effect was evaluated using dose–volume histograms, monitor units (MUs), and dose to organs at risk (OARs) for a definite volume corresponding to the dose–volume constraint in lung SBRT. The times required for dose calculations using each parameter set were compared for clinical practicality.

Results

Larger grid sizes caused a dose increase to the structures and required higher MUs to achieve the target coverage. The discrete beam arrangements at each angular increment led to over- and under-estimated OARs doses due to the undulating dose distribution. When a 2° angular increment was used in both studies, a 4-mm grid size changed the dose variation by up to 3–4% (50 cGy) for the heart and the spinal cord, while a 3-mm grid size produced a dose difference of <1% (12 cGy) in all tested OARs. When a 3-mm grid size was employed, angular increments of 6° and 10° caused maximum dose variations of 3% (23 cGy) and 10% (61 cGy) in the spinal cord, respectively, while a 4° increment resulted in a dose difference of <1% (8 cGy) in all cases except for that of one patient. The 3-mm grid size and 4° angular increment enabled a 78% savings in computation time without making any critical sacrifices to dose accuracy.

Conclusions

A parameter set with a 3-mm grid size and a 4° angular increment is found to be appropriate for predicting patient dose distributions with a dose difference below 1% while reducing the computation time by more than half for lung SBRT using DCAT.
Appendix
Available only for authorised users
Literature
2.
go back to reference Timmerman R, Abdulrahman R, Kavanagh BD, Meyer J: Lung cancer: a model for implementing stereotactic body radiation therapy into practice. In IMRT, IGRT, SBRT: advanced in the treatment planning and delivery of radiotherapy. Edited by: Meyer JL, Kavanagh JA, Purdy JA, Timmerman R. Basel: Karger; 2007:367-385. Timmerman R, Abdulrahman R, Kavanagh BD, Meyer J: Lung cancer: a model for implementing stereotactic body radiation therapy into practice. In IMRT, IGRT, SBRT: advanced in the treatment planning and delivery of radiotherapy. Edited by: Meyer JL, Kavanagh JA, Purdy JA, Timmerman R. Basel: Karger; 2007:367-385.
3.
go back to reference Solberg TD, Boedeker KL, Fogg R, Selch MT, DeSalles AA: Dynamic arc radiosurgery field shaping: a comparison with static field conformal and noncoplanar circular arcs. Int J Radiat Oncol Biol Phys 2001, 49: 1481-1491. 10.1016/S0360-3016(00)01537-6CrossRefPubMed Solberg TD, Boedeker KL, Fogg R, Selch MT, DeSalles AA: Dynamic arc radiosurgery field shaping: a comparison with static field conformal and noncoplanar circular arcs. Int J Radiat Oncol Biol Phys 2001, 49: 1481-1491. 10.1016/S0360-3016(00)01537-6CrossRefPubMed
4.
go back to reference Holt A, van Vliet-Vroegindeweij C, Mans A, Belderbos JS, Damen EM: Volumetric-modulated arc therapy for stereotactic body radiotherapy of lung tumors: a comparison with intensity-modulated radiotherapy techniques. Int J Radiat Oncol Biol Phys 2011, 81: 1560-1567. 10.1016/j.ijrobp.2010.09.014CrossRefPubMed Holt A, van Vliet-Vroegindeweij C, Mans A, Belderbos JS, Damen EM: Volumetric-modulated arc therapy for stereotactic body radiotherapy of lung tumors: a comparison with intensity-modulated radiotherapy techniques. Int J Radiat Oncol Biol Phys 2011, 81: 1560-1567. 10.1016/j.ijrobp.2010.09.014CrossRefPubMed
5.
go back to reference Buyyounouski MK, Balter P, Lewis B, D’Ambrosio DJ, Dilling TJ, Miller RC, Schefter T, Tome W, Harris EE, Price RA Jr, et al.: Stereotactic body radiotherapy for early-stage non-small-cell lung cancer: report of the ASTRO emerging technology committee. Int J Radiat Oncol Biol Phys 2010, 78: 3-10. 10.1016/j.ijrobp.2010.04.010CrossRefPubMed Buyyounouski MK, Balter P, Lewis B, D’Ambrosio DJ, Dilling TJ, Miller RC, Schefter T, Tome W, Harris EE, Price RA Jr, et al.: Stereotactic body radiotherapy for early-stage non-small-cell lung cancer: report of the ASTRO emerging technology committee. Int J Radiat Oncol Biol Phys 2010, 78: 3-10. 10.1016/j.ijrobp.2010.04.010CrossRefPubMed
6.
