Evaluation of a Subject specific dual-transmit approach for improving B1 field homogeneity in cardiovascular magnetic resonance at 3T

Authors: Ramkumar Krishnamurthy, Amol Pednekar, Marc Kouwenhoven, Benjamin Cheong, Raja Muthupillai

Published in: Journal of Cardiovascular Magnetic Resonance | Issue 1/2013

Login to get access

Abstract

Background

Radiofrequency (RF) shading artifacts degrade image quality while performing cardiovascular magnetic resonance (CMR) at higher field strengths. In this article, we sought to evaluate the effect of local RF (B1 field) shimming by using a dual-source–transmit RF system for cardiac cine imaging and to systematically evaluate the effect of subject body type on the B1 field with and without local RF shimming.

Methods

We obtained cardiac images from 37 subjects (including 11 patients) by using dual-transmit 3T CMR. B1 maps with and without subject-specific local RF shimming (exploiting the independent control of transmit amplitude and phase of the 2 RF transmitters) were obtained. Metrics quantifying B1 field homogeneity were calculated and compared with subject body habitus.

Results

Local RF shimming across the region encompassed by the heart increased the mean flip angle (μ) in that area (88.5 ± 15.2% vs. 81.2 ± 13.3%; P = 0.0014), reduced the B1 field variation by 42.2 ± 13%, and significantly improved the percentage of voxels closer to μ (39% and 82% more voxels were closer to ± 10% and ± 5% of μ, respectively) when compared with no RF shimming. B1 homogeneity was independent of subject body type (body surface area [BSA], body mass index [BMI] or anterior-posterior/right-left patient width ratio [AP/RL]). Subject specific RF (B1) shimming with a dual-transmit system improved local RF homogeneity across all body types.

