Cardiac function and myocardial perfusion immediately following maximal treadmill exercise inside the MRI room

Authors: Mihaela Jekic, Eric L Foster, Michelle R Ballinger, Subha V Raman, Orlando P Simonetti

Published in: Journal of Cardiovascular Magnetic Resonance | Issue 1/2008

Login to get access

Abstract

Treadmill exercise stress testing is an essential tool in the prevention, detection, and treatment of a broad spectrum of cardiovascular disease. After maximal exercise, cardiac images at peak stress are typically acquired using nuclear scintigraphy or echocardiography, both of which have inherent limitations. Although CMR offers superior image quality, the lack of MRI-compatible exercise and monitoring equipment has prevented the realization of treadmill exercise CMR.
It is critical to commence imaging as quickly as possible after exercise to capture exercise-induced cardiac wall motion abnormalities. We modified a commercial treadmill such that it could be safely positioned inside the MRI room to minimize the distance between the treadmill and the scan table. We optimized the treadmill exercise CMR protocol in 20 healthy volunteers and successfully imaged cardiac function and myocardial perfusion at peak stress, followed by viability imaging at rest. Imaging commenced an average of 30 seconds after maximal exercise. Real-time cine of seven slices with no breath-hold and no ECG-gating was completed within 45 seconds of exercise, immediately followed by stress perfusion imaging of three short-axis slices which showed an average time to peak enhancement within 57 seconds of exercise. We observed a 3.1-fold increase in cardiac output and a myocardial perfusion reserve index of 1.9, which agree with reported values for healthy subjects at peak stress. This study successfully demonstrates in-room treadmill exercise CMR in healthy volunteers, but confirmation of feasibility in patients with heart disease is still needed.
Appendix
Available only for authorised users
Literature
1.
go back to reference Goldhammer S, Scherf D: Elektrokardiographische unterschungen bei kranken mit angina pectoris ("ambulatorischer Typus"). Ztschr f klin Med. 1932, 122: 134- Goldhammer S, Scherf D: Elektrokardiographische unterschungen bei kranken mit angina pectoris ("ambulatorischer Typus"). Ztschr f klin Med. 1932, 122: 134-
2.
go back to reference Bruce RA, Blackmon JR, Jones JW, Strait G: Exercising Testing in Adult Normal Subjects and Cardiac Patients. Pediatrics. 1963, 32 (Suppl): 742-756.PubMed Bruce RA, Blackmon JR, Jones JW, Strait G: Exercising Testing in Adult Normal Subjects and Cardiac Patients. Pediatrics. 1963, 32 (Suppl): 742-756.PubMed
3.
go back to reference Myers J, Voodi L, Umann T, Froelicher VF: A survey of exercise testing: methods, utilization, interpretation, and safety in the VAHCS. J Cardiopulm Rehabil. 2000, 20: 251-8. 10.1097/00008483-200007000-00007.CrossRefPubMed Myers J, Voodi L, Umann T, Froelicher VF: A survey of exercise testing: methods, utilization, interpretation, and safety in the VAHCS. J Cardiopulm Rehabil. 2000, 20: 251-8. 10.1097/00008483-200007000-00007.CrossRefPubMed
4.
go back to reference Lear SA, Brozic A, Myers JN, Ignaszewski A: Exercise stress testing. An overview of current guidelines. Sports Med. 1999, 27: 285-312. 10.2165/00007256-199927050-00002.CrossRefPubMed Lear SA, Brozic A, Myers JN, Ignaszewski A: Exercise stress testing. An overview of current guidelines. Sports Med. 1999, 27: 285-312. 10.2165/00007256-199927050-00002.CrossRefPubMed
5.
go back to reference Ashley EA, Myers J, Froelicher V: Exercise testing in clinical medicine. Lancet. 2000, 356: 1592-7. 10.1016/S0140-6736(00)03138-X.CrossRefPubMed Ashley EA, Myers J, Froelicher V: Exercise testing in clinical medicine. Lancet. 2000, 356: 1592-7. 10.1016/S0140-6736(00)03138-X.CrossRefPubMed
6.