go back to reference Benedict SH, Yenice KM, Followill D, Galvin JM, Hinson W, Kavanagh B, Keall P, Lovelock M, Meeks S, Papiez L, et al.: Stereotactic body radiation therapy: the report of AAPM task group 101. Med Phys 2010, 37: 4078-4101. 10.1118/1.3438081CrossRefPubMed Benedict SH, Yenice KM, Followill D, Galvin JM, Hinson W, Kavanagh B, Keall P, Lovelock M, Meeks S, Papiez L, et al.: Stereotactic body radiation therapy: the report of AAPM task group 101. Med Phys 2010, 37: 4078-4101. 10.1118/1.3438081CrossRefPubMed
7.
go back to reference Takeda A, Kunieda E, Sanuki N, Ohashi T, Oku Y, Sudo Y, Iwashita H, Ooka Y, Aoki Y, Shigematsu N, Kubo A: Dose distribution analysis in stereotactic body radiotherapy using dynamic conformal multiple arc therapy. Int J Radiat Oncol Biol Phys 2009, 74: 363-369. 10.1016/j.ijrobp.2008.08.012CrossRefPubMed Takeda A, Kunieda E, Sanuki N, Ohashi T, Oku Y, Sudo Y, Iwashita H, Ooka Y, Aoki Y, Shigematsu N, Kubo A: Dose distribution analysis in stereotactic body radiotherapy using dynamic conformal multiple arc therapy. Int J Radiat Oncol Biol Phys 2009, 74: 363-369. 10.1016/j.ijrobp.2008.08.012CrossRefPubMed
8.
go back to reference Wiggenraad RG, Petoukhova AL, Versluis L, van Santvoort JP: Stereotactic radiotherapy of intracranial tumors: a comparison of intensity-modulated radiotherapy and dynamic conformal arc. Int J Radiat Oncol Biol Phys 2009, 74: 1018-1026. 10.1016/j.ijrobp.2008.09.057CrossRefPubMed Wiggenraad RG, Petoukhova AL, Versluis L, van Santvoort JP: Stereotactic radiotherapy of intracranial tumors: a comparison of intensity-modulated radiotherapy and dynamic conformal arc. Int J Radiat Oncol Biol Phys 2009, 74: 1018-1026. 10.1016/j.ijrobp.2008.09.057CrossRefPubMed
9.
go back to reference Morales-Paliza MA, Coffey CW, Ding GX: Evaluation of the dynamic conformal arc therapy in comparison to intensity-modulated radiation therapy in prostate, brain, head-and-neck and spine tumors. J Appl Clin Med Phys 2011, 12: 3197. Morales-Paliza MA, Coffey CW, Ding GX: Evaluation of the dynamic conformal arc therapy in comparison to intensity-modulated radiation therapy in prostate, brain, head-and-neck and spine tumors. J Appl Clin Med Phys 2011, 12: 3197.
10.
go back to reference Ding M, Newman F, Kavanagh BD, Stuhr K, Johnson TK, Gaspar LE: Comparative dosimetric study of three-dimensional conformal, dynamic conformal arc, and intensity-modulated radiotherapy for brain tumor treatment using novalis system. Int J Radiat Oncol Biol Phys 2009, 75: S82-S86. Ding M, Newman F, Kavanagh BD, Stuhr K, Johnson TK, Gaspar LE: Comparative dosimetric study of three-dimensional conformal, dynamic conformal arc, and intensity-modulated radiotherapy for brain tumor treatment using novalis system. Int J Radiat Oncol Biol Phys 2009, 75: S82-S86.
11.
go back to reference Niemierko A, Goitein M: The influence of the size of the grid used for dose calculation on the accuracy of dose estimation. Med Phys 1989, 16: 239-247. 10.1118/1.596419CrossRefPubMed Niemierko A, Goitein M: The influence of the size of the grid used for dose calculation on the accuracy of dose estimation. Med Phys 1989, 16: 239-247. 10.1118/1.596419CrossRefPubMed
12.
go back to reference Feygelman V, Zhang G, Stevens C: Initial dosimetric evaluation of SmartArc - a novel VMAT treatment planning module implemented in a multi-vendor delivery chain. J Appl Clin Med Phys 2010, 11: 3169.PubMed Feygelman V, Zhang G, Stevens C: Initial dosimetric evaluation of SmartArc - a novel VMAT treatment planning module implemented in a multi-vendor delivery chain. J Appl Clin Med Phys 2010, 11: 3169.PubMed
13.
