Dosimetric precision of an ion beam tracking system

Authors: Christoph Bert, Alexander Gemmel, Nami Saito, Naved Chaudhri, Dieter Schardt, Marco Durante, Gerhard Kraft, Eike Rietzel

Published in: Radiation Oncology | Issue 1/2010

Login to get access

Abstract

Background

Scanned ion beam therapy of intra-fractionally moving tumors requires motion mitigation. GSI proposed beam tracking and performed several experimental studies to analyse the dosimetric precision of the system for scanned carbon beams.

Methods

A beam tracking system has been developed and integrated in the scanned carbon ion beam therapy unit at GSI. The system adapts pencil beam positions and beam energy according to target motion.
Motion compensation performance of the beam tracking system was assessed by measurements with radiographic films, a range telescope, a 3D array of 24 ionization chambers, and cell samples for biological dosimetry. Measurements were performed for stationary detectors and moving detectors using the beam tracking system.

Results

All detector systems showed comparable data for a moving setup when using beam tracking and the corresponding stationary setup. Within the target volume the mean relative differences of ionization chamber measurements were 0.3% (1.5% standard deviation, 3.7% maximum). Film responses demonstrated preserved lateral dose gradients. Measurements with the range telescope showed agreement of Bragg peak depth under motion induced range variations. Cell survival experiments showed a mean relative difference of -5% (-3%) between measurements and calculations within the target volume for beam tracking (stationary) measurements.