go back to reference Tavel ME, Shaar C: Relation between the electrocardiographic stress test and degree and location of myocardial ischemia. Am J Cardiol. 1999, 84: 119-24. 10.1016/S0002-9149(99)00219-2.CrossRefPubMed Tavel ME, Shaar C: Relation between the electrocardiographic stress test and degree and location of myocardial ischemia. Am J Cardiol. 1999, 84: 119-24. 10.1016/S0002-9149(99)00219-2.CrossRefPubMed
7.
go back to reference Lee TH, Boucher CA: Clinical practice. Noninvasive tests in patients with stable coronary artery disease. N Engl J Med. 2001, 344: 1840-5. 10.1056/NEJM200106143442406.CrossRefPubMed Lee TH, Boucher CA: Clinical practice. Noninvasive tests in patients with stable coronary artery disease. N Engl J Med. 2001, 344: 1840-5. 10.1056/NEJM200106143442406.CrossRefPubMed
8.
go back to reference Garber AM, Solomon NA: Cost-effectiveness of alternative test strategies for the diagnosis of coronary artery disease. Ann Intern Med. 1999, 130: 719-28.CrossRefPubMed Garber AM, Solomon NA: Cost-effectiveness of alternative test strategies for the diagnosis of coronary artery disease. Ann Intern Med. 1999, 130: 719-28.CrossRefPubMed
9.
go back to reference Klem I, Heitner JF, Shah DJ, Sketch MH, Behar V, Weinsaft J, Cawley P, Parker M, Elliott M, Judd RM, Kim RJ: Improved detection of coronary artery disease by stress perfusion cardiovascular magnetic resonance with the use of delayed enhancement infarction imaging. J Am Coll Cardiol. 2006, 47: 1630-8. 10.1016/j.jacc.2005.10.074.CrossRefPubMed Klem I, Heitner JF, Shah DJ, Sketch MH, Behar V, Weinsaft J, Cawley P, Parker M, Elliott M, Judd RM, Kim RJ: Improved detection of coronary artery disease by stress perfusion cardiovascular magnetic resonance with the use of delayed enhancement infarction imaging. J Am Coll Cardiol. 2006, 47: 1630-8. 10.1016/j.jacc.2005.10.074.CrossRefPubMed
10.
go back to reference Nagel E, Lehmkuhl HB, Bocksch W, Klein C, Vogel U, Frantz E, Ellmer A, Dreysse S, Fleck E: Noninvasive Diagnosis of Ischemia-Induced Wall Motion Abnormalities With the Use of High-Dose Dobutamine Stress MRI: Comparison With Dobutamine Stress Echocardiography. Circulation. 1999, 99: 763-770.CrossRefPubMed Nagel E, Lehmkuhl HB, Bocksch W, Klein C, Vogel U, Frantz E, Ellmer A, Dreysse S, Fleck E: Noninvasive Diagnosis of Ischemia-Induced Wall Motion Abnormalities With the Use of High-Dose Dobutamine Stress MRI: Comparison With Dobutamine Stress Echocardiography. Circulation. 1999, 99: 763-770.CrossRefPubMed
11.
go back to reference Schwitter J: MR-IMPACT: Magnetic Resonance Imaging for Myocardial Perfusion Assessment in Coronary Artery Disease Trial. European Society of Cardiology Congress. 2005, Stockholm, Sweden: European Society of Cardiology Schwitter J: MR-IMPACT: Magnetic Resonance Imaging for Myocardial Perfusion Assessment in Coronary Artery Disease Trial. European Society of Cardiology Congress. 2005, Stockholm, Sweden: European Society of Cardiology
12.