go back to reference Chung H, Jin H, Palta J, Suh TS, Kim S: Dose variations with varying calculation grid size in head and neck IMRT. Phys Med Biol 2006, 51: 4841-4856. 10.1088/0031-9155/51/19/008CrossRefPubMed Chung H, Jin H, Palta J, Suh TS, Kim S: Dose variations with varying calculation grid size in head and neck IMRT. Phys Med Biol 2006, 51: 4841-4856. 10.1088/0031-9155/51/19/008CrossRefPubMed
14.
go back to reference Leavitt DD, Tobler M, Gaffney D, Zhang P, Moeller J: Comparison of interpolated vs. calculated micromultileaf settings in dynamic conformal arc treatment. Med Dosim 2000, 25: 17-21. 10.1016/S0958-3947(99)00035-7CrossRefPubMed Leavitt DD, Tobler M, Gaffney D, Zhang P, Moeller J: Comparison of interpolated vs. calculated micromultileaf settings in dynamic conformal arc treatment. Med Dosim 2000, 25: 17-21. 10.1016/S0958-3947(99)00035-7CrossRefPubMed
15.
go back to reference Grebe G, Pfaender M, Roll M, Luedemann L, Wurm RE: Dynamic arc radiosurgery and radiotherapy: commissioning and verification of dose distributions. Int J Radiat Oncol Biol Phys 2001, 49: 1451-1460. 10.1016/S0360-3016(01)01436-5CrossRefPubMed Grebe G, Pfaender M, Roll M, Luedemann L, Wurm RE: Dynamic arc radiosurgery and radiotherapy: commissioning and verification of dose distributions. Int J Radiat Oncol Biol Phys 2001, 49: 1451-1460. 10.1016/S0360-3016(01)01436-5CrossRefPubMed
16.
go back to reference Kim S, Kim T, Ko SJ, Serago C, Smith A, Vallow LA, Peterson JL, Lee R: Negative margin technique - a novel planning strategy to improve dose conformation in SBRT using dynamic conformal arc delivery. J Appl Clin Med Phys 2013, 14: 4283. Kim S, Kim T, Ko SJ, Serago C, Smith A, Vallow LA, Peterson JL, Lee R: Negative margin technique - a novel planning strategy to improve dose conformation in SBRT using dynamic conformal arc delivery. J Appl Clin Med Phys 2013, 14: 4283.
17.
go back to reference Ohtakara K, Hayashi S, Hoshi H: Characterisation of dose distribution in linear accelerator-based intracranial stereotactic radiosurgery with the dynamic conformal arc technique: consideration of the optimal method for dose prescription and evaluation. Br J Radiol 2012, 85: 69-76. 10.1259/bjr/20905396PubMedCentralCrossRefPubMed Ohtakara K, Hayashi S, Hoshi H: Characterisation of dose distribution in linear accelerator-based intracranial stereotactic radiosurgery with the dynamic conformal arc technique: consideration of the optimal method for dose prescription and evaluation. Br J Radiol 2012, 85: 69-76. 10.1259/bjr/20905396PubMedCentralCrossRefPubMed
18.
go back to reference Papanikolaou N, Battista JJ, Boyer AL, Kappas C, Klein E, Mackie TR, Sharpe M, Van Dyk J: Tissue Inhomogeneity Corrections for Megavoltage Photon Beams. Madison WI: American Association in Physicists in Medicine Report No.85; 2004. Papanikolaou N, Battista JJ, Boyer AL, Kappas C, Klein E, Mackie TR, Sharpe M, Van Dyk J: Tissue Inhomogeneity Corrections for Megavoltage Photon Beams. Madison WI: American Association in Physicists in Medicine Report No.85; 2004.
19.
go back to reference Svensson GK, Baily NA, Loevinger R, Morton RJ, Moyer RF, Purdy JA, Shalek RJ, Wootton P, Wright KA: Determination of Absorbed Dose in a Patient Irradiated by Beams of X or Gamma Rays in Radiotherapy Procedures. International Commission on Radiation Units; 1976.21. Physical aspects of quality assurance in radiation therapy. American Association in Physicists in Medicine Report No.13; 1994. Svensson GK, Baily NA, Loevinger R, Morton RJ, Moyer RF, Purdy JA, Shalek RJ, Wootton P, Wright KA: Determination of Absorbed Dose in a Patient Irradiated by Beams of X or Gamma Rays in Radiotherapy Procedures. International Commission on Radiation Units; 1976.21. Physical aspects of quality assurance in radiation therapy. American Association in Physicists in Medicine Report No.13; 1994.