Conclusions

The beam tracking system has been successfully integrated. Full functionality has been validated dosimetrically in experiments with several detector types including biological cell systems.
Appendix
Available only for authorised users
Literature
1.
go back to reference Kraft G: Tumor Therapy with Heavy Charged Particles. Prog Part Nucl Phys 2000, 45: s473-s544. 10.1016/S0146-6410(00)00112-5CrossRef Kraft G: Tumor Therapy with Heavy Charged Particles. Prog Part Nucl Phys 2000, 45: s473-s544. 10.1016/S0146-6410(00)00112-5CrossRef
2.
go back to reference Debus J, Haberer T, Schulz-Ertner D, Jäkel O, Wenz F, Enghardt W, Schlegel W, Kraft G, Wannenmacher M: Carbon ion irradiation of skull base tumors at GSI. First clinical results and future perspectives. Strahlenther Onkol 2000, 176: 211-216. 10.1007/s000660050002CrossRefPubMed Debus J, Haberer T, Schulz-Ertner D, Jäkel O, Wenz F, Enghardt W, Schlegel W, Kraft G, Wannenmacher M: Carbon ion irradiation of skull base tumors at GSI. First clinical results and future perspectives. Strahlenther Onkol 2000, 176: 211-216. 10.1007/s000660050002CrossRefPubMed
3.
go back to reference Phillips MH, Pedroni E, Blattmann H, Boehringer T, Coray A, Scheib S: Effects of respiratory motion on dose uniformity with a charged particle scanning method. Phys Med Biol 1992, 37: 223-233. 10.1088/0031-9155/37/1/016CrossRefPubMed Phillips MH, Pedroni E, Blattmann H, Boehringer T, Coray A, Scheib S: Effects of respiratory motion on dose uniformity with a charged particle scanning method. Phys Med Biol 1992, 37: 223-233. 10.1088/0031-9155/37/1/016CrossRefPubMed
4.
go back to reference Bert C, Grözinger SO, Rietzel E: Quantification of interplay effects of scanned particle beams and moving targets. Phys Med Biol 2008, 53: 2253-2265. 10.1088/0031-9155/53/9/003CrossRefPubMed Bert C, Grözinger SO, Rietzel E: Quantification of interplay effects of scanned particle beams and moving targets. Phys Med Biol 2008, 53: 2253-2265. 10.1088/0031-9155/53/9/003CrossRefPubMed
5.
go back to reference Minohara S, Kanai T, Endo M, Noda K, Kanazawa M: Respiratory gated irradiation system for heavy-ion radiotherapy. Int J Radiat Oncol 2000, 47: 1097-1103. 10.1016/S0360-3016(00)00524-1CrossRef Minohara S, Kanai T, Endo M, Noda K, Kanazawa M: Respiratory gated irradiation system for heavy-ion radiotherapy. Int J Radiat Oncol 2000, 47: 1097-1103. 10.1016/S0360-3016(00)00524-1CrossRef
6.
go back to reference Keall PJ, Kini VR, Vedam SS, Mohan R: Motion adaptive x-ray therapy: a feasibility study. Phys Med Biol 2001, 46: 1-10. 10.1088/0031-9155/46/1/301CrossRefPubMed Keall PJ, Kini VR, Vedam SS, Mohan R: Motion adaptive x-ray therapy: a feasibility study. Phys Med Biol 2001, 46: 1-10. 10.1088/0031-9155/46/1/301CrossRefPubMed
7.
go back to reference Bert C, Saito N, Schmidt A, Chaudhri N, Schardt D, Rietzel E: Target motion tracking with a scanned particle beam. Med Phys 2007, 34: 4768-4771. 10.1118/1.2815934CrossRefPubMed Bert C, Saito N, Schmidt A, Chaudhri N, Schardt D, Rietzel E: Target motion tracking with a scanned particle beam. Med Phys 2007, 34: 4768-4771. 10.1118/1.2815934CrossRefPubMed
8.
go back to reference Schweikard A, Glosser G, Bodduluri M, Murphy MJ, Adler JR: Robotic motion compensation for respiratory movement during radiosurgery. Comput Aided Surg 2000, 5: 263-277. 10.3109/10929080009148894CrossRefPubMed Schweikard A, Glosser G, Bodduluri M, Murphy MJ, Adler JR: Robotic motion compensation for respiratory movement during radiosurgery. Comput Aided Surg 2000, 5: 263-277. 10.3109/10929080009148894CrossRefPubMed
9.
go back to reference Mori S, Wolfgang J, Lu HM, Schneider R, Choi NC, Chen GTY: Quantitative Assessment of Range Fluctuations in Charged Particle Lung Irradiation. Int J Radiat Oncol 2008, 70: 253-261. 10.1016/j.ijrobp.2007.08.049CrossRef Mori S, Wolfgang J, Lu HM, Schneider R, Choi NC, Chen GTY: Quantitative Assessment of Range Fluctuations in Charged Particle Lung Irradiation. Int J Radiat Oncol 2008, 70: 253-261. 10.1016/j.ijrobp.2007.08.049CrossRef
10.