Conclusion

With or without RF shimming, cardiac B1 field homogeneity does not depend on body type, as characterized by BMI, BSA, and AP/RL. For all body types studied, cardiac B1 field homogeneity was significantly improved by performing local RF shimming with 2 independent RF-transmit channels. This finding indicates the need for subject-specific RF shimming.
Appendix
Available only for authorised users
Literature
1.
go back to reference Franklin KM, Dale BM, Merkle EM: Improvement in B1-inhomogeneity artifacts in the abdomen at 3T MR imaging using a radiofrequency cushion. J Magn Reson Imaging: JMRI. 2008, 27 (6): 1443-47. 10.1002/jmri.21164.CrossRefPubMed Franklin KM, Dale BM, Merkle EM: Improvement in B1-inhomogeneity artifacts in the abdomen at 3T MR imaging using a radiofrequency cushion. J Magn Reson Imaging: JMRI. 2008, 27 (6): 1443-47. 10.1002/jmri.21164.CrossRefPubMed
2.
go back to reference Kim D, Oesingmann N, McGorty K: Hybrid adiabatic-rectangular pulse train for effective saturation of magnetization within the whole heart at 3 T. Magn Reson Med. 2009, 62 (6): 1368-78. 10.1002/mrm.22140.PubMedCentralCrossRefPubMed Kim D, Oesingmann N, McGorty K: Hybrid adiabatic-rectangular pulse train for effective saturation of magnetization within the whole heart at 3 T. Magn Reson Med. 2009, 62 (6): 1368-78. 10.1002/mrm.22140.PubMedCentralCrossRefPubMed
3.
go back to reference Nezafat R, et al: B1-insensitive T2 preparation for improved coronary magnetic resonance angiography at 3 T. Magn Reson Med. 2006, 55 (4): 858-64. 10.1002/mrm.20835.CrossRefPubMed Nezafat R, et al: B1-insensitive T2 preparation for improved coronary magnetic resonance angiography at 3 T. Magn Reson Med. 2006, 55 (4): 858-64. 10.1002/mrm.20835.CrossRefPubMed
4.
go back to reference Sung K, Nayak KS: Measurement and characterization of RF nonuniformity over the heart at 3T using body coil transmission. Journal of magnetic resonance imaging: JMRI. 2008, 27 (3): 643-48. 10.1002/jmri.21253.CrossRefPubMed Sung K, Nayak KS: Measurement and characterization of RF nonuniformity over the heart at 3T using body coil transmission. Journal of magnetic resonance imaging: JMRI. 2008, 27 (3): 643-48. 10.1002/jmri.21253.CrossRefPubMed
5.
go back to reference Sung K, Nayak KS: Design and use of tailored hard-pulse trains for uniformed saturation of myocardium at 3 Tesla. Magn Reson Med. 2008, 60 (4): 997-1002. 10.1002/mrm.21765.CrossRefPubMed Sung K, Nayak KS: Design and use of tailored hard-pulse trains for uniformed saturation of myocardium at 3 Tesla. Magn Reson Med. 2008, 60 (4): 997-1002. 10.1002/mrm.21765.CrossRefPubMed
6.
go back to reference Sung K, Nayak KS: B1+ compensation in 3T cardiac imaging using short 2DRF pulses. Magn Reson Med. 2008, 59 (3): 441-46. 10.1002/mrm.21443.CrossRefPubMed Sung K, Nayak KS: B1+ compensation in 3T cardiac imaging using short 2DRF pulses. Magn Reson Med. 2008, 59 (3): 441-46. 10.1002/mrm.21443.CrossRefPubMed
7.
go back to reference Tang L, Hue YK, Ibrahim TS: Studies of RF shimming techniques with minimization of RF power deposition and their associated temperature changes. Concepts Magn Reson: Part B. Magn Reson Eng. 2011, 39B (1): 11-25. 10.1002/cmr.b.20185.CrossRef Tang L, Hue YK, Ibrahim TS: Studies of RF shimming techniques with minimization of RF power deposition and their associated temperature changes. Concepts Magn Reson: Part B. Magn Reson Eng. 2011, 39B (1): 11-25. 10.1002/cmr.b.20185.CrossRef
9.
go back to reference Wen H, et al: The intrinsic signal-to-noise ratio in human cardiac imaging at 1.5, 3, and 4 T. J Magn Reson. 1997, 125 (1): 65-71. 10.1006/jmre.1996.1072.PubMedCentralCrossRefPubMed Wen H, et al: The intrinsic signal-to-noise ratio in human cardiac imaging at 1.5, 3, and 4 T. J Magn Reson. 1997, 125 (1): 65-71. 10.1006/jmre.1996.1072.PubMedCentralCrossRefPubMed
10.
go back to reference Zhai Z, et al: A study of the relationship between B1-field uniformity, body aspect ratio and SAR for whole-body RF shimming at 3.0T. Proc Int Soc Magn Reson Med. 2009, 3045: 3045-Abstract number: 3045 Zhai Z, et al: A study of the relationship between B1-field uniformity, body aspect ratio and SAR for whole-body RF shimming at 3.0T. Proc Int Soc Magn Reson Med. 2009, 3045: 3045-Abstract number: 3045
11.
go back to reference Harvey RP, Cecilia P, Simons J: B1 shimming performance versus channel/Mode count. Proc Int Soc Magn Reson Med. 2010, Abstract number: 1486 Harvey RP, Cecilia P, Simons J: B1 shimming performance versus channel/Mode count. Proc Int Soc Magn Reson Med. 2010, Abstract number: 1486
12.
go back to reference Willinek WA, et al: Dual-source parallel radiofrequency excitation body MR imaging compared with standard MR imaging at 3.0 T: initial clinical experience. Radiology. 2010, 256 (3): 966-75. 10.1148/radiol.10092127.CrossRefPubMed Willinek WA, et al: Dual-source parallel radiofrequency excitation body MR imaging compared with standard MR imaging at 3.0 T: initial clinical experience. Radiology. 2010, 256 (3): 966-75. 10.1148/radiol.10092127.CrossRefPubMed
13.
go back to reference Mueller A, Kouwenhoven M, Naehle CP: Dual-source radiofrequency transmission with patient-adaptive local radiofrequency shimming for 3.0-T cardiac MR imaging: initial experience. Radiology. 2012, 263: 77-85. 10.1148/radiol.11110347.CrossRefPubMed Mueller A, Kouwenhoven M, Naehle CP: Dual-source radiofrequency transmission with patient-adaptive local radiofrequency shimming for 3.0-T cardiac MR imaging: initial experience. Radiology. 2012, 263: 77-85. 10.1148/radiol.11110347.CrossRefPubMed
14.
go back to reference Mosteller RD: Simplified calculation of body-surface area. N Engl J Med. 1987, 317 (17): 1098-PubMed Mosteller RD: Simplified calculation of body-surface area. N Engl J Med. 1987, 317 (17): 1098-PubMed
15.
go back to reference Cunningham CH, Pauly JM, Nayak KS: Saturated double-angle method for rapid B1+ mapping. Magn Reson Med. 2006, 55 (6): 1326-33. 10.1002/mrm.20896.CrossRefPubMed Cunningham CH, Pauly JM, Nayak KS: Saturated double-angle method for rapid B1+ mapping. Magn Reson Med. 2006, 55 (6): 1326-33. 10.1002/mrm.20896.CrossRefPubMed
16.
go back to reference Schar M, Vonken EJ, Stuber M: Simultaneous B(0)- and B(1)+−map acquisition for fast localized shim, frequency, and RF power determination in the heart at 3 T. Magn Reson Med. 2010, 63 (2): 419-26. 10.1002/mrm.22234.PubMedCentralCrossRefPubMed Schar M, Vonken EJ, Stuber M: Simultaneous B(0)- and B(1)+−map acquisition for fast localized shim, frequency, and RF power determination in the heart at 3 T. Magn Reson Med. 2010, 63 (2): 419-26. 10.1002/mrm.22234.PubMedCentralCrossRefPubMed
17.
go back to reference Alvarez-Linera J: 3T MRI: advances in brain imaging. Eur J Radiol. 2008, 67 (3): 415-26. 10.1016/j.ejrad.2008.02.045.CrossRefPubMed Alvarez-Linera J: 3T MRI: advances in brain imaging. Eur J Radiol. 2008, 67 (3): 415-26. 10.1016/j.ejrad.2008.02.045.CrossRefPubMed
18.
go back to reference Machann J, Schlemmer HP, Schick F: Technical challenges and opportunities of whole-body magnetic resonance imaging at 3T. Physica medica: PM: Int J Devoted Appl Phys Med Biol. 2008, 24 (2): 63-70. Machann J, Schlemmer HP, Schick F: Technical challenges and opportunities of whole-body magnetic resonance imaging at 3T. Physica medica: PM: Int J Devoted Appl Phys Med Biol. 2008, 24 (2): 63-70.
19.
go back to reference Oshinski JN DJ, Sharma P, Gharib AM, Pettigrew RI: Cardiovascular magnetic resonance at 3.0 T: Current state of the art. J Cardiovasc Magn Reson. 2010, 12: 55-68. 10.1186/1532-429X-12-55.PubMedCentralCrossRefPubMed Oshinski JN DJ, Sharma P, Gharib AM, Pettigrew RI: Cardiovascular magnetic resonance at 3.0 T: Current state of the art. J Cardiovasc Magn Reson. 2010, 12: 55-68. 10.1186/1532-429X-12-55.PubMedCentralCrossRefPubMed
20.
go back to reference Singerman RW, et al: Simulation of B1 field distribution and intrinsic signal-to-noise in cardiac MRI as a function of static magnetic field. J Magn Reson. 1997, 125 (1): 72-83. 10.1006/jmre.1996.1073.CrossRefPubMed Singerman RW, et al: Simulation of B1 field distribution and intrinsic signal-to-noise in cardiac MRI as a function of static magnetic field. J Magn Reson. 1997, 125 (1): 72-83. 10.1006/jmre.1996.1073.CrossRefPubMed
21.
22.
go back to reference Vernickel P, et al: Eight-channel transmit/receive body MRI coil at 3T. Magn Reson Med. 2007, 58 (2): 381-89. 10.1002/mrm.21294.CrossRefPubMed Vernickel P, et al: Eight-channel transmit/receive body MRI coil at 3T. Magn Reson Med. 2007, 58 (2): 381-89. 10.1002/mrm.21294.CrossRefPubMed
23.
go back to reference Setsompop K, Wald L, Alagappan V, Gagoski B, Hebrank F, Fontius U, Schmitt F, Adalsteinsson E: Parallel RF transmission with eight channels at 3 Tesla. Magn Reson Med. 2006, 56 (5): 1163-71. 10.1002/mrm.21042.CrossRefPubMed Setsompop K, Wald L, Alagappan V, Gagoski B, Hebrank F, Fontius U, Schmitt F, Adalsteinsson E: Parallel RF transmission with eight channels at 3 Tesla. Magn Reson Med. 2006, 56 (5): 1163-71. 10.1002/mrm.21042.CrossRefPubMed
Metadata
Title
Evaluation of a Subject specific dual-transmit approach for improving B1 field homogeneity in cardiovascular magnetic resonance at 3T
Authors
Ramkumar Krishnamurthy
Amol Pednekar
Marc Kouwenhoven
Benjamin Cheong
Raja Muthupillai
Publication date
01-12-2013
Publisher
BioMed Central
Published in
Journal of Cardiovascular Magnetic Resonance / Issue 1/2013
Electronic ISSN: 1532-429X
DOI
https://doi.org/10.1186/1532-429X-15-68

Other articles of this Issue 1/2013

Journal of Cardiovascular Magnetic Resonance 1/2013 Go to the issue