Cardiac-respiratory self-gated cine ultra-short echo time (UTE) cardiovascular magnetic resonance for assessment of functional cardiac parameters at high magnetic fields

Authors: Verena Hoerr, Nina Nagelmann, Arno Nauerth, Michael T Kuhlmann, Jörg Stypmann, Cornelius Faber

Published in: Journal of Cardiovascular Magnetic Resonance | Issue 1/2013

Login to get access

Abstract

Background

To overcome flow and electrocardiogram-trigger artifacts in cardiovascular magnetic resonance (CMR), we have implemented a cardiac and respiratory self-gated cine ultra-short echo time (UTE) sequence. We have assessed its performance in healthy mice by comparing the results with those obtained with a self-gated cine fast low angle shot (FLASH) sequence and with echocardiography.

Methods

2D self-gated cine UTE (TE/TR = 314 μs/6.2 ms, resolution: 129 × 129 μm, scan time per slice: 5 min 5 sec) and self-gated cine FLASH (TE/TR = 3 ms/6.2 ms, resolution: 129 × 129 μm, scan time per slice: 4 min 49 sec) images were acquired at 9.4 T. Volume of the left and right ventricular (LV, RV) myocardium as well as the end-diastolic and -systolic volume was segmented manually in MR images and myocardial mass, stroke volume (SV), ejection fraction (EF) and cardiac output (CO) were determined. Statistical differences were analyzed by using Student t test and Bland-Altman analyses.

Results

Self-gated cine UTE provided high quality images with high contrast-to-noise ratio (CNR) also for the RV myocardium (CNRblood-myocardium = 25.5 ± 7.8). Compared to cine FLASH, susceptibility, motion, and flow artifacts were considerably reduced due to the short TE of 314 μs. The aortic valve was clearly discernible over the entire cardiac cycle. Myocardial mass, SV, EF and CO determined by self-gated UTE were identical to the values measured with self-gated FLASH and showed good agreement to the results obtained by echocardiography.

Conclusions

Self-gated UTE allows for robust measurement of cardiac parameters of diagnostic interest. Image quality is superior to self-gated FLASH, rendering the method a powerful alternative for the assessment of cardiac function at high magnetic fields.
Appendix
Available only for authorised users
Literature
1.
go back to reference Niendorf T, Sodickson DK, Krombach GA, Schulz-Menger J: Toward cardiovascular MRI at 7 T: clinical needs, technical solutions and research promises. Eur Radiol. 2010, 20: 2806-2816. 10.1007/s00330-010-1902-8.PubMedCentralCrossRefPubMed Niendorf T, Sodickson DK, Krombach GA, Schulz-Menger J: Toward cardiovascular MRI at 7 T: clinical needs, technical solutions and research promises. Eur Radiol. 2010, 20: 2806-2816. 10.1007/s00330-010-1902-8.PubMedCentralCrossRefPubMed
2.
go back to reference Moser E, Stahlberg F, Ladd ME, Trattnig S: 7-T MR—from research to clinical applications?. NMR Biomed. 2012, 25: 695-716. 10.1002/nbm.1794.CrossRefPubMed Moser E, Stahlberg F, Ladd ME, Trattnig S: 7-T MR—from research to clinical applications?. NMR Biomed. 2012, 25: 695-716. 10.1002/nbm.1794.CrossRefPubMed
3.
go back to reference Bun SS, Kober F, Jacquier A, Espinosa L, Kalifa J, Bonzi MF, Kopp F, Lalevee N, Zaffran S, Deharo JC, Cozzone PJ, Bernard M: Value of in vivo T2 measurement for myocardial fibrosis assessment in diabetic mice at 11.75 T. Invest Radiol. 2012, 47: 319-323. 10.1097/RLI.0b013e318243e062.CrossRefPubMed Bun SS, Kober F, Jacquier A, Espinosa L, Kalifa J, Bonzi MF, Kopp F, Lalevee N, Zaffran S, Deharo JC, Cozzone PJ, Bernard M: Value of in vivo T2 measurement for myocardial fibrosis assessment in diabetic mice at 11.75 T. Invest Radiol. 2012, 47: 319-323. 10.1097/RLI.0b013e318243e062.CrossRefPubMed
4.
