Towards real-time cardiovascular magnetic resonance guided transarterial CoreValve implantation: in vivo evaluation in swine

Authors: Philipp Kahlert, Nina Parohl, Juliane Albert, Lena Schäfer, Renate Reinhardt, Gernot M Kaiser, Ian McDougall, Brad Decker, Björn Plicht, Raimund Erbel, Holger Eggebrecht, Mark E Ladd, Harald H Quick

Published in: Journal of Cardiovascular Magnetic Resonance | Issue 1/2012

Login to get access

Abstract

Background

Real-time cardiovascular magnetic resonance (rtCMR) is considered attractive for guiding TAVI. Owing to an unlimited scan plane orientation and an unsurpassed soft-tissue contrast with simultaneous device visualization, rtCMR is presumed to allow safe device navigation and to offer optimal orientation for precise axial positioning. We sought to evaluate the preclinical feasibility of rtCMR-guided transarterial aortic valve implatation (TAVI) using the nitinol-based Medtronic CoreValve bioprosthesis.

Methods

rtCMR-guided transfemoral (n = 2) and transsubclavian (n = 6) TAVI was performed in 8 swine using the original CoreValve prosthesis and a modified, CMR-compatible delivery catheter without ferromagnetic components.

Results

rtCMR using TrueFISP sequences provided reliable imaging guidance during TAVI, which was successful in 6 swine. One transfemoral attempt failed due to unsuccessful aortic arch passage and one pericardial tamponade with subsequent death occurred as a result of ventricular perforation by the device tip due to an operating error, this complication being detected without delay by rtCMR. rtCMR allowed for a detailed, simultaneous visualization of the delivery system with the mounted stent-valve and the surrounding anatomy, resulting in improved visualization during navigation through the vasculature, passage of the aortic valve, and during placement and deployment of the stent-valve. Post-interventional success could be confirmed using ECG-triggered time-resolved cine-TrueFISP and flow-sensitive phase-contrast sequences. Intended valve position was confirmed by ex-vivo histology.