go back to reference Armstrong WF, Zoghbi WA: Stress echocardiography: current methodology and clinical applications. J Am Coll Cardiol. 2005, 45: 1739-47. 10.1016/j.jacc.2004.12.078.CrossRefPubMed Armstrong WF, Zoghbi WA: Stress echocardiography: current methodology and clinical applications. J Am Coll Cardiol. 2005, 45: 1739-47. 10.1016/j.jacc.2004.12.078.CrossRefPubMed
13.
go back to reference Tavel ME: Stress Testing in Cardiac Evaluation: Current Concepts With Emphasis on the ECG 10.1378/chest.119.3.907. Chest. 2001, 119: 907-925. 10.1378/chest.119.3.907.CrossRefPubMed Tavel ME: Stress Testing in Cardiac Evaluation: Current Concepts With Emphasis on the ECG 10.1378/chest.119.3.907. Chest. 2001, 119: 907-925. 10.1378/chest.119.3.907.CrossRefPubMed
14.
go back to reference Bruce RA, Hornsten TR: Exercise stress testing in evaluation of patients with ischemic heart disease. Prog Cardiovasc Dis. 1969, 11: 371-90. 10.1016/0033-0620(69)90027-9.CrossRefPubMed Bruce RA, Hornsten TR: Exercise stress testing in evaluation of patients with ischemic heart disease. Prog Cardiovasc Dis. 1969, 11: 371-90. 10.1016/0033-0620(69)90027-9.CrossRefPubMed
15.
go back to reference Roger VL, Jacobsen SJ, Pellikka PA, Miller TD, Bailey KR, Gersh BJ: Prognostic value of treadmill exercise testing: a population-based study in Olmsted County, Minnesota. Circulation. 1998, 98: 2836-41.CrossRefPubMed Roger VL, Jacobsen SJ, Pellikka PA, Miller TD, Bailey KR, Gersh BJ: Prognostic value of treadmill exercise testing: a population-based study in Olmsted County, Minnesota. Circulation. 1998, 98: 2836-41.CrossRefPubMed
16.
go back to reference Fletcher GF, Balady GJ, Amsterdam EA, Chaitman B, Eckel R, Fleg J, Froelicher VF, Leon AS, Piña IL, Rodney R, Simons-Morton DA, Williams MA, Bazzarre T: Exercise standards for testing and training: a statement for healthcare professionals from the American Heart Association. Circulation. 2001, 104: 1694-740. 10.1161/hc3901.095960.CrossRefPubMed Fletcher GF, Balady GJ, Amsterdam EA, Chaitman B, Eckel R, Fleg J, Froelicher VF, Leon AS, Piña IL, Rodney R, Simons-Morton DA, Williams MA, Bazzarre T: Exercise standards for testing and training: a statement for healthcare professionals from the American Heart Association. Circulation. 2001, 104: 1694-740. 10.1161/hc3901.095960.CrossRefPubMed
18.
go back to reference Cheng CP, Herfkens RJ, Taylor CA: Inferior vena caval hemodynamics quantified in vivo at rest and during cycling exercise using magnetic resonance imaging. Am J Physiol Heart Circ Physiol. 2003, 284: H1161-7.CrossRefPubMed Cheng CP, Herfkens RJ, Taylor CA: Inferior vena caval hemodynamics quantified in vivo at rest and during cycling exercise using magnetic resonance imaging. Am J Physiol Heart Circ Physiol. 2003, 284: H1161-7.CrossRefPubMed
19.
go back to reference Cheng CP, Herfkens RJ, Taylor CA: Abdominal aortic hemodynamic conditions in healthy subjects aged 50–70 at rest and during lower limb exercise: in vivo quantification using MRI. Atherosclerosis. 2003, 168: 323-31. 10.1016/S0021-9150(03)00099-6.CrossRefPubMed Cheng CP, Herfkens RJ, Taylor CA: Abdominal aortic hemodynamic conditions in healthy subjects aged 50–70 at rest and during lower limb exercise: in vivo quantification using MRI. Atherosclerosis. 2003, 168: 323-31. 10.1016/S0021-9150(03)00099-6.CrossRefPubMed
20.