20.
go back to reference Han T, Mikell JK, Salehpour M, Mourtada F: Dosimetric comparison of Acuros XB deterministic radiation transport method with Monte Carlo and model-based convolution methods in heterogeneous media. Med Phys 2011, 38: 2651-2664. 10.1118/1.3582690PubMedCentralCrossRefPubMed Han T, Mikell JK, Salehpour M, Mourtada F: Dosimetric comparison of Acuros XB deterministic radiation transport method with Monte Carlo and model-based convolution methods in heterogeneous media. Med Phys 2011, 38: 2651-2664. 10.1118/1.3582690PubMedCentralCrossRefPubMed
21.
go back to reference Gagne IM, Zavgorodni S: Evaluation of the analytical anisotropic algorithm in an extreme water-lung interface phantom using Monte Carlo dose calculations. J Appl Clin Med Phys 2007, 8: 33-46. Gagne IM, Zavgorodni S: Evaluation of the analytical anisotropic algorithm in an extreme water-lung interface phantom using Monte Carlo dose calculations. J Appl Clin Med Phys 2007, 8: 33-46.
22.
go back to reference Carrasco P, Jornet N, Duch MA, Weber L, Ginjaume M, Eudaldo T, Jurado D, Ruiz A, Ribas M: Comparison of dose calculation algorithms in phantoms with lung equivalent heterogeneities under conditions of lateral electronic disequilibrium. Med Phys 2004, 31: 2899-2911. 10.1118/1.1788932CrossRefPubMed Carrasco P, Jornet N, Duch MA, Weber L, Ginjaume M, Eudaldo T, Jurado D, Ruiz A, Ribas M: Comparison of dose calculation algorithms in phantoms with lung equivalent heterogeneities under conditions of lateral electronic disequilibrium. Med Phys 2004, 31: 2899-2911. 10.1118/1.1788932CrossRefPubMed
23.
go back to reference Dempsey JF, Romeijn HE, Li JG, Low DA, Palta JR: A fourier analysis of the dose grid resolution required for accurate IMRT fluence map optimization. Med Phys 2005, 32: 380-388. 10.1118/1.1843354CrossRefPubMed Dempsey JF, Romeijn HE, Li JG, Low DA, Palta JR: A fourier analysis of the dose grid resolution required for accurate IMRT fluence map optimization. Med Phys 2005, 32: 380-388. 10.1118/1.1843354CrossRefPubMed
24.
go back to reference Webb S, McQuaid D: Some considerations concerning volume-modulated arc therapy: a stepping stonetowards a general theory. Phys Med Biol 2009, 54: 4345-4360. 10.1088/0031-9155/54/14/001CrossRefPubMed Webb S, McQuaid D: Some considerations concerning volume-modulated arc therapy: a stepping stonetowards a general theory. Phys Med Biol 2009, 54: 4345-4360. 10.1088/0031-9155/54/14/001CrossRefPubMed
25.
go back to reference Otto K: Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys 2008, 35: 310-317. 10.1118/1.2818738CrossRefPubMed Otto K: Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys 2008, 35: 310-317. 10.1118/1.2818738CrossRefPubMed
26.
go back to reference Berbeco RI, Pope CJ, Jiang SB: Measurement of the interplay effect in lung IMRT treatment using EDR2 films. J Appl Clin Med Phys 2006, 7: 33-42.PubMed Berbeco RI, Pope CJ, Jiang SB: Measurement of the interplay effect in lung IMRT treatment using EDR2 films. J Appl Clin Med Phys 2006, 7: 33-42.PubMed
27.
go back to reference Lee CM, Watson GA, Leavitt DD: Dynamic collimator optimization compared with fixed collimator angle in arc-based stereotactic radiotherapy: a dosimetric analysis. Neurosurg Focus 2005, 19: E12.PubMed Lee CM, Watson GA, Leavitt DD: Dynamic collimator optimization compared with fixed collimator angle in arc-based stereotactic radiotherapy: a dosimetric analysis. Neurosurg Focus 2005, 19: E12.PubMed
Metadata
Title
Optimal set of grid size and angular increment for practical dose calculation using the dynamic conformal arc technique: a systematic evaluation of the dosimetric effects in lung stereotactic body radiation therapy
Authors
Ji-Yeon Park
Siyong Kim
Hae-Jin Park
Jeong-Woo Lee
Yeon-Sil Kim
Tae-Suk Suh
Publication date
01-12-2014
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2014
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/1748-717X-9-5

Other articles of this Issue 1/2014

Radiation Oncology 1/2014 Go to the issue