go back to reference Bohl S, Lygate CA, Barnes H, Medway D, Stork LA, Schulz-Menger J, Neubauer S, Schneider JE: Advanced methods for quantification of infarct size in mice using three-dimensional high-field late gadolinium enhancement MRI. Am J Physiol Heart Circ Physiol. 2009, 296: H1200-H1208. 10.1152/ajpheart.01294.2008.PubMedCentralCrossRefPubMed Bohl S, Lygate CA, Barnes H, Medway D, Stork LA, Schulz-Menger J, Neubauer S, Schneider JE: Advanced methods for quantification of infarct size in mice using three-dimensional high-field late gadolinium enhancement MRI. Am J Physiol Heart Circ Physiol. 2009, 296: H1200-H1208. 10.1152/ajpheart.01294.2008.PubMedCentralCrossRefPubMed
5.
go back to reference Hankiewicz JH, Lewandowski ED: Improved cardiac tagging resolution at ultra-high magnetic field elucidates transmural differences in principal strain in the mouse heart and reduced stretch in dilated cardiomyopathy. J Cardiovasc Magn Reson. 2007, 9: 883-890. 10.1080/10976640701693683.CrossRefPubMed Hankiewicz JH, Lewandowski ED: Improved cardiac tagging resolution at ultra-high magnetic field elucidates transmural differences in principal strain in the mouse heart and reduced stretch in dilated cardiomyopathy. J Cardiovasc Magn Reson. 2007, 9: 883-890. 10.1080/10976640701693683.CrossRefPubMed
6.
go back to reference Neuberger T, Greiser A, Nahrendorf M, Jakob PM, Faber C, Webb AG: 23Na microscopy of the mouse heart in vivo using density-weighted chemical shift imaging. Magn Reson Mater Phys Biol Med. 2004, 17: 196-200. 10.1007/s10334-004-0048-6.CrossRef Neuberger T, Greiser A, Nahrendorf M, Jakob PM, Faber C, Webb AG: 23Na microscopy of the mouse heart in vivo using density-weighted chemical shift imaging. Magn Reson Mater Phys Biol Med. 2004, 17: 196-200. 10.1007/s10334-004-0048-6.CrossRef
7.
go back to reference Aguor EN, Arslan F, van de Kolk CW, Nederhoff MG, Doevendans PA, van Echteld CJ, Pasterkamp G, Strijkers GJ: Quantitative T 2* assessment of acute and chronic myocardial ischemia/reperfusion injury in mice. Magn Reson Mater Phys Biol Med. 2012, 25: 369-379. 10.1007/s10334-012-0304-0.CrossRef Aguor EN, Arslan F, van de Kolk CW, Nederhoff MG, Doevendans PA, van Echteld CJ, Pasterkamp G, Strijkers GJ: Quantitative T 2* assessment of acute and chronic myocardial ischemia/reperfusion injury in mice. Magn Reson Mater Phys Biol Med. 2012, 25: 369-379. 10.1007/s10334-012-0304-0.CrossRef
8.
go back to reference Bovens SM, te Boekhorst BC, den Ouden K, van de Kolk KW, Nauerth A, Nederhoff MG, Pasterkamp G, ten Hove M, van Echteld CJ: Evaluation of infarcted murine heart function: comparison of prospectively triggered with self-gated MRI. NMR Biomed. 2011, 24: 307-315. 10.1002/nbm.1593.CrossRefPubMed Bovens SM, te Boekhorst BC, den Ouden K, van de Kolk KW, Nauerth A, Nederhoff MG, Pasterkamp G, ten Hove M, van Echteld CJ: Evaluation of infarcted murine heart function: comparison of prospectively triggered with self-gated MRI. NMR Biomed. 2011, 24: 307-315. 10.1002/nbm.1593.CrossRefPubMed
9.