Conclusions

Our study shows that rtCMR-guided TAVI using the commercial CoreValve prosthesis in conjunction with a modified delivery system is feasible in swine, allowing improved procedural guidance including immediate detection of complications and direct functional assessment with reduction of radiation and omission of contrast media.
Appendix
Available only for authorised users
Literature
1.
go back to reference Leon MB, Smith CR, Mack M, Miller DC, Moses JW, Svensson LG, Tuzcu EM, Webb JG, Fontana GP, Makkar RR: Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery. N Engl J Med. 2010, 363: 1597-07. 10.1056/NEJMoa1008232.CrossRefPubMed Leon MB, Smith CR, Mack M, Miller DC, Moses JW, Svensson LG, Tuzcu EM, Webb JG, Fontana GP, Makkar RR: Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery. N Engl J Med. 2010, 363: 1597-07. 10.1056/NEJMoa1008232.CrossRefPubMed
2.
go back to reference Thomas M, Schymik G, Walther T, Himbert D, Lefevre T, Treede H, Eggebrecht H, Rubino P, Michev I, Lange R: Thirty-day results of the SAPIEN aortic bioprosthesis european outcome (SOURCE) Registry: A European registry of transcatheter aortic valve implantation using the Edwards SAPIEN valve. Circulation. 2010, 122: 62-9. 10.1161/CIRCULATIONAHA.109.907402.CrossRefPubMed Thomas M, Schymik G, Walther T, Himbert D, Lefevre T, Treede H, Eggebrecht H, Rubino P, Michev I, Lange R: Thirty-day results of the SAPIEN aortic bioprosthesis european outcome (SOURCE) Registry: A European registry of transcatheter aortic valve implantation using the Edwards SAPIEN valve. Circulation. 2010, 122: 62-9. 10.1161/CIRCULATIONAHA.109.907402.CrossRefPubMed
3.
go back to reference Tamburino C, Capodanno D, Ramondo A, Petronio AS, Ettori F, Santoro G, Klugmann S, Bedogni F, Maisano F, Marzocchi A: Incidence and predictors of early and late mortality after transcatheter aortic valve implantation in 663 patients with severe aortic stenosis. Circulation. 2011, 123: 299-08. 10.1161/CIRCULATIONAHA.110.946533.CrossRefPubMed Tamburino C, Capodanno D, Ramondo A, Petronio AS, Ettori F, Santoro G, Klugmann S, Bedogni F, Maisano F, Marzocchi A: Incidence and predictors of early and late mortality after transcatheter aortic valve implantation in 663 patients with severe aortic stenosis. Circulation. 2011, 123: 299-08. 10.1161/CIRCULATIONAHA.110.946533.CrossRefPubMed
4.
go back to reference Geisbusch S, Bleiziffer S, Mazzitelli D, Ruge H, Bauernschmitt R, Lange R: Incidence and management of CoreValve dislocation during transcatheter aortic valve implantation. Circ Cardiovasc Interv. 2010, 3: 531-6. 10.1161/CIRCINTERVENTIONS.110.944983.CrossRefPubMed Geisbusch S, Bleiziffer S, Mazzitelli D, Ruge H, Bauernschmitt R, Lange R: Incidence and management of CoreValve dislocation during transcatheter aortic valve implantation. Circ Cardiovasc Interv. 2010, 3: 531-6. 10.1161/CIRCINTERVENTIONS.110.944983.CrossRefPubMed
5.
go back to reference Abdel-Wahab M, Zahn R, Sherif M, Schneider S, Gerckens U, Schuler G, Sievert H, Eggebrecht H, Senges J, Richardt G: In-hospital outcome of aortic regurgitation after TAVI: results from the prospective multicentre German TAVI registry. Eur Heart J. 2010, 21: 160- Abdel-Wahab M, Zahn R, Sherif M, Schneider S, Gerckens U, Schuler G, Sievert H, Eggebrecht H, Senges J, Richardt G: In-hospital outcome of aortic regurgitation after TAVI: results from the prospective multicentre German TAVI registry. Eur Heart J. 2010, 21: 160-
6.