go back to reference Taylor CA, Cheng CP, Espinosa LA, Tang BT, Parker D, Herfkens RJ: In vivo quantification of blood flow and wall shear stress in the human abdominal aorta during lower limb exercise. Ann Biomed Eng. 2002, 30: 402-8. 10.1114/1.1476016.CrossRefPubMed Taylor CA, Cheng CP, Espinosa LA, Tang BT, Parker D, Herfkens RJ: In vivo quantification of blood flow and wall shear stress in the human abdominal aorta during lower limb exercise. Ann Biomed Eng. 2002, 30: 402-8. 10.1114/1.1476016.CrossRefPubMed
21.
go back to reference Niezen RA, Doornbos J, van der Wall EE, de Roos A: Measurement of aortic and pulmonary flow with MRI at rest and during physical exercise. J Comput Assist Tomogr. 1998, 22: 194-201. 10.1097/00004728-199803000-00006.CrossRefPubMed Niezen RA, Doornbos J, van der Wall EE, de Roos A: Measurement of aortic and pulmonary flow with MRI at rest and during physical exercise. J Comput Assist Tomogr. 1998, 22: 194-201. 10.1097/00004728-199803000-00006.CrossRefPubMed
22.
go back to reference Hjortdal VE, Emmertsen K, Stenbøg E, Fründ T, Schmidt MR, Kromann O, Sørensen K, Pedersen EM: Effects of exercise and respiration on blood flow in total cavopulmonary connection: a real-time magnetic resonance flow study. Circulation. 2003, 108: 1227-31. 10.1161/01.CIR.0000087406.27922.6B.CrossRefPubMed Hjortdal VE, Emmertsen K, Stenbøg E, Fründ T, Schmidt MR, Kromann O, Sørensen K, Pedersen EM: Effects of exercise and respiration on blood flow in total cavopulmonary connection: a real-time magnetic resonance flow study. Circulation. 2003, 108: 1227-31. 10.1161/01.CIR.0000087406.27922.6B.CrossRefPubMed
23.
go back to reference Gibbons RJ, Balady GJ, Bricker JT, Chaitman BR, Fletcher GF, Froelicher VF, Mark DB, McCallister BD, Mooss AN, O'Reilly MG, Winters WL, Gibbons RJ, Antman EM, Alpert JS, Faxon DP, Fuster V, Gregoratos G, Hiratzka LF, Jacobs AK, Russell RO, Smith SC: ACC/AHA 2002 guideline update for exercise testing: summary article. A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee to Update the 1997 Exercise Testing Guidelines). J Am Coll Cardiol. 2002, 40: 1531-40. 10.1016/S0735-1097(02)02164-2.CrossRefPubMed Gibbons RJ, Balady GJ, Bricker JT, Chaitman BR, Fletcher GF, Froelicher VF, Mark DB, McCallister BD, Mooss AN, O'Reilly MG, Winters WL, Gibbons RJ, Antman EM, Alpert JS, Faxon DP, Fuster V, Gregoratos G, Hiratzka LF, Jacobs AK, Russell RO, Smith SC: ACC/AHA 2002 guideline update for exercise testing: summary article. A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee to Update the 1997 Exercise Testing Guidelines). J Am Coll Cardiol. 2002, 40: 1531-40. 10.1016/S0735-1097(02)02164-2.CrossRefPubMed
24.
go back to reference Miyamura M, Honda Y: Oxygen intake and cardiac output during treadmill and bicycle exercise. J Appl Physiol. 1972, 32: 185-a-188. Miyamura M, Honda Y: Oxygen intake and cardiac output during treadmill and bicycle exercise. J Appl Physiol. 1972, 32: 185-a-188.