go back to reference Grözinger SO, Bert C, Haberer T, Kraft G, Rietzel E: Motion compensation with a scanned ion beam: a technical feasibility study. Radiat Oncol 2008, 3: 34. 10.1186/1748-717X-3-34PubMedCentralCrossRefPubMed Grözinger SO, Bert C, Haberer T, Kraft G, Rietzel E: Motion compensation with a scanned ion beam: a technical feasibility study. Radiat Oncol 2008, 3: 34. 10.1186/1748-717X-3-34PubMedCentralCrossRefPubMed
11.
go back to reference Saito N, Bert C, Chaudhri N, Gemmel A, Schardt D, Rietzel E: Speed and accuracy of a beam tracking system for treatment of moving targets with scanned ion beams. Phys Med Biol 2009, 54: 4849-4862. 10.1088/0031-9155/54/16/001CrossRefPubMed Saito N, Bert C, Chaudhri N, Gemmel A, Schardt D, Rietzel E: Speed and accuracy of a beam tracking system for treatment of moving targets with scanned ion beams. Phys Med Biol 2009, 54: 4849-4862. 10.1088/0031-9155/54/16/001CrossRefPubMed
13.
go back to reference Puck TT, Marcus PI: A rapid method for viable cell titration and clone production with Hela cells in tissue culture: The use of X-irradiated cells to supply conditioning factors. Proc Natl Acad Sci USA 1955, 41: 432-437. 10.1073/pnas.41.7.432PubMedCentralCrossRefPubMed Puck TT, Marcus PI: A rapid method for viable cell titration and clone production with Hela cells in tissue culture: The use of X-irradiated cells to supply conditioning factors. Proc Natl Acad Sci USA 1955, 41: 432-437. 10.1073/pnas.41.7.432PubMedCentralCrossRefPubMed
14.
go back to reference Gemmel A, Hasch B, Ellerbrock M, Weyrather WK, Kramer M: Biological dose optimization with multiple ion fields. Phys Med Biol 2008, 53: 6991-7012. 10.1088/0031-9155/53/23/022CrossRefPubMed Gemmel A, Hasch B, Ellerbrock M, Weyrather WK, Kramer M: Biological dose optimization with multiple ion fields. Phys Med Biol 2008, 53: 6991-7012. 10.1088/0031-9155/53/23/022CrossRefPubMed
15.
go back to reference Karger CP, Jäkel O, Hartmann GH: A system for three-dimensional dosimetric verification of treatment plans in intensity-modulated radiotherapy with heavy ions. Med Phys 1999, 26: 2125-2132. 10.1118/1.598728CrossRefPubMed Karger CP, Jäkel O, Hartmann GH: A system for three-dimensional dosimetric verification of treatment plans in intensity-modulated radiotherapy with heavy ions. Med Phys 1999, 26: 2125-2132. 10.1118/1.598728CrossRefPubMed
16.
go back to reference Sihver L, Schardt D, Kanai T: Depth-dose distributions of high-energy carbon, oxygen and neon beams in water. Jpn J Med Phys 1998, 18: 1-21. Sihver L, Schardt D, Kanai T: Depth-dose distributions of high-energy carbon, oxygen and neon beams in water. Jpn J Med Phys 1998, 18: 1-21.
17.
go back to reference Jäkel O, Jacob C, Schardt D, Karger CP, Hartmann GH: Relation between carbon ion ranges and x-ray CT numbers. Med Phys 2001, 28: 701-703. 10.1118/1.1357455CrossRefPubMed Jäkel O, Jacob C, Schardt D, Karger CP, Hartmann GH: Relation between carbon ion ranges and x-ray CT numbers. Med Phys 2001, 28: 701-703. 10.1118/1.1357455CrossRefPubMed
18.
go back to reference Krämer M, Jäkel O, Haberer T, Kraft G, Schardt D, Weber U: Treatment planning for heavy-ion radiotherapy: physical beam model and dose optimization. Phys Med Biol 2000, 45: 3299-3317. 10.1088/0031-9155/45/11/313CrossRefPubMed Krämer M, Jäkel O, Haberer T, Kraft G, Schardt D, Weber U: Treatment planning for heavy-ion radiotherapy: physical beam model and dose optimization. Phys Med Biol 2000, 45: 3299-3317. 10.1088/0031-9155/45/11/313CrossRefPubMed
19.
go back to reference Krämer M, Scholz M: Rapid calculation of biological effects in ion radiotherapy. Phys Med Biol 2006, 51: 1959-1970. 10.1088/0031-9155/51/8/001CrossRefPubMed Krämer M, Scholz M: Rapid calculation of biological effects in ion radiotherapy. Phys Med Biol 2006, 51: 1959-1970. 10.1088/0031-9155/51/8/001CrossRefPubMed
20.
go back to reference Spielberger B, Krämer M, Scholz M, Kraft G: Three-dimensional dose verification in complex particle radiation fields based on X-ray films. Nucl Instrum Meth B 2003, 209: 277-282. 10.1016/S0168-583X(02)02012-8CrossRef Spielberger B, Krämer M, Scholz M, Kraft G: Three-dimensional dose verification in complex particle radiation fields based on X-ray films. Nucl Instrum Meth B 2003, 209: 277-282. 10.1016/S0168-583X(02)02012-8CrossRef
21.