go back to reference Ferreira ND, Caeiro D, Adao L, Oliveira M, Goncalves H, Ribeiro J, Teixeira M, Albuquerque A, Primo J, Braga P: Incidence and predictors of permanent pacemaker requirement after transcatheter aortic valve implantation with a self-expanding bioprosthesis. Pacing Clin Electrophysiol. 2010, 33: 1364-72. 10.1111/j.1540-8159.2010.02870.x.CrossRefPubMed Ferreira ND, Caeiro D, Adao L, Oliveira M, Goncalves H, Ribeiro J, Teixeira M, Albuquerque A, Primo J, Braga P: Incidence and predictors of permanent pacemaker requirement after transcatheter aortic valve implantation with a self-expanding bioprosthesis. Pacing Clin Electrophysiol. 2010, 33: 1364-72. 10.1111/j.1540-8159.2010.02870.x.CrossRefPubMed
7.
go back to reference Janosi RA, Kahlert P, Plicht B, Bose D, Wendt D, Thielmann M, Jakob H, Eggebrecht H, Erbel R, Buck T: Guidance of percutaneous transcatheter aortic valve implantation by real-time three-dimensional transesophageal echocardiography - A single-center experience. Minim Invasive Ther Allied Technol. 2009, 18: 142-8. 10.1080/13645700902920833.CrossRefPubMed Janosi RA, Kahlert P, Plicht B, Bose D, Wendt D, Thielmann M, Jakob H, Eggebrecht H, Erbel R, Buck T: Guidance of percutaneous transcatheter aortic valve implantation by real-time three-dimensional transesophageal echocardiography - A single-center experience. Minim Invasive Ther Allied Technol. 2009, 18: 142-8. 10.1080/13645700902920833.CrossRefPubMed
8.
go back to reference Tzikas A, Schultz C, Van Mieghem NM, de Jaegere PP, Serruys PW: Optimal projection estimation for transcatheter aortic valve implantation based on contrast-aortography: validation of a Prototype Software. Catheter Cardiovasc Interv. 2010, 76: 602-7. 10.1002/ccd.22641.CrossRefPubMed Tzikas A, Schultz C, Van Mieghem NM, de Jaegere PP, Serruys PW: Optimal projection estimation for transcatheter aortic valve implantation based on contrast-aortography: validation of a Prototype Software. Catheter Cardiovasc Interv. 2010, 76: 602-7. 10.1002/ccd.22641.CrossRefPubMed
9.
go back to reference John M, Liao R, Zheng Y, Nottling A, Boese J, Kirschstein U, Kempfert J, Walther T: System to guide transcatheter aortic valve implantations based on interventional C-arm CT imaging. Med Image Comput Comput Assist Interv. 2010, 13: 375-82.PubMed John M, Liao R, Zheng Y, Nottling A, Boese J, Kirschstein U, Kempfert J, Walther T: System to guide transcatheter aortic valve implantations based on interventional C-arm CT imaging. Med Image Comput Comput Assist Interv. 2010, 13: 375-82.PubMed
10.
go back to reference Zheng Y, John M, Liao R, Boese J, Kirschstein U, Georgescu B, Zhou SK, Kempfert J, Walther T, Brockmann G: Automatic aorta segmentation and valve landmark detection in C-arm CT: application to aortic valve implantation. Med Image Comput Comput Assist Interv. 2010, 13: 476-83.PubMed Zheng Y, John M, Liao R, Boese J, Kirschstein U, Georgescu B, Zhou SK, Kempfert J, Walther T, Brockmann G: Automatic aorta segmentation and valve landmark detection in C-arm CT: application to aortic valve implantation. Med Image Comput Comput Assist Interv. 2010, 13: 476-83.PubMed
11.
go back to reference Bagur R, Webb JG, Nietlispach F, Dumont E, De LR, Doyle D, Masson JB, Gutierrez MJ, Clavel MA, Bertrand OF: Acute kidney injury following transcatheter aortic valve implantation: predictive factors, prognostic value, and comparison with surgical aortic valve replacement. Eur Heart J. 2010, 31: 865-4. 10.1093/eurheartj/ehp552.PubMedCentralCrossRefPubMed Bagur R, Webb JG, Nietlispach F, Dumont E, De LR, Doyle D, Masson JB, Gutierrez MJ, Clavel MA, Bertrand OF: Acute kidney injury following transcatheter aortic valve implantation: predictive factors, prognostic value, and comparison with surgical aortic valve replacement. Eur Heart J. 2010, 31: 865-4. 10.1093/eurheartj/ehp552.PubMedCentralCrossRefPubMed