go back to reference Heijman E, de Graaf W, Niessen P, Nauerth A, van Eys G, de Graaf L, Nicolay K, Strijkers GJ: Comparison between prospective and retrospective triggering for mouse cardiac MRI. NMR Biomed. 2007, 20: 439-447. 10.1002/nbm.1110.CrossRefPubMed Heijman E, de Graaf W, Niessen P, Nauerth A, van Eys G, de Graaf L, Nicolay K, Strijkers GJ: Comparison between prospective and retrospective triggering for mouse cardiac MRI. NMR Biomed. 2007, 20: 439-447. 10.1002/nbm.1110.CrossRefPubMed
10.
go back to reference Lai P, Larson AC, Park J, Carr JC, Li D: Respiratory self-gated four-dimensional coronary MR angiography: a feasibility study. Magn Reson Med. 2008, 59: 1378-1385. 10.1002/mrm.21617.PubMedCentralCrossRefPubMed Lai P, Larson AC, Park J, Carr JC, Li D: Respiratory self-gated four-dimensional coronary MR angiography: a feasibility study. Magn Reson Med. 2008, 59: 1378-1385. 10.1002/mrm.21617.PubMedCentralCrossRefPubMed
11.
go back to reference O'Brien KR, Gabriel RS, Greiser A, Cowan BR, Young AA, Kerr AJ: Aortic valve stenotic area calculation from phase contrast cardiovascular magnetic resonance: the importance of short echo time. J Cardiovasc Magn Reson. 2009, 11: 49-60. 10.1186/1532-429X-11-49.PubMedCentralCrossRefPubMed O'Brien KR, Gabriel RS, Greiser A, Cowan BR, Young AA, Kerr AJ: Aortic valve stenotic area calculation from phase contrast cardiovascular magnetic resonance: the importance of short echo time. J Cardiovasc Magn Reson. 2009, 11: 49-60. 10.1186/1532-429X-11-49.PubMedCentralCrossRefPubMed
12.
go back to reference Robson MD, Gatehouse PD, Bydder M, Bydder GM: Magnetic resonance: An introduction to ultrashort TE (UTE) imaging. J Comput Assist Tomogr. 2003, 27: 825-846. 10.1097/00004728-200311000-00001.CrossRefPubMed Robson MD, Gatehouse PD, Bydder M, Bydder GM: Magnetic resonance: An introduction to ultrashort TE (UTE) imaging. J Comput Assist Tomogr. 2003, 27: 825-846. 10.1097/00004728-200311000-00001.CrossRefPubMed
13.
go back to reference Rahmer J, Börnert P, Dries SP: Assessment of anterior cruciate ligament reconstruction using 3D ultrashort echo-time MR imaging. J Magn Reson Imaging. 2009, 29: 443-448. 10.1002/jmri.21653.CrossRefPubMed Rahmer J, Börnert P, Dries SP: Assessment of anterior cruciate ligament reconstruction using 3D ultrashort echo-time MR imaging. J Magn Reson Imaging. 2009, 29: 443-448. 10.1002/jmri.21653.CrossRefPubMed
14.
go back to reference Takahashi M, Togao O, Obara M, van Cauteren M, Ohno Y, Doi S, Kuro-o M, Malloy C, Hsia CC, Dimitrov I: Ultra-short echo time (UTE) MR imaging of the lung: comparison between normal and emphysematous lungs in mutant mice. J Magn Reson Imaging. 2010, 32: 326-333. 10.1002/jmri.22267.PubMedCentralCrossRefPubMed Takahashi M, Togao O, Obara M, van Cauteren M, Ohno Y, Doi S, Kuro-o M, Malloy C, Hsia CC, Dimitrov I: Ultra-short echo time (UTE) MR imaging of the lung: comparison between normal and emphysematous lungs in mutant mice. J Magn Reson Imaging. 2010, 32: 326-333. 10.1002/jmri.22267.PubMedCentralCrossRefPubMed
15.