25.
go back to reference Armstrong WF, Pellikka PA, Ryan T, Crouse L, Zoghbi WA: Stress echocardiography: recommendations for performance and interpretation of stress echocardiography. Stress Echocardiography Task Force of the Nomenclature and Standards Committee of the American Society of Echocardiography. J Am Soc Echocardiogr. 1998, 11: 97-104. 10.1016/S0894-7317(98)70132-4.CrossRefPubMed Armstrong WF, Pellikka PA, Ryan T, Crouse L, Zoghbi WA: Stress echocardiography: recommendations for performance and interpretation of stress echocardiography. Stress Echocardiography Task Force of the Nomenclature and Standards Committee of the American Society of Echocardiography. J Am Soc Echocardiogr. 1998, 11: 97-104. 10.1016/S0894-7317(98)70132-4.CrossRefPubMed
26.
go back to reference Salustri A, Pozzoli MM, Hermans W, Ilmer B, Cornel JH, Reijs AE, Roelandt JR, Fioretti PM: Relationship between exercise echocardiography and perfusion single-photon emission computed tomography in patients with single-vessel coronary artery disease. Am Heart J. 1992, 124: 75-83. 10.1016/0002-8703(92)90922-I.CrossRefPubMed Salustri A, Pozzoli MM, Hermans W, Ilmer B, Cornel JH, Reijs AE, Roelandt JR, Fioretti PM: Relationship between exercise echocardiography and perfusion single-photon emission computed tomography in patients with single-vessel coronary artery disease. Am Heart J. 1992, 124: 75-83. 10.1016/0002-8703(92)90922-I.CrossRefPubMed
27.
go back to reference Iliceto S, D'Ambrosio G, Sorino M, Papa A, Amico A, Ricci A, Rizzon P: Comparison of postexercise and transesophageal atrial pacing two-dimensional echocardiography for detection of coronary artery disease. Am J Cardiol. 1986, 57: 547-53. 10.1016/0002-9149(86)90832-5.CrossRefPubMed Iliceto S, D'Ambrosio G, Sorino M, Papa A, Amico A, Ricci A, Rizzon P: Comparison of postexercise and transesophageal atrial pacing two-dimensional echocardiography for detection of coronary artery disease. Am J Cardiol. 1986, 57: 547-53. 10.1016/0002-9149(86)90832-5.CrossRefPubMed
28.
go back to reference Dymond DS, Foster C, Grenier RP, Carpenter J, Schmidt DH: Peak exercise and immediate postexercise imaging for the detection of left ventricular functional abnormalities in coronary artery disease. Am J Cardiol. 1984, 53: 1532-7. 10.1016/0002-9149(84)90574-5.CrossRefPubMed Dymond DS, Foster C, Grenier RP, Carpenter J, Schmidt DH: Peak exercise and immediate postexercise imaging for the detection of left ventricular functional abnormalities in coronary artery disease. Am J Cardiol. 1984, 53: 1532-7. 10.1016/0002-9149(84)90574-5.CrossRefPubMed
29.
go back to reference Kellman P, Epstein FH, McVeigh ER: Adaptive sensitivity encoding incorporating temporal filtering (TSENSE). Magn Reson Med. 2001, 45: 846-52. 10.1002/mrm.1113.CrossRefPubMed Kellman P, Epstein FH, McVeigh ER: Adaptive sensitivity encoding incorporating temporal filtering (TSENSE). Magn Reson Med. 2001, 45: 846-52. 10.1002/mrm.1113.CrossRefPubMed
30.
go back to reference Kettenbach J, Kacher DF, Kanan AR, Rostenberg B, Fairhurst J, Stadler A, Kienreich K, Jolesz FA: Intraoperative and interventional MRI: recommendations for a safe environment. Minim Invasive Ther Allied Technol. 2006, 15: 53-64. 10.1080/13645700600640774.CrossRefPubMed Kettenbach J, Kacher DF, Kanan AR, Rostenberg B, Fairhurst J, Stadler A, Kienreich K, Jolesz FA: Intraoperative and interventional MRI: recommendations for a safe environment. Minim Invasive Ther Allied Technol. 2006, 15: 53-64. 10.1080/13645700600640774.CrossRefPubMed
31.