12.
go back to reference O'Brien KR, Gabriel RS, Greiser A, Cowan BR, Young AA, Kerr AJ: Aortic valve stenotic area calculation from phase contrast cardiovascular magnetic resonance: the importance of short echo time. J Cardiovasc Magn Reson. 2009, 11: 49-10.1186/1532-429X-11-49.PubMedCentralCrossRefPubMed O'Brien KR, Gabriel RS, Greiser A, Cowan BR, Young AA, Kerr AJ: Aortic valve stenotic area calculation from phase contrast cardiovascular magnetic resonance: the importance of short echo time. J Cardiovasc Magn Reson. 2009, 11: 49-10.1186/1532-429X-11-49.PubMedCentralCrossRefPubMed
13.
go back to reference Kozerke S, Schwitter J, Pedersen EM, Boesiger P: Aortic and mitral regurgitation: quantification using moving slice velocity mapping. J Magn Reson Imaging. 2001, 14: 106-12. 10.1002/jmri.1159.CrossRefPubMed Kozerke S, Schwitter J, Pedersen EM, Boesiger P: Aortic and mitral regurgitation: quantification using moving slice velocity mapping. J Magn Reson Imaging. 2001, 14: 106-12. 10.1002/jmri.1159.CrossRefPubMed
14.
go back to reference Kozerke S, Hasenkam JM, Pedersen EM, Boesiger P: Visualization of flow patterns distal to aortic valve prostheses in humans using a fast approach for cine 3D velocity mapping. J Magn Reson Imaging. 2001, 13: 690-8. 10.1002/jmri.1097.CrossRefPubMed Kozerke S, Hasenkam JM, Pedersen EM, Boesiger P: Visualization of flow patterns distal to aortic valve prostheses in humans using a fast approach for cine 3D velocity mapping. J Magn Reson Imaging. 2001, 13: 690-8. 10.1002/jmri.1097.CrossRefPubMed
15.
go back to reference Kozerke S, Hasenkam JM, Nygaard H, Paulsen PK, Pedersen EM, Boesiger P: Heart motion-adapted MR velocity mapping of blood velocity distribution downstream of aortic valve prostheses: initial experience. Radiology. 2001, 218: 548-55.CrossRefPubMed Kozerke S, Hasenkam JM, Nygaard H, Paulsen PK, Pedersen EM, Boesiger P: Heart motion-adapted MR velocity mapping of blood velocity distribution downstream of aortic valve prostheses: initial experience. Radiology. 2001, 218: 548-55.CrossRefPubMed
16.
go back to reference Quick HH, Kuehl H, Kaiser G, Hornscheidt D, Mikolajczyk KP, Aker S, Debatin JF, Ladd ME: Interventional MRA using actively visualized catheters, TrueFISP, and real-time image fusion. Magn Reson Med. 2003, 49: 129-7. 10.1002/mrm.10334.CrossRefPubMed Quick HH, Kuehl H, Kaiser G, Hornscheidt D, Mikolajczyk KP, Aker S, Debatin JF, Ladd ME: Interventional MRA using actively visualized catheters, TrueFISP, and real-time image fusion. Magn Reson Med. 2003, 49: 129-7. 10.1002/mrm.10334.CrossRefPubMed
17.
go back to reference Kahlert P, Eggebrecht H, Plicht B, Kraff O, McDougall I, Decker B, Erbel R, Ladd ME, Quick HH: Towards real-time cardiovascular magnetic resonance-guided transarterial aortic valve implantation: in vitro evaluation and modification of existing devices. J Cardiovasc Magn Reson. 2010, 12: 58-10.1186/1532-429X-12-58.PubMedCentralCrossRefPubMed Kahlert P, Eggebrecht H, Plicht B, Kraff O, McDougall I, Decker B, Erbel R, Ladd ME, Quick HH: Towards real-time cardiovascular magnetic resonance-guided transarterial aortic valve implantation: in vitro evaluation and modification of existing devices. J Cardiovasc Magn Reson. 2010, 12: 58-10.1186/1532-429X-12-58.PubMedCentralCrossRefPubMed