go back to reference Nielsen HT, Gold GE, Olcott EW, Pauly JM, Nishimura DG: Ultrashort echo-time 2D time-of-flight MR angiography using a half-pulse excitation. Magn Reson Med. 1999, 41: 591-599. 10.1002/(SICI)1522-2594(199903)41:3<591::AID-MRM23>3.0.CO;2-R.CrossRefPubMed Nielsen HT, Gold GE, Olcott EW, Pauly JM, Nishimura DG: Ultrashort echo-time 2D time-of-flight MR angiography using a half-pulse excitation. Magn Reson Med. 1999, 41: 591-599. 10.1002/(SICI)1522-2594(199903)41:3<591::AID-MRM23>3.0.CO;2-R.CrossRefPubMed
16.
go back to reference Girard OM, Du J, Agemy L, Sugahara KN, Kotamraju VR, Ruoslahti E, Bydder GM, Mattrey RF: Optimization of iron oxide nanoparticle detection using ultrashort echo time pulse sequences: comparison of T(1), T(2)*, and synergistic T(1) - T(2)* contrast mechanisms. Magn Reson Med. 2011, 65: 1649-1660. 10.1002/mrm.22755.PubMedCentralCrossRefPubMed Girard OM, Du J, Agemy L, Sugahara KN, Kotamraju VR, Ruoslahti E, Bydder GM, Mattrey RF: Optimization of iron oxide nanoparticle detection using ultrashort echo time pulse sequences: comparison of T(1), T(2)*, and synergistic T(1) - T(2)* contrast mechanisms. Magn Reson Med. 2011, 65: 1649-1660. 10.1002/mrm.22755.PubMedCentralCrossRefPubMed
17.
go back to reference Strobel K, Hoerr V, Schmid F, Wachsmuth L, Löffler B, Faber C: Early detection of lung inflammation: Exploiting T(1) -effects of iron oxide particles using UTE MRI. Magn Reson Med. 2012, 68: 1924-1931. 10.1002/mrm.24180.CrossRefPubMed Strobel K, Hoerr V, Schmid F, Wachsmuth L, Löffler B, Faber C: Early detection of lung inflammation: Exploiting T(1) -effects of iron oxide particles using UTE MRI. Magn Reson Med. 2012, 68: 1924-1931. 10.1002/mrm.24180.CrossRefPubMed
18.
go back to reference Hiba B, Richard N, Janier M, Croisille P: Cardiac and respiratory double self-gated cine MRI in the mouse at 7 T. Magn Reson Med. 2006, 55: 506-513. 10.1002/mrm.20815.CrossRefPubMed Hiba B, Richard N, Janier M, Croisille P: Cardiac and respiratory double self-gated cine MRI in the mouse at 7 T. Magn Reson Med. 2006, 55: 506-513. 10.1002/mrm.20815.CrossRefPubMed
19.
go back to reference Hiba B, Richard N, Thibault H, Janier M: Cardiac and respiratory self-gated cine MRI in the mouse: comparison between radial and rectilinear techniques at 7T. Magn Reson Med. 2007, 58: 745-753. 10.1002/mrm.21355.CrossRefPubMed Hiba B, Richard N, Thibault H, Janier M: Cardiac and respiratory self-gated cine MRI in the mouse: comparison between radial and rectilinear techniques at 7T. Magn Reson Med. 2007, 58: 745-753. 10.1002/mrm.21355.CrossRefPubMed
21.
go back to reference Stypmann J, Engelen MA, Troatz C, Rothenburger M, Eckardt L, Tiemann K: Echocardiographic assessment of global left ventricular function in mice. Lab Anim. 2009, 43: 127-137. 10.1258/la.2007.06001e.CrossRefPubMed Stypmann J, Engelen MA, Troatz C, Rothenburger M, Eckardt L, Tiemann K: Echocardiographic assessment of global left ventricular function in mice. Lab Anim. 2009, 43: 127-137. 10.1258/la.2007.06001e.CrossRefPubMed
22.