go back to reference Fischer SE, Wickline SA, Lorenz CH: Novel real-time R-wave detection algorithm based on the vectorcardiogram for accurate gated magnetic resonance acquisitions. Magn Reson Med. 1999, 42: 361-70. 10.1002/(SICI)1522-2594(199908)42:2<361::AID-MRM18>3.0.CO;2-9.CrossRefPubMed Fischer SE, Wickline SA, Lorenz CH: Novel real-time R-wave detection algorithm based on the vectorcardiogram for accurate gated magnetic resonance acquisitions. Magn Reson Med. 1999, 42: 361-70. 10.1002/(SICI)1522-2594(199908)42:2<361::AID-MRM18>3.0.CO;2-9.CrossRefPubMed
32.
go back to reference Nagel E, Klein C, Paetsch I, Hettwer S, Schnackenburg B, Wegscheider K, Fleck E: Magnetic resonance perfusion measurements for the noninvasive detection of coronary artery disease. Circulation. 2003, 108: 432-7. 10.1161/01.CIR.0000080915.35024.A9.CrossRefPubMed Nagel E, Klein C, Paetsch I, Hettwer S, Schnackenburg B, Wegscheider K, Fleck E: Magnetic resonance perfusion measurements for the noninvasive detection of coronary artery disease. Circulation. 2003, 108: 432-7. 10.1161/01.CIR.0000080915.35024.A9.CrossRefPubMed
33.
go back to reference Ogawa T, Spina RJ, Martin WH, Kohrt WM, Schechtman KB, Holloszy JO, Ehsani AA: Effects of aging, sex, and physical training on cardiovascular responses to exercise. Circulation. 1992, 86: 494-503.CrossRefPubMed Ogawa T, Spina RJ, Martin WH, Kohrt WM, Schechtman KB, Holloszy JO, Ehsani AA: Effects of aging, sex, and physical training on cardiovascular responses to exercise. Circulation. 1992, 86: 494-503.CrossRefPubMed
34.
go back to reference Rerkpattanapipat P, Gandhi SK, Darty SN, Williams RT, Davis AD, Mazur W, Clark HP, Little WC, Link KM, Hamilton CA, Hundley WG: Feasibility to detect severe coronary artery stenoses with upright treadmill exercise magnetic resonance imaging. Am J Cardiol. 2003, 92: 603-6. 10.1016/S0002-9149(03)00734-3.CrossRefPubMed Rerkpattanapipat P, Gandhi SK, Darty SN, Williams RT, Davis AD, Mazur W, Clark HP, Little WC, Link KM, Hamilton CA, Hundley WG: Feasibility to detect severe coronary artery stenoses with upright treadmill exercise magnetic resonance imaging. Am J Cardiol. 2003, 92: 603-6. 10.1016/S0002-9149(03)00734-3.CrossRefPubMed
35.
go back to reference Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, Pennell DJ, Rumberger JA, Ryan T, Verani MS, American Heart Association Writing Group on Myocardial Segmentation and Registration for Cardiac Imaging: Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: a statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation. 2002, 105: 539-42. 10.1161/hc0402.102975.CrossRefPubMed Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, Pennell DJ, Rumberger JA, Ryan T, Verani MS, American Heart Association Writing Group on Myocardial Segmentation and Registration for Cardiac Imaging: Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: a statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation. 2002, 105: 539-42. 10.1161/hc0402.102975.CrossRefPubMed
36.