18.
go back to reference Kahlert P, Parohl N, Albert J, Schäfer L, Reinhardt R, Kaiser GM, McDougall I, Decker B, Plicht B, Erbel R: Real-Time Magnetic Resonance Imaging-Guided Transarterial Aortic Valve Implantation: In-Vivo Evaluation in Swine. Research Correspondence. J Am Coll Cardiol. 2012, 59: 192-3. 10.1016/j.jacc.2011.09.046.CrossRefPubMed Kahlert P, Parohl N, Albert J, Schäfer L, Reinhardt R, Kaiser GM, McDougall I, Decker B, Plicht B, Erbel R: Real-Time Magnetic Resonance Imaging-Guided Transarterial Aortic Valve Implantation: In-Vivo Evaluation in Swine. Research Correspondence. J Am Coll Cardiol. 2012, 59: 192-3. 10.1016/j.jacc.2011.09.046.CrossRefPubMed
19.
go back to reference Kaiser GM, Breuckmann F, Aker S, Eggebrecht H, Kuehl H, Erbel R, Fruhauf NR, Broelsch CE, Quick HH: Anesthesia for cardiovascular interventions and magnetic resonance imaging in pigs. J Am Assoc Lab Anim Sci. 2007, 46: 30-3.PubMed Kaiser GM, Breuckmann F, Aker S, Eggebrecht H, Kuehl H, Erbel R, Fruhauf NR, Broelsch CE, Quick HH: Anesthesia for cardiovascular interventions and magnetic resonance imaging in pigs. J Am Assoc Lab Anim Sci. 2007, 46: 30-3.PubMed
20.
go back to reference Kos S, Huegli R, Hofmann E, Quick HH, Kuehl H, Aker S, Kaiser GM, Borm PJ, Jacob AL, Bilecen D: First magnetic resonance imaging-guided aortic stenting and cava filter placement using a polyetheretherketone-based magnetic resonance imaging-compatible guidewire in swine: proof of concept. Cardiovasc Intervent Radiol. 2009, 32: 514-21. 10.1007/s00270-008-9483-5.CrossRefPubMed Kos S, Huegli R, Hofmann E, Quick HH, Kuehl H, Aker S, Kaiser GM, Borm PJ, Jacob AL, Bilecen D: First magnetic resonance imaging-guided aortic stenting and cava filter placement using a polyetheretherketone-based magnetic resonance imaging-compatible guidewire in swine: proof of concept. Cardiovasc Intervent Radiol. 2009, 32: 514-21. 10.1007/s00270-008-9483-5.CrossRefPubMed
21.
go back to reference Bonatti J, Vassiliades T, Nifong W, Jakob H, Erbel R, Fosse E, Werkkala K, Sutlic Z, Bartel T, Friedrich G: How to build a cath-lab operating room. Heart Surg Forum. 2007, 10: E344-E8. 10.1532/HSF98.20070709.CrossRefPubMed Bonatti J, Vassiliades T, Nifong W, Jakob H, Erbel R, Fosse E, Werkkala K, Sutlic Z, Bartel T, Friedrich G: How to build a cath-lab operating room. Heart Surg Forum. 2007, 10: E344-E8. 10.1532/HSF98.20070709.CrossRefPubMed
22.
go back to reference Eggebrecht H, Kuhl H, Kaiser GM, Aker S, Zenge MO, Stock F, Breuckmann F, Grabellus F, Ladd ME, Mehta RH: Feasibility of real-time magnetic resonance-guided stent-graft placement in a swine model of descending aortic dissection. Eur Heart J. 2006, 27: 613-20. 10.1093/eurheartj/ehi732.CrossRefPubMed Eggebrecht H, Kuhl H, Kaiser GM, Aker S, Zenge MO, Stock F, Breuckmann F, Grabellus F, Ladd ME, Mehta RH: Feasibility of real-time magnetic resonance-guided stent-graft placement in a swine model of descending aortic dissection. Eur Heart J. 2006, 27: 613-20. 10.1093/eurheartj/ehi732.CrossRefPubMed
23.
go back to reference Khawaja MZ, Rajani R, Cook A, Khavandi A, Moynagh A, Chowdhary S, Spence MS, Brown S, Khan SQ, Walker N: Permanent Pacemaker Insertion After CoreValve Transcatheter Aortic Valve Implantation: Incidence and Contributing Factors (the UK CoreValve Collaborative). Circulation. 2011, 123: 951-60. 10.1161/CIRCULATIONAHA.109.927152.CrossRefPubMed Khawaja MZ, Rajani R, Cook A, Khavandi A, Moynagh A, Chowdhary S, Spence MS, Brown S, Khan SQ, Walker N: Permanent Pacemaker Insertion After CoreValve Transcatheter Aortic Valve Implantation: Incidence and Contributing Factors (the UK CoreValve Collaborative). Circulation. 2011, 123: 951-60. 10.1161/CIRCULATIONAHA.109.927152.CrossRefPubMed