go back to reference O'Sullivan JD: A fast sinc function gridding algorithm for fourier inversion in computer tomography. IEEE Trans Med Imaging. 1985, 4: 200-207.CrossRefPubMed O'Sullivan JD: A fast sinc function gridding algorithm for fourier inversion in computer tomography. IEEE Trans Med Imaging. 1985, 4: 200-207.CrossRefPubMed
23.
go back to reference Jackson JI, Meyer CH, Nishimura DG, Macovski A: Selection of a convolution function for Fourier inversion using gridding [computerised tomography application]. IEEE Trans Med Imaging. 1991, 10: 473-478. 10.1109/42.97598.CrossRefPubMed Jackson JI, Meyer CH, Nishimura DG, Macovski A: Selection of a convolution function for Fourier inversion using gridding [computerised tomography application]. IEEE Trans Med Imaging. 1991, 10: 473-478. 10.1109/42.97598.CrossRefPubMed
24.
go back to reference Beatty PJ, Nishimura DG, Pauly JM: Rapid gridding reconstruction with a minimal oversampling ratio. IEEE Trans Med Imaging. 2005, 24: 799-808.CrossRefPubMed Beatty PJ, Nishimura DG, Pauly JM: Rapid gridding reconstruction with a minimal oversampling ratio. IEEE Trans Med Imaging. 2005, 24: 799-808.CrossRefPubMed
25.
go back to reference Fries P, Massmann A, Seidel R, Müller A, Stroeder J, Custodis F, Reil J, Schneider G, Buecker A: Comparison of retrospectively self-gated and prospectively triggered FLASH sequences for cine imaging of the aorta in mice at 9.4 Tesla. Invest Radiol. 2012, 47: 259-266. 10.1097/RLI.0b013e31823d3eb6.CrossRefPubMed Fries P, Massmann A, Seidel R, Müller A, Stroeder J, Custodis F, Reil J, Schneider G, Buecker A: Comparison of retrospectively self-gated and prospectively triggered FLASH sequences for cine imaging of the aorta in mice at 9.4 Tesla. Invest Radiol. 2012, 47: 259-266. 10.1097/RLI.0b013e31823d3eb6.CrossRefPubMed
27.
go back to reference Robson MD, Gatehouse PD: Consequences of T2 relaxation during half-pulse slice selection for ultrashort TE imaging. Magn Reson Med. 2010, 64: 610-615.PubMed Robson MD, Gatehouse PD: Consequences of T2 relaxation during half-pulse slice selection for ultrashort TE imaging. Magn Reson Med. 2010, 64: 610-615.PubMed
28.
go back to reference Uecker M, Zhang S, Voit D, Karaus A, Merboldt KD, Frahm J: Real-time MRI at a resolution of 20 ms. NMR Biomed. 2010, 23: 986-994. 10.1002/nbm.1585.CrossRefPubMed Uecker M, Zhang S, Voit D, Karaus A, Merboldt KD, Frahm J: Real-time MRI at a resolution of 20 ms. NMR Biomed. 2010, 23: 986-994. 10.1002/nbm.1585.CrossRefPubMed
29.
go back to reference Niebergall A, Zhang S, Kunay E, Keydana G, Job M, Uecker M, Frahm J: Real-time MRI of speaking at a resolution of 33 ms: undersampled radial FLASH with nonlinear inverse reconstruction. Magn Reson Med. 2013, 69: 477-485. 10.1002/mrm.24276.CrossRefPubMed Niebergall A, Zhang S, Kunay E, Keydana G, Job M, Uecker M, Frahm J: Real-time MRI of speaking at a resolution of 33 ms: undersampled radial FLASH with nonlinear inverse reconstruction. Magn Reson Med. 2013, 69: 477-485. 10.1002/mrm.24276.CrossRefPubMed
30.