go back to reference Wintersperger BJ, Reeder SB, Nikolaou K, Dietrich O, Huber A, Greiser A, Lanz T, Reiser MF, Schoenberg SO: Cardiac CINE MR imaging with a 32-channel cardiac coil and parallel imaging: impact of acceleration factors on image quality and volumetric accuracy. J Magn Reson Imaging. 2006, 23: 222-7. 10.1002/jmri.20484.CrossRefPubMed Wintersperger BJ, Reeder SB, Nikolaou K, Dietrich O, Huber A, Greiser A, Lanz T, Reiser MF, Schoenberg SO: Cardiac CINE MR imaging with a 32-channel cardiac coil and parallel imaging: impact of acceleration factors on image quality and volumetric accuracy. J Magn Reson Imaging. 2006, 23: 222-7. 10.1002/jmri.20484.CrossRefPubMed
37.
go back to reference Alam M, Hedman A, Nordlander R, Samad B: Right ventricular function before and after an uncomplicated coronary artery bypass graft as assessed by pulsed wave Doppler tissue imaging of the tricuspid annulus. Am Heart J. 2003, 146: 520-6. 10.1016/S0002-8703(03)00313-2.CrossRefPubMed Alam M, Hedman A, Nordlander R, Samad B: Right ventricular function before and after an uncomplicated coronary artery bypass graft as assessed by pulsed wave Doppler tissue imaging of the tricuspid annulus. Am Heart J. 2003, 146: 520-6. 10.1016/S0002-8703(03)00313-2.CrossRefPubMed
38.
go back to reference David JS, Tousignant CP, Bowry R: Tricuspid annular velocity in patients undergoing cardiac operation using transesophageal echocardiography. J Am Soc Echocardiogr. 2006, 19: 329-34. 10.1016/j.echo.2005.09.013.CrossRefPubMed David JS, Tousignant CP, Bowry R: Tricuspid annular velocity in patients undergoing cardiac operation using transesophageal echocardiography. J Am Soc Echocardiogr. 2006, 19: 329-34. 10.1016/j.echo.2005.09.013.CrossRefPubMed
39.
go back to reference Marcu CB, Beek AM, Van Rossum AC: Cardiovascular magnetic resonance imaging for the assessment of right heart involvement in cardiac and pulmonary disease. Heart Lung Circ. 2006, 15: 362-70. 10.1016/j.hlc.2006.08.003.CrossRefPubMed Marcu CB, Beek AM, Van Rossum AC: Cardiovascular magnetic resonance imaging for the assessment of right heart involvement in cardiac and pulmonary disease. Heart Lung Circ. 2006, 15: 362-70. 10.1016/j.hlc.2006.08.003.CrossRefPubMed
40.
go back to reference Raman SV, Cook SC, McCarthy B, Ferketich AK: Usefulness of multidetector row computed tomography to quantify right ventricular size and function in adults with either tetralogy of Fallot or transposition of the great arteries. Am J Cardiol. 2005, 95: 683-6. 10.1016/j.amjcard.2004.11.014.CrossRefPubMed Raman SV, Cook SC, McCarthy B, Ferketich AK: Usefulness of multidetector row computed tomography to quantify right ventricular size and function in adults with either tetralogy of Fallot or transposition of the great arteries. Am J Cardiol. 2005, 95: 683-6. 10.1016/j.amjcard.2004.11.014.CrossRefPubMed
Metadata
Title
Cardiac function and myocardial perfusion immediately following maximal treadmill exercise inside the MRI room
Authors
Mihaela Jekic
Eric L Foster
Michelle R Ballinger
Subha V Raman
Orlando P Simonetti
Publication date
01-12-2008
Publisher
BioMed Central
Published in
Journal of Cardiovascular Magnetic Resonance / Issue 1/2008
Electronic ISSN: 1532-429X
DOI
https://doi.org/10.1186/1532-429X-10-3

Other articles of this Issue 1/2008

Journal of Cardiovascular Magnetic Resonance 1/2008 Go to the issue