go back to reference Weyrather WK, Ritter S, Scholz M, Kraft G: RBE for carbon track-segment irradiation in cell lines of differing repair capacity. Int J Radiat Biol 1999, 75: 1357-1364. 10.1080/095530099139232CrossRefPubMed Weyrather WK, Ritter S, Scholz M, Kraft G: RBE for carbon track-segment irradiation in cell lines of differing repair capacity. Int J Radiat Biol 1999, 75: 1357-1364. 10.1080/095530099139232CrossRefPubMed
22.
go back to reference Elsässer T, Krämer M, Scholz M: Accuracy of the local effect model for the prediction of biologic effects of carbon ion beams in vitro and in vivo. Int J Radiat Oncol Biol Phys 2008, 71: 866-872.CrossRefPubMed Elsässer T, Krämer M, Scholz M: Accuracy of the local effect model for the prediction of biologic effects of carbon ion beams in vitro and in vivo. Int J Radiat Oncol Biol Phys 2008, 71: 866-872.CrossRefPubMed
23.
go back to reference Karger CP, Debus J, Kuhn S, Hartmann GH: Three-dimensional accuracy and interfractional reproducibility of patient fixation and positioning using a stereotactic head mask system. Int J Radiat Oncol Biol Phys 2001, 49: 1493-1504. 10.1016/S0360-3016(00)01562-5CrossRefPubMed Karger CP, Debus J, Kuhn S, Hartmann GH: Three-dimensional accuracy and interfractional reproducibility of patient fixation and positioning using a stereotactic head mask system. Int J Radiat Oncol Biol Phys 2001, 49: 1493-1504. 10.1016/S0360-3016(00)01562-5CrossRefPubMed
24.
go back to reference Chaudhri N, Saito N, Bert C, Franczak B, Steidl P, Durante M, Rietzel E, Schardt D: Ion-optical studies for a range adaptation method in ion beam therapy using a static wedge degrader combined with magnetic beam deflection. Phys Med Biol 2010, 55: 3499-3513. 10.1088/0031-9155/55/12/015CrossRefPubMed Chaudhri N, Saito N, Bert C, Franczak B, Steidl P, Durante M, Rietzel E, Schardt D: Ion-optical studies for a range adaptation method in ion beam therapy using a static wedge degrader combined with magnetic beam deflection. Phys Med Biol 2010, 55: 3499-3513. 10.1088/0031-9155/55/12/015CrossRefPubMed
25.
go back to reference Haasbeek CJ, Lagerwaard FJ, Cuijpers JP, Slotman BJ, Senan S: Is adaptive treatment planning required for stereotactic radiotherapy of stage I non-small-cell lung cancer? Int J Radiat Oncol Biol Phys 2007, 67: 1370-1374. 10.1016/j.ijrobp.2006.11.018CrossRefPubMed Haasbeek CJ, Lagerwaard FJ, Cuijpers JP, Slotman BJ, Senan S: Is adaptive treatment planning required for stereotactic radiotherapy of stage I non-small-cell lung cancer? Int J Radiat Oncol Biol Phys 2007, 67: 1370-1374. 10.1016/j.ijrobp.2006.11.018CrossRefPubMed
26.
go back to reference Sawant A, Smith RL, Venkat RB, Santanam L, Cho B, Poulsen P, Cattell H, Newell J, Parikh P, Keall P: Geometric Accuracy and Latency of an Integrated 4D IMRT Delivery System using Real-time Internal Position Monitoring and Dynamic MLC Tracking. Int J Radiat Oncol 2008, 72: S27-S28.CrossRef Sawant A, Smith RL, Venkat RB, Santanam L, Cho B, Poulsen P, Cattell H, Newell J, Parikh P, Keall P: Geometric Accuracy and Latency of an Integrated 4D IMRT Delivery System using Real-time Internal Position Monitoring and Dynamic MLC Tracking. Int J Radiat Oncol 2008, 72: S27-S28.CrossRef
27.
go back to reference Lin T, Tang X, Sandhu A, Jiang S: Fluoroscopic lung tumor tracking. Medical Physics AAPM 2008. abstract 8551 Lin T, Tang X, Sandhu A, Jiang S: Fluoroscopic lung tumor tracking. Medical Physics AAPM 2008. abstract 8551
28.
go back to reference Balter JM, Wright JN, Newell LJ, Friemel B, Dimmer S, Cheng Y, Wong J, Vertatschitsch E, Mate TP: Accuracy of a wireless localization system for radiotherapy. Int J Radiat Oncol Biol Phys 2005, 61: 933-937. 10.1016/j.ijrobp.2004.11.009CrossRefPubMed Balter JM, Wright JN, Newell LJ, Friemel B, Dimmer S, Cheng Y, Wong J, Vertatschitsch E, Mate TP: Accuracy of a wireless localization system for radiotherapy. Int J Radiat Oncol Biol Phys 2005, 61: 933-937. 10.1016/j.ijrobp.2004.11.009CrossRefPubMed
Metadata
Title
Dosimetric precision of an ion beam tracking system
Authors
Christoph Bert
Alexander Gemmel
Nami Saito
Naved Chaudhri
Dieter Schardt
Marco Durante
Gerhard Kraft
Eike Rietzel
Publication date
01-12-2010
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2010
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/1748-717X-5-61

Other articles of this Issue 1/2010

Radiation Oncology 1/2010 Go to the issue