go back to reference Vakil P, Ansari SA, Hurley MC, Bhat H, Batjer HH, Bendok BR, Eddleman CS, Carroll TJ: Magnetization spoiling in radial FLASH contrast-enhanced MR digital subtraction angiography. J Magn Reson Imaging. 2012, 36: 249-258. 10.1002/jmri.23630.PubMedCentralCrossRefPubMed Vakil P, Ansari SA, Hurley MC, Bhat H, Batjer HH, Bendok BR, Eddleman CS, Carroll TJ: Magnetization spoiling in radial FLASH contrast-enhanced MR digital subtraction angiography. J Magn Reson Imaging. 2012, 36: 249-258. 10.1002/jmri.23630.PubMedCentralCrossRefPubMed
31.
go back to reference Schneider JE, Cassidy PJ, Lygate C, Tyler DJ, Wiesmann F, Grieve SM, Hulbert K, Clarke K, Neubauer S: Fast, high-resolution in vivo cine magnetic resonance imaging in normal and failing mouse hearts on a vertical 11.7 T system. J Magn Reson Imaging. 2003, 18: 691-701. 10.1002/jmri.10411.CrossRefPubMed Schneider JE, Cassidy PJ, Lygate C, Tyler DJ, Wiesmann F, Grieve SM, Hulbert K, Clarke K, Neubauer S: Fast, high-resolution in vivo cine magnetic resonance imaging in normal and failing mouse hearts on a vertical 11.7 T system. J Magn Reson Imaging. 2003, 18: 691-701. 10.1002/jmri.10411.CrossRefPubMed
32.
go back to reference Cassidy PJ, Schneider JE, Grieve SM, Lygate C, Neubauer S, Clarke K: Assessment of motion gating strategies for mouse magnetic resonance at high magnetic fields. J Magn Reson Imaging. 2004, 19: 229-237. 10.1002/jmri.10454.CrossRefPubMed Cassidy PJ, Schneider JE, Grieve SM, Lygate C, Neubauer S, Clarke K: Assessment of motion gating strategies for mouse magnetic resonance at high magnetic fields. J Magn Reson Imaging. 2004, 19: 229-237. 10.1002/jmri.10454.CrossRefPubMed
33.
go back to reference Tyler DJ, Lygate CA, Schneider JE, Cassidy PJ, Neubauer S, Clarke K: CINE-MR imaging of the normal and infarcted rat heart using an 11.7 T vertical bore MR system. J Cardiovasc Magn Reson. 2006, 8: 327-333. 10.1080/10976640500451903.CrossRefPubMed Tyler DJ, Lygate CA, Schneider JE, Cassidy PJ, Neubauer S, Clarke K: CINE-MR imaging of the normal and infarcted rat heart using an 11.7 T vertical bore MR system. J Cardiovasc Magn Reson. 2006, 8: 327-333. 10.1080/10976640500451903.CrossRefPubMed
34.
go back to reference Herold V, Mörchel P, Faber C, Rommel E, Haase A, Jakob PM: In vivo quantitative three-dimensional motion mapping of the murine myocardium with PC-MRI at 17.6 T. Magn Reson Med. 2006, 55: 1058-1064. 10.1002/mrm.20866.CrossRefPubMed Herold V, Mörchel P, Faber C, Rommel E, Haase A, Jakob PM: In vivo quantitative three-dimensional motion mapping of the murine myocardium with PC-MRI at 17.6 T. Magn Reson Med. 2006, 55: 1058-1064. 10.1002/mrm.20866.CrossRefPubMed
35.