24.
go back to reference Kuehne T, Yilmaz S, Meinus C, Moore P, Saeed M, Weber O, Higgins CB, Blank T, Elsaesser E, Schnackenburg B: magnetic resonance imaging-guided transcatheter implantation of a prosthetic valve in aortic valve position: feasibility study in swine. J Am Coll Cardiol. 2004, 44: 2247-9. 10.1016/j.jacc.2004.09.007.CrossRefPubMed Kuehne T, Yilmaz S, Meinus C, Moore P, Saeed M, Weber O, Higgins CB, Blank T, Elsaesser E, Schnackenburg B: magnetic resonance imaging-guided transcatheter implantation of a prosthetic valve in aortic valve position: feasibility study in swine. J Am Coll Cardiol. 2004, 44: 2247-9. 10.1016/j.jacc.2004.09.007.CrossRefPubMed
25.
go back to reference McVeigh ER, Guttman MA, Lederman RJ, Li M, Kocaturk O, Hunt T, Kozlov S, Horvath KA: Real-time interactive MRI-guided cardiac surgery: aortic valve replacement using a direct apical approach. Magn Reson Med. 2006, 56: 958-64. 10.1002/mrm.21044.PubMedCentralCrossRefPubMed McVeigh ER, Guttman MA, Lederman RJ, Li M, Kocaturk O, Hunt T, Kozlov S, Horvath KA: Real-time interactive MRI-guided cardiac surgery: aortic valve replacement using a direct apical approach. Magn Reson Med. 2006, 56: 958-64. 10.1002/mrm.21044.PubMedCentralCrossRefPubMed
26.
go back to reference Horvath KA, Mazilu D, Guttman M, Zetts A, Hunt T, Li M: Midterm results of transapical aortic valve replacement via real-time magnetic resonance imaging guidance. J Thorac Cardiovasc Surg. 2010, 139: 424-30. 10.1016/j.jtcvs.2009.08.005.PubMedCentralCrossRefPubMed Horvath KA, Mazilu D, Guttman M, Zetts A, Hunt T, Li M: Midterm results of transapical aortic valve replacement via real-time magnetic resonance imaging guidance. J Thorac Cardiovasc Surg. 2010, 139: 424-30. 10.1016/j.jtcvs.2009.08.005.PubMedCentralCrossRefPubMed
27.
go back to reference Horvath KA, Mazilu D, Kocaturk O, Li M: Transapical aortic valve replacement under real-time magnetic resonance imaging guidance: experimental results with balloon-expandable and self-expanding stents. Eur J Cardiothorac Surg. 2011, 39: 822-8. 10.1016/j.ejcts.2010.09.030.PubMedCentralCrossRefPubMed Horvath KA, Mazilu D, Kocaturk O, Li M: Transapical aortic valve replacement under real-time magnetic resonance imaging guidance: experimental results with balloon-expandable and self-expanding stents. Eur J Cardiothorac Surg. 2011, 39: 822-8. 10.1016/j.ejcts.2010.09.030.PubMedCentralCrossRefPubMed
28.
go back to reference Koos R, Altiok E, Mahnken AH, Neizel M, Dohmen G, Marx N, Kuhl H, Hoffmann R: Evaluation of aortic root for definition of prosthesis size by magnetic resonance imaging and cardiac computed tomography: Implications for transcatheter aortic valve implantation. Int J Cardiol. 2011, PMID: 21315460, doi: 10.1016/j.ijcard.2011.01.044, Koos R, Altiok E, Mahnken AH, Neizel M, Dohmen G, Marx N, Kuhl H, Hoffmann R: Evaluation of aortic root for definition of prosthesis size by magnetic resonance imaging and cardiac computed tomography: Implications for transcatheter aortic valve implantation. Int J Cardiol. 2011, PMID: 21315460, doi: 10.1016/j.ijcard.2011.01.044,
29.
go back to reference Glockner JF, Johnston DL, McGee KP: Evaluation of cardiac valvular disease with MR imaging: qualitative and quantitative techniques. Radiographics. 2003, 23: e9-10.1148/rg.e9.CrossRefPubMed Glockner JF, Johnston DL, McGee KP: Evaluation of cardiac valvular disease with MR imaging: qualitative and quantitative techniques. Radiographics. 2003, 23: e9-10.1148/rg.e9.CrossRefPubMed
30.
go back to reference Gatehouse PD, Keegan J, Crowe LA, Masood S, Mohiaddin RH, Kreitner KF, Firmin DN: Applications of phase-contrast flow and velocity imaging in cardiovascular MRI. Eur Radiol. 2005, 15: 2172-84. 10.1007/s00330-005-2829-3.CrossRefPubMed Gatehouse PD, Keegan J, Crowe LA, Masood S, Mohiaddin RH, Kreitner KF, Firmin DN: Applications of phase-contrast flow and velocity imaging in cardiovascular MRI. Eur Radiol. 2005, 15: 2172-84. 10.1007/s00330-005-2829-3.CrossRefPubMed
31.
go back to reference Nitz WR, Oppelt A, Renz W, Manke C, Lenhart M, Link J: On the heating of linear conductive structures as guide wires and catheters in interventional MRI. J Magn Reson Imaging. 2001, 13: 105-14. 10.1002/1522-2586(200101)13:1<105::AID-JMRI1016>3.0.CO;2-0.CrossRefPubMed Nitz WR, Oppelt A, Renz W, Manke C, Lenhart M, Link J: On the heating of linear conductive structures as guide wires and catheters in interventional MRI. J Magn Reson Imaging. 2001, 13: 105-14. 10.1002/1522-2586(200101)13:1<105::AID-JMRI1016>3.0.CO;2-0.CrossRefPubMed
32.
go back to reference Mekle R, Zenge MO, Ladd ME, Quick HH, Hofmann E, Scheffler K, Bilecen D: Initial in vivo studies with a polymer-based MR-compatible guide wire. J Vasc Interv Radiol. 2009, 20: 1384-9. 10.1016/j.jvir.2009.07.001.CrossRefPubMed Mekle R, Zenge MO, Ladd ME, Quick HH, Hofmann E, Scheffler K, Bilecen D: Initial in vivo studies with a polymer-based MR-compatible guide wire. J Vasc Interv Radiol. 2009, 20: 1384-9. 10.1016/j.jvir.2009.07.001.CrossRefPubMed
33.
go back to reference Krueger S, Schmitz S, Weiss S, Wirtz D, Linssen M, Schade H, Kraemer N, Spuentrup E, Krombach G, Buecker A: An MR guidewire based on micropultruded fiber-reinforced material. Magn Reson Med. 2008, 60: 1190-6. 10.1002/mrm.21743.CrossRefPubMed Krueger S, Schmitz S, Weiss S, Wirtz D, Linssen M, Schade H, Kraemer N, Spuentrup E, Krombach G, Buecker A: An MR guidewire based on micropultruded fiber-reinforced material. Magn Reson Med. 2008, 60: 1190-6. 10.1002/mrm.21743.CrossRefPubMed
34.
go back to reference Grube E, Naber C, Abizaid A, Sousa E, Mendiz O, Lemos P, Kalil FR, Mangione J, Buellesfeld L: Feasibility of transcatheter aortic valve implantation without balloon pre-dilation a pilot study. JACC Cardiovasc Interv. 2011, 4: 751-7. 10.1016/j.jcin.2011.03.015.CrossRefPubMed Grube E, Naber C, Abizaid A, Sousa E, Mendiz O, Lemos P, Kalil FR, Mangione J, Buellesfeld L: Feasibility of transcatheter aortic valve implantation without balloon pre-dilation a pilot study. JACC Cardiovasc Interv. 2011, 4: 751-7. 10.1016/j.jcin.2011.03.015.CrossRefPubMed
Metadata
Title
Towards real-time cardiovascular magnetic resonance guided transarterial CoreValve implantation: in vivo evaluation in swine
Authors
Philipp Kahlert
Nina Parohl
Juliane Albert
Lena Schäfer
Renate Reinhardt
Gernot M Kaiser
Ian McDougall
Brad Decker
Björn Plicht
Raimund Erbel
Holger Eggebrecht
Mark E Ladd
Harald H Quick
Publication date
01-12-2012
Publisher
BioMed Central
Published in
Journal of Cardiovascular Magnetic Resonance / Issue 1/2012
Electronic ISSN: 1532-429X
DOI
https://doi.org/10.1186/1532-429X-14-21

Other articles of this Issue 1/2012

Journal of Cardiovascular Magnetic Resonance 1/2012 Go to the issue