go back to reference Herold V, Wellen J, Ziener CH, Weber T, Hiller KH, Nordbeck P, Rommel E, Haase A, Bauer WR, Jakob PM, Sarkar SK: In vivo comparison of atherosclerotic plaque progression with vessel wall strain and blood flow velocity in apoE(−/−) mice with MR microscopy at 17.6 T. Magn Reson Mater Phys Biol Med. 2009, 22: 159-166. 10.1007/s10334-008-0160-0.CrossRef Herold V, Wellen J, Ziener CH, Weber T, Hiller KH, Nordbeck P, Rommel E, Haase A, Bauer WR, Jakob PM, Sarkar SK: In vivo comparison of atherosclerotic plaque progression with vessel wall strain and blood flow velocity in apoE(−/−) mice with MR microscopy at 17.6 T. Magn Reson Mater Phys Biol Med. 2009, 22: 159-166. 10.1007/s10334-008-0160-0.CrossRef
36.
go back to reference Herold V, Parczyk M, Mörchel P, Ziener CH, Klug G, Bauer WR, Rommel E, Jakob PM: In vivo measurement of local aortic pulse-wave velocity in mice with MR microscopy at 17.6 Tesla. Magn Reson Med. 2009, 61: 1293-1299. 10.1002/mrm.21957.CrossRefPubMed Herold V, Parczyk M, Mörchel P, Ziener CH, Klug G, Bauer WR, Rommel E, Jakob PM: In vivo measurement of local aortic pulse-wave velocity in mice with MR microscopy at 17.6 Tesla. Magn Reson Med. 2009, 61: 1293-1299. 10.1002/mrm.21957.CrossRefPubMed
37.
go back to reference Parczyk M, Herold V, Klug G, Bauer WR, Rommel E, Jakob PM: Regional in vivo transit time measurements of aortic pulse wave velocity in mice with high-field CMR at 17.6 Tesla. J Cardiovasc Magn Reson. 2010, 12: 72-80. 10.1186/1532-429X-12-72.PubMedCentralCrossRefPubMed Parczyk M, Herold V, Klug G, Bauer WR, Rommel E, Jakob PM: Regional in vivo transit time measurements of aortic pulse wave velocity in mice with high-field CMR at 17.6 Tesla. J Cardiovasc Magn Reson. 2010, 12: 72-80. 10.1186/1532-429X-12-72.PubMedCentralCrossRefPubMed
38.
go back to reference Schneider JE, Hulbert KJ, Lygate CA, Ten Hove M, Cassidy PJ, Clarke K, Neubauer S: Long-term stability of cardiac function in normal and chronically failing mouse hearts in a vertical-bore MR system. Magn Reson Mater Phys Biol Med. 2004, 17: 162-169. 10.1007/s10334-004-0056-6.CrossRef Schneider JE, Hulbert KJ, Lygate CA, Ten Hove M, Cassidy PJ, Clarke K, Neubauer S: Long-term stability of cardiac function in normal and chronically failing mouse hearts in a vertical-bore MR system. Magn Reson Mater Phys Biol Med. 2004, 17: 162-169. 10.1007/s10334-004-0056-6.CrossRef
39.
go back to reference Heijman E, Aben JP, Penners C, Niessen P, Guillaume R, van Eys G, Nicolay K, Strijkers GJ: Evaluation of manual and automatic segmentation of the mouse heart from CINE MR images. J Magn Reson Imaging. 2008, 27: 86-93. 10.1002/jmri.21236.CrossRefPubMed Heijman E, Aben JP, Penners C, Niessen P, Guillaume R, van Eys G, Nicolay K, Strijkers GJ: Evaluation of manual and automatic segmentation of the mouse heart from CINE MR images. J Magn Reson Imaging. 2008, 27: 86-93. 10.1002/jmri.21236.CrossRefPubMed
40.
go back to reference Stuckey DJ, Carr CA, Tyler DJ, Clarke K: Cine-MRI versus two-dimensional echocardiography to measure in vivo left ventricular function in rat heart. NMR Biomed. 2008, 21: 765-772. 10.1002/nbm.1268.CrossRefPubMed Stuckey DJ, Carr CA, Tyler DJ, Clarke K: Cine-MRI versus two-dimensional echocardiography to measure in vivo left ventricular function in rat heart. NMR Biomed. 2008, 21: 765-772. 10.1002/nbm.1268.CrossRefPubMed
41.
go back to reference Bucholz E, Ghaghada K, Qi Y, Mukundan S, Johnson GA: Four-dimensional MR microscopy of the mouse heart using radial acquisition and liposomal gadolinium contrast agent. Magn Reson Med. 2008, 60: 111-118. 10.1002/mrm.21618.PubMedCentralCrossRefPubMed Bucholz E, Ghaghada K, Qi Y, Mukundan S, Johnson GA: Four-dimensional MR microscopy of the mouse heart using radial acquisition and liposomal gadolinium contrast agent. Magn Reson Med. 2008, 60: 111-118. 10.1002/mrm.21618.PubMedCentralCrossRefPubMed
42.
go back to reference Bucholz E, Ghaghada K, Qi Y, Mukundan S, Rockman HA, Johnson GA: Cardiovascular phenotyping of the mouse heart using a 4D radial acquisition and liposomal Gd-DTPA-BMA. Magn Reson Med. 2010, 63: 979-987. 10.1002/mrm.22259.PubMedCentralCrossRefPubMed Bucholz E, Ghaghada K, Qi Y, Mukundan S, Rockman HA, Johnson GA: Cardiovascular phenotyping of the mouse heart using a 4D radial acquisition and liposomal Gd-DTPA-BMA. Magn Reson Med. 2010, 63: 979-987. 10.1002/mrm.22259.PubMedCentralCrossRefPubMed
43.
go back to reference Coolen BF, Abdurrachim D, Motaal AG, Nicolay K, Prompers JJ, Strijkers GJ: High frame rate retrospectively triggered Cine MRI for assessment of murine diastolic function. Magn Reson Med. 2013, 69: 648-656. 10.1002/mrm.24287.CrossRefPubMed Coolen BF, Abdurrachim D, Motaal AG, Nicolay K, Prompers JJ, Strijkers GJ: High frame rate retrospectively triggered Cine MRI for assessment of murine diastolic function. Magn Reson Med. 2013, 69: 648-656. 10.1002/mrm.24287.CrossRefPubMed
44.
go back to reference Suttie JJ, Delabarre L, Pitcher A, van de Moortele PF, Dass S, Snyder CJ, Francis JM, Metzger GJ, Weale P, Ugurbil K, Neubauer S, Robson M, Vaughan T: 7 Tesla (T) human cardiovascular magnetic resonance imaging using FLASH and SSFP to assess cardiac function: validation against 1.5 T and 3 T. NMR Biomed. 2012, 25: 27-34. 10.1002/nbm.1708.PubMedCentralCrossRefPubMed Suttie JJ, Delabarre L, Pitcher A, van de Moortele PF, Dass S, Snyder CJ, Francis JM, Metzger GJ, Weale P, Ugurbil K, Neubauer S, Robson M, Vaughan T: 7 Tesla (T) human cardiovascular magnetic resonance imaging using FLASH and SSFP to assess cardiac function: validation against 1.5 T and 3 T. NMR Biomed. 2012, 25: 27-34. 10.1002/nbm.1708.PubMedCentralCrossRefPubMed
Metadata
Title
Cardiac-respiratory self-gated cine ultra-short echo time (UTE) cardiovascular magnetic resonance for assessment of functional cardiac parameters at high magnetic fields
Authors
Verena Hoerr
Nina Nagelmann
Arno Nauerth
Michael T Kuhlmann
Jörg Stypmann
Cornelius Faber
Publication date
01-12-2013
Publisher
BioMed Central
Published in
Journal of Cardiovascular Magnetic Resonance / Issue 1/2013
Electronic ISSN: 1532-429X
DOI
https://doi.org/10.1186/1532-429X-15-59

Other articles of this Issue 1/2013

Journal of Cardiovascular Magnetic Resonance 1/2013 Go